K231434 · Terrats Medical SL · NHA · Aug 14, 2023 · Dental
Device Facts
Record ID
K231434
Device Name
DESS Dental Smart Solutions
Applicant
Terrats Medical SL
Product Code
NHA · Dental
Decision Date
Aug 14, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Indications for Use
DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.
Device Story
DESS Dental Smart Solutions are multi-unit dental abutments (straight, 17°, 30°) designed to attach to endosseous dental implants; used for multi-unit restorations like bridges and bars. Input: physical interface with specific OEM implant bodies (Zimmer/Biomet 3i). Transformation: mechanical coupling via abutment screws to provide a stable prosthetic platform. Output: 4.8mm diameter prosthetic platform for restoration attachment. Used in dental clinics by clinicians. Benefits: enables screw-retained prosthetic rehabilitation for edentulous patients. Provided sterile or non-sterile for clinical use.
Clinical Evidence
Bench testing only. No clinical data included. Evidence includes mechanical testing (ISO 14801) for fatigue strength, MR safety analysis (ASTM standards), biocompatibility (ISO 10993-5), and sterilization validation (ISO 17665, ISO 11137).
Technological Characteristics
Materials: Titanium alloy (Ti-6Al-4V) per ASTM F136. Design: Multi-unit abutments (straight, 17°, 30° angulation) with 4.8mm prosthetic platform. Connectivity: Mechanical interface with internal/external hex dental implants. Sterilization: Moist heat (non-sterile components) or gamma irradiation (sterile components).
Indications for Use
Indicated for patients requiring prosthetic restoration support in the maxillary or mandibular arch using endosseous dental implants. Compatible with specific Zimmer and Biomet 3i implant systems.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
Terrats Medical SL % Kevin Thomas Vice President & Director of Regulatory Affairs PaxMed International, LLC 12264 EL Camino Real Suite 400 San Diego, California 92130
Re: K231434
Trade/Device Name: DESS Dental Smart Solutions Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: May 17, 2023 Received: May 17, 2023
Dear Kevin Thomas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Andrew I. Steen -S
Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K231434
Device Name
DESS Dental Smart Solutions
Indications for Use (Describe)
DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.
| Implant Systems Compatibility | Implant Body Ø, mm | Implant Platform Ø, mm |
|-------------------------------|--------------------|------------------------|
| Internal Hex Connections | | |
| Zimmer 3.1mmD Eztetic™ | 3.1 | 2.9 |
| Zimmer Screw-Vent® | 3.7 | 3.5 |
| | 4.7 | 4.5 |
| Zimmer Tapered Screw-Vent® | 3.7, 4.1 | 3.5 |
| | 4.7 | 4.5 |
| | 6.0 | 5.7 |
| TSX™ Implant System | 3.7, 4.1, 4.7 | 3.5 |
| | 5.4, 6.0 | 4.5 |
| TSX™ Implant System, 3.1mmD | 3.1 | 2.9 |
| External Hex Connections | | |
| Biomet 3i OSSEOTITE® Implants | 3.25 | 3.4 |
| | 3.75 | 4.1 |
| | 4.0 | 4.1 |
#### Implant Systems Compatibility
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary K231434 Terrats Medical SL DESS Dental Smart Solutions August 7, 2023
#### ADMINISTRATIVE INFORMATION
| Manufacturer Name | Terrats Medical SL<br>Carrer Mogoda, 75-99<br>Barberà del Vallès 08210<br>Barcelona, Spain |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Telephone | +34 935 646 006 |
| Official Contact | Roger Terrats, CEO |
| Representative/Consultant | Kevin A. Thomas, PhD<br>Floyd G. Larson, MS, MBA<br>PaxMed International, LLC<br>12264 El Camino Real, Suite 400<br>San Diego, CA 92130<br>Telephone +1 858-792-1235<br>Fax +1 858-792-1236<br>Email kthomas@paxmed.com<br>flarson@paxmed.com |
#### DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name | DESS Dental Smart Solutions |
|------------------------|---------------------------------------------------------------------------------------------|
| Common Name | Dental implant abutment |
| Regulation Number | 21 CFR 872.3630 |
| Regulation Name | Endosseous dental implant abutment |
| Regulatory Class | Class II |
| Product Code | NHA |
| Classification Panel | Dental |
| Reviewing Office | Office of Health Technology 1 (Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices) |
| Reviewing Division | Division of Dental and ENT Devices |
#### PREDICATE DEVICE INFORMATION
Primary Predicate Device K230143, DESS Dental Smart Solutions, Terrats Medical SL
Reference Devices K142082, Zimmer 3.1mmD Dental Implant System, Zimmer Dental, Inc. K013227, Screw Vent Implant; Tapered Screw Vent Implant, Sulzer Dental, Inc. K072589, Tapered Screw-Vent Implant, 4.1mmD, Zimmer Dental, Inc. K220978, TSX™ Implants, Biomet 3i LLC K063286, OSSEOTITE® Dental Implants, Implant Innovations, Inc. K111216, OSSEOTTE® 2 - Dental Implants, Biomet 3i, Inc. K212538, DESS Dental Implants, Terrats Medical SL
{5}------------------------------------------------
K170588. DESS Dental Implants. Terrats Medical SL K22269, DESS Dental Implants, Terrats Medical SL K213063, TLX SRAs and TLX Gold Abutments, Straumann USA, LLC
The reference devices K142082, K013227, K072589, K220978, K063286, K111216 are for OEM implant body clearances. The reference device K212538 is for sterilization, packaging, and shelf life for devices provided sterile to the end user. The reference device K170588 is for compatibility with the Zimmer Screw-Vent", Zimmer Tapered Screw-Vent®, and Biomet 3i OSSEOTITE® implants. The reference device K22269 is for referenced moist heat sterilization and biocompatibility data. The reference device K213063 is for the technological characteristic of multiunit abutments provided sterile.
## INDICATIONS FOR USE STATEMENT
DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.
| Implant Systems Compatibility | Implant Body Ø, mm | Implant Platform Ø, mm |
|-------------------------------|--------------------|------------------------|
| Internal Hex Connections | | |
| Zimmer 3.1mmD Eztetic™ | 3.1 | 2.9 |
| Zimmer Screw-Vent® | 3.7 | 3.5 |
| | 4.7 | 4.5 |
| Zimmer Tapered Screw-Vent® | 3.7, 4.1 | 3.5 |
| | 4.7 | 4.5 |
| | 6.0 | 5.7 |
| TSX™ Implant System | 3.7, 4.1, 4.7 | 3.5 |
| | 5.4, 6.0 | 4.5 |
| TSX™ Implant System, 3.1mmD | 3.1 | 2.9 |
| External Hex Connections | | |
| Biomet 3i OSSEOTITE® Implants | 3.25 | 3.4 |
| | 3.75 | 4.1 |
| | 4.0 | 4.1 |
| | | Implant Systems Compatibility | |
|--|--|-------------------------------|--|
|--|--|-------------------------------|--|
## SUBJECT DEVICE DESCRIPTION
The purpose of this submission is to add components to the DESS Dental Smart Solutions system, which includes abutments compatible with a variety of original equipment manufacturers (OEM) of dental implants.
This submission adds abutments, designated DESS Multi-Unit Abutments, for implant lines from Zimmer Dental and Biomet 31 (now collectively, ZimVie Dental). The subject device abutments include Multi-Unit Abutments in straight, 17° angled, and 30° angled designs, which are compatible with implants having internal hex connections. This submission is also to change how previously cleared devices are provided; the change is from previously provided nonsterile to now provided sterile. All abutments are provided with the appropriate abutment screw (if applicable) for attachment to the corresponding implant. The subject device is only intended for multi-unit restorations such as bridges and bars.
A summary of the subject device abutment designs and the compatible OEM implants is provided in the table Summary of Subject Device Multi-Unit Abutment Designs on the following page.
{6}------------------------------------------------
| | Subject Device Multi-Unit Abutments | | | | |
|-----------------------------------------------------------|-------------------------------------|------------------------------------|------------------------|-------------------------------------------------------|-----------------------------------------------------------|
| Connection and<br>Compatible Implant Line | Angulation | Implant-Abutment<br>Platform Ø, mm | Gingival Height,<br>mm | New Components,<br>Provided Sterile<br>and Nonsterile | Previously Cleared<br>Components,<br>Now provided Sterile |
| Internal Hex Connections | | | | | |
| Zimmer 3.1mmD Eztetic™<br>TSXTM Implant System,<br>3.1mmD | 0° | 2.9 | 1, 2, 3, 4, 5 | X | |
| | 17° | | 2.5, 3.5 | X | |
| | 30° | | 3.5, 4.5 | X | |
| Zimmer Screw-Vent®<br>TSXTM Implant System | 0° | 3.5, 4.5 | 1, 2, 3, 4, 5 | X | X |
| | 17° | | 2.5, 3.5 | X | X |
| | 30° | | 3.5, 4.5 | X | X |
| Zimmer Tapered Screw-Vent® | 0° | 3.5, 4.5, 5.7 | 1, 2, 3, 4, 5 | X | X |
| | 17° | | 2.5, 3.5 | X | X |
| | 30° | | 3.5, 4.5 | X | X |
| External Hex Connections | | | | | |
| Biomet 3i OSSEOTITE®<br>Implants | 17° | 3.4 | 2, 3, 4 | X | |
| | 30° | 4.1 | 3, 4, 5 | X | |
## Summary of Subject Device Multi-Unit Abutment Designs
Subject device components that will be provided sterile include:
new (not previously cleared) multi-unit abutment screw compatible with Zimmer 3.1mmD Eztetic and TSX™ Implant System, 3.1mmD implants; new multi-unit abutments compatible with Screw-Vent", Tapered Screw-Vent", and TSX™ implants with implant platform diameters of 4.5 mm; new multi-unit abutments compatible with Biomet 3i OSSEOTITE® implants; previously cleared multi-unit abutments compatible with Screw-Vent", Tapered Screw-Vent®, and TSX™ implants with implant platform diameters of 3.5 mm and 4.5 mm; and previously cleared prosthetic components compatible with all subject device multi-unit abutments.
Subject device components that will be provided non-sterile include:
new (not previously cleared) multi-unit abutment screw compatible with Zimmer 3.1mmD Eztetic and TSX™ Implant System, 3.1mmD implants; new multi-unit abutments with Screw-Vent", and TSX™ implants with implant platform diameters of 4.5 mm and new multi-unit abutments compatible with Biomet 3i OSSEOTITE® implants.
The design dimensions and tolerances of subject device abutments and screws for the Zimmer 3.1mmD Eztetic and TSX™ Implant Systems have been established on the basis of a contractual agreement and working relationship between ZimVie and Terrats Medical SL to ensure that the abutments are designed to fit the corresponding implants. Compatibility of the subject abutments and screws for Screw-Vent®, Tapered Screw-Vent®, and TSX™ implant lines was provided in the prior Terrats Medical SL submission K170588. Compatibility of the subject abutments and screws for the Biomet 3i OSSEOTITE® external hex implants was provided in the prior Terrats Medical SL submission K170588.
## Multi-Unit Abutments
The Multi-Unit Abutments are designed for attachment of multi-unit screw-retained restorations and are provided in three (3) designs, straight, angled 30°. The design of the straight Multi-Unit Abutments is similar to that of straight Multi-Unit Abutments cleared in the primary predicate device K230143 with the exception of the implant connections and platform diameters. The straight Multi-Unit Abutment is provided only in a non-engaging. threaded design that attaches directly to the implant. All straight Multi-Unit Abutments are provided with a prosthetic platform diameter of 4.8 mm. Straight Multi-Unit Abutments are provided only for the compatible internal hex implants listed in the table above. The gingival height of the straight Multi-Unit Abutment ranges from 1 mm to 5 mm in 1 mm increments.
{7}------------------------------------------------
510(k) Summary Page 4 of 7
The angled Multi-Unit Abutments are provided in an engaging design that requires an abutment screw, with angulations of 17° and 30°. The angled Multi-Unit Abutments are provided for the compatible internal hex connection implants and the compatible external hex connection implants listed in the table above (on page 3). All angled Multi-Unit Abutments are provided with a prosthetic platform diameter of 4.8 mm, and with gingival heights from 2.5 mm to 4.5 mm. The designs of the angled Multi-Unit Abutments are similar to those of the angled Multi-Unit Abutments cleared in the primary predicate device K230143.All Multi-Unit Abutments, straight and angled, are manufactured from titanium alloy (Ti-6A1-4V) conforming to ASTM F136.
#### Screws
This submission includes two (2) abutment screws to be used with the subject device abutment screw for the compatible internal hex implants with a platform diameter of 2.9 mm; and another abutment screw for all compatible external hex implants (with platforms of 3.4 mm). The screws have a hex or hexalobular instrument interface and are manufactured from titanium alloy (Ti-6Al-4V) conforming to ASTM F136.
## PERFORMANCE DATA
Non-clinical data submitted or referenced to demonstrate substantial equivalence included:
- provided in this submission was non-clinical analysis to evaluate the metallic subject devices and compatible . dental implants in the MR environment according to ASTM F2052 (magnetically induced displacement force), ASTM F2213 (magnetically induced torque), ASTM F2182 (RF induced heating), and ASTM F2119 (image artifact), and the FDA guidance document Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment (issued May 2021);
- provided in this submission was engineering analysis to demonstrate that the subject device abutments compatible ● with the Screw-Vent", Tapered Screw-Vent", and TSX™ implants do not create a new worst-case construct in terms of mechanical testing according to ISO 14801;
- . provided in this submission was mechanical testing conducted according to ISO 14801 to support the performance of the subject device abutments compatible with the Zimmer 3.1mmD Eztetic and TSX™ implants, and the performance of the subject device abutments compatible with the Biomet 3i OSSEOTITE® Implants;
- referenced from K22269 was moist heat sterilization for subject devices provided non-sterile to the end user, ● validated to a sterility assurance level of 10° by the overkill method according to ANSI/AAMI/ISO 17665-1 and ANSI/AAMI/ISO TIR 17665-2: analysis showed that the subject devices do not create a new worst case for moist heat sterilization:
- referenced from K212538 was gamma irradiation sterility assurance level of 10° by ● selecting and substantiating a 25 kGy dose using method VDmax25, according to ISO 11137-1 and ISO 11137-2; bacterial endotoxin testing including Limulus amebocyte lysate (LAL) test according to ANSI/AAMI ST72 to demonstrate that all sterile product meets a limit of < 20 EU/device; and shelf life testing of samples after accelerated aging equivalent to five (5) years of real time aging according to ASTM F1980, with testing of the packaging sterile barrier and sterility testing of product;
- . referenced from K22269 was biocompatibility testing according to ISO 10993-5 (cytotoxicity) for the abutment material ASTM F136; and
- . referenced from K170588 was reverse engineering compatibility data for the Zimmer Screw-Vent", Zimmer Tapered Screw-Vent®, Zimmer TSX™, and Biomet 3i OSSEOTITE® implant lines.
No clinical data were included in this submission.
## EQUIVALENCE TO MARKETED DEVICES
All abutment screws are similar or identical in design, materials, and technological characteristics to those cleared in the primary predicate device K230143, except for threads and lengths that accommodate the new compatibilities.
{8}------------------------------------------------
Subiect device components that are provided non-sterile are to be sterilized by the same moist heat evelerenced from the primary predicate K230143. The subject devices that are provided non-sterile are packaged in either a PETG blister pack or a PET bag, the same packaging referenced from the primary predicate K230143.
Subject device components that are provided sterile by gamma irradiation are packaged in a PETG blister with a Tyvek® Iid. This is the same sterilization, packaging, and 5-year shelf life as validated in the reference device K212538.
The risks associated with use of the subject device angled multi-unit abutments in combination with the compatible implants are mitigated by mechanical testing performed according to ISO 14801.
#### CONCLUSION
The subject device, the primary predicate device, and the reference devices have the same intended use, have similar technological characteristics, and are materials. The subject device and the primary predicate device encompass the same range of physical dimensions, are packaged in similar materials, and are to be sterilized using similar methods. The data included in this submission demonstrate substantial equivalence to the predicate devices listed above.
{9}------------------------------------------------
| Subject Device | Indications for Use Statement | | |
|---------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------|------------------------|
| K231434<br>DESS Dental Smart<br>Solutions<br><br>Terrats Medical SL | DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in maxillary or mandibular arch to provide support for prosthetic restorations. | | |
| | Implant Systems Compatibility | Implant Body Ø, mm | Implant Platform Ø, mm |
| | Internal Hex Connections | | |
| | Zimmer 3.1mmD Eztetic™ | 3.1 | 2.9 |
| | | 3.7 | 3.5 |
| | Zimmer Screw-Vent® | 4.7 | 4.5 |
| | | 3.7, 4.1 | 3.5 |
| | Zimmer Tapered Screw-Vent® | 4.7 | 4.5 |
| | | 6.0 | 5.7 |
| | | 3.7, 4.1, 4.7 | 3.5 |
| | TSX™ Implant System | 5.4, 6.0 | 4.5 |
| | TSX™ Implant System, 3.1mmD | 3.1 | 2.9 |
| | External Hex Connections | | |
| | | 3.25 | 3.4 |
| | Biomet 3i OSSEOTITE® Implants | 3.75 | 4.1 |
| | | 4.0 | 4.1 |
| Primary Predicate<br>Device | Indications for Use Statement | | |
| K230143<br>DESS Dental Smart<br>Solutions<br><br>Terrats Medical SL | DESS Multi-Unit Abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations. | | |
| | Compatible Implant Systems | Implant Body Ø, mm | Implant Platform Ø, mm |
| | Internal Hex Connection | | |
| | | 3.7 | 3.5 |
| | Legacy1 | 4.2 | 3.5 |
| | | 4.7 | 4.5 |
| | | 3.7 | 3.5 |
| | Legacy2, simplyLegacy2, | 4.2 | 3.5 |
| | Legacy3, simplyLegacy3, | 4.7 | 4.5 |
| | Legacy4 | 5.2 | 4.5 |
| | Internal Conical Connection | | |
| | | 3.2 | 3.0 |
| | | 3.7 | 3.0 |
| | InterActive | 4.3 | 3.4 |
| | | 5.0 | 3.4 |
| | | 3.2 | 3.0 |
| | | 3.7 | 3.0 |
| | | 4.2 | 3.0 |
| | Simply Iconic™ | 4.7 | 3.0 |
| | | 4.7 | 3.4 |
| Terrats Medical SL | | 5.2 | 3.4 |
# Table of Substantial Equivalence – Indications for Use Statement
{10}------------------------------------------------
| | Subject Device | Primary Predicate Device | Reference Device |
|--------------------------------------------|------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| | K231434<br>DESS Dental Smart Solutions<br>Terrats Medical SL | K230143<br>DESS Dental Smart Solutions<br>Terrats Medical SL | K213063<br>TLX SRAs and TLX Gold<br>Abutments<br>Straumann USA, LLC |
| Reason for Predicate /<br>Reference Device | Not applicable | Designs, materials, manufacturing | Reference for multi-unit<br>abutments provided sterile |
| Product Codes | NHA | NHA | NHA |
| Intended Use | Functional and esthetic<br>rehabilitation of the edentulous<br>mandible or maxilla | Functional and esthetic<br>rehabilitation of the edentulous<br>mandible or maxilla | Functional and esthetic<br>rehabilitation of the edentulous<br>mandible or maxilla |
| Abutment Designs | | | |
| Abutment Types | Multi-Unit | Multi-Unit | Multi-Unit |
| Prosthesis Attachment | Screw Retained | Screw Retained | Screw Retained |
| Restoration | Multi-unit | Multi-unit | Multi-unit |
| Prosthetic Interface<br>Connections | Internal hex, External hex | Internal hex, Internal conical | Internal conical |
| Abutment/Implant Platform<br>Diameter | 2.9 - 5.7 mm | 3.0 - 4.5 mm | TLX SRA: 6 mm<br>TLX Gold: 4.0 (NT), 5.0 (RT),<br>and 7.0 (WT) |
| Prosthetic Platform Diameter | 4.8 mm | 4.8 mm | TLX SRA: 4.6 mm<br>TLX Gold: not provided in<br>510(k) Summary |
| Gingival Height | 1 mm - 5 mm | 1 mm - 5 mm | Not provided in 510(k)<br>Summary |
| Abutment Angulation, degrees | Straight (0°), 17°, 30° | Straight (0°), 17°, 30° | TLX SRA: 17°, 30°<br>TLX Gold: 0°, and up to 30° |
| Abutment Material | Ti-6Al-4V ELI | Ti-6Al-4V ELI | TLX SRA: Ti-6Al-7Nb<br>TLX Gold: Ceramicor® |
| Abutment Screw Material | Ti-6Al-4V ELI | Ti-6Al-4V ELI | Ti-6Al-7Nb |
| How Provided | | | |
| Sterilization | Non-sterile, and sterile by gamma<br>irradiation | Non-sterile | TLX SRA: Sterile by gamma<br>irradiation<br>TLX Gold: non-sterile |
| Usage - All Components | Single patient, single use | Single patient, single use | Single patient, single use |
# Table of Substantial Equivalence – Technological Characteristics
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.