SonoBlock; SonoBlock II

K241954 · PAJUNK GmbH Medizintechnologie · BSP · Sep 11, 2024 · Anesthesiology

Device Facts

Record IDK241954
Device NameSonoBlock; SonoBlock II
ApplicantPAJUNK GmbH Medizintechnologie
Product CodeBSP · Anesthesiology
Decision DateSep 11, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5150
Device ClassClass 2
AttributesTherapeutic

Intended Use

SonoBlock and SonoBlock II Needles equipped with Cornerstone reflectors are used to puncture the tissue in order to gam entry and inject local anaesthetics to induce regional anaesthesia. Warning: SonoBlock and SonoBlock II Needles are not intended for RF ablation or any other type of ablation procedure.

Device Story

SonoBlock and SonoBlock II are single-use anesthesia conduction needles used for peripheral nerve blocks under ultrasound guidance. Needles feature Cornerstone reflectors to enhance ultrasound visibility. Device is used by clinicians in clinical settings. Operation involves manual tissue puncture and injection of local anesthetics. SonoBlock II needles utilize direct injection molding for hub-to-needle bonding, while SonoBlock needles may use gluing. Devices are supplied sterile (EtO) and are compatible with NRFit (ISO 80369-6) or Luer (ISO 80369-7) connectors. Output is the physical delivery of anesthetic to the target nerve site, guided by ultrasound visualization of the needle tip. Benefits include improved needle localization and precise anesthetic delivery for regional anesthesia.

Clinical Evidence

Bench testing only. Performance verified via ISO 9626 (bending rigidity, breakage), ISO 7864 (hub-to-needle bonding force), and ISO 80369-series (leakage, stress cracking, axial load). Sterilization validated per ISO 11135 (SAL 10^-6). Biocompatibility confirmed per ISO 10993-1. Shelf life validated for 5 years.

Technological Characteristics

Stainless steel (1.4301) tubing; hub materials include ABS or Copolyester (Tritan MX731); Parylene (NanoLine) and silicone coating; Cornerstone ultrasound reflectors; NRFit or Luer connectors. Sterilization via Ethylene Oxide. Manufacturing includes direct injection molding or gluing. Complies with ISO 9626, ISO 7864, ISO 80369-series, ISO 10993-series, and ISO 11135-series.

Indications for Use

Indicated for patients requiring regional anesthesia via peripheral nerve blocks. Not for RF ablation or other ablation procedures.

Regulatory Classification

Identification

An anesthesia conduction needle is a device used to inject local anesthetics into a patient to provide regional anesthesia.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. # September 11, 2024 PAJUNK GmbH Medizintechnologie Chiara Meyer Specialist Regulatory Affairs Karl-Hall-Str. 1 Pajunkstr. 1 Geisingen, BW 78187 Germany # Re: K241954 Trade/Device Name: SonoBlock; SonoBlock II Regulation Number: 21 CFR 868.5150 Regulation Name: Anesthesia Conduction Needle Regulatory Class: Class II Product Code: BSP Dated: July 3, 2024 Received: July 3, 2024 # Dear Chiara Meyer: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). {1}------------------------------------------------ Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. {2}------------------------------------------------ See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, **Bradley Q. Quinn -S** Bradley Quinn Assistant Director DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K241954 Device Name SonoBlock and SonoBlock II Indications for Use (Describe) SonoBlock and SonoBlock II Needles equipped with Cornerstone reflectors are used to puncture the tissue in order to gam entry and inject local anaesthetics to induce regional anaesthesia. Warning: SonoBlock and SonoBlock II Needles are not intended for RF ablation or any other type of ablation procedure. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| | <span style="font-size:100%;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> | |-------------------------------------------------------------------------------------| | <span style="font-size:100%;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the word "PAJUNK" in a sans-serif font. The letters are a teal color, and there is a registered trademark symbol after the "K". The background is white. # 510_k # 510(k) Summary # Registrations USA 510(k) Summary as required by 21 CFR 807.92(c). # Date of Preparation: 2024-09-06 # Document Control Number: K241954 | 510(k) owner: | PAJUNK® GmbH Medizintechnologie<br>Karl-Hall-Str. 1/ Pajunkstr. 1<br>78187 Geisingen<br>Baden-Wuerttemberg, Germany<br>Fon: +49(0)7704-9291-586<br>Fax: +49(0)7704-9291-605<br>Establishment Registration Number:<br>9611612 | |----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Submitter Information/ production site: | PAJUNK® GmbH Medizintechnologie<br>Karl-Hall-Str. 1/ Pajunkstr. 1<br>78187 Geisingen<br>Baden-Wuerttemberg, Germany<br>Fon: +49(0)7704-9291-586<br>Fax: +49(0)7704-9291-605<br>Establishment Registration Number:<br>9611612 | | Manufacturer/ Submitter Contact: | Christian G. H. Quass<br>Director Regulatory Affairs, Safety Officer<br>Fon: +49(0)7704-9291-586<br>Fax: +49(0)7704-9291-602 | | | E-Mail: christian.quass@pajunk.com | | Alternative Manufacturer/ Submitter Contact: | Chiara Meyer<br>Regulatory Affairs Specialist<br>Fon: +49(0)7704-9291-627<br>Fax: +49(0)7704-9291-602<br>E-Mail: chiara.meyer@pajunk.com | | USA Contact: | PAJUNK MEDICAL SYSTEMS<br>4575 Marconi Dr.<br>Alpharetta, GA 30005<br>Registration Number: 3004076349<br>FEI Number*: 3004076349 | {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the word "PAJUNK" in a teal color. The letters are large and bold, and they appear to be a sans-serif font. There is a registered trademark symbol to the right of the letter K. | US-Official Correspondent | Marco Wohnig | |---------------------------|-----------------------------------------------| | | President | | | PAJUNK Medical Systems | | | 4575 Marconi Dr. | | | Alpharetta, GA 30005 | | | Phone: 001-770-4936832 | | | E-Mail: Marco.Wohnig@pajunk-usa.com | | Contract Sterilizer: | Sterigenics Germany GmbH | | | Kasteler straße 45 | | | 65203 Wiesbaden | | | Germany, Hessen | | | Establishment Registration Number: 3002807090 | | 2nd Contract Sterilizer: | HA2 MEDIZINTECHNIK GMBH | | | Am Bahndamm 11 | | | Halberstadt Saxony-Anhalt, DE 38820 | | | Registration Number: 3009039068 | | | FEI Number*: 3009039068 | # Device Name and Classification | Device Name: | SonoBlock, SonoBlock II | |------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Premarket Notification Number: | K241954 | | Classification Name: | Anesthesia Conduction Needle | | Classification Reference: | 21 CFR § 868.5150 | | Product Code: | BSP | | Establishment Registration Number: | 9611612 | | Regulatory Class: | II | | Panel: | Anesthesiology | | Sterilization Method: | Ethylene Oxide<br>disposable device, supplied sterile to the end user and non-<br>sterile intended to be sterilized prior to use to re-packagers/<br>medical device manufacturers | {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the word "PAJUNK" in a sans-serif font. The letters are a teal color. There is a registered trademark symbol to the right of the K. | Contract Sterilizer: | Sterigenics Germany GmbH<br>Kasteler straße 45<br>65203 Wiesbaden<br>Germany, Hessen<br>Establishment Registration Number: 3002807090 | |---------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------| | Additional Contract Sterilizer: | HA2 MEDIZINTECHNIK GMBH<br>Am Bahndamm 11<br>Halberstadt Saxony-Anhalt, DE 38820<br>Registration Number: 3009039068<br>FEI Number*: 3009039068 | # Changes to Device: Predicate Device | Device Name: | SONO-SERIES CORNERSTONE TECHNIQUE | |------------------------------------|-----------------------------------| | Premarket Notification Number: | K111374 | | Trade Name | SonoPlex, SonoPlex STIM, SonoLong | | Classification Name: | SonoPlex, SonoPlex STIM, SonoLong | | Classification Reference: | 21 CFR § 868.5150 | | Product Code: | BSP | | Establishment Registration Number: | 9611612 | {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the word "PAJUNK" in a sans-serif font. The word is in a teal color. There is a registered trademark symbol to the right of the K. | Panel: | Anesthesiology | |-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Sterilization Method: | Ethylene Oxide<br>disposable device, supplied sterile to the end user and<br>non-sterile intended to be sterilized prior to use to re-<br>packagers/ medical device manufacturers | | Contract Sterilizer: | Sterigenics Germany GmbH<br>Kasteler straße 45<br>65203 Wiesbaden<br>Germany, Hessen<br>Establishment Registration Number: 3002807090 | The subject device, the SonoBlock II nerve block needle is a single-use anaesthesia conducting nerve block needle intended for performing peripheral nerve blocks. The SonoBlock and SonoBlock II nerve block needle is intended for use under ultrasound guidance. Visibility under ultrasound is enhanced by CornerStone reflectors as cleared by FDA in several Premarket Notification Submissions. The predicate device, the SonoPlex II nerve block needle is a single-use device intended for use peripheral nerve blocks. The SonoPlex and SonoPlex II nerve block needle is intended for use under ultrasound guidance. Visibility under ultrasound is enhanced by CornerStone reflectors. Additionally an electrical stimulus may be applied to the needle via a cable and connector to assist the physician pinpoint the area of application (also referred to as Dual Guidance). K111374 in 2011 has been filed and cleared for ultrasound enhanced Parylene coated needles (Cornerstones) optionally equipped with an electrical cable for optional application of an electrical stimulus. S these needles are optionally used for Dual Guidance: Ultrasound guidance and verification via electrical stimulus (while landmark technique without ultrasound and electrical stimulus is also widespread). Since 2011 some alterations have taken place after clearance: - The SonoPlex STIM needles (Ultrasound and stimulation) have been renamed to SonoPlex - Needles formerly named SonoPlex (Ultrasound only) have been renamed to SonoBlock - Hub-to-needle technology formerly cleared as gluing (2K and UV) has been optionally amended by direct injection moulding - For injection moulding the material of the hub has been changed - By direct overmoulding, the tube and where applicable the cable are directly moulded to the hub and one screw connection is substituted by this manufacturing step. Each of these alterations followed FDA´s Guidance "Deciding When to Submit a 510(k) for a Change to an {8}------------------------------------------------ 510(k) Summary Existing Device". None of these alterations did result in additional hazards to the indications for use and the clinical technique have not been impacted. Each of the alterations has been diligently verified and validated during design process and has been accompanied by risk management measures during manufacturing. None of the alterations did impact the risk status or the labelling. Safety and effectiveness as well as efficacy of the device remain unaltered. The SonoBlock and SonoBlock II has been added to K111374 SonoPlex II as an Add to file following FDA´s Guidance "Deciding When to Submit a 510(k) for a Change to an Existing Device". Both devices share the indications for use (Peripheral nerve block) and only differ by offering options for localization techniques. The SonoBlock II-series is directly injection moulded instead of glued (as the predicate is). This change does neither raise additional risk in performance nor new additional biocompatibility issues. Since this modification/ additional device option does neither state a significant change nor does it pose any new risk to the patient this has been documented as an Add-to-file. This Premarket Notification is intended to file a standalone 510(k) for the SonoBlock II needles and to inform the Agency about additional options to be added to the device´s manufacturing process. Subject to this 510(k) is - . the validation of an additional optional sterilization service provider (HA2 Medizintechnik, Halberstadt, Establishment Registration number 3009039068) - the validation of an additional optional packaging material (medical paper). Both, new sterilization service provider and additional packaging material, have been added to the Design History file and the Device master record in order to be able to package and sterilize the device using additional processes as an additional option. Both devices, the predicate SonoPlex II as well as the subject SonoBlock and SonoBlock II are manufactured by PAJUNK GmbH Medizintechnologie, the sponsor of this submission, in Geisingen, Germany. Both devices, the predicate SonoPlex II as well as the subject SonoBlock/ SonoBlock II share the same indications for use. The safety and effectiveness in localization options is substantially equivalent. Apart from stimulation option both devices, the predicate device as well as the subject device, share identical required specifications for device performance as laid down in ISO 9626, ISO 7864, ISO 80369-series, ISO 10993-series and ISO 11135-series. Both devices, the predicate device as well as the subject device are manufactured in the same regulatory framework, e. g. Quality Management System (compliant with 21cfr820 and ISO 13485) using the same risk management system according to ISO 14971. The intended use as well as the basic technical specification of the needles which is relevant to clinical use is identical to the predicate devices and has been cleared in 510(k)s sent in earlier by the sponsor. The clinical technique, the indications for use, the technical specification, the materials used, the sterility status (validation and sterility assurance level) as well as the biocompatibility status is identical. None of the performance criteria is adversely affected by the device modification as it is subject to this submission. {9}------------------------------------------------ Image /page/9/Picture/0 description: The image shows the word "PAJUNK" in a bold, sans-serif font. The letters are a teal color. A registered trademark symbol is located to the right of the "K". This section of the submission contains the following information: # Contents | 1 Determination of Substantial Equivalence | 7 | |--------------------------------------------|----| | 2 Technical Description | 7 | | 3 Substantial Equivalence Discussion | 10 | | 4 Performance Testing | 11 | | 5 Sterilization | 16 | | 6 Shelf Life | 16 | | 7 Biocompatibility | 16 | | 8 Technology Characteristics | 17 | | 9 Conclusion | 17 | {10}------------------------------------------------ Image /page/10/Picture/1 description: The image shows the word "PAJUNK" in a sans-serif font. The color of the text is a shade of teal. A horizontal line is present in the negative space of the letter "P". A registered trademark symbol is present to the right of the letter "K". # 510(k) Summary # 1 Determination of Substantial Equivalence ### Intended Use Predicate Device (510(K) Predicate) Needles/ cannulas equipped with Cornerstone reflectors as covered by the PAJUNK® Sonoseries SonoPlex STIM are used to puncture the tissue in order to gain entry and inject local anaesthetics to induce regional anaesthesia. An electrical stimulus may be applied to the needle via a cable and connector to assist the physician pinpoint the area of application. #### Warning: The PAJUNK GmbH Medizintechnologie needles of the Sono-series are not intended for RF ablation or any other type of ablation procedure. #### Intended Use Subject Device SonoBlock and SonoBlock II Needles equipped with Cornerstone reflectors are used to puncture the tissue in order to gain entry and inject local anaesthetics to induce regional anaesthesia. Warning: SonoBlock and SonoBlock II Needles are not intended for RF ablation or any other type of ablation procedure. #### Discussion The Intended Use of the Predicate device and of the subject device is substantially equivalent. The Subject device offers localization via ultrasound as it is state of the art and well established. The Predicate Device additionally offers electrical stimulus for localization (Dual Guidance). Both methods are state of the art and substantially equivalent. #### Conclusion: substantially equivalent # 2 Technical Description #### Substantial Equivalence: Tabulatory Outline {11}------------------------------------------------ Image /page/11/Picture/0 description: The image shows the word "PAJUNK" in a sans-serif font. The letters are a teal color. There is a registered trademark symbol to the right of the K. | Characteristics | Subject Device<br>SonoBlock II<br>Manufacturer:<br>PAJUNK® GmbH<br>Medizintechnologie,<br>Geisingen | Subject Device<br>SonoBlock<br>Manufacturer:<br>PAJUNK® GmbH<br>Medizintechnologie,<br>Geisingen | Predicate Device<br>K111374<br>SonoPlex II<br>Manufacturer:<br>PAJUNK® GmbH<br>Medizintechnologie,<br>Geisingen | Predicate Device<br>K111374<br>SonoPlex<br>Manufacturer:<br>PAJUNK® GmbH<br>Medizintechnologie,<br>Geisingen | |----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Picture | | | | | | Needle Tubing | Stainless Steel<br>1.4301 | Stainless Steel<br>1.4301 | Stainless Steel<br>1.4301 | Stainless Steel 1.4301 | | Tip | Tuohy, Quincke,<br>Chiba, Facette | Tuohy, Quincke,<br>Chiba, Facette | Tuohy, Quincke,<br>Chiba, Facette | Tuohy, Quincke,<br>Chiba, Facette | | Hub | Lustran/Guardian<br>ABS 348 white<br>012002 (No Patient<br>Contact) | Copolyester (Tritan<br>MX731)<br>(Indirect Patient<br>Contact) | Lustran/Guardian<br>ABS 348 white<br>012002 (No Patient<br>Contact) | Lustran/Guardian ABS<br>348 white 012002 (No<br>Patient Contact) | | Coating | Parylene<br>(NanoLine®),<br>Silicone (Dow<br>Corning 360 Medical<br>Fluid 12500 cSt),<br>Silicone thinner<br>(Dow Corning Q7-<br>9180 Silicone Fluid<br>1.0 cSt) | Parylene<br>(NanoLine®), Silicone<br>(Dow Corning 360<br>Medical Fluid 12500<br>cSt), Silicone thinner<br>(Dow Corning Q7-<br>9180 Silicone Fluid<br>1.0 cSt) | Parylene<br>(NanoLine®), Silicone<br>(Dow Corning 360<br>Medical Fluid 12500<br>cSt), Silicone thinner<br>(Dow Corning Q7-<br>9180 Silicone Fluid<br>1.0 cSt) | Parylene (NanoLine®),<br>Silicone (Dow Corning<br>360 Medical Fluid<br>12500 cSt), Silicone<br>thinner (Dow Corning<br>Q7-9180 Silicone<br>Fluid 1.0 cSt) | | US-<br>enhancement | Cornerstone | Cornerstone | Cornerstone | Cornerstone | | Graduation | Tampapur TPU | Tampapur TPU | Tampapur TPU | Tampapur TPU | | Bonding<br>Technology<br>Needle-to-Hub | Directly molded | glued | Directly molded | glued | | Characteristics | Subject Device<br>SonoBlock II<br>Manufacturer:<br>PAJUNK® GmbH<br>Medizintechnologie,<br>Geisingen | Subject Device<br>SonoBlock<br>Manufacturer:<br>PAJUNK® GmbH<br>Medizintechnologie,<br>Geisingen | Predicate Device<br>K111374<br>SonoPlex II<br>Manufacturer:<br>PAJUNK® GmbH<br>Medizintechnologie,<br>Geisingen | Predicate Device<br>K11374<br>SonoPlex<br>Manufacturer:<br>PAJUNK® GmbH<br>Medizintechnologie,<br>Geisingen | | Glue | n.a. | (Epoxy adhesive -<br>Araldite 2011),<br>alternatively: UV<br>adhesive (Loctite AA<br>3921) | n.a. | (Epoxy adhesive -<br>Araldite 2011),<br>alternatively: UV<br>adhesive (Loctite AA<br>3921) | | Injection tube | PVC (MED7536), PC<br>(HP4) | PVC (MED7536), PC<br>(HP4), ABS (Terlux<br>2802) | PVC (MED7536), PC<br>(HP4) | PVC (MED7536), PC<br>(HP4), ABS (Terlux<br>2802) | | Cable | n.a. | n.a. | PVC, Copper tin-<br>plated, brass tin-<br>plated, ABS<br>(Guardian 348 -<br>RAL9003) or Brass<br>tin-plated, Copper<br>tin-plated, cable<br>(tinned wire), ABS<br>PA-757 RAL 9003 /<br>RAL1018<br>2mm plug | PVC, Copper tin-<br>plated, brass tin-<br>plated, ABS (Guardian<br>348 - RAL9003) or<br>Brass tin-plated,<br>Copper tin-plated,<br>cable (tinned wire),<br>ABS PA-757 RAL 9003<br>/ RAL1018<br>2mm plug | | Diameter | 20G - 25G | 20G - 25G | 20G - 25G | 20G - 25G | | Length | 40mm - 120 mm | 40mm - 120 mm | 40mm - 120 mm | 40mm - 120 mm | | Connectivity | ISO 80369-6 (NRFit)<br>ISO 80369-7 (LUER) | ISO 80369-6 (NRFit)<br>ISO 80369-7 (LUER) | ISO 80369-6 (NRFit)<br>ISO 80369-7 (LUER) | ISO 80369-6 (NRFit)<br>ISO 80369-7 (LUER) | | Packaging | Tyvek, Foil, chromo-<br>duplex board<br>Alternatively: Paper,<br>Foil | Tyvek, Foil, chromo-<br>duplex board<br>Alternatively: Paper,<br>Foil | Tyvek, Foil, chromo-<br>duplex board | Tyvek, Foil, chromo-<br>duplex board | | Characteristics | Subject Device<br>SonoBlock II<br>Manufacturer:<br>PAJUNK® GmbH<br>Medizintechnologie,<br>Geisingen | Subject Device<br>SonoBlock<br>Manufacturer:<br>PAJUNK® GmbH<br>Medizintechnologie,<br>Geisingen | Predicate Device<br>K111374<br>SonoPlex II<br>Manufacturer:<br>PAJUNK® GmbH<br>Medizintechnologie,<br>Geisingen | Predicate Device<br>K111374<br>SonoPlex<br>Manufacturer:<br>PAJUNK® GmbH<br>Medizintechnologie,<br>Geisingen | | Biocompatibility | Complies with ISO10993-series as required by « Use of International Standard ISO 10993-1,<br>"Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk<br>management process" », Docket Number : FDA-2013-D-0350 | | | | | Sterilization<br>Method | Ethyleneoxide | Ethyleneoxide | Ethyleneoxide | Ethyleneoxide | | Sterility<br>Assurance Level | SAL=10-6 | SAL=10-6 | SAL=10-6 | SAL=10-6 | | Sterilization<br>Service Provider | Sterigenics,<br>Wiesbaden<br>HA2 Medizintechnik,<br>Halberstadt | Sterigenics,<br>Wiesbaden<br>HA2 Medizintechnik,<br>Halberstadt | Sterigenics,<br>Wiesbaden | Sterigenics,<br>Wiesbaden | | Shelf Life | 60 month from<br>sterilization | 60 month from<br>sterilization | 60 month from<br>sterilization | 60 month from<br>sterilization | {12}------------------------------------------------ Image /page/12/Picture/0 description: The image features the word "PAJUNK" in a bold, sans-serif font. The color of the text is a shade of teal. To the left of the word "PAJUNK" is a horizontal line with a pointed end, resembling a stylized needle or arrow. A circled "R" symbol is located to the upper right of the letter "K", indicating that the brand name is registered. {13}------------------------------------------------ Image /page/13/Picture/0 description: The image shows the word "PAJUNK" in a sans-serif font. The color of the text is a shade of green. There is a registered trademark symbol to the right of the letter K. # 3 Substantial Equivalence Discussion The Predicate devices, the SonoPlex II needles, are equipped with a cable for additionally pinpointing the needles position via an electrical stimulus called dual guidance (ultrasound and electrical stimulation). This is why a cable is attached to the hub in addition. The SonoBlock II needles are equipped with ultrasound enhancing cornerstones without any cable since there is no stimulation. Both, ultrasound position guidance as well as ultrasound position guidance and electrical stimulation are state of the art. Both techniques are safe and effective. Both, the subject device as well as the predicate device can be either directly injection moulded or glued. Both manufacturing technologies are safe and the processes are running stable. Both technologies are standard technologies. The predicate devices, the SonoPlex needles, are packed with Tyvek and foil and sterilized with Ethyleneoxide at Sterigenics, Germany. The subject devices, the SonoBlock needles, are packed with Tyvek and foil as well as optionally with Medical {14}------------------------------------------------ Image /page/14/Picture/1 description: The image shows the word "PAJUNK" in a sans-serif font. The letters are a teal color. There is a registered trademark symbol to the right of the letter K. Paper and foil. Furthermore, the subject devices are sterilized at Sterigenics in Wiesbaden as well as optionally at HA2 Medizintechnik in Halberstadt. The main differences which are subject to this premarket submission are - the addition of a sterilization facility and a) - the addition of an alternative packaging material. b) In order to verify substantial equivalence of the sterilization process a sterilization according ISO11135 has been performed and successfully accomplished. The sterility assurance level of 10-6 has been successfully validated at both facilities: Sterigenics and HA2. In order to verify substantial equivalence of the packaging and shelf life tests have been performed. Both, medical paper and Tyvek have been validated and verified to allow sterilization using EtO and to maintain sterility after a shelf life of 60 month after sterilization. # 4 Performance Testing The needles subject to this premarket submission are subject to performance testing prior to release as well as after shelf life scenario of 5 years. There is no impact of the change subject to this submission (additional sterilization provider, additional optional packaging) on the performance of the needle as defined in the user requirements specification. Especially the hub to needle bonding requirements remain unaffected. The processes are still valid and reliable. The test listed below are standard tests performed on a regular basis as well as type tests performed after design transfer. Tests are performed to comply with the international standards listed below in the respectively most current version: ISO 9626 Stainless steel needle tubing for manufacture of medical devices ISO 7864 Sterile hypodermic needles for single use - Requirements and test methods ISO 80369-6 Small bore connectors for liquids and gases in healthcare applications - Part 6: Connectors for neuraxial applications ISO 80369-7 Small-bore connectors for liguids and gases in healthcare applications - Part 7: Connectors for intravascular or hypodermic applications ISO 80369-20 Small-bore connectors for liquids and gases in healthcare applications - Part 20: Common test methods ISO 9626 defines the quality and properties of the stainless steel tubing remains unaltered and therefore ongoing compliance with this standard remains unaffected by the change. There is no additional/ new risk of breakage or bending. ISO 80369-series defines the type of connector which also is not affected by the change. The technical equivalency of LUER and NRFit has been verified and cleared in several submissions before and therefore is in {15}------------------------------------------------ compliance. The test reports related to this standard series also do not directly contribute to demonstration of substantial equivalence in this submission. There is no additional risk in this regard. #### ISO 7864 defines properties of the needle. Those properties are not directly linked to the material. Section 4.12 of this standard is directly linked to the stability of the connection of the needle and the risk of detachment oft he needle from the hub. The acceptance criteria have to be met non regarding the material oft he hub and the needle or the type of connection technology. So the test reports for hub-to-needle bondage are required to demonstrate substantial equivalence. The test reports attached to this e-mail subjecting worst case needles demonstrate compliance with the standard (69N required, an average of 104,9N and 196,01N met in result) as well as process stability when manufacturing the needles either with PC Tritan or ABS, either glued or directly moulded. ## Needle: stability test bending rigidity Reason for test: The needle has to demonstrate bending stability and resistance against breakage in order to resist forces reasonably assumed to be applied to the needle in situ under the defined intended use Procedure of test: The test procedure is defined by international standard EN ISO 9626: Stainless steel needle tubing for manufacture of medical devices. Acceptance Criteria for bending rigidity/ deflection: | Item code<br>SonoBlock II | Max deflection [mm] | Item code SonoBlock II<br>NRFit® | Max deflection [mm] | |---------------------------|---------------------|----------------------------------|---------------------| | 001280-71 | 0,52 | 001260-74 | 0,52 | | 001280-74 | 0,52 | 001260-89 | 0,52 | | 001280-77 | 0,38 | 001260-71 | 0,52 | | 001280-95 | 0,48 | 001260-77 | 0,38 | | 001281-71 | 0,52 | | | | 001281-74 | 0,52 | | | | 001281-77 | 0,38 | | | | 001281-72 | 0,48 | | | | 001281-95 | 0,48 | | | {16}------------------------------------------------ Image /page/16/Picture/0 description: The image shows the word "PAJUNK" in a teal sans-serif font. The "A" in "PAJUNK" is stylized with a horizontal line through the middle, resembling a needle. A small registered trademark symbol is located to the upper right of the "K". | Item code<br>SonoBlock | Max deflection [mm] | Item code<br>SonoBlock | Max deflection [mm] | |------------------------|---------------------|------------------------|---------------------| | 001180-77 | 0,38 | 001180-31G | 0,48 | | 001180-74 | 0,52 | | | | 001180-71 | 0,52 | | | | 001180-81 | 0,37 | | | | 001181-77 | 0,38 | | | | 001181-74 | 0,52 | | | Results: The bending rigidity of the subject needles is in compliance with the standard. #### Test successfully passed #### Needle: Breakage Reason for test: The needle has to demonstrate stability in order to resist forces reasonably assumed to be applied to the needle in situ under the defined intended use. Procedure of test: The test procedure is defined by international standard EN ISO 9626: Stainless steel needle tubing for manufacture of medical devices Acceptance criteria: no breakage during visual inspection after 20 complete cycles of reversal of force at a rate of 0,5 Hz #### Test successfully passed #### Needle: stability test bonding to hub Reason for test: The needle has to demonstrate stability at the bonding of the hub in order to resist forces reasonably assumed to be applied to the needle in situ under the defined intended use. Procedure of test: The test procedure is defined by international standard EN ISO 7864: Sterile hypodermic needles for single use Acceptance Criteria for bonding to hub: {17}------------------------------------------------ Image /page/17/Picture/0 description: The image shows the word "PAJUNK" in a sans-serif font. The letters are a teal color. There is a registered trademark symbol to the right of the K. | Item code<br>SonoBlock II | Force min [N] | Item code SonoBlock II<br>NRFit® | Force min [N] | |---------------------------|---------------|----------------------------------|---------------| | 001280-71 | 40 | 001260-74 | 40 | | 001280-74 | 40 | 001260-89 | 40 | | 001280-77 | 44 | 001260-71 | 40 | | 001280-95 | 54 | 001260-77 | 44 | | 001281-71 | 40 | | | | 001281-74 | 40 | | | | 001281-77 | 44 | | | | 001281-72 | 54 | | | | 001281-95 | 54 | | | | Item code<br>SonoBlock | Force min [N] | Item code<br>SonoBlock | Force min [N] | |------------------------|---------------|------------------------|---------------| | 001180-77 | 44 | 001180-31G | 40 | | 001180-74 | 40 | | | | 001180-71 | 40 | | | | 001180-81 | 22 | | | | 001181-77 | 44 | | | | 001181-74 | 40 | | | {18}------------------------------------------------ Results: The bending rigidity of the subject needles is in compliance with the standard. #### Test successfully passed Needle: Penetration force Reason for test: The needles have to demonstrate less trauma when applied with the patient under the intended use. ISO 7864 Sterile hypodermic needles for single use recommends penetration force testing only without giving a normative test method. Procedure of test: According to international European standard EN 13097. Pass/ Fail criteria: -none- objective comparison only. Results: The subject needles comply with defined penetration/ insertion forces. ### Test successfully passed Hub/ Connection: Performance tests defined in ISO 80369-20 for compliance with ISO 80369-7 Reason for test: The hub of the tubing has to demonstrate tightness and stability in order to resist forces reasonably assumed to be applied in situ under the defined intended use. Procedure of tests: The test procedure is defined by international standard ISO 80369-20 and specific requirements and acceptance criteria are defined in the substandards ISO 80369-6 (for NRFit) and ISO 80369-7 (for LUER) Tests conducted: 80369-6_6.1_FluidLeakage 80369-6 6.2 AirLeakage 80369-6_6.3_StressCracking 80369-6_6.4_SeparationAxialLoad 80369-6_6.5_Unscrewing 80369-6 6.6 Overriding 80369-7_7.1_FluidLeakage 80369-7_7.2_AirLeakage 80369-7 7.3 StressCracking 80369-7_7.4_SeparationAxialLoad 80369-7_7.5_Unscrewing 80369-7_7.6_Overriding Results: The subject needles respectively the hubs of the tubes which are part of the needles comply with defined requirements of the standard. ## Test successfully passed {19}------------------------------------------------ Image /page/19/Picture/0 description: The image shows the word "PAJUNK" in a sans-serif font. The color of the text is a shade of teal. A registered trademark symbol is located to the right of the letter "K". # 5 Sterilization The contract sterilizer and the sterilizing process are identical to the process and sterilizer used for all PAUNK® - manufactured devices which are already cleared for market or exempt. #### Sterilization parameters are: | SAL | 10-6 | |-------------|-----------------------| | Type of gas | Ethylene Oxide 99,99% | Sterilization has been validated according to ISO 11135-1 Overkill Approach (1 sublethal cycle, 1 full cycle) Residuals of EO and ECH are in compliance with ISO 10993-7. Cleaning and Sterilization method, which ensures an SAL of 10-6 as well as compliance with limits for chemical burden, bioburden, pyroburden (i.e. LAL) and EtO-residuals as well as shelf life have been validated and are safe and effective. The limits listed below are met by each device: Limits for Residuals: 25ppm = 25μg/(g/device) of Ehyleneoxide (EO); 25ppm = 25μg/(g/device) Ethylene chlorhydrine # 6 Shelf Life Efficacy of sterile product´s lifecycle has been validated using process most challenging worst case devices. Sterility tests have been performed using process most challenging worst case devices with similar characteristics made from the same material after 5 years. The devices were found to be sterile after 5 years, the sterile barrier system is efficient. Performance of the device (LUER connection, stability of bonding connections, needle´s bending rigidity) has been tested. There is no decrease in performance after 5 years. Shelf-life is set to 5 years. # 7 Biocompatibility All products comply with ISO 10993-1 and with FDA´s guidance Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". Therefore, based on sterilization and residuals validation the kits also are considered to be biocompatible. {20}------------------------------------------------ Image /page/20/Picture/0 description: The image shows the word "PAJUNK" in a sans-serif font. The letters are a teal color. There is a registered trademark symbol to the right of the letter K. # 8 Technology Characteristics Both, the subject device and the predicate device, consist of identical components. # 9 Conclusion The comparison between the predicate device and the subject device of this submission focussed on the validated sterilization process and the shelf life testing demonstrates that the subject device can be optionally in addition packed using medical paper and can be optionally sterilized at the alternative sterilization service provider at an SAL of 10-6. Based on current in-process testing regularly conducted, safe performance of the SonoBlock II needle is demonstrated.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...