K060266 · Dade Behring, Inc. · JIS · Mar 9, 2006 · Clinical Chemistry
Device Facts
Record ID
K060266
Device Name
DIMENSION IRON CALIBRATOR
Applicant
Dade Behring, Inc.
Product Code
JIS · Clinical Chemistry
Decision Date
Mar 9, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1150
Device Class
Class 2
Indications for Use
The IRON Calibrator is an in vitro diagnostic product intended to be used to calibrate the IRON method for the Dimension® clinical chemistry system.
Device Story
Dimension® Iron Calibrator is an aqueous solution of iron wire dissolved in dilute HCl, provided in a 6-ampule kit across three concentration levels (0, 50, 1075 µg/dL). It is used in clinical chemistry laboratories to calibrate the IRON assay on Dimension® clinical chemistry systems. By establishing a calibration curve traceable to NIST SRM 937, the device enables the system to accurately quantify iron levels in patient samples. Healthcare providers use the resulting assay measurements to assess patient iron status and inform clinical decision-making. The calibrator ensures analytical accuracy and consistency across the assay range.
Clinical Evidence
No clinical data. Bench testing only. Stability testing performed over 13 months comparing product stored at 2-8°C vs -20°C. Stress testing included hot/cold shipping simulations and freeze/thaw cycles. Drift criteria established: ≤5 µg/dL (Level 1), ≤5% CV (Level 2), and ≤8% CV (Level 3).
Technological Characteristics
Aqueous solution of iron wire in dilute HCl. Traceable to NIST SRM 937. Three-level calibrator kit. Stability testing per CEN 13640. Risk management per ISO 14971:2000.
Indications for Use
Indicated for use as an in vitro diagnostic calibrator for the IRON method on the Dimension® clinical chemistry system. For prescription use only.
Regulatory Classification
Identification
A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
Dimension® IRN/TIBC Calibrator (DC21) (k944093)
Related Devices
K983548 — FERRITIN CALIBRATOR · Dade Behring, Inc. · Oct 27, 1998
K994114 — DIMENSION TOTAL IRON BINDING CAPACITY (IBCT) CALIBRATOR · Dade Behring, Inc. · Jan 20, 2000
K061251 — DIMENSION VISTA TOTAL IRON BINDING CAPACITY CALIBRATOR · Dade Behring, Inc. · May 31, 2006
K052002 — IRON STANDARD · Roche Diagnostics Corp. · Aug 22, 2005
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k060266
B. Purpose for Submission:
Clearance of a new calibrator
C. Measurand:
Total Iron
D. Type of Test:
Not applicable.
E. Applicant:
Dade Behring, Inc.
F. Proprietary and Established Names:
Dimension® IRON Calibrator (IRON Cal-DC85)
G. Regulatory Information:
1. Regulation section:
§862.1150, Calibrator, Primary
2. Classification:
Class II
3. Product code:
JIS
4. Panel:
Clinical Chemistry (75)
H. Intended Use:
1. Intended use(s):
See “Indication(s) for use” below.
2. Indication(s) for use:
The IRON Calibrator is an *in vitro* diagnostic product intended to be used to calibrate the IRON method for the Dimension® clinical chemistry system.
3. Special conditions for use statement(s):
For prescription use only
4. Special instrument requirements:
Dimension® clinical chemistry system
I. Device Description:
The IRON calibrator is an aqueous solution of iron wire dissolved in a dilute solution of HCl. The kit consists of 6 ampules, two at each of three levels.
{1}
2
J. Substantial Equivalence Information:
1. Predicate device name(s):
Dimension® IRN/TIBC Calibrator (DC21)
2. Predicate 510(k) number(s):
k944093
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended Use | In vitro diagnostic devices for calibrating iron assays | In vitro diagnostic devices for calibrating iron assays |
| Matrix | Aqueous solutions of iron wire dissolved in a dilute solutionof HCl | Aqueous solutions of iron wire dissolved in a dilute solutionof HCl |
| Detection | Bi-chromatic endpoint measurement (600 and 700 nm) | Bi-chromatic endpoint measurement (600 and 700 nm) |
| Assay Methodology | Chromatic | Chromatic |
| Levels | Multi-point with 3 levels | Multi-point with 3 levels |
| Traceability | NIST SRM 937 | NIST SRM 937 |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Target Concentrations | 0, 500 and 1075 ug/dL | 50, 500 and 1000 ug/dL |
K. Standard/Guidance Document Referenced (if applicable):
1) Continuous Quality Improvement: Essential Management Approaches; Approved Guideline (GP 22-A)
2) Stability Testing of In Vitro Diagnostic Reagents (13640)
3) Medical devices - Application of risk management to medical devices (14971:2000)
4) Medical Devices - Symbols to be used with medical device labels, labeling and information to be supplied (15223)
5) Guidance for Industry and FDA Staff - Use of Symbols on Labels and in Labeling of In Vitro Diagnostic Devices Intended for Professional Use
6) Guidance for Industry - Abbreviated 510(k) Submissions for In Vitro Diagnostic Calibrators; Final
L. Test Principle:
Not applicable.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
{2}
Not applicable.
b. Linearity/assay reportable range:
Not applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Calibrator set points are traceable to NIST SRM 937.
Calibrator shelf life was determined by comparing results of three different lots of the product stored at 2-8°C with product stored at -20°C over 13 months. Product was stored at 2-8 °C throughout testing cycle and tested at month 0, 1, 2, 4, 6, 7, 10, 12, 13, 18, 24, 25 (using 5 replicates per calibrator level). The control material was stored at -20 °C and tested at the same frequency. Stress testing was also performed using one lot of product stored at 2 -8 °C throughout testing cycle and tested at days 7, then months 6,7,12 and 13 (using 5 replicates per calibrator level) under hot (stressed shipping) conditions ((a) Store product at 30°C for 36 h, 35°C for 12h, then 45°C for 4 h then stored long term at 4°C. (b) Store product at 25°C for 48h, then 30°C for 6 h then store long term at 4°C. (c) Store product at 15°C for 54h, then store long term at 4°C.) and cold (stressed shipping) conditions (three freeze/thaw cycles then stored long-term @ 4°C). Calibrator shelf life was determined by comparing results of the product stored at 4°C with product stored at -20°C to ensure that analytical system drift is dissociated with calibrator drift. The drift at each test level must meet the following sponsor-set criteria:
- Less than or equal to 5ug/dL for Level 1 (0 µg/dL)
- Less than or equal to 5% CV Level 2 (500 µg/dL) and
- Less than or equal to 8% CV for level 3 (1075 µg/dL).
Using this criteria, the sponsor claims the time frame which the criteria are met and applies that time frame as the stability for unopened product stored at 2-8°C.
The Dimension® IRON Calibrator is in a single-use glass ampule product. According to the Dimension® IRON Calibrator Instructions for Use, opened ampules should be used immediately and any portion not used should be discarded. As such, no “Opened Product” stability studies were performed.
The target concentrations will be 0, 500 and 1075 µg/dL. This strategy supports a preference by the customer to have an assay range that is never lower than the lowest level of calibrator and never higher than the highest level of calibrator in order to be in compliance accrediting agencies and demonstrate linearity with each lot of reagent. The intermediate level was chosen based on a level to achieve the best linear curve fit.
Assay Range: 5 – 1000 µg/dL
{3}
Calibrator Target Levels: 0, 500, 1075 µg/dL
(These levels may vary based on lot to lot assignment.)
d. Detection limit:
Not applicable.
e. Analytical specificity:
Not applicable.
f. Assay cut-off:
Not applicable.
2. Comparison studies:
a. Method comparison with predicate device:
Not applicable.
b. Matrix comparison:
Not applicable.
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable.
b. Clinical specificity:
Not applicable.
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable.
4. Clinical cut-off:
Not applicable.
5. Expected values/Reference range:
Not applicable.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.