INOVO, INC. ACCUPULSE SINGLE LUMEN CONSERVING REGULATOR
Applicant
Inovo, Inc.
Product Code
NFB · Anesthesiology
Decision Date
Jul 18, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5905
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Inovo AccuPulse Single Lumen Pneumatic Conserving Regulator is used to deliver a prescribed flow of medical-grade oxygen to the patient while conserving gas from a high-pressure oxygen cylinder, by sensing the patient inhalation cycle and supplying gas only during that phase of breathing.
Device Story
AccuPulse Conserving Regulator is a high-pressure oxygen regulator and conserving device; compact unit attaches to oxygen cylinders via CGA 870 style yoke. Device senses patient inhalation via pressure-sensitive diaphragm (0.10 to 0.35 cm H2O); triggers sliding valve to release oxygen pulse during inspiration. Pneumatic timing controls pulse duration; delivers 12 cc/lpm at flows from 1 to 6 LPM. Protects diaphragm by isolating it from oxygen flow via sliding valve mechanism. Includes continuous flow mode for device failure. Operates at breathing rates up to 40 BPM; skips breaths at higher rates. Used in clinical or home settings to extend oxygen cylinder duration; reduces oxygen waste compared to continuous-flow regulators. Healthcare providers prescribe flow settings; patients use device to receive oxygen therapy.
Clinical Evidence
No clinical data. Substantial equivalence established via bench testing including ASTM G175 promoted ignition, hydrostatic testing, proof pressure testing, conservation testing, flow regulation testing, and environmental testing.
Technological Characteristics
Compact aluminum body with brass high-pressure zones. Pneumatic sensing/actuation via pressure-sensitive diaphragm and sliding valve. CGA 870 style yoke inlet. Pressure reduction to 22 PSIG. Mechanical/pneumatic operation; no software or electronics. ASTM G175 compliant.
Indications for Use
Indicated for patients requiring supplemental medical-grade oxygen therapy delivered from high-pressure cylinders. Device conserves oxygen by sensing inhalation and delivering gas only during the inspiratory phase.
Regulatory Classification
Identification
A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.
{0}------------------------------------------------
## 510(K) SUMMARY
This summary of 510(k) safety and effectiveness is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR Part 807.92.
1060751 The assigned 510(k) number is:
#### 1. Submitter's Identification:
Inovo, Inc. 2975 S. Horseshoe Drive Naples, Fl. 34104
Date Summary Prepared: March 10, 2006
Contact Person: Mr. Michael T. Dildine Quality Assurance Manager
#### 2. Name of the Device:
Inovo, Inc. AccuPulse Conserving Requlator
#### 3. Class and Product Code
Class II , Product Code NFB , Anesthesiology Panel
#### 4. Predicate Device Information:
- 1. Precision Medical Inc. Easy Pulse 5 (K032018)
- 2. Chad Oxypneumatic Cypress Model 511 (K020475)
- 3. Inovo Economizer (K031983)
#### 5. Device Description:
The Inovo, Inc. AccuPulse Conserving Regulator is a high pressure oxygen regulator and conserving device that is combined into a single compact unit. It is designed to extend the use time of oxygen cylinders. The AccuPulse senses the start of inhalation and immediately releases a short "pulse" of oxygen to the patient. Since all of the "pulse" of oxygen finds it way deep into the lungs, less oxygen is required to accomplish the same effect as traditional continuous-flow oxygen requlators.
The AccuPulse Conserving Regulator contains an integral regulator with a CGA 870 style yoke. The regulator portion reduced the pressure to about 22PSIG. The AccuPulse then delivers oxygen to the patient by sensing the beginning of inhalation using a diaphragm, which opens a sliding valve for a specific period of time controlled by pneumatic timing. The unit is designed to deliver 12 cc/lpm of oxygen to the patient at flows from 1 LPM to 6 LPM.
{1}------------------------------------------------
The AccuPulse also has a continuous mode that delivers selectable continuous flow in the event of device failure.
Patient inspiration is detected by a pressure-sensitive diaphragm set by the manufacturer in the range of 0.10 to 0.35 cm H2O. The device is capable of delivering a bolus of oxygen at the beginning of the patient's inhalation at breathing rates up to 40 breaths per minute (BPM). Patients with more rapid breathing rates than 40 BPM will cause the device to "skip" breaths.
#### Intended Use: 6.
The AccuPulse Conserving Requlator is used to deliver a prescribed flow of medical grade oxygen to the patient while conserving gas from a highpressure cylinder, by sensing the patient inhalation cycle and supplying gas only during that phase of breathing.
#### 7. Comparison to Predicate Devices:
The Inovo, Inc. AccuPulse Conserving Regulator is substantially equivalent to the Precision Medical Inc. Easy Pulse 5 (K032018), the Chad Oxyoneumatic Cypress Model 511 (K020475) and the Inovo Economizer (K031983). These predicate devices were cleared with the same indications for use as our device.
Similarities include features of the device, input operating pressures, outlet pressure, selectable outlet flows, oxygen conservation, inlet configuration, outlet connection (except the Economizer, K031983, which has a dual-lumen outlet connection), gauge, and trigger point. A major feature is a compact aluminum body with all brass in high pressure zones.
Similarities include features of the device attached as Exhibit #2.
There is one technological characteristic difference between the AccuPulse and the predicate devices:
## Timing
The AccuPulse Conserving Regulator has a sliding valve which connects the outflow of oxygen to the outlet port. Upon patient inhalation, sliding valve closes the passage between the sensing diaphragm and the outlet port and then opens a passage to allow oxygen to the patient. The valve does not allow any oxygen to contact the sensing diaphragm, thereby protecting the diaphragm. In addition, the valve does not allow the unit to pulse a second time until a defined interval has passed.
There is no effect on safety and effectiveness.
{2}------------------------------------------------
### Discussion of Non-Clinical Test Performed for Determination of Substantial 8. Equivalence are as follows:
Performance Testing Included:
- Promoted Ignition ASTM G175 .
- Hydrostatic Test .
- Proof Pressure Test .
- Conservation Test .
- Flow Regulation Test .
- Environmental Testing .
#### ல் Discussion of Clinical Test Performed
Not Applicable
#### 10. Conclusions:
The subject device, the Inovo, Inc. AccuPulse Conserving Regulator has the same intended use as the predicate devices, the Inovo Economizer (K031983), Precision Medical (K032018) and the Chad (K020475). Moreover, bench testing contained in our submission and non-clinical testing supplied demonstrates that there are no differences in their performance characteristics, thereby not raising any new questions of safety and effectiveness. Thus, the Inovo, Inc. AccuPulse Conserving Regulator is substantially equivalent to the predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 1 8 2006
Mr. Michael T. Dildine Quality Assurance Manager Inovo, Incorporated 2975 S. Horseshoe Drive, Suite 600 Naples, Florida 34104
Re: K060751
Trade/Device Name: Inovo, Inc. AccuPulse Single Lumen Pneumatic Conserving Regulator Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator (IPPB) Regulatory Class: II Product Code: NFB Dated: July 12, 2006 Received: July 12, 2006
Dear Mr. Dildine:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Dildine
Please be advised that FDA's issuance of a substantial equivalence determination does not Please be advised that FDA \$ Issualics or a subscriptives with other requirements
mean that FDA has made a determination that your devices with other Federal agencies mean that FDA has made a decemination that your and ministered by other Federal agencies.
of the Act or any Federal statutes and regulations administers byt not limited to: r of the Act of any rederal statues and reguirements, including, but not limited to: registration
You must comply with all the Act strements, and monufacturing practice You must comply with an the Act 3 requirements and 801); good manufacturing practice
and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing (21 CFR Par and listing (21 CFR Fall 807), labeling (21 CFR Pat 820); of CFR Part 820); and if
requirements as set forth in the quality systems (QS) regulation (21 CFR Partisers 531, 542 requirements as set form in the quality systems (Sections 53) -542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 5 10(k)
the success of the same of the same in a consisted in a I his letter will anow you to begin mailer.ing your and equivalence of your device to a premarket notification. The PDA initing of substantler of a consisted on the spermits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). A If you desire specific advice for your de roo on one of the regulation of the regulation please contact inc Office or Comphanes and (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the may outain other general mionnational and Consumer Assistance at its toll-free Dribber (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Chiu Lin, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Device Name: Inovo, Inc. AccuPulse Single Lumen Pneumatic Conserving Regulator
Indication For Use:
The Inovo AccuPulse Single Lumen Pneumatic Conserving Regulator is used to deliver a prescribed flow of medical-grade oxygen to the patient while conserving gas from a high-pressure oxygen cylinder, by sensing the patient inhalation cycle and supplying gas only during that phase of breathing.
Prescription Use X (Per 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
W Mak fn AAb
Vision Sign-Off) I sion of Anesthesiology, General Hospital. Control, Dental Dental Devi
Page 1 of 1
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .