Peer-reviewed clinical literature (ABACUS trial); Peer-reviewed clinical literature (Barnard et al. study)
The sponsor cites published clinical studies to demonstrate that the bolus calculator feature facilitates the optimization of glycemic control, reduces fear of hypoglycemia, and improves patient confidence in diabetes management in patients on multiple daily insulin injection therapy.
ABACUS trial; Clinical trial (referenced in literature)
Patients with suboptimal glycemic control on multiple daily insulin injection (MDI) therapy
Not applicable for this study
Glycemic control
Barnard et al. study; Clinical study (referenced in literature)
Patients with type 1 diabetes mellitus treated with multiple daily insulin injections
Not applicable for this study
Fear of hypoglycemia, confidence in dosage accuracy
Indications for Use
The ACCU-CHEK Aviva Expert System is indicated as an aid in the treatment of insulin-requiring diabetes. The ACCU-CHEK Aviva Expert System consists of the ACCU-CHEK Aviva Expert Meter, ACCU-CHEK Aviva Plus test strips, ACCU-CHEK Aviva control solutions, and ACCU-CHEK Bolus Advisor. The ACCU-CHEK Aviva Expert System is intended to facilitate the optimization of glycemic control in patients who are trained in multiple daily insulin injection therapy and are under the supervision of healthcare professionals experienced in managing insulin treated patients. The ACCU-CHEK Aviva Expert blood glucose monitoring system is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips. The ACCU-CHEK Aviva Expert blood glucose monitoring system is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes. The ACCU-CHEK Aviva Expert blood glucose monitoring system is intended to be used by a single person and should not be shared. The ACCU-CHEK Aviva Expert blood glucose monitoring system should not be used for the diagnosis or screening of diabetes or for neonatal use. Alternative site testing should NOT be used with the ACCU-CHEK Aviva Expert blood glucose monitoring system. The ACCU-CHEK Aviva Expert System is intended for prescription home use only. The ACCU-CHEK Aviva Expert meter is also indicated for the calculation of an insulin dose or carbohydrate intake based on user-entered data. The ACCU-CHEK Bolus Advisor, as a component of the ACCU-CHEK Aviva Expert meter, is intended for use in providing insulin dose recommendations in response to blood glucose, health events, and carbohydrate input. The ACCU-CHEK Bolus Advisor is intended to provide direction for insulin adjustment within the scope of a preplanned treatment program from a healthcare professional. Before its use, a physician or healthcare professional must prescribe the ACCU-CHEK Aviva Expert System and provide the patient-specific target blood glucose, insulin-to-carbohydrate ratio, and insulin sensitivity parameters to be programmed into the ACCU-CHEK Bolus Advisor. Once programmed, a patient must consult with his/her physician or healthcare professional before making any changes to these ACCU-CHEK Bolus Advisor settings.
Device Story
System measures capillary whole blood glucose via amperometric assay using mutant GDH-PQQ enzyme; provides quantitative glucose results. Includes optional Bolus Advisor for insulin dose/carbohydrate intake calculation. Used at home by patients with diabetes; requires physician-programmed parameters (target glucose, insulin-to-carbohydrate ratio, insulin sensitivity). Input includes measured glucose, user-entered carbohydrate intake, health events, time of day, and active insulin. Output provides insulin dose recommendations based on preplanned treatment program. Benefits include optimized glycemic control through automated dosing guidance. Meter is for single-patient use only.
Clinical Evidence
Performance testing cleared under k131366. Method comparison: 48 samples <75 mg/dL (100% within 10 mg/dL); 252 samples ≥75 mg/dL (99.2% within 20%). Precision: Repeatability SD ≤5.6 mg/dL (CV ≤2.9%); Reproducibility SD ≤5.1 mg/dL (CV ≤2.4%). Literature cited for clinical benefit: ABACUS trial (Ziegler et al.) and Barnard et al. study on hypoglycemia fear/confidence.
Technological Characteristics
Amperometric glucose monitoring system; uses mutant Q-GDH PQQ enzyme. Hematocrit range 10-65%; altitude up to 10,000 ft. Measuring range 20-600 mg/dL. Sample volume 0.6 µL. Test time 5 seconds. Connectivity via wired data transfer (no wireless). Code key calibration. Operating temp 14-38°C.
Indications for Use
Indicated for insulin-requiring diabetes patients on multiple daily insulin injection therapy. Not for neonates, critically ill patients, or patients using NPH/intermediate-acting insulin. Contraindicated for diagnosis/screening of diabetes.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
ACCU-CHEK Aviva Expert Blood Glucose Monitoring system (k131366)
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number:
k142089
B. Purpose for Submission:
Addition of insulin bolus calculator description information to labeling
C. Measurand:
Capillary whole blood glucose
D. Type of Test:
Quantitative, amperometric assay, glucose dehydrogenase (mutant GDH-PQQ)
E. Applicant:
Roche Diagnostics Corporation
F. Proprietary and Established Names:
ACCU-CHEK Aviva Expert Blood Glucose Monitoring System
G. Regulatory Information:
| Regulation | Name | Class | Product Code | Panel |
| --- | --- | --- | --- | --- |
| 21 § 862.1345 | Glucose dehydrogenase, glucose test system | II | LFR | (75) Chemistry |
| 21 § 868.1890 | Calculator, drug dose | II | NDC | General and Plastic Surgery |
H. Intended Use:
1. Intended use(s):
See indications for use below.
2. Indication(s) for use:
The ACCU-CHEK Aviva Expert System is indicated as an aid in the treatment of insulin-requiring diabetes. The ACCU-CHEK Aviva Expert System consists of the ACCU-CHEK Aviva Expert Meter, ACCU-CHEK Aviva Plus test strips, ACCU-CHEK Aviva control solutions, and ACCU-CHEK Bolus Advisor. The ACCU-CHEK Aviva Expert System is intended to facilitate the optimization of glycemic control in patients who are trained in multiple daily insulin injection therapy and are under the supervision of healthcare professionals experienced in managing insulin treated patients.
The ACCU-CHEK Aviva Expert blood glucose monitoring system is intended to be used
{1}
for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips. The ACCU-CHEK Aviva Expert blood glucose monitoring system is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes. The ACCU-CHEK Aviva Expert blood glucose monitoring system is intended to be used by a single person and should not be shared. The ACCU-CHEK Aviva Expert blood glucose monitoring system should not be used for the diagnosis or screening of diabetes or for neonatal use. Alternative site testing should NOT be used with the ACCU-CHEK Aviva Expert blood glucose monitoring system. The ACCU-CHEK Aviva Expert System is intended for prescription home use only.
The ACCU-CHEK Aviva Expert meter is also indicated for the calculation of an insulin dose or carbohydrate intake based on user-entered data. The ACCU-CHEK Bolus Advisor, as a component of the ACCU-CHEK Aviva Expert meter, is intended for use in providing insulin dose recommendations in response to blood glucose, health events, and carbohydrate input. The ACCU-CHEK Bolus Advisor is intended to provide direction for insulin adjustment within the scope of a preplanned treatment program from a healthcare professional. Before its use, a physician or healthcare professional must prescribe the ACCU-CHEK Aviva Expert System and provide the patient-specific target blood glucose, insulin-to-carbohydrate ratio, and insulin sensitivity parameters to be programmed into the ACCU-CHEK Bolus Advisor. Once programmed, a patient must consult with his/her physician or healthcare professional before making any changes to these ACCU-CHEK Bolus Advisor settings.
3. Special conditions for use statement(s):
- For in vitro diagnostic use only
- Capillary fingerstick blood only
- Not for testing on neonates
- Do not use for diagnosis or screening of diabetes mellitus
- Inaccurate results may occur in severe dehydration as a result of diabetic ketoacidosis or due to hyperglycemic non-ketotic syndrome, hypotension, shock, decompensated heart failure NYHA Class IV, or peripheral arterial occlusive disease.
- Not to be used for patients who are critically ill
- For single-patient use only
- Not to be used with patients taking Neutral Protamine Hagedorn (NPH) or any other intermediate-acting insulin.
- For prescription home use only
- Patient-specific target blood glucose, insulin-to-carbohydrate ratio, and insulin sensitivity parameters should be provided by a physician or healthcare professional.
- Consultation with a physician or healthcare professional should occur before making any changes to the ACCU-CHEK Bolus Advisor settings.
4. Special instrument requirements: ACCU-CHEK Aviva Expert Meter
{2}
3
I. Device Description:
The ACCU-CHEK Aviva Expert Blood Glucose Monitoring System consists of the following:
- ACCU-CHEK Aviva Expert meter
- ACCU-CHEK Bolus Advisor (a component of the Aviva Expert meter)
- ACCU-CHEK Aviva Plus test strips (k101299)
- ACCU-CHEK Aviva control solutions (k043474)
- ACCU-CHEK FastClix lancing device
The ACCU-CHEK Aviva Expert blood glucose monitoring system (cleared under k131366) is a blood glucose monitoring system that makes use of the ACCU-CHEK Aviva Expert meter, ACCU-CHEK Aviva Plus test strips and the ACCU-CHEK Aviva control solutions. The ACCU-CHEK Aviva Expert system provides the user with the ability to measure blood glucose levels when a sample of capillary blood is applied to the test strip. The meter also provides an optional insulin bolus calculator (the ACCU-CHEK Bolus Advisor) designed for use by individuals with diabetes who require insulin. The insulin bolus calculator algorithm is identical to the bolus calculator algorithm that was cleared for this system (k131366). This bolus calculator is meant to be used by patients with diabetes on multiple daily insulin injection (MDI) therapy. The insulin bolus calculation function is optional in that a user can simply obtain a blood glucose value through capillary blood testing and does not need to use the insulin bolus calculator portion of the system if it is not desired. In order to calculate the appropriate bolus of insulin, the ACCU-CHEK Bolus Advisor takes into account the measured blood glucose, the target blood glucose, the carbohydrate intake, the insulin-to-carbohydrate ratio, the insulin sensitivity, health events (such as exercise), the time of day, and the active insulin. Before using the ACCU-CHEK Aviva Expert system, a physician or healthcare professional must provide the patient-specific target blood glucose, insulin-to-carbohydrate ratio, and insulin sensitivity parameters.
J. Substantial Equivalence Information:
1. Predicate device name(s):
ACCU-CHEK Aviva Expert Blood Glucose Monitoring System
2. Predicate 510(k) number(s):
k131366
3. Comparison with predicate:
| Item | Device- ACCU-CHEK Aviva Expert System (k142089) | Predicate- Device- ACCU-CHEK Aviva Expert System (k131366) |
| --- | --- | --- |
| Indications for Use | Same | Quantitative measurement of glucose (sugar) in capillary blood from the finger tip, indicated for diabetes management by calculating an insulin dose |
{3}
| Item | Device- ACCU-CHEK Aviva Expert System (k142089) | Predicate- Device- ACCU-CHEK Aviva Expert System (k131366) |
| --- | --- | --- |
| | | or carbohydrate intake. |
| Insulin bolus calculator | Same | Insulin bolus calculator for insulin dosing calculations. |
| Test Strip | Same | Aviva Plus Test Strip |
| Test Principle | Same | Amperometric Detection |
| Enzyme | Same | Mutant Q-GDH PQQ |
| Hematocrit Range | Same | 10 to 65% |
| Maximum Altitude | Same | 10,000 feet |
| Measuring Range | Same | 20 – 600 mg/dL |
| Sample Volume | Same | 0.6 mcL |
| Test Time | Same | 5 seconds |
| Operating Temperature and Relative Humidity | Same | 14 to 38°C (57 to 100°F), 10-80% r.h. |
| Coding | Same | Code key insertion; lot-specific code key provided with each box of Aviva Plus test strips |
| Closed and Open Test Strip Shelf Life Stability | Same | 18 months |
| Control Solutions | Same | Aqueous, 2 levels, uses ACCU-CHEK Aviva Control Solutions |
| Differences | | |
| --- | --- | --- |
| Item | Device- ACCU-CHEK Aviva Expert System (k142089) | Predicate- Device- ACCU-CHEK Aviva Expert System (k131366) |
| Labeling | Addition of an insulin bolus calculator description to meter manuals | |
# K. Standard/Guidance Document Referenced (if applicable): N/A
# L. Test Principle:
When an ACCU-CHEK Aviva Plus test strip is inserted into the ACCU-CHEK Aviva Expert meter, a small alternating current (AC) is applied until the application of blood causes a spike in the conductivity to be observed at the measurement and sample sufficiency electrodes – both are used to assure an adequate sample has been applied. The instrument then applies a series of AC voltages at four frequencies and reads the AC responses. These carry information about the sample type and environmental temperature; they also allow the system to perform various internal quality checks. After the AC measures are completed, a
{4}
small (DC) voltage is applied and current is observed which is proportionate to the glucose. The AC and DC information are then combined to provide a hematocrit and temperature compensated glucose result.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
a. Precision/Reproducibility:
As established in k111353.
b. Linearity/assay reportable range:
As established in k111353.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
As established in k111353. Traceability is to NIST SRM 917 and was established under k101299. ACCU-CHEK Aviva Plus test strip stability was evaluated under k101299. Closed and open vial test strip stability is 18 months when stored at 36-86°F (2-30°C). ACCUCHEK Aviva control solutions stability and value assignment was evaluated under k043474. Closed vial stability at 36-90°F is 24 months from the date of manufacture. Open vial stability is 3 months from the date of opening.
d. Detection limit:
As established in k111353. The reportable range is 20-600 mg/dL.
e. Analytical specificity:
As established in k111353.
The sponsor has the following limitations in their labeling:
- Lipemic samples (triglycerides) in excess of 1800 mg/dL may produce elevated results.
- Blood concentrations of galactose >15 mg/dL will cause overestimation of blood glucose results.
- Intravenous administration of ascorbic acid which results in blood concentrations of ascorbic acid >3 mg/dL will cause overestimation of blood glucose results.
f. Assay cut-off:
Not applicable.
### 2. Comparison studies:
a. Method comparison with predicate device:
As established in k111353.
b. Matrix comparison:
Not applicable. The device is for fingerstick capillary blood only.
### 3. Clinical studies:
a. Clinical Sensitivity:
{5}
Not applicable.
b. Clinical specificity:
Not applicable.
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable.
4. Clinical cut-off:
Not applicable.
5. Expected values/Reference range:
The sponsor included the following expected values in the labeling:
The fasting glucose level for a non-diabetic adult is below 100 mg/dL.¹,²
Two hours after meals the normal blood glucose level for a non-diabetic is less than 140 mg/dL.¹
¹American Diabetes Association: Diagnosis and Classification of Diabetes Mellitus (Position Statement). Diabetes Care 34 (Supp. 1) S66, 2011.
²Tietz Fundamentals of Clinical Chemistry, 6ᵗʰ Edition, Edited by Burtis CA and Ashwood ED, W. B. Saunders Co., Philadelphia, PA, 2008, p. 849.
N. Instrument Name:
ACCU-CHECK Aviva Expert Meter
O. System Descriptions:
1. Modes of Operation:
Each test strip is single use and must be replaced with a new strip for additional readings. Before using the ACCU-CHEK Bolus Calculator on the ACCU-CHEK Aviva Expert system, a physician or healthcare professional must provide the patient-specific target blood glucose, insulin-to-carbohydrate ratio, and insulin sensitivity parameters. Labeling is provided for the Bolus Calculator functions.
Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device?
Yes ☐ X or No ☐
Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission?
Yes ☐ or No ☐ X
2. Software:
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
Yes ☐ X or No ☐
{6}
The applicant has provided documentation that indicates the device was designed and developed under good software life-cycle processes.
3. Specimen Identification:
There is no sample identification function with this device. Samples are applied directly to the test strip as they are collected.
4. Specimen Sampling and Handling:
The glucose test is intended to be used with capillary whole blood from the finger. The sample is applied directly to the test strip and testing is performed immediately. Sample storage is not required.
5. Calibration:
Each lot of test strips contains a code key specific to the lot and the key must be inserted into the meter prior to using a new lot of strips.
6. Quality Control:
Two levels of aqueous glucose control solutions are available with this system (Level 1, Level 2) but must be purchased separately. The meter automatically recognizes control materials as separate from patient samples and identifies them as controls on the display screen. Control results are not included in patient glucose averaging. Recommendations on when to test the control materials and troubleshooting steps if the controls are outside of specifications are provided in the labeling. An acceptable range for each control level is printed on the test strip vial label.
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
1. Hematocrit study: As established in k111353, the claimed hematocrit range is 10-65%.
2. Altitude study: As established in k111353, the claimed altitude is 10,000 ft.
3. Infection Control Study: As established in k111353. The ACCU-CHECK Aviva Expert Blood Glucose Monitoring System is for single patient use only and validation testing supported the use of Super Sani-Cloth (EPA reg. no. 9480-4) for 260 cleaning and disinfection cycles to simulate 5 years of single-patient meter use. Labeling was reviewed for adequate instructions for the validated cleaning and disinfection procedures.
4. Environmental testing for temperature and relative humidity: As established in k111353. Operating temperature and humidity ranges are from 57-100°F (14-38°C), 10-80% RH.
5. Sample volume requirements: As established in k111353. The minimum sample volume is 6 µL.
6. EMC and electrical safety: As established in k111353.
{7}
7. Usability study: As established in k111353.
8. Readability study: As established in k111353.
9. Customer support is available 24 hrs/day, 7 days per week by calling 1-800-688-4578.
**Q. Proposed Labeling:**
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
**R. Conclusion:**
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
8
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.