INOGEN SATELLITE CONSERVER, MODEL SC-100

K052563 · Medical Electronic Devices, Inc. · NFB · Oct 24, 2005 · Anesthesiology

Device Facts

Record IDK052563
Device NameINOGEN SATELLITE CONSERVER, MODEL SC-100
ApplicantMedical Electronic Devices, Inc.
Product CodeNFB · Anesthesiology
Decision DateOct 24, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5905
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Inogen Satellite Conserver is intended for use to conserve oxygen for patients prescribed supplemental oxygen and use nasal cannulas as part of an oxygen delivery system including an oxygen concentrator, compressed oxygen cylinders, or liquid oxygen system.

Device Story

Inogen Satellite Conserver is a battery-operated, microprocessor-controlled oxygen conserver. Device connects to oxygen supply sources (concentrators, cylinders, liquid oxygen) via supply tube. Breath sensor detects onset of inhalation; normally closed valve triggers delivery of oxygen bolus equivalent to 1-5 liters per minute constant flow. Used by patients prescribed supplemental oxygen. Device optimizes oxygen usage by delivering gas only during inhalation phase. Healthcare providers use device to manage patient oxygen therapy delivery. Benefits include extended duration of oxygen supply systems.

Clinical Evidence

No clinical data. Bench testing only. Performance testing conducted to verify device functionality with compressed oxygen cylinders and liquid oxygen systems. Previous extensive software, safety, mechanical, and EMI/EMC testing for the original device (K033197) remains applicable as hardware and software are unchanged.

Technological Characteristics

Battery-operated, microprocessor-controlled oxygen conserver. Features breath sensor and normally closed valve. Connectivity via oxygen supply tube to external oxygen sources. Identical to predicate (K033197) in electronic circuitry, control software, and enclosure. No hardware or software changes.

Indications for Use

Indicated for patients prescribed supplemental oxygen requiring nasal cannula delivery via oxygen concentrator, compressed oxygen cylinder, or liquid oxygen system.

Regulatory Classification

Identification

A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Medical Electronic Devices Corp. Image /page/0/Picture/1 description: The image contains a sequence of handwritten alphanumeric characters. The sequence starts with the letter 'K', followed by the number '0', then '5', '2', '5', '6', and ends with the number '3'. The characters are written in a dark, bold font, which makes them stand out against the white background. # OCT 2 4 2005 # 510(K) Summary ## Medical Electronic Devices Corporation ## Inogen Satellite Conserver SC-100 #### Submitter Information: Medical Electronic Devices Corporation 2807 Oregon Court, D6 Torrance, California 90503 | Submitter's Name: | Thomas Wenzel | |-------------------|----------------| | Phone: | (310) 618-0306 | | FAX: | (310)618-1034 | Date Submitted: 9/9/05 Device Name: Proprietary Name: Inogen Satellite Conserver Common Name: Oxygen Conserver Classification Name: Non Continuous Ventilator (IPPB) ## Predicate Device Equivalence: Substantial equivalence is claimed to the Inogen Satellite Conserver SC-100 (K033197), CHAD Therapeutics Oxymatic Model 411 (K003455), and Invacare Venture IDD Oxygen Conserving Device (K002284). ## Device Description: The Inogen Satellite Conserver is a battery operated electronic device that is microprocessor controlled and contains a breath sensor and normally closed valve. The device is connected to an oxygen supply source and upon detecting the beginning of an inhalation; the device delivers a bolus of oxygen that is equivalent in most users, depending on the flow setting, to 1 to 5 liters per minute constant flow. The Inogen Satellite Conserver can be used with bottled oxygen systems and as an accessory to the Inogen One Oxygen Concentrator (K032818). The device uses these systems as its oxygen supply and is connected by an oxygen supply tube. {1}------------------------------------------------ #### Indications For Use: The Inogen Satellite Conserver is intended for use to conserve oxygen for patients prescribed supplemental oxygen and use nasal cannulas as part of an oxygen delivery system including an oxygen concentrator, compressed oxygen cylinders, or liquid oxygen system. #### Comparison of Technological Characteristics: The Inogen Satellite Conserver has the same technological characteristics as the predicate device, i.e. Inogen Satellite Conserver (K033197). The electronic circuitry, control software, and device enclosure of the new device are identical to the Inogen Satellite Conserver (K033197). This 510(K) is a change in the Indications for Use only, adding compressed oxygen cylinders and liquid oxygen systems as additional sources of oxygen. #### Summary of Testing: Testing of the original Inogen Satellite Conserver contained extensive software validation testing, safety testing, mechanical testing, performance testing, and EMI/EMC testing. Since there was no change to the device's hardware or software these tests are applicable to the new device. Comprehensive performance testing with compressed oxygen cylinders and liquid oxygen systems was done to ensure that the device functioned within its specifications. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of an eagle with three lines representing its wings. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" is arranged in a circular fashion around the eagle symbol. , Public Health Service OCT 2 4 2005 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. Thomas Wenzel President Medical Electronic Devices, Incorporated 2807 Oregon Court, Unit D6 Torrance, California 90503 Re: K052563 Trade/Device Name: Inogen Satellite Conserver Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator (IPPB) Regulatory Class: II Product Code: NFB Dated: September 9, 2005 Received: September 19, 2005 Dear Mr. Wenzel: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ #### Page 2 - Mr. Wenzel Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. C Sincerely yours, Susan Turner Shin-Ling, Ph.D. Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use 510(k) Number (if known): Device Name: Inogen Satellite Conserver Indications For Use: Intended for use to conserve oxygen for patients prescribed supplemental oxygen and use nasal cannulas as part of an oxygen delivery system including an oxygen concentrator, compressed oxygen cylinder, or liquid oxygen system. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Devices Page 1 of ____________________________________________________________________________________________________________________________________________________________________ 510(k) Number:
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...