RX EPIDURAL NEEDLE

K053318 · Epimed International, Inc. · BSP · Feb 17, 2006 · Anesthesiology

Device Facts

Record IDK053318
Device NameRX EPIDURAL NEEDLE
ApplicantEpimed International, Inc.
Product CodeBSP · Anesthesiology
Decision DateFeb 17, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5150
Device ClassClass 2
AttributesTherapeutic

Intended Use

The RX Epidural Needle is intended for the epidural placement, directly or through an epidural catheter, of anesthetic agents to elicit regional anesthesia.

Device Story

RX Epidural Needle is a sterile, single-use, disposable device consisting of a stainless steel cannula with a ground beveled distal tip and a molded plastic hub. A stylet with a custom-trimmed shaft and plastic hub is included. Used by clinicians for epidural placement of anesthetic agents to elicit regional anesthesia. Device facilitates delivery of anesthesia directly or through an epidural catheter.

Clinical Evidence

No clinical data; bench testing only.

Technological Characteristics

Stainless steel cannula with ground beveled distal tip; molded plastic hub; includes stylet; sterile, single-use, disposable; available in various lengths and gauges.

Indications for Use

Indicated for the epidural placement of anesthetic agents to elicit regional anesthesia, either directly or via an epidural catheter.

Regulatory Classification

Identification

An anesthesia conduction needle is a device used to inject local anesthetics into a patient to provide regional anesthesia.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # Section II ## 510(K) Summary ## Company Information: Epimed International, Inc. 141 Sal Landrio Drive Johnstown, NY 12095 (518) 725-0209 Contact: Christopher B. Lake Manager of QA/RA ### Date Prepared: October 31, 2005 ### Trade Name: RX Epidural Needle #### Common Name: Epidural Needle ## Product Class/Classification: Class II 868.5150 Needle, Conduction, Anesthetic ## Predicate Device(s): Epimed International Epidural Needle (K030562) #### Description: The RX Epidural Needle consists of a stainless steel cannula with a ground beveled distal tip. A plastic hub is molded onto the proximal end of the cannula. A stylet is also provided with the device which consists of a custom trimmed shaft and a molded plastic hub. The RX Epidural Needle will be provided as a sterile, single use, disposable device. The RX Epidural Needle will be available in a variety of lengths and gauges. {1}------------------------------------------------ #### Intended Use: The RX Epidural Needle is intended for the epidural placement, directly or through an epidural catheter, of anesthetic agents to elicit regional anesthesia. #### Comparison to Predicate: The RX Epidural Needle has similar physical and technical characteristics to the Epimed International Epidural Needle marketed under K030562. ## Conclusion: The comparison to the predicate device demonstrates that the RX Epidural Needle is safe and effective and is substantially equivalent to the predicate device. Epimed International, Inc. Signature Christopher B. Lake Manager of Quality Assurance/Regulatory Affairs {2}------------------------------------------------ Image /page/2/Picture/1 description: The image is a black and white seal for the U.S. Department of Health and Human Services. The seal features the department's name in a circular arrangement around a symbol. The symbol consists of a stylized caduceus, which is a staff with two snakes entwined around it, often used as a symbol of medicine and healthcare. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 FEB 1 7 2006 Mr. Christopher Lake Manager of Quality Assurance and Regulatory Affairs Epimed International, Inc. 141 Sal Landrio Dr. Crossroads Business Park Johnstown, NY 12095 Re: K053318 Trade/Device Name: RX Epidural Needle Regulation Number: 868.5150 Regulation Name: Anethesia conduction needle Regulatory Class: II Product Code: BSP Dated: November 28, 2005 Received: December 6, 2005 Dear Mr. Lake: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ #### Page 2 -- Mr. Lake Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Susan Tunne for Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use 510(k) Number (if known): unk. KOS 3318 Device Name: Rx Epidural Needle Indications For Use: The RX Epidural Needle is intended for the epidural placement, directly or through an epidural catheter, of anesthetic agents to elicit regional anesthesia. Prescription Use __ × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Cee Sulion ology General Hospital, Control. Dental Devices K093318 Page 1 of ____________________________________________________________________________________________________________________________________________________________________
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...