The Respond OC Conserving Regulator (130-0800) is intended for prescription use only, to be used as part of a portable oxygen delivery system for patients that require supplemental oxygen in their home and for ambulatory use. It may also be used to deliver a constant flow of oxygen to the patient.
Device Story
Respond OC Conserving Regulator (130-0800) is a dual-mode, non-electric, pneumatic high-pressure oxygen regulator and conserving device. It attaches to ambulatory oxygen cylinders via a CGA 870 yoke. Input: high-pressure oxygen (300–3000 psig). Operation: internal regulator reduces pressure to 22 psig; sensing diaphragm detects patient's negative pressure during inhalation to trigger a pneumatic timing mechanism, releasing a bolus of oxygen via a standard nasal cannula. Device also provides a continuous flow mode (2, 3, 4 LPM). Used in home and ambulatory environments by patients. Output: pulsed oxygen bolus (conserve mode) or constant flow (regulator mode). Benefits: extends cylinder ambulatory time up to 6:1. Healthcare providers use the device to manage supplemental oxygen delivery; output affects patient mobility and oxygen therapy adherence.
Clinical Evidence
Bench testing only. Testing included biocompatibility (ISO 18562-2, ISO 18562-3), accelerated aging, flow comparison, flow regulation, conservation testing, promoted ignition burst pressure (ASTM G 175-24), environmental testing, altitude testing, flow response, and impact testing. Results met all pre-defined acceptance criteria and demonstrated performance similar to the predicate.
Technological Characteristics
Aluminum unibody construction with brass high-pressure components. CGA 870 yoke inlet. Pneumatic sensing diaphragm trigger. No electrical power source. Nominal outlet pressure 22 psi. Operates with 300–3000 psig input. Supports 1-5 LPM conserve mode and 2, 3, 4 LPM continuous flow. Ignition sensitivity tested per ASTM G 175-24.
Indications for Use
Indicated for patients requiring supplemental oxygen in home and ambulatory settings. No contraindications.
Regulatory Classification
Identification
A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 24, 2025
Responsive Respiratory
% Paul Dryden
President
ProMedic, LLC
131 Bay Point Dr. NE
St. Petersburg, Florida 33704
Re: K250322
Trade/Device Name: Respond OC Conserving Regulator (130-0800)
Regulation Number: 21 CFR 868.5905
Regulation Name: Noncontinuous Ventilator (IPPB)
Regulatory Class: Class II
Product Code: NFB
Dated: June 22, 2025
Received: June 23, 2025
Dear Paul Dryden:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250322 - Paul Dryden
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.
{2}
K250322 - Paul Dryden
Page 3
See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Bradley Q. Quinn -S
Bradley Quinn
Assistant Director
DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K250322 | |
| Device Name Respond OC Conserving Regulator (130-0800) | |
| Indications for Use (Describe) The Respond OC Conserving Regulator (130-0800) is intended for prescription use only, to be used as part of a portable oxygen delivery system for patients that require supplemental oxygen in their home and for ambulatory use. It may also be used to deliver a constant flow of oxygen to the patient. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K250322
# 510(k) Summary
Page 1 of 6
Date Prepared: 22-Jul-25
Sponsor: Responsive Respiratory, Inc.
6244 Lemay Ferry Rd.
Saint Louis, MO 63129
Steve Bannon, President
(866) 333-4030
Submission Correspondent: Paul Dryden - ProMedic, LLC
Proprietary or Trade Name: Respond OC Conserving Regulator (130-0800)
Classification Name: Conserver, Oxygen
Product Code: NFB
CFR: 21 CFR 868.5905
Proprietary or Trade Name: Chad Therapeutics Bonsai Model 800 – K070740
Classification Name: Conserver, Oxygen
Product Code: NFB
CFR: 21 CFR 868.5905
## Device Description
The Respond OC Conserving Regulator (130-0800) is a dual mode high pressure oxygen regulator and conserving device that allows for either a continuous or conserve flow of medical grade oxygen to the patient. The device is designed with a CGA 870 style yoke for use with ambulatory oxygen cylinders. The integrated regulator reduces cylinder pressure to 22psig (+/-3psi) to the sensing diaphragm which allows the Respond OC to sense the start of inhalation by the patient and release a controlled amount of oxygen by pneumatic timing in short bursts into the lungs via an oxygen cannula. The Respond OC supports a delivered oxygen equivalency of 1 to 5 LPM in the conserve mode, which extends the ambulatory time of cylinders up to 6:1 based on the cylinder pressure and selected setting. The continuous flow mode offers settings of 2, 3 and 4 LPM if the patient requires a constant flow of oxygen.
## Principle of Operation
The Respond OC Conserving Regulator (130-0800) is a dual mode device allowing for both continuous and conserve flows. Oxygen flow rates are adjusted by setting an external flow control knob to the prescribed rate. In continuous mode, the Respond OC functions as a standard continuous flow regular, similar to traditional click-style regulators. In conserve mode, the unit acts as a conserving device, which increases the ambulatory time over that of continuous flow systems. The flow of oxygen is controlled by the patient's inhalation, which triggers a sensing diaphragm within the control module to release a bolus of oxygen directly into the lungs at the beginning of the inhalation cycle.
## Indications for Use:
The Respond OC Conserving Regulator (130-0800) is intended for prescription use only, to be used as part of a portable oxygen delivery system for patients that require supplemental oxygen in their home and for ambulatory use. It may also be used to deliver a constant flow of oxygen to the patient.
## Patient Population:
Patients that require supplemental oxygen in their home and for ambulatory use.
{5}
K250322
510(k) Summary
Page 2 of 6
## Environments of use:
Home and ambulatory use.
## Substantial Equivalence Discussion
### Indications for Use –
The indications for use for the subject and predicate devices are similar. We have updated the Indications for Use language to mirror more recent submissions.
**Discussion** – The indications for use between the subject and predicate device are similar.
### Technology and construction –
The Respond OC Conserving Regulator (130-0800) is a combination high pressure oxygen regulator and conserving device. The unit operates similarly to other oxygen conservers currently on the market today wherein the device contains a traditional oxygen regulator coupled with a conserver. It is constructed with a CGA 870 yoke for use with high pressure oxygen cylinders and utilizes an aluminum unibody design with brass componentry in the high pressure areas. The device requires no electrical power source. Operating pressure range is 300 – 3000 psig with a nominal outlet pressure of 22 psi.
The device uses a single lumen cannula attached to a single outlet port that delivers oxygen to the patient. In regulator (steady flow) mode, the device features selectable flows of 1-5 LPM equivalence in 1 LPM increments. Flow rates are adjustable by a click-style flow control knob that allows adjustment between both steady flow or conserve flows without having to turn the device or flow of oxygen off. In the conserve mode, the flow of oxygen is triggered and controlled by the patient respiratory cycle. Upon inhalation, the sensing diaphragm within the control module is triggered to release a bolus of oxygen to the patient. No oxygen is delivered during exhalation.
The technology utilized within the device is substantially equivalent to the predicate device – Chad Therapeutics Bonsai. Both units are pneumatic devices designed for use with CGA 870 connections on high pressure oxygen cylinders and are combination devices offering conserve and continuous flow modes of oxygen delivery to the patient. Each device utilizes a click-style knob to select delivery modes and oxygen flow settings. In conserve mode, both devices utilize a sensing diaphragm to sense the inhalation/exhalation cycle and deliver a bolus of oxygen to the patient. Further, both units include similar materials of construction, input operating pressures, outlet pressures, inlet configuration, outlet connection, gauges, inhalation trigger points, oxygen conservation percentages and weight.
The functional differences between the units include body styles and the length of oxygen bolus delivery during inhalation. Since pulse is a fixed time and the breath rate is variable, the percent of the breath that the pulse uses is a function of the breath rate.
- At 15 BPM, the total breath is 4 seconds. With an assumed 1:1 I:E ratio, the inhale time is 2 seconds
- At 20 BPM, the total breath is 3 seconds. With an assumed 1:1 I:E ratio, the inhale time is 1.5 seconds
- At 30 BPM, the total breath is 2 seconds and the same assumed 1:1 I:E ratio, the inhale time decreases to 1 second.
- Using the above examples, the differential between the delivery and length of the bolus fluctuates between .4 second difference at 15 BPM, to less than .2 seconds at 30 BPM.
**Discussion** – The differences do not raise different questions of safety and effectiveness.
{6}
K250322
510(k) Summary
Page 3 of 6
## Environment of Use
The environment of use is identical.
**Discussion** – The environments of use are similar to the predicate – Chad Therapeutics Bonsai Model 800 – K070740.
## Patient Population
The patient population of the subject device and predicate is identical.
**Discussion** – The patient population is equivalent to the predicate device – Chad Therapeutics Bonsai Model 800 – K070740.
## Non-Clinical Testing Summary
Performance testing demonstrated that the subject device met its acceptance criteria. Testing included:
### Biocompatibility
- ISO 18562-2 – Emissions of Particulate Matter
- ISO 18562-3 – Emissions of VOCs with Toxicological Risk Assessment
### Bench testing
- Accelerated Aging
- Flow Comparison
- Flow Regulation Test
- Conservation Testing
- Promoted Ignition Burst Pressure Test
- Environmental Testing
- Altitude Test
- Flow Response Test
- Impact Test
### Bench testing
Performance testing demonstrated that the subject device performs similarly to the predicate device.
**Discussion** – The test results are similar to the predicate and within pre-defined acceptance criteria.
### Biocompatibility
**Discussion** – The subject device was found to meet the applicable requirements for biocompatibility safety for the intended population.
## Discussion of Differences
The subject device indications for use are similar to the predicate. We have removed the “modified” and “unmodified” language from the predicate device. Indications for Use. Also, we have reorganized the language to detail that the device also delivers a constant flow of oxygen, which is the same as the predicate but was not specified in its indications for use statement. We have also removed "i.e., for prescription use only."
The functional differences between the units include body styles and the length of oxygen bolus delivery during inhalation. Since pulse is a fixed time and the breath rate is variable, the percent of the breath that the pulse uses is a function of the breath rate.
- At 15 BPM, the total breath is 4 seconds. With an assumed 1:1 I:E ratio, the inhale time is 2 seconds
- At 20 BPM, the total breath is 3 seconds. With an assumed 1:1 I:E ratio, the inhale time is 1.5 seconds
{7}
K250322
510(k) Summary
Page 4 of 6
- At 30 BPM, the total breath is 2 seconds and the same assumed 1:1 I:E ratio, the inhale time decreases to 1 second.
- Using the above examples, the differential between the delivery and length of the bolus fluctuates between .4 second difference at 15 BPM, to less than .2 seconds at 30 BPM.
**Discussion** – The differences do not raise different questions of safety and effectiveness.
## Substantial Equivalence Conclusion
Through testing, the subject device has demonstrated it is substantially equivalent to the predicate device – Chad Therapeutics Bonsai Model 800 – K070740.
{8}
K250322
510(k) Summary
Page 5 of 6
| Comparative Characteristics | Subject Device Respond OC | Predicate Device Bonsai Model 800 K070740 | Similarities / Differences |
| --- | --- | --- | --- |
| Manufacturer | Responsive Respiratory, Inc. | Chad Therapeutics | |
| Prescription Device | Yes | Yes | Similar |
| Device classification and Product Code | NFB 868.5905 | NFB 868.5905 | Similar |
| Indications for Use | The Respond OC Conserving Regulator is intended for prescription use only, to be used as part of a portable oxygen delivery system for patients that require supplemental oxygen in their home and for ambulatory use. It may also be used to deliver a constant flow of oxygen to the patient. | The Chad Therapeutics Bonsai (modified Chad Therapeutics OXYPNEUMATIC CYPRESS Model 511) is intended for use in the same manner as the unmodified device, i.e., for prescription use only, to be used as part of a portable oxygen delivery system for patients that require supplemental oxygen in their home and for ambulatory use. | Similar
The subject device indications for use are similar to the predicate. We have removed the “modified” and “unmodified” language from the predicate device Indications for Use. Also, we have reorganized the language to detail that the device also delivers a constant flow of oxygen, which is the same as the predicate but was not specified in its indications for use statement. We have also removed "i.e., for prescription use only." |
| Contraindications | None | None | Similar |
| Patient Population | Patients that require supplemental oxygen in their home and for ambulatory use | Patients that require supplemental oxygen in their home and for ambulatory use | Similar |
| Environment of Use | Home and ambulatory use | Home and ambulatory use | Similar |
| Oxygen Supply | Compressed oxygen cylinder 300 to 3,000 psig | Compressed oxygen cylinder 500 to 3,000 psig | Similar |
| Cylinder Configuration | CGA 870 Valve | CGA 870 Valve | Similar |
| Technology | Delivers a pulsed dose of oxygen to patient by sensing inhalation utilizing a standard nasal cannula of 7ft or less in length. | Delivers a pulsed dose of oxygen to patient by sensing inhalation utilizing a standard nasal cannula of 7ft or less in length. | Similar |
| Pneumatic Design – Non Electric | Yes | Yes | Similar |
| Features | | | |
| Cannula Length | Uses standard single lumen cannula up to 7ft in length | Uses standard single lumen cannula up to 7ft in length | Similar
Note: nasal cannula is user supplied |
{9}
K250322
510(k) Summary
Page 6 of 6
| Comparative Characteristics | Subject Device Respond OC | | Predicate Device Bonsai Model 800 K070740 | | Similarities / Differences |
| --- | --- | --- | --- | --- | --- |
| Design | 2-in-1 design allows patients to switch from conservation mode to standard continuous regulator mode. | | Delivers a uniform oxygen pulse at increments. Can be switched to continuous mode. | | Similar |
| Trigger Method | Inspiratory Effort, Negative Pressure | | Inspiratory Effort, Negative Pressure | | Similar |
| Conserve Mode | Senses a breath and delivers in first 1/3 of breath cycle | | Senses a breath/delivers oxygen in the first half of the inspiratory cycle | | Similar
Both devices deliver oxygen within a similar time period in the early stages of the inhalation cycle |
| Selectable Outlet Flow | Continuous – 2, 3, 4 lpm | | Continuous – 2 and 4 lpm | | Similar |
| | Conserve – 1-5 lpm | | Conserve – 1-6 lpm | | Similar |
| Oxygen Bolus Size (ml)
At 20 BPM | Conserve 1 lpm | 11 ml | Conserve 1 lpm | 11 ml | Similar |
| | Conserve 2 lpm | 24 ml | Conserve 2 lpm | 29 ml | |
| | Conserve 3 lpm | 37 ml | Conserve 3 lpm | 41 ml | |
| | Conserve 4 lpm | 46 ml | Conserve 4 lpm | 50 ml | |
| | Conserve 5 lpm | 54 ml | Conserve 5 lpm | 56 ml | |
| | | | Conserve 6 lpm | 58 ML | |
| Outlet Configuration | Cannula Hose Barb | | Cannula Hose Barb | | Similar |
| Regulator Outlet Pressure | 22 PSI | | 25 PSI | | Similar |
| Breaths Per Minute (bpm) | Up to 35 | | 14 to 40 | | Similar |
| Savings Ration | 6:1 | | 6:1 | | Similar |
| Ignition Sensitivity and Fault Tolerance | Passed ASTM G 175-24 | | Passed ASTM G 175-03 | | Similar |
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.