K233316 · Terrats Medical SL · NHA · Jan 5, 2024 · Dental
Device Facts
Record ID
K233316
Device Name
DESS Dental Smart Solutions
Applicant
Terrats Medical SL
Product Code
NHA · Dental
Decision Date
Jan 5, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K233316 · Jan 5, 2024
DESS Dental Smart Solutions
Terrats Medical SL
Published clinical literature (Woods, Delfino, and Rajan, 2021)
The document uses a published literature review to evaluate the subject devices in the MR environment, specifically addressing magnetically induced displacement force and torque.
MR safety; Literature review; Magnetic resonance environment
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices; Literature review / Scientific rationale; Follow-up/Duration: Not applicable; Study Period: Not applicable
Not applicable; Sample Size: Not applicable; Number of Sites: Not applicable
Not applicable for this study
Magnetically induced displacement force and torque in the MR environment
Indications for Use
DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.
Device Story
DESS Dental Smart Solutions abutments (healing, temporary, multi-unit) and screws provide support for prosthetic restorations in the maxillary or mandibular arch. Components are used in conjunction with various OEM endosseous dental implants. The device is used in a clinical setting by dental professionals. This submission updates the sterilization status of previously cleared components to include sterile-to-the-end-user options via gamma irradiation. The device functions as a mechanical interface between the implant and the final prosthesis. Benefits include enabling functional and esthetic rehabilitation for edentulous patients. The healthcare provider selects the appropriate abutment based on the specific implant system and clinical requirements.
Clinical Evidence
No clinical data were included in this submission. Substantial equivalence is supported by non-clinical bench testing, including MR environment displacement/torque analysis, gamma irradiation validation (SAL 10^-6), bacterial endotoxin testing (ANSI/AAMI ST72, < 20 EU/device), and accelerated aging shelf-life testing (ASTM F1980).
Technological Characteristics
Materials: Ti-6Al-4V alloy (ASTM F136). Surface: Anodization and SelectGrip®. Screws: Ti-6Al-4V alloy with DLC coating. Sterilization: Gamma irradiation (25 kGy) or non-sterile. Packaging: PETG blister with Tyvek® lid. Connectivity: None (mechanical).
Indications for Use
Indicated for patients requiring prosthetic restoration support in the maxillary or mandibular arch using endosseous dental implants. Compatible with specific OEM implant systems listed in the device documentation.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 5, 2024
Terrats Medical SL % Kevin Thomas Vice President & Director of Regulatory Affairs PaxMed International, LLC 12264 El Camino Real. Suite 400 San Diego, California 92130
Re: K233316
Trade/Device Name: DESS Dental Smart Solutions Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: January 2, 2024 Received: January 2, 2024
Dear Kevin Thomas:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{3}------------------------------------------------
3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Andrew I. Steen -S
Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K233316
Device Name
DESS Dental Smart Solutions
Indications for Use (Describe)
DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.
| Compatible Implant System | Implant Body Diameter, mm | Implant Platform Name |
|-----------------------------------------------|------------------------------------|-----------------------|
| Ankylos C/X | 3.5, 4.5, 5.5 | 2.52 |
| | 3.0 | 3.0 |
| | 3.6 | 3.6 |
| Astra Tech EV | 4.2 | 4.2 |
| | 4.8 | 4.8 |
| | 5.4 | 5.4 |
| Astra Tech OsseoSpeed™ | 3.5, 4.0 | 3.5/4.0 |
| | 4.5, 5.0 | 4.5/5.0 |
| BioHorizons Internal | 3.8, 4.6 | 3.5 |
| | 4.6, 5.8 | 4.5 |
| | 5.8 | 5.7 |
| Biomet 3i Certain® | 3.25 | 3.4 |
| | 4.0 | 4.1 |
| | 5.0 | 5.0 |
| Biomet 3i OSSEOTITE® | 3.25 | 3.4 |
| | 3.75, 4.0 | 4.1 |
| | 5.0 | 5.0 |
| Camlog | 3.8 | 3.8 |
| | 4.3 | 4.3 |
| Dentium SuperLine | 3.6, 4.0, 4.5, 5.0, 6.0, 7.0 | 3.3 |
| FRIADENT XiVE® | 3.4 | 3.4 |
| | 3.8 | 3.8 |
| | 4.5 | 4.5 |
| Neodent Grand Morse | 3.5, 3.75, 4.0, 4.3, 5.0, 6.0, 7.0 | Grand Morse (GM) |
| NobelActive® | 3.5 | NP |
| NobelReplace Conical<br>NobelParallel Conical | 4.3, 5.0 | RP |
| NobelReplace® | 3.5 | NP |
| | 4.3 | RP |
| | 5.0 | WP |
| Nobel Brånemark System® | 3.3 | NP |
| | 3.75, 4.0 | RP |
| | 5.0 | WP |
| Osstem TS | 3.5 | Mini |
| Hiossen TS | 4.0, 4.5, 5.0, 6.0, 7.0 | Regular |
| Straumann BLX | 3.5, 3.75, 4.0, 4.5 | RB |
| | 5.0, 5.5, 6.5 | WB |
| Straumann® Bone Level | 3.3 | NC |
| | 4.1, 4.8 | RC |
| Straumann® Tissue Level | 3.3, 4.1, 4.8 | RN |
| | 4.8 | WN |
| Zimmer Screw Vent® | 3.3, 3.7, 4.1 | 3.5 |
| | 4.7 | 4.5 |
| Tapered Screw-Vent® | 6.0 | 5.7 |
| TSXTM Implant System | 3.7, 4.1, 4.7 | 3.5 |
| | 5.4, 6.0 | 4.5 |
## Compatible Implant Systems
{5}------------------------------------------------
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
PSC Publishing Services (301) 443-6740 EF
{6}------------------------------------------------
# 510(k) Summary K233316 Terrats Medical SL DESS Dental Smart Solutions January 2, 2024
## ADMINISTRATIVE INFORMATION
| Manufacturer Name | Terrats Medical SL<br>Carrer Mogoda, 75-99<br>Barberà del Vallès 08210<br>Barcelona, Spain |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Telephone | +34 935 646 006 |
| Official Contact | Roger Terrats, CEO |
| Representative/Consultant | Kevin A. Thomas, PhD<br>Floyd G. Larson, MS, MBA<br>PaxMed International, LLC<br>12264 El Camino Real, Suite 400<br>San Diego, CA 92130<br>Telephone +1 858-792-1235<br>Fax +1 858-792-1236<br>Email kthomas@paxmed.com<br>flarson@paxmed.com |
#### DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name | DESS Dental Smart Solutions |
|------------------------|---------------------------------------------------------------------------------------------|
| Common Name | Dental implant abutment |
| Regulation Number | 21 CFR 872.3630 |
| Regulation Name | Endosseous dental implant abutment |
| Regulatory Class | Class II |
| Product Code | NHA |
| Classification Panel | Dental |
| Reviewing Office | Office of Health Technology 1 (Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices) |
| Reviewing Division | Division of Dental and ENT Devices |
# PREDICATE DEVICE INFORMATION
Primary Predicate Device K170588, DESS Dental Smart Solutions, Terrats Medical SL
Reference Devices K191986, DESS Dental Smart Solutions, Terrats Medical SL K212628, DESS Dental Smart Solutions, Terrats Medical SL K222288, DESS Dental Smart Solutions, Terrats Medical SL K212538, DESS Dental Implants, Terrats Medical SL K231434, DESS Dental Smart Solutions, Terrats Medical SL
{7}------------------------------------------------
Reference Devices for OEM implant body clearances
- K140347, ANKYLOS C/X Implant System, Dentsply International Inc.
- K120414. OsseoSpeed™ Plus, Astra Tech AB
- K111287, Astra Tech Implant System Plus, Astra Tech AB
- K101732, Astra Tech Implant System, Astra Tech AB
- K042429, BioHorizons The Prodigy System™ Endosseous Implants, BioHorizons Implant Systems, Inc.
- K071638, BioHorizons Tapered Internal Implant System, BioHorizons Implant Systems, Inc.
- K093321, BioHorizons Laser-Lok 3.0 Implant System, BioHorizons Implant Systems, Inc.
- K063341, 3i OSSEOTITE® Certain® Dental Implants, Implant Innovations, Inc.
- K083496, CAMLOG Implant System Modified Implants and Abutments, Altatec GmbH
- K160965, SuperLine, Dentium Co., Ltd.
- K073075, FRIADENT Implant Systems (FRIALIT® plus Dental Implant System, XiVE® S plus Dental Implant System, XiVE® TG plus Dental Implant System, ANKYLOS® plus Dental Implant System), Dentsply International Inc.
- K163194, Neodent Implant System GM Line, JJGC Indústria e Comércio de Materiais Dentários SA
- K180536, Neodent Implant System GM Line, JJGC Indústria e Comércio de Materiais Dentários S.A.
- K201225, Neodent Implant System GM Helix Implants 7.0, JJGC Indústria e Comércio de Materiais Dentários S.A.
- K142260, NobelActive®, Nobel Biocare AB
- K173418, NobelParallel™ Conical Connection, Nobel Biocare AB
- K050705, TiUnite Implants®, Nobel Biocare AB
- K050406, NOBELSPEEDY™ Implants, Nobel Biocare USA LLC
- K050258, Groovy Implants, Nobel Biocare AB
- K023113, Replace TiUnite Endosseous Implant, Nobel Biocare USA, Inc.
- K161604, OSSTEM Implant System, OSSTEM Implant Co., Ltd.
- K173961, Straumann® BLX Implant System, Institut Straumann AG
- K181703, Straumann® BLX Line Extension Implants, SRAs and Anatomic Abutments, Institut Straumann AG
- K191256, Straumann BLX Ø3.5 mm Implants, Institut Straumann AG
- K210855, Straumann BLX Implant System, Institut Straumann AG
- K212533, BLX WB Ø5.0 (L18), Ø5.5 and Ø6.5 mm (L14 and L16) Implants, Institut Straumann AG
- K140878, Straumann® Bone Level Tapered Implants, Straumann USA, LLC
- K130222, Straumann® Dental Implant System SLActive and Roxolid Product Families, Straumann USA, LLC
- K013227. Screw Vent Implant: Tapered Screw Vent Implant. Sulzer Dental. Inc.
- K072589, Tapered Screw-Vent Implant, 4.1mmD, Zimmer Dental, Inc.
- K220978, TSXTM Implants, Biomet 3i LLC
The primary predicate device K170588, and the reference devices K191986, K212628, and K222288 are in support of substantial equivalence of the subject designs, material, manufacturing, and biocompatibility. The reference device K212538 is in support of substantial equivalence of the subject device sterilization, packaging, and shelf life.
# INDICATIONS FOR USE STATEMENT
DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.
| Compatible Implant System | Implant Body Diameter, mm | Implant Platform Name |
|---------------------------|---------------------------|-----------------------|
| Ankylos C/X | 3.5, 4.5, 5.5 | 2.52 |
| | 3.0 | 3.0 |
| | 3.6 | 3.6 |
| Astra Tech EV | 4.2 | 4.2 |
| | 4.8 | 4.8 |
| | 5.4 | 5.4 |
| Astra Tech OsseoSpeed™ | 3.5, 4.0 | 3.5/4.0 |
| | 4.5, 5.0 | 4.5/5.0 |
| BioHorizons Internal | 3.8, 4.6 | 3.5 |
| | 4.6, 5.8 | 4.5 |
| | 5.8 | 5.7 |
#### Compatible Implant Systems
{8}------------------------------------------------
| Compatible Implant System | Implant Body Diameter, mm | Implant Platform Name |
|-----------------------------------------------|------------------------------------|-----------------------|
| Biomet 3i Certain® | 3.25 | 3.4 |
| | 4.0 | 4.1 |
| | 5.0 | 5.0 |
| Biomet 3i OSSEOTITE® | 3.25 | 3.4 |
| | 3.75, 4.0 | 4.1 |
| | 5.0 | 5.0 |
| Camlog | 3.8 | 3.8 |
| | 4.3 | 4.3 |
| Dentium SuperLine | 3.6, 4.0, 4.5, 5.0, 6.0, 7.0 | 3.3 |
| FRIADENT XIVE® | 3.4 | 3.4 |
| | 3.8 | 3.8 |
| | 4.5 | 4.5 |
| Neodent Grand Morse | 3.5, 3.75, 4.0, 4.3, 5.0, 6.0, 7.0 | Grand Morse (GM) |
| NobelActive® | 3.5 | NP |
| NobelReplace Conical<br>NobelParallel Conical | 4.3, 5.0 | RP |
| NobelReplace® | 3.5 | NP |
| | 4.3 | RP |
| | 5.0 | WP |
| Nobel Brånemark System® | 3.3 | NP |
| | 3.75, 4.0 | RP |
| | 5.0 | WP |
| Osstem TS | 3.5 | Mini |
| Hiossen TS | 4.0, 4.5, 5.0, 6.0, 7.0 | Regular |
| Straumann BLX | 3.5, 3.75, 4.0, 4.5 | RB |
| | 5.0, 5.5, 6.5 | WB |
| Straumann® Bone Level | 3.3 | NC |
| | 4.1, 4.8 | RC |
| Straumann® Tissue Level | 3.3, 4.1, 4.8 | RN |
| | 4.8 | WN |
| Zimmer Screw Vent® | 3.3, 3.7, 4.1 | 3.5 |
| Tapered Screw-Vent® | 4.7 | 4.5 |
| | 6.0 | 5.7 |
| TSX™ Implant System | 3.7, 4.1, 4.7 | 4.5 |
| | 5.4, 6.0 | 4.5 |
## SUBJECT DEVICE DESCRIPTION
The purpose of this submission is to expand the DESS Dental Smart Solutions abutment system by a change in sterilization status to provide products sterile to the end user that were previously cleared to be provided non-sterile. The subject device abutments and abutment screws were cleared previously to be provided non-sterile to the end user in K170588, K191986, K212628, and K22288. All subject device components will now be provided sterile.
The subject device components include Healing Abutments, Multi-Unit Abutments (0, 17°, and 30°), and abutment screws. A summary of the subject components is provided in the following table.
| | Subject Device Components | | | | |
|------------------------|---------------------------|------------------------|-----------------------------|-----------------------------------------|--------|
| OEM Implant Lines | Healing<br>Abutments | Temporary<br>Abutments | Munti-Unit<br>Abutments, 0° | Munti-Unit<br>Abutments, 17° and<br>30° | Screws |
| Ankylos C/X | X | X | | | X |
| Astra Tech EV | X | X | X | X | X |
| Astra Tech OsseoSpeed™ | X | X | X | X | X |
| BioHorizons Internal | X | | X | | |
| Biomet 3i Certain® | X | X | | | X |
| Biomet 3i OSSEOTITE® | X | X | X | | X |
| Camlog | X | X | | | X |
{9}------------------------------------------------
| | Subject Device Components | | | | |
|-------------------------|---------------------------|------------------------|-----------------------------|-----------------------------------------|--------|
| OEM Implant Lines | Healing<br>Abutments | Temporary<br>Abutments | Munti-Unit<br>Abutments, 0° | Munti-Unit<br>Abutments, 17° and<br>30° | Screws |
| Dentium SuperLine | | | X | | |
| FRIADENT XiVE® | X | X | | | X |
| Neodent Grand Morse | | | X | X | |
| NobelActive® | X | X | X | X | X |
| NobelReplace® | X | X | X | | X |
| Nobel Brånemark System® | X | X | X | | X |
| Osstem TS | | X | | | X |
| Hiossen TS | | X | | | X |
| Straumann BLX | X | | X | X | |
| Straumann® Bone Level | X | X | X | X | X |
| Straumann® Tissue Level | X | X | X | | X |
| Zimmer Screw Vent® | X | X | | | X |
| Tapered Screw-Vent® | X | X | | | X |
| TSX™ Implant System | X | X | | | X |
Healing Abutments are compatible with the OEM implant lines as listed in this summary, and as protectors (healing attachments) for use with Straumann Converter Abutment for Straumann Tissue Level implants) and for DESS Multi-Unit Abutments. The Healing Abutments are provided in a range of gingival heights from 1 mm to 6.5 mm, and in a range of coronal (maximum) diameters of 3.5 mm to 7 mm.
Temporary Abutments are compatible with the OEM implant lines as listed in this summary and are available for single-unit and multiple-unit restorations. Temporary Abutments are provided in a range of gingival heights from 0.35 mm to 2.5 mm, and in a range of prosthetic platform diameters from 4.5 mm to 6.8 mm.
Multi-Unit Abutments are compatible with the OEM implant lines as listed in this summary in straight (00) and angled (17° and 30°) designs. Multi-Unit are provided in gingival heights from 1 mm to 6 mm for the straight (0°) designs, 2.5 mm to 4.5 mm for angled 17º designs, and 3.5 mm for the angled 30º designs.
Abutment screws are compatible with the OEM implant lines as listed in this summary. There are no changes to the design of the screws cleared previously in K170588. K212628. K191986, and K222288.
The subject device abutments and abutment screws have the same OEM implant compatibilities as the identical screws cleared in K170588, K212628, K191986, and K222288.
# PERFORMANCE DATA
Non-clinical data submitted or referenced to demonstrate substantial equivalence included:
- . non-clinical analysis was performed to evaluate the subject devices (including all abutments, abutment screw, and materials) in the MR environment using scientific rationale and published literature (TO Woods, JG Delfino, and S Rajan, "Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices," Journal of Testing and Evaluation, Volume 49, No. 2, 2021, pp. 783-795); the analysis addressed parameters per the FDA guidance Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment (issued May 2021) including magnetically induced displacement force and torque;
- referenced from K212538 was gamma irradiation validation to a sterility assurance level of 10° by selecting and substantiating a 25 kGy dose using method VDmax25, according to ISO 11137-1 and ISO 11137-2; bacterial endotoxin testing including Limulus amebocyte lysate (LAL) test according to ANSI/AAMI ST72 to demonstrate that all sterile product meets a limit of < 20 EU/device; and shelf life testing of samples after
{10}------------------------------------------------
accelerated aging equivalent to five (5) years of real time aging according to ASTM F1980, with testing of the packaging sterile barrier and sterility testing of product;
- . referenced from K170588 was moist heat sterilization validated by the overkill method to a sterility assurance level (SAL) of 106 according to ISO 17665-1 and ISO/TR 17665-2;
- referenced from K170588, K191986, K212628, and K222288 was biocompatibility of the subject device ● components:
- . referenced from K170588. K191986. K212628. and K222288 was compatibility of the subject abutnents and the OEM implants listed in the Indications for Use Statement.
No clinical data were included in this submission.
# EQUIVALENCE TO MARKETED DEVICES
All subject device components are identical in design, material, and technological characteristics to those of the primary predicate device K170588, and the reference devices K191986, K2126288.
The subject device abutments have similar or identical ranges abutment-implant platform diameter, prosthetic platform diameter, gingival height, and angulation as the components cleared in the predicate device K170588, and the reference devices K191986. K212628, and K222288. The subject device screws are identical to the screws previously cleared in K170588, K191986, K212628, and K222288.
The subject device abutments have the same compatibilities as previously cleared in the primary predicate device K170588 and in the reference devices K191986, K212628, and K222288.
All subject device components are provided sterile by gamma irradiation and are packaged in a PETG blister with a Tyvek® lid. The gamma sterilization, packaging, and 5-year shelf life for sterile devices is the same as that validated in the reference device K212538. Subject devices that are provided non-sterile are packaged in PETG blister packs or PET bags.
# CONCLUSION
The subject device, the primary predicate device, and the reference devices have the same intended use, identical or nearly identical technological characteristics, and are material. The subject device and the primary predicate device encompass the same of physical dimensions, are packaged in similar materials, and are to be sterilized using similar methods. The data included in this submission demonstrate substantial equivalence to the predicate devices listed above.
{11}------------------------------------------------
510(k) Summary
Page 6 of 6
## Table of Substantial Equivalence
| Comparison | Subject Device | Primary Predicate Devices | Reference Device | Reference Device | Reference Device |
|---------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | DESS Dental Smart Solutions<br>Terrats Medical SL | K170588<br>DESS Dental Smart Solutions<br>Terrats Medical SL | K191986<br>DESS Dental Smart Solutions<br>Terrats Medical SL | K212628<br>DESS Dental Smart Solutions<br>Terrats Medical SL | K222288<br>DESS Dental Smart Solutions<br>Terrats Medical SL |
| Product Code | NHA | NHA | NHA | NHA | NHA |
| Reason for predicate/reference | n/a | Abutment designs, OEM compatibilities, manufacturing | Abutment designs, OEM compatibilities, manufacturing | Abutment designs, OEM compatibilities, manufacturing | Abutment designs, OEM compatibilities, manufacturing |
| Intended Use | Functional and esthetic rehabilitation of the edentulous<br>mandible or maxilla | Functional and esthetic rehabilitation of the edentulous<br>mandible or maxilla | Functional and esthetic rehabilitation of the edentulous<br>mandible or maxilla | Functional and esthetic rehabilitation of the edentulous<br>mandible or maxilla | Functional and esthetic rehabilitation of the edentulous<br>mandible or maxilla |
| Indications for Use Statement | DESS Dental Smart Solutions abutments are intended to be<br>used in conjunction with endosseous dental implants in the<br>maxillary or mandibular arch to provide support for<br>prosthetic restorations.<br>The complete Indications for Use Statement with OEM<br>implant compatibilities is provided in this Summary. | DESS Dental Smart Solutions abutments are intended to be<br>used in conjunction with endosseous dental implants in the<br>maxillary or mandibular arch to provide support for<br>prosthetic restorations.<br>All digitally designed custom abutments for use with<br>TiBase or Pre-milled Blank are to be sent to a Terrats<br>Medical validated milling center for manufacture.<br>The complete Indications for Use Statement with OEM<br>implant compatibilities is provided in the 510(k) Summary<br>for K170588. | DESS Dental Smart Solutions abutments are intended to be<br>used in conjunction with endosseous dental implants in the<br>maxillary or mandibular arch to provide support for<br>prosthetic restorations.<br>All digitally designed custom abutments for use with Ti<br>Base abutments or Pre-milled (Blank) abutments are to be<br>sent to a Terrats Medical validated milling center for<br>manufacture.<br>The complete Indications for Use Statement with OEM<br>implant compatibilities is provided in the 510(k) Summary<br>for K1191986. | DESS Dental Smart Solutions abutments are intended to be<br>used in conjunction with endosseous dental implants in the<br>maxillary or mandibular arch to provide support for<br>prosthetic restorations.<br>All digitally designed custom abutments for use with DESS<br>Bases or Blanks are to be sent to a Terrats Medical<br>validated milling center for manufacture.<br>The complete Indications for Use Statement with OEM<br>implant compatibilities is provided in the 510(k) Summary<br>for K212628. | DESS Dental Smart Solutions abutments are intended to be<br>used in conjunction with endosseous dental implants in the<br>maxillary or mandibular arch to provide support for<br>prosthetic restorations.<br>All digitally designed custom abutments for use with Ti<br>Base abutments or Pre-milled Blank abutments are to be<br>sent to a Terrats Medical validated milling center for<br>manufacture.<br>The complete Indications for Use Statement with OEM<br>implant compatibilities is provided in the 510(k) Summary<br>for K222288. |
| Designs | Healing Abutments | Healing Abutments | Healing Abutments | Healing Abutments | Healing Abutments |
| | Temporary Abutments | Temporary Abutments | Temporary Abutments | | Temporary Abutments |
| | Multi-unit Abutments, Straight | Multi-unit Abutments, Straight | | Multi-unit Abutments, Straight | Multi-unit Abutments, Straight |
| | Multi-unit Abutments, Angled 17° and 30° | | Multi-unit Abutments, Angled 17° and 30° | | Multi-unit Abutments, Angled 17° and 30° |
| | Prosthetic components for Multi-unit Abutments | Prosthetic components for Multi-unit Abutments | | | |
| | Screws | Screws | Screws | Screws | Screws |
| Prosthesis Attachment | Cement-retained<br>Screw-retained | Cement-retained<br>Screw-retained | Cement-retained<br>Screw-retained | Cement-retained<br>Screw-retained | Cement-retained<br>Screw-retained |
| Restoration | Single-unit<br>Multi-unit | Single-unit<br>Multi-unit | Single-unit<br>Multi-unit | Single-unit<br>Multi-unit | Single-unit<br>Multi-unit |
| Abutment/Implant Platform Ø, mm | 2.52 - 5.7 | 3.4 - 5.7 | 2.52 - 6.0 | 2.3 - 6.0 | 2.52 - 6.5 |
| Prosthetic Platform Ø, mm | 4.5 - 6.8 | 4.5 - 6.8 | 4.5 - 6.5 | 4.5 - 6.5 | 4.5 - 6.5 |
| Abutment Angle | 0°, 17°, 30° | 0° | 0°, 17°, 30° | 0° | 0°, 17°, 30° |
| Material | | | | |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.