DEVILBISS PD1000

K020329 · Sunrise Medical Hhg, Inc. · NFB · Jun 19, 2002 · Anesthesiology

Device Facts

Record IDK020329
Device NameDEVILBISS PD1000
ApplicantSunrise Medical Hhg, Inc.
Product CodeNFB · Anesthesiology
Decision DateJun 19, 2002
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5905
Device ClassClass 2
AttributesTherapeutic

Intended Use

The DeVilbiss PD1000 Compact Conserver is intended as a delivery device for medial-grade oxygen from high-pressure oxygen cylinders. This is an ambulatory device, which allows patients to ambulate Joneer than they would with a continuous flow regulator on the same cylinder.

Device Story

The DeVilbiss PD1000 is an ambulatory oxygen conserving device designed to extend the use time of high-pressure oxygen cylinders. It attaches to cylinders via an integral pressure regulator (CGA 870 style). The device uses a pressure switch to detect the start of patient inhalation (sensitivity 0.1 to 0.25 cm H2O). Upon detection, a microprocessor-controlled valve releases a pulsed bolus of oxygen at the beginning of the inhalation cycle. It supports flow settings from 1.0 to 6.0 lpm. If batteries fail or the device malfunctions, a continuous flow back-up mode provides 2 lpm. LED indicators provide battery status and valve activation feedback. The device is intended for patient self-use during ambulation. By delivering oxygen only during inhalation, it reduces total oxygen consumption compared to continuous flow systems, allowing patients to remain mobile for longer periods.

Clinical Evidence

No clinical data provided. Substantial equivalence is based on non-clinical bench testing comparing pulse volumes, valve timing, and waveform characteristics against the predicate device.

Technological Characteristics

Ambulatory oxygen conserver; integral pressure regulator (CGA 870); pressure switch sensing (0.1-0.25 cm H2O); microprocessor-controlled valve; powered by 2 AA alkaline or NiMH batteries; LED status indicators; continuous flow back-up mode (2 lpm); fixed valve timing; no audible alarms.

Indications for Use

Indicated for patients requiring supplemental oxygen therapy from high-pressure cylinders; intended for ambulatory use to extend oxygen supply duration.

Regulatory Classification

Identification

A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K020329 Image /page/0/Picture/1 description: The image shows a logo for Sunrise. The logo consists of a stylized sun with horizontal lines representing the sun's rays. The word "SUNRISE" is written in bold, sans-serif letters below the sun symbol. The logo is black and white. JUN 1 9 2002 Sunrise Medical HHG, Inc. Respiratory Products Division 100 DeVilbiss Drive P.O. Box 635 Somerset, PA 15501 USA Tel: (814) 443-4881 Fax: (814) 443-7572 510(k) Summary Submitter's Name: Brian Hershey Address: Sunrise Medical HHG Inc. Respiratory Products Division 100 DeVilbiss Drive P.O. Box 635 Somerset, PA 15501-0635 Phone #: 814-443-7577 Fax #: 814-443-7571 Contact Person: Brian Hershey Date Summary Prepared: April 22, 2002 DeVilbiss PD1000 Name of Device: Oxygen Conserver Common Name: Classification Name: VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Legally Marketed Device to which Equivalence is Claimed: DeVilbiss EX2000 Pulse Dose Manufacturer: Sunrise Medical HHG, Inc. Respiratory Products Division Predicate Device 510(k) Number: K961126 ### Device Description: The DeVilbiss PD1000 compact conserving device is designed to extend the use time of oxygen cylinders. The device uses PulseDose technology to accomplish the oxygen savings. PulseDose senses the start of inhalation and instantly releases a short "pulsed" dose at the very beginning of the inhalation cycle. Since all of the "pulsed" oxygen finds its way deep into the lungs, less oxygen is required to accomplish the same effect than with traditional continuous flow oxygen systems. The PD1000 compact conserving device contains an integral pressure regulator with CGA 870 style yoke. The pressure regulator reduces the cylinder pressure to 19 - 25 psig. The PD1000 then delivers oxygen to the {1}------------------------------------------------ patient by sensing inhalation via a pressure switch and opening a two-way valve for a specified period of time controlled by a microprocessor. The unit is designed to deliver 16.5 cc/lpm of oxygen to the patient at 1.0. 1.5. 2.0, 2.5, 3.0, 4.0, 5.0 and 6.0 flow settings. The unit is powered by two "AA" alkaline or nickel metal hydride batteries. The PD1000 has a continuous flow back-up mode that will deliver 2 ipm continuous to the patient in the event of a device failure or dead batteries. The device has two LED's that indicate battery status and valve activation. When the unit is first turned on, either a red or green LED will flash to indicate the unit is turned on and the battery status. Each time the valve is activated during an inhalation, the red or green LED will flash to indicate the valve was activated and the battery status. The device will operate approximately 4 -- 8 hours when the red LED first begins to flash. When the battery voltage drops below approximately 1.9 vdc the red LED will remain on continuously and the valve will no longer operate. The continuous flow back-up mode can still be used to provide oxygen therapy to the patient when the batteries are no longer able to operate the unit. The ability of the PD1000 to detect patient inspirations is based on the sensitivity of the pressure switch which is set within a range of .1 to .25 cm H2O. The device is capable of delivering a bolus of oxygen at the beginning of each inhalation up to 40 breaths per minute (every 1.5 second minimum delay between breaths will cause the device to "skip" breaths at rates greater than 40 breaths per minute. ### Intended Use of the Device: The DeVilbiss PD1000 Compact Conserving Device is intended as a delivery device for medical-grade oxygen from high-pressure oxygen cylinders. This is an ambulatory device, which allows patients to ambulate longer than they would with a continuous flow regulator on the same cylinder. ### Technological Characteristics: The DeVilbiss PD1000 Compact Conserving Device has three key technological characteristic differences as compared to the predicate DeVilbiss EX2000 PulseDose Conserver. ### A. Timing ミ The PD1000 Compact Conserving Device has standard fixed valve times that are programmed into the microprocessor during assembly. This is different from the EX2000 PulseDose conserver in that the valve timing is adjusted during assembly through calibration of the pulse volumes at each dose setting. Although the valve open timing is set differently between the units, they provide comparable pulse volumes. There is no effect on safety and effectiveness. #### B. Breath Detection The PD1000 Compact Conserving Device utilizes a pressure switch to detect patient inhalation. This is different from the EX2000 PulseDose conserver which uses a pressure transducer to sense the patient inhalation. Both types of sensors provide adequate sensitivity to detect .1 to .25 cm of H2O vacuum. The pressure switch was selected for the PD1000 in order to reduce the amount of electronics required and gain a more robust sensor. There is no effect on safety and effectiveness. - C. Alarms The PD1000 Compact Conserving Device does not have audible alarms unlike the EX2000 PulseDose conserver. The PD1000 operates like the predicate mechanical conservers that are noted in the table of {2}------------------------------------------------ comparisons. The device is designed for ambulatory use, allowing patients to ambulate longer than they would with a continuous flow regulator on the same cylinder. There is no effect on safety and effectiveness #### Performance Data: The pulse volumes of the PD1000 were compared to the pulse volumes of the EX2000 predicate device. The pulse volume in cubic centimeters, valve "ON" time and waveform of the pulse volume or "bolus" was recorded for each setting of the PD1000. This data was also obtained for the same settings on the EX2000. The data indicates that the PD1000 pulse volumes match the corresponding pulse volumes of the EX2000. See appendix section D.1 for EX2000 / PD1000 Pulse volume comparisons. ## Clinical Performance Data: The DeVilbiss PD1000 Compact Conserving Device contains no new technologies, intended uses or new features that require clinical performance evaluations. #### Conclusions: The non-clinical test data concludes that the units are equivalent. {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle or bird with three heads in profile, suggesting a sense of unity or collaboration. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird image. The logo is black and white. ### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 # JUN 1 9 2002 Sunrise Medical HHG, Inc. c/o Mr. Brian Hershey Respiratory Products Division 100 DeVilbiss Drive P.O. Box 635 Somerset. PA 15501 Re: K020329 DeVilbiss PD1000 Regulation Number: 868.5905 Regulation Name: Ventilator, Non-continuous Regulatory Class: II (two) Product Code: NFB Dated: April 22, 2002 Received: April 24, 2002 Dear Mr. Hershey: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 - Mr. Brian Hershey Please be advised that FDA's issuance of a substantial equivalence determination does not mean t that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours. N. DeGorter Donna-Bea Tillman, Ph.D. Acting Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ Page I of 1 ### 510(k) Number (if known): K020329 Device Name: DeVilbiss PD1000 Indications For Use: The DeVilbiss PD1000 Compact Conserver is intended as a delivery device for medial-grade oxygen from high-pressure oxygen cylinders. This is an ambulatory device, which allows patients to ambulate Joneer than they would with a continuous flow regulator on the same cylinder. ## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Division of Cardiovascular & Respiratory Devices 510(k) Number K020329 (Optional Format 3-10-98) **Prescription Use** (Per 21 CFR 801.109)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...