OXY-SERVE II OXYGEN CONSERVING REGULATOR

K051988 · Superior Products, Inc. · NFB · Sep 21, 2005 · Anesthesiology

Device Facts

Record IDK051988
Device NameOXY-SERVE II OXYGEN CONSERVING REGULATOR
ApplicantSuperior Products, Inc.
Product CodeNFB · Anesthesiology
Decision DateSep 21, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5905
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Oxy-Serve II Oxygen Conserving Regulator is intended to be used in home, respiratory, pulmonary, and skilled care facilities by ambulatory patients who have been diagnosed as requiring and have been prescribed oxygen administration by a physician. The Conserver is used in conjunction with portable oxygen cylinders to extend the usable cylinder life by delivering oxygen only during the inhalation portion of the patients breathing cycle.

Device Story

Oxy-Serve II is a pneumatic oxygen conserving pressure regulator; integrates with portable oxygen cylinders. Input: patient inhalation detected via sensing diaphragm through dual lumen cannula. Operation: delivers oxygen only during inhalation phase; flow ceases during exhalation. Features indexable control knob (1-6 LPM, 0.5 LPM increments); switchable to continuous flow mode (bypasses sensing diaphragm). Used by ambulatory patients in home or clinical settings. Benefits: extends useful life of oxygen cylinders by conserving gas during exhalation.

Clinical Evidence

Bench testing only. Performance, mechanical, and environmental testing conducted to ensure device meets intended performance criteria.

Technological Characteristics

Aluminum construction with brass high-pressure chamber components. Pneumatic operation; no electrical power required. Dual lumen cannula interface. Indexable control knob for 11 flow rates (1-6 LPM). Features sensing diaphragm for inhalation-triggered flow and manual switch for continuous flow mode.

Indications for Use

Indicated for ambulatory patients diagnosed as requiring prescribed oxygen administration. Used in home, respiratory, pulmonary, and skilled care facilities.

Regulatory Classification

Identification

A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ SEP 2 1 2005 K051988 Image /page/0/Picture/2 description: The image shows the logo for Superior Products, Inc. The logo includes a diamond shape with the letters "S" and "P" inside. The text below the logo includes the company name, address, and phone and fax numbers. The address is 3786 Ridge Road Cleveland, Ohio 44144, and the phone number is (216) 651-9400. ## 510(k) SUMMARY - Submitter Information: 1. Superior Products, Inc. 3786 Ridge Rd. Cleveland, Ohio 44144 | Contact Person: | Ronald Johnston | |-----------------|-----------------| | Phone | 216-651-9400 | | Fax | 216-651-4071 | - June 1, 2005 2. Date Prepared: - 3. Name of Device: | Trade Name: | Oxy-Serve II Oxygen Conserving Regulator | |----------------------|---------------------------------------------------| | Common Name: | Oxygen Conserver | | Classification Name: | Non-continuous ventilator (IPPB), 21 CFR 868.5905 | | Regulatory Class: | II (two) | | Product Code: | NFB | #### 4. Predicate Device Information: 1. K010747 - OPC Oxygen Conserving Regulator - Western Medica, Corp. #### 5. Device Description: The Oxy-Serve II Oxygen Conserving Regulator is a high pressure oxygen regulator integrated with a conserving device to deliver oxygen during the inhalation phase of a patients breathing cycle, thereby extending the useful life of an oxygen cylinder. The Oxy-Serve II is constructed of aluminum with brass high pressure chamber components. The device is pneumatic and requires no electrical power source. An indexable control knob selects from 11 flow rates ranging from 1 to 6 LPM in .5 LPM increments. Two outlet ports are connectable using a dual lumen cannula. One port senses the inhalation of the patient thus initiating the selected flow to the patient during the full inhalation cycle. Flow ceases at the end of the inhalation cycle. The second port delivers the oxygen. The device can operate as a conserving regulator, or as a standard continuous flow regulator operable by sliding a switch on the face of the regulator housing. In continuous mode the sensing diaphragm control component is bypassed and a continuous flow of oxygen is delivered at the flow selected by the indexable control knob through the outlet port. {1}------------------------------------------------ ### Indications for Use: 6. The Oxy-Serve II Oxygen Conserving Regulator is intended as an oxygen conserving pressure regulator for ambulatory patients. The Conserving Regulator delivers a prescribed flow of oxygen only during the inhalation cycle, conserving oxygen during the exhalation cycle. ### 7. Intended Use: The Oxy-Serve II Oxygen Conserving Regulator is intended to be used in home, respiratory, pulmonary, and skilled care facilities by ambulatory patients who have been diagnosed as requiring and have been prescribed oxygen administration by a physician. The Conserver is used in conjunction with portable oxygen cylinders to extend the usable cylinder life by delivering oxygen only during the inhalation portion of the patients breathing cycle. #### 8. Substantial Equivalence: The Oxy-Serve II Oxygen Conserving Regulator is substantially equivalent to K010747 -OPC Oxygen Conserving Regulator - Western Medica. The device has the same technological characteristics (i.e., materials, energy source, mode of operation) as the predicate device and functions similarly through the use of a pressure regulator and flow sensing diaphragms to control the delivery of the oxygen. #### ರಿ. Performance Testing: The Oxy-Serve II Oxygen Conserving Regulator has been subjected to performance, mechanical, and environmental testing to insure the device meets the performance criteria as intended. #### 10. Conclusions: The Oxy-Serve II Oxygen Conserving Regulator, having the same intended use as the predicate device, and having been proven through lab testing to have similar performance characteristics, as well as passing extensive safety testing, is substantially equivalent to the legally marketed predicate device and is safe and effective for its intended use. {2}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features the department's name encircling a stylized caduceus symbol. The caduceus is a staff with two snakes coiled around it, often associated with medicine and healthcare. The seal is simple and monochromatic. Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 SEP 2 1 2005 Mr. Ronald Johnson Director of Engineering Superior Products, Incorporated 3786 Ridge Road Cleveland, Ohio 44144 Re: K051988 Trade/Device Name: Oxy-Serve II Oxygen Conserving Regulator Regulation Number: 21 CFR 868.5905 Regulation Name: Non-Continuous Ventilator (IPPB) Regulatory Class: II Product Code: NFB Dated: September 12, 2005 Received: September 13, 2005 Dear Mr. Johnston: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket the Fouchar F USA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include controls for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting (1 wr 1) it is a may be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 – Mr. Johnston Please be advised that FDA's issuance of a substantial equivalence determination does not Please be advised that I DA s issualited or a bour device complies with other requirements mean that IDA has made a decemand regulations administered by other Federal agencies. of the Act of ally rederal bate and rightenents, including, but not limited to: registration You must comply with an the Her Pier 801); good manufacturing practice allo listing (21 CFR Part 807), abolity systems (QS) regulation (21 CFR Part 820); and if requirents as set form in the quality ajones of the moved provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will anow you to objail made of substantial equivalence of your device to a premits in the confication. The PDF results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), If you desire specific aurres for your at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division Inay other general meetiment on one Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Chris Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use 510(k) Number (if known): OS 1988 Device Name: _________________________________________________________________________________________________________________________________________________________________ Indications For Use: The Oxy-Serve II Oxygen Conserving Regulator is intended as an oxygen conserving The Oxy-Selve if Oxygon Ochserving Regulation Regulator delivers a pressure "regulator" for "announcery" parents" since "conserving oxygen during the exhalation cycle. Prescription Use × (Per 21 CFR 801 Subpart D) AND/OR Over-The-Counter-Use __ (21 CFR 807 Subpart C) (Please DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Clur (Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Devices 510(k) Number.*_*_ Page 1 of 1
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...