DIMENSION TOTAL TRIIODOTHYRONINE CALIBRATOR (RC414)
K032697 · Dade Behring, Inc. · JIS · Nov 7, 2003 · Clinical Chemistry
Device Facts
Record ID
K032697
Device Name
DIMENSION TOTAL TRIIODOTHYRONINE CALIBRATOR (RC414)
Applicant
Dade Behring, Inc.
Product Code
JIS · Clinical Chemistry
Decision Date
Nov 7, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1150
Device Class
Class 2
Intended Use
The T3 Calibrator is intended for use in the calibration of Total Triiodothyronine (T3) method on the Dimension® clinical chemistry system with the Heterogeneous Immunoassay Module.
Device Story
The Dimension® Total Triiodothyronine (T3) Calibrator (RC414) is a liquid, multi-level calibrator kit consisting of 10 vials (two each at five levels). Level 1 contains stripped human serum; levels 2-5 contain L-triiodothyronine in a stripped human serum base. The device is used in a clinical laboratory setting to calibrate the T3 assay on the Dimension® clinical chemistry system equipped with a Heterogeneous Immunoassay Module. By establishing a calibration curve, the device enables the system to accurately quantify T3 concentrations in patient samples, supporting clinical diagnostic decision-making regarding thyroid function.
Clinical Evidence
bench testing only
Technological Characteristics
Liquid calibrator kit; 5 levels (0, 1, 2, 4, 6.5 ng/mL); matrix: stripped human serum base; analyte: L-triiodothyronine. Designed for use with Dimension® clinical chemistry system Heterogeneous Immunoassay Module.
Indications for Use
Indicated for use as a calibrator for the Total Triiodothyronine (T3) assay on the Dimension® clinical chemistry system with the Heterogeneous Immunoassay Module.
Regulatory Classification
Identification
A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
K061885 — DIMENSION VISTA SYSTEM TOTAL TRIIODOTHYRONINE CALIBRATOR, MODEL KC250 · Dade Behring, Inc. · Aug 14, 2006
K100778 — DIMENSION EXL LOCI THYROID CALIBRATOR, MODEL RC 610A · Siemens Healthcare Diagnostics · May 5, 2010
Submission Summary (Full Text)
{0}------------------------------------------------
K032697
# Summary of Safety and Effectiveness Information
This summary of 510(k)safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| Submitter's Name: | Lorraine Piestrak<br>Dade Behring Inc.<br>P.O. Box 6101<br>Newark, DE 19714-6101 |
|--------------------------|----------------------------------------------------------------------------------|
| Date of Preparation: | August 29, 2003 |
| Name of Product: | Dimension® Total Triiodothyronine (T3) Calibrator (RC414) |
| FDA Classification Name: | Calibrator |
| Predicate Device: | Opus Total T3 Calibrator (K953160) |
Device Description: The Dimension® Total Triiodothyronine (T3) Calibrator (RC414) is a liquid product. The kit consists of 10 vials, two each at levels 1 through 5. Level 1 vials contain 2 mL of stripped human serum. Vials for levels 2 through 5, contain 1 mL with concentrations of L-triiodothyronine in a stripped human serum base.
Intended use: The Dimension® Total Triiodothyronine (T3) Calibrator (RC414) is intended for use in the calibration of the Total Triiodothyronine (T3) method on the Dimension® clinical chemistry system with the Heterogeneous Immunoassay Module.
# Comparison to Predicate Device:
| | Dimension®<br>T3 Calibrator | Opus<br>Total T3 Calibrator |
|--------------|-------------------------------------------------------|-------------------------------------------------------|
| Intended Use | Calibrator | Calibrator |
| Analyte | T3 | T3 |
| Matrix | stripped human serum base | stripped human serum base |
| Form | liquid | liquid |
| Volume | 2 mL per vial @ level 1<br>1 mL per vial @ levels 2-5 | 2 mL per vial @ level 1<br>1 mL per vial @ levels 2-6 |
| Levels | 5 levels<br>@ 0, 1, 2, 4, 6.5 ng/ml | 6 levels<br>@ 0, 0.5, 1, 2, 4, 6 ng/ml |
Comments on Substantial Equivalence: Both the proposed Dade Behring Dimension® Total Triiodothyronine (T3) Calibrator (RC414) and the existing Opus Total T3 calibrator are in vitro diagnostic products intended for calibrating T3 assays.
Conclusion: The Dimension® Total Triiodothyronine (T3) Calibrator (RC414) is substantially equivalent to the Opus Total T3 Calibrators based on the comparison discussed above.
Lorraine Piestrak
Fraine Piestrak Regulatory Affairs and Compliance Manager August 29, 2003
# 0000005
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" is arranged in a circular fashion around the eagle image.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Ms. Larraine Piestrak Regulatory Affairs and Compliance Manager Dade Behring, Inc. Chemistry/Immunochemistry Glasgow Business Community P.O. Box 6101 - Building 500 Newark, Delaware 19714
NOV - 7 2003
Re: k032697
Trade/Device Name: Dimension® Total Triiodothyronine Calibrator (RC414) Regulation Number: 21 CFR § 862.1150 Regulation Name: Calibrator Regulatory Class: II Product Code: JIS Dated: August 29, 2003 Received: September 2, 2003
Dear Ms. Piestrak:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
{2}------------------------------------------------
Page 2 -
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
iteven Sutman
Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### Indications For Use Statement
Device Name: Dimension® Total Triiodothyronine Calibrator (RC414)
# Indications for Use:
The Dimension® Total Triiodothyronine (T3) Calibrator (RC414) is intended for use in the calibration of the Total Triiodothyronine (T3) method on the Dimension® clinical chemistry system with the Heterogeneous Immunoassay Module.
Lorraine Piestrak
Regulatory Affairs and Compliance Manager
August 29, 2003
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
**Prescription Use**
(Per 21 CFR 801.109)
OR
Over-the-counter Use
Carol Benson / Jean Cooper, DVM
Division Sign-Off
\$10(k) k032697
00004
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.