OsseoSpeed™ Plus implants of the Astra Tech Implant System Plus are intended to be used: · To replace missing teeth in single or multiple unit applications within the mandible or maxilla · For immediate placement in extraction sites and partially or completely healed alveolar ridge situations · For both one- and two-stage surgical procedures · Especially well in soft bone applications where implants with other implant surface treatments may be less effective · Together with immediate loading protocol in all indications, except in single tooth situations in soft bone (type IV) where implant stability may be difficult to obtain and immediate loading may not be appropriate. Abutments: Astra Tech Implant System Plus abutments are intended to be used in conjunction with the Astra Tech Implant System Plus in fully edentulous or partially edentulous maxillary and/or mandibular arches to provide support for crowns, bridges or overdentures.
Device Story
Root-form dental implants and abutments; used for tooth replacement in mandible or maxilla. Implants feature MicroThread neck design and OsseoSpeed surface; new 6-keyway anti-rotation interface for abutment positioning. Abutments include cover screws, healing abutments, and prosthetic components (cylindrical, asymmetric, angled). Used by dental clinicians in clinical settings. Preoperative planning supported by transparent radiographic guide sheets (1.0X-1.8X magnification). Device provides structural support for crowns, bridges, or overdentures; facilitates osseointegration. Benefits include restoration of dental function and aesthetics.
Clinical Evidence
Bench testing only; performance evaluated per ISO 14801 standards to demonstrate mechanical integrity and substantial equivalence.
Technological Characteristics
Root-form, self-tapping, threaded dental implants. Features MicroThread neck design and OsseoSpeed surface. Anti-rotation interface uses 6-keyway cross-section. Abutments available in various diameters, heights, and angles. Materials consistent with previously cleared predicates. Sterilization methods and packaging are equivalent to predicate systems.
Indications for Use
Indicated for patients requiring tooth replacement in mandible or maxilla, including immediate placement in extraction sites or healed ridges. Suitable for one- or two-stage surgery and soft bone applications. Contraindicated for immediate loading in single-tooth soft bone (type IV) situations where stability is insufficient.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
CAMLOG Implant System Modified Implants and Abutments (K083496)
Submission Summary (Full Text)
{0}------------------------------------------------
## 510(k) Summary
SEP 2 6 2011
# Astra Tech AB Astra Tech Implant System Plus
## K11287
September 26, 2011
## ADMINISTRATIVE INFORMATION
| Manufacturer Name: | Astra Tech AB |
|----------------------------|---------------------------------|
| | Aminogatan 1, P.O. Box 14 |
| | Mölndal, Sweden SE-431-21 |
| | Telephone: +46 31 776 30 00 |
| | Fax: +46 31 776 30 10 |
| Official Contact: | Christina Lewing |
| | Head of Regulatory Affairs |
| Representative/Consultant: | Linda K. Schulz, BSDH, RDH |
| | Kevin A. Thomas, PhD |
| | PaxMed International, LLC |
| | 11234 El Camino Real, Suite 200 |
| | San Diego, CA 92130 |
| | Telephone: +1 (858) 792-1235 |
| | Fax: +1 (858) 792-1236 |
| | email: lschulz@paxmed.com |
| | kthomas@paxmed.com |
## DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name: | Astra Tech Implant System Plus |
|-----------------------------|---------------------------------------------------------------------------------------------------|
| Component Name: | OsseoSpeed™ Plus implants and abutments |
| Common Name: | Dental implant and abutment |
| Classification Regulations: | Implant, endosseous, root form<br>Endosseous dental implant abutment<br>Class II, 21 CFR 872.3640 |
| Product Code: | DZE, NHA |
| Classification Panel: | Dental Products Panel |
| Reviewing Branch: | Dental Devices Branch |
{1}------------------------------------------------
#### 510(k) Summary
### INTENDED USE
OsseoSpeed™ Plus implants of the Astra Tech Implant System Plus are intended to be used:
- · To replace missing teeth in single or multiple unit applications within the mandible or maxilla
- · For immediate placement in extraction sites and partially or completely healed alveolar ridge situations
- · For both one- and two-stage surgical procedures
- · Especially well in soft bone applications where implants with other implant surface treatments may be less effective
- · Together with immediate loading protocol in all indications, except in single tooth situations in soft bone (type IV) where implant stability may be difficult to obtain and immediate loading may not be appropriate.
#### Abutments:
Astra Tech Implant System Plus abutments are intended to be used in conjunction with the Astra Tech Implant System Plus in fully edentulous or partially edentulous maxillary and/or mandibular arches to provide support for crowns, bridges or overdentures.
#### DEVICE DESCRIPTION
OsseoSpeed Plus implants are straight, self-tapping, threaded, root-form dental implants provided in multiple diameters and lengths with the previously cleared OsseoSpeed™ surface. Implants are available in diameters of 3.6 mm, 4.2 mm and lengths of 9 mm, 11 mm and 13 mm. The neck portion of the implant includes the MicroThread™ design. A new anti-rotation interface with 6 keyway cross section connections allows for six abutment positions. All other features and procedures of OsseoSpeed Plus implants are the same as presently cleared OsseoSpeed implants.
OsseoSpeed Plus abutment components consist of cover screws, healing abutments and abutments with corresponding screws. Cover screws and healing abutments are provided in sizes to match the implant platform. Healing abutments are provided in sizes to match the implant platform and various gingival diameters and gingival heights. OsseoSpeed Plus abutments have three design types, cylindrical, asymmetric and angled (20°). Each abutment design type is provided in three platform diameters (3.6, 4.2 and 4.8 mm), two prosthetic margin diameters (4.5 and 5.5 mm or 5.5 and 7.0 mm) and two gingival heights (low = 1.5- 2.5 mm and high = 2.5-3.5 mm) for a total of eleven abutment design combinations.
To assist in preoperative treatment planning of position and direction of implant placement, transparent radiographic implant guide sheets are provided with magnifications from 1.0X to 1.8X. The radiographic implant guides are usable with all OsseoSpeed Plus implants.
{2}------------------------------------------------
### EQUIVALENCE TO MARKETED DEVICE
Astra Tech AB demonstrated that, for the purposes of FDA 's regulation of medical devices, the Astra Tech Implant System is substantially equivalent in indications and design principles to the following predicate devices, each of which has been determined by FDA to be substantially equivalent to pre-amendment devices:
Astra Tech AB, Astra Tech Implant System cleared under K101732,
Astra Tech AB, Astra Tech Implant System cleared under K091239.
Astra Tech Inc., Fixture MicroThread™ OsseoSpeed™ cleared under K053384, and
Altatec GmbH, CAMLOG Implant System Modified Implants and Abutments cleared under K083496.
The subject device and the predicate devices have intended use, have the same technological characteristics and are made of the same materials. They encompass the same range of physical dimensions, including diameter and length of the implants, and diameter, height and angle of the abutments. The subject and predicate devices are packaged in similar materials and sterilized using similar methods. The technological change from the Astra Tech Implant System OsseoSpeed implants is the new anti-rotation feature of keyway cross section connections which is similar to the cam anti-rotation mechanism of the CAMLOG Implant System. Any differences in the technological characteristics do not raise new issues of safety or efficacy.
Performance testing was provided to demonstrate substantial equivalence and included methods described in ISO 14801.
The Astra Tech Implant System has the following similarities to the predicate devices:
- . has the same intended use.
- . uses the same operating principle,
- . incorporates the same basic design.
- . incorporates the same materials, and
- . has similar packaging and is sterilized using the same materials and processes.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the emblem.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Astra Tech AB C/O Ms. Linda K. Schulz Regulatory Affairs Paxmed International, LLC 11234 Camino Real. Suite 200 San Diego, California 92130
SEP 2 2011
Re: K111287
Trade/Device Name: Astra Tech Implant System Plus Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: DZE, NHA Product Code: 11 Dated: September 13. 2011 Received: September 14, 2011
Dear Ms. Schulz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2- Ms. Schulz
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
ff you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm 115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go 10 http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{5}------------------------------------------------
## Indications for Use
510(k) Number: K111287
Device Name: Astra Tech Implant System Plus
Indications for Use:
Implants:
OsseoSpeed™ Plus implants of the Astra Tech Implant System Plus are intended to be used:
- · To replace missing teeth in single or multiple unit applications within the mandible or maxilla
- · For immediate placement in extraction sites and partially or completely healed alveolar ridge situations
- · For both one- and two-stage surgical procedures
- · Especially well in soft bone applications where implants with other implant surface treatments may be less effective
- · Together with immediate loading protocol in all indications, except in single tooth situations in soft bone (type IV) where implant stability may be difficult to obtain and immediate loading may not be appropriate.
Abutments:
Astra Tech Implant System Plus abutments are intended to be used in conjunction with the Astra Tech Implant System Plus in fully edentulous or partially edentulous maxillary and/or mandibular arches to provide support for crowns, bridges or overdentures.
Prescription Use X (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH. Office of Device Evaluation (ODE)
Susan Ruse
Page 1 of 1
Division Sign-Off)
Division of Anesthesiology, General Hospital
fection Control, Dental Devices
510(k) Number: K111287
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.