Straumann® implants and abutments are intended for use in oral implantation to provide a support structure for connected prosthetic devices. Straumann® dental implants and abutments are intended for oral implantation to provide a support structure for connected prosthetic devices. Copings are indirectly connected to the endosseous dental implant and are indicated for use as an aid in prosthetic rehabilitations.
Device Story
TLX SRAs and TLX Gold Abutments are endosseous dental implant abutments used to support prosthetic restorations (crowns, bridges, full-arch). TLX SRAs are screw-retained components made of Ti-6Al-7Nb (TAN), provided sterile via gamma radiation, and available in 17° and 30° angulations. TLX Gold Abutments consist of a Ceramicor® alloy base with a TAN basal screw and POM-C modeling aid; they are provided non-sterile for end-user sterilization. Both systems connect to Straumann TLX dental implants via the TorcFit connection. Used in clinical dental settings by trained professionals, these devices provide the structural interface between the implant and the final prosthesis. They benefit patients by enabling stable, functional, and esthetic tooth replacement.
Clinical Evidence
Bench testing only. Dynamic fatigue testing performed per ISO 14801. Gold Abutments tested in saline at 2 million cycles; SRA Abutments tested in ambient air at 5 million cycles. Both demonstrated performance equivalent to predicates without failure. Biocompatibility evaluated per ISO 10993-1; no new issues identified.
Technological Characteristics
Materials: Ti-6Al-7Nb (TAN), Ceramicor® alloy, POM-C. Connection: TorcFit. Sterilization: Gamma radiation (SRAs) or end-user sterilization (Gold Abutments). Fatigue testing: ISO 14801. Biocompatibility: ISO 10993-1. Dimensions: 3.5mm height (SRAs); 4mm minimum post height, 0.7mm minimum wall thickness (Gold Abutments).
Indications for Use
Indicated for edentulous or partially edentulous patients requiring functional and esthetic oral rehabilitation with crowns, bridges, or full-arch prostheses. Copings are indicated as an aid in prosthetic rehabilitations.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
Predicate Devices
Straumann TLX Novaloc and Cementable Abutments (K203355)
Reference Devices
Straumann® BLX Line Extension Implants, SRAs, and Anatomic Abutments (K181703)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services - USA. To the right of this symbol, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
4/8/2022
Institut Straumann AG % Jennifer Jackson Director, Regulatory Affairs Straumann USA, LLC 60 Minuteman Road Andover, Massachusetts 01810
Re: K213063
Trade/Device Name: TLX SRAs and TLX Gold Abutments Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: March 9, 2022 Received: March 11, 2022
Dear Jennifer Jackson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Andrew Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K213063
Device Name TLX SRAs and TLX Gold Abutments
#### Indications for Use (Describe) TLX SRAs
Straumann® abutments are indicated to be placed into Straumann® dental implants to provide a support structure for the functional and esthetic oral rehabilitation of edentulous patients with crowns, bridges or full-arch prostheses.
### TLX Gold Abutments
Straumann® abutments are indicated to be placed into Straumann® dental implants to provide a support structure for the functional and esthetic oral rehabilitation of edentulous patients with crowns, bridges or full-arch prostheses.
Copings are indirectly connected to the endosseous dental implant and are indicated for use as an aid in prosthetic rehabilitations.
| Type of Use (Select one or both, as applicable) | | | | | | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------|---------------------------------------------------------------------------|
| <table border="0"><tr><td><table border="0"><tr><td><div align="left"> <input checked="checked" type="checkbox"/> </div></td><td><div align="left">Residential Use (Part 9, OBC 9.36) Select One</div></td></tr></table></td><td><table border="0"><tr><td><div align="left"> <input type="checkbox"/> </div></td><td><div align="left">Day Care Centre Use (Part 9, OBC 9.36) Select One</div></td></tr></table></td></tr></table> | <table border="0"><tr><td><div align="left"> <input checked="checked" type="checkbox"/> </div></td><td><div align="left">Residential Use (Part 9, OBC 9.36) Select One</div></td></tr></table> | <div align="left"> <input checked="checked" type="checkbox"/> </div> | <div align="left">Residential Use (Part 9, OBC 9.36) Select One</div> | <table border="0"><tr><td><div align="left"> <input type="checkbox"/> </div></td><td><div align="left">Day Care Centre Use (Part 9, OBC 9.36) Select One</div></td></tr></table> | <div align="left"> <input type="checkbox"/> </div> | <div align="left">Day Care Centre Use (Part 9, OBC 9.36) Select One</div> |
| <table border="0"><tr><td><div align="left"> <input checked="checked" type="checkbox"/> </div></td><td><div align="left">Residential Use (Part 9, OBC 9.36) Select One</div></td></tr></table> | <div align="left"> <input checked="checked" type="checkbox"/> </div> | <div align="left">Residential Use (Part 9, OBC 9.36) Select One</div> | <table border="0"><tr><td><div align="left"> <input type="checkbox"/> </div></td><td><div align="left">Day Care Centre Use (Part 9, OBC 9.36) Select One</div></td></tr></table> | <div align="left"> <input type="checkbox"/> </div> | <div align="left">Day Care Centre Use (Part 9, OBC 9.36) Select One</div> | |
| <div align="left"> <input checked="checked" type="checkbox"/> </div> | <div align="left">Residential Use (Part 9, OBC 9.36) Select One</div> | | | | | |
| <div align="left"> <input type="checkbox"/> </div> | <div align="left">Day Care Centre Use (Part 9, OBC 9.36) Select One</div> | | | | | |
> Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## TLX SRAs and TLX Gold Abutments
510(k) Summary
# 510(k) Summary
## Submitter's Contact Information
| Submitter: | Straumann USA, LLC<br>60 Minuteman Road<br>Andover, MA 01810<br>Registration No.: 1222315 Owner/Operator No.: 9005052<br>On the behalf of:<br>Institut Straumann AG<br>Peter Merian-Weg 12<br>4052 Basel, Switzerland |
|-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Jennifer M. Jackson, MS<br>Director of Regulatory Affairs and Quality<br>Phone Number: +1-978-747-2509<br>Fax Number: +1-978-747-0023 |
| Prepared By &<br>Alternate Contact: | Yu-Ting Chang<br>International Regulatory Affairs Manager<br>Institut Straumann AG<br>Phone number: +41 61 965 1318 |
| Date of Submission: | April 8, 2022 |
## Name of the Device
| Trade Names: | TLX SRAs and TLX Gold Abutments |
|------------------------|-------------------------------------|
| Common Name: | Endosseous Dental Implant Abutments |
| Classification Name: | Endosseous Dental Implant Abutments |
| Regulation Number: | §872.3630 |
| Device Classification: | II |
| Product Code(s): | NHA |
| Classification Panel: | Dental |
| Proprietary Name: | TLX SRAs and TLX Gold Abutments |
{5}------------------------------------------------
## TLX SRAs and TLX Gold Abutments
### 510(k) Summary
### Predicate Device(s)
Primary Predicate:
- K203355 Straumann TLX Novaloc and Cementable Abutments Cementable . Abutments
Reference Devices:
- K181703 Straumann® BLX Line Extension Implants, SRAs, and Anatomic Abutments . - SRAs
- K200597 BLX Gold Abutments .
### Device Description
#### TLX SRAs
Straumann® TLX SRAs are intended to be placed on TLX Straumann dental implants to provide support for screw-retained single or multi-unit restorations. They are identical in design to the reference devices BLX SRAs cleared as per K181703, except for the laser marking and the anodization. The TLX SRAs have no anodization, but additional laser marking at the connection. The TLX SRAs are provided sterile by gamma radiation. The TLX SRAs are made of TAN (Ti-6Al-7Nb) and are available in angulations of 17° and 30°. The TLX SRAs are secured in the implant through a basal screw made of TAN. The SRAs are provided with a transfer piece (Pin).
### TLX Gold Abutments
Straumann TLX Gold Abutments consist of a Ceramicor® alloy base, which is fixed to the implant by means of a basal screw made of Ti-6AI-7Nb, Titanium Aluminum Niobium (TAN). A modelling aid made of POM-C is attached onto the abutment by friction fit when delivered in the package. There are two versions of TLX Gold Abutments: one is for single crown restoration, and the other is for multiunit restoration such as bar or bridge. The TLX Gold Abutments for crown contain the TorcFit connection which provides the rotational indexation. On the other hand, the TLX Gold Abutments for bar/bridge do not have the TorcFit indexation feature since it is used in multiunit restorations. Both versions are available in three different platform diameters, NT (Narrow, TorcFit™), RT (Regular, TorcFit™) and WT (Wide, TorcFit™), allowing the abutments to sit on the shoulders of the corresponding TLX implants. The TLX Gold
{6}------------------------------------------------
## TLX SRAs and TLX Gold Abutments
### 510(k) Summary
Abutments have a maximum angulation of 30°, a minimum post height of 4 mm, and a minimum wall thickness of 0.7 mm.
### Intended Use
### TLX SRAs
Straumann® implants and abutments are intended for use in oral implantation to provide a support structure for connected prosthetic devices.
### TLX Gold Abutments
Straumann® dental implants and abutments are intended for oral implantation to provide a support structure for connected prosthetic devices.
Copings are indirectly connected to the endosseous dental implant and are indicated for use as an aid in prosthetic rehabilitations.
### Indications for Use
### TLX SRAs
Straumann® abutments are indicated to be placed into Straumann® dental implants to provide a support structure for the functional and esthetic oral rehabilitation of edentulous or partially edentulous patients with crowns, bridges, or full-arch prostheses.
### TLX Gold Abutments
Straumann® abutments are indicated to be placed into Straumann® dental implants to provide a support structure for the functional and esthetic oral rehabilitation of edentulous or partially edentulous patients with crowns, bridges, or full-arch prostheses.
Copings are indirectly connected to the endosseous dental implant and are indicated for use as an aid in prosthetic rehabilitations.
### Technological Characteristics
The technological characteristics of the subject devices are compared to the primary predicate and reference devices in the following tables:
{7}------------------------------------------------
# TLX SRAs and TLX Gold Abutments
510(k) Summary
### TLX SRAs
| FEATURE | PROPOSED DEVICE | PRIMARY PREDICATE DEVICE | REFERENCE DEVICE | EQUIVALENCE DISCUSSION |
|--------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K Number | K213063<br>Straumann TLX SRAs | K203355<br>Straumann TLX Cementable<br>Abutments | K181703<br>Straumann BLX SRAs | |
| Indications<br>for Use | Straumann® abutments are<br>indicated to be placed into<br>Straumann® dental implants to<br>provide a support structure for<br>the functional and esthetic oral<br>rehabilitation of edentulous or<br>partially edentulous patients with<br>crowns, bridges, or full-arch<br>prostheses. | Prosthetic components directly<br>connected to the endosseous<br>dental implant are intended for<br>use as an aid in prosthetic<br>rehabilitations. Final abutments<br>may be placed into occlusion for<br>implants with sufficient primary<br>stability and with appropriate<br>occlusal loading to restore<br>chewing function or for implants<br>that are fully osseointegrated | Prosthetic components directly<br>or indirectly connected to the<br>endosseous dental implant are<br>intended for use as an aid in<br>prosthetic rehabilitations.<br>Temporary components can be<br>used prior to the insertion of the<br>final components to maintain,<br>stabilize and shape the soft<br>tissue during the healing phase;<br>they may not be placed into<br>occlusion. Final abutments may<br>be placed into occlusion when<br>the implant is fully<br>osseointegrated. BLX Temporary<br>Abutments have a maximum<br>duration of usage of 180 days. | Equivalent.<br>All allow the abutment to be<br>place/connected to dental<br>implants and the purpose is to<br>aid/support the prosthetic<br>rehabilitation.<br>The subject device provides<br>more detail regarding the type of<br>rehabilitation These are also<br>implicitly covered under primary<br>predicate and reference device.<br>The final abutment placement<br>recommendation of the subject<br>device was moved from the<br>Indications for Use Statement to<br>the Caution/Precautions section<br>of the Instructions for Use since<br>the loading of the implant body is<br>inherent to the Indications for<br>Use of the overarching implant<br>body clearance.<br>Therefore, the scope is the same<br>among subject, primary<br>predicate, and reference<br>devices. |
| Material | Ti-6Al-7Nb (TAN) | Ti-6Al-7Nb (TAN) | Ti-6Al-7Nb (TAN) | Identical |
| Implant to<br>Abutment<br>Connection | TLX TorcFit | TLX TorcFit | BLX TorcFit | Identical<br>Identical to the primary predicate<br>device. |
| FEATURE | PROPOSED DEVICE | PRIMARY PREDICATE DEVICE | REFERENCE DEVICE | EQUIVALENCE DISCUSSION |
| K Number | K213063<br>Straumann TLX SRAs | K203355<br>Straumann TLX Cementable<br>Abutments | K181703<br>Straumann BLX SRAs | |
| Platform<br>Diameter | Ø4.6 mm | 4.0 (NT), 5.0 (RT) and 7.0 (WT) | Ø4.6 mm | Identical<br>Identical to the reference device |
| Platform | NT | NT, RT, and WT | RB/WB | Identical<br>Identical to the primary predicate<br>device. |
| Heights | 3.5 mm | Straight: 5,7 and 6 mm<br>Angled: 6 mm | 1.5, 2.5, 3.5 and 4.5 mm | Identical<br>Identical to the reference device |
| Angulation | 17° and 30° | 0° and 15 | 0°, 17° and 30° | Identical<br>Identical to the reference device |
| Sterilization<br>Method | Sterile by gamma radiation | Non-sterile | Sterile by gamma radiation | Identical<br>Identical to the reference device |
| Prosthesis<br>type | Screw-retained | Cemented-retained | Screw-retained | Identical<br>Identical to the reference device |
| Surface | Not anodized | Not anodized | Anodized | Identical<br>Identical to the primary predicate<br>device. |
| Shelf life | 5 years | Without defined shelf life | 5 years | Identical<br>Identical to the reference device |
| Packaging | Sterile barrier system (blister)<br>protected by folding box | Non-sterile blister | Sterile barrier system (blister)<br>protected by folding box | Identical<br>Identical to the reference device |
| FEATURE | PROPOSED DEVICE | PRIMARY PREDICATE DEVICE | REFERENCE DEVICE | EQUIVALENCE DISCUSSION |
| K Number | K213063<br>Straumann TLX Gold<br>Abutments | K203355<br>Straumann TLX Cementable<br>Abutments | K200597<br>BLX Gold Abutments | |
| Indications<br>for Use | Straumann® abutments are<br>indicated to be placed into<br>Straumann® dental implants to<br>provide a support structure for<br>the functional and esthetic oral<br>rehabilitation of edentulous or<br>partially edentulous patients with<br>crowns, bridges or full-arch<br>prostheses.<br>Copings are indirectly connected<br>to the endosseous dental implant<br>and are indicated for use as an<br>aid in prosthetic rehabilitations. | Prosthetic components directly<br>connected to the endosseous<br>dental implant are intended for<br>use as an aid in prosthetic<br>rehabilitations. Final abutments<br>may be placed into occlusion for<br>implants with sufficient primary<br>stability and with appropriate<br>occlusal loading to restore<br>chewing function or for implants<br>that are fully osseointegrated | Straumann abutments are<br>intended to be placed into<br>Straumann dental implants to<br>provide support for prosthetic<br>reconstructions such as crowns,<br>bridges, and overdentures. | Equivalent.<br>All allow the abutment to be<br>placed/connected to dental<br>implants for the purpose of<br>aiding/supporting the prosthetic<br>rehabilitation.<br>The subject device Indications<br>for Use provides more detail<br>concerning the rehabilitation. The<br>indications are also covered<br>under the primary predicate and<br>reference devices Indications for<br>Use.<br>The final abutment placement<br>recommendation for the subject<br>devices was moved from the<br>Indications for Use Statement to<br>the Caution/Precautions section<br>of the Instructions for Use since<br>the loading of the implant body is<br>inherent to the Indications for<br>Use of the overarching implant<br>body clearance.<br>The scope of the Indications for<br>Use is the same among subject,<br>primary predicate, and reference<br>devices. |
| Material | Ceramicor® (base)<br>TAN (screw)<br>POM (modeling aid) | Ti-6Al-7Nb (TAN) | Ceramicor® (base)<br>TAN (screw)<br>POM (modeling aid) | Identical<br>Identical to the reference device. |
| FEATURE | PROPOSED DEVICE | PRIMARY PREDICATE DEVICE | REFERENCE DEVICE | EQUIVALENCE DISCUSSION |
| K Number | K213063<br>Straumann TLX Gold<br>Abutments | K203355<br>Straumann TLX Cementable<br>Abutments | K200597<br>BLX Gold Abutments | |
| Abutment<br>Shape in<br>Coronal<br>Portion | Axial screw access hole for<br>placement of abutment into<br>implant | Axial screw access hole for<br>placement of abutment into<br>implant | Axial screw access hole for<br>placement of abutment into<br>implant | Identical<br>Identical to the primary predicate<br>and reference device. |
| Abutment to<br>Implant<br>Connection | TLX TorcFit | TLX TorcFit | BLX TorcFit | Identical<br>Identical to the primary predicate<br>device. |
| Platform<br>Diameter<br>(mm) | 4.0 (NT), 5.0 (RT) and 7.0 (WT) | 4.0 (NT), 5.0 (RT) and 7.0 (WT) | RB/WB: 3.8 and 4.5<br>WB: 4.5 and 5.5 | Identical<br>Identical to the primary predicate<br>device. |
| Post Height<br>of Stock<br>Component | NT (3 mm), RT (3.6 mm) and WT<br>(4 mm) | Straight: 5.7 and 6 mm<br>Angled: 6 mm | RW/WB and WB: 2.6 mm | Equivalent<br>The post heights are equivalent<br>to the reference device. In both<br>cases, the post provides the<br>initial support for casting. |
| Wall<br>thickness | 0.7*…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.