K050705 · Nobel Biocare AB · DZE · May 19, 2005 · Dental
Device Facts
Record ID
K050705
Device Name
TIUNITE IMPLANTS
Applicant
Nobel Biocare AB
Product Code
DZE · Dental
Decision Date
May 19, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3640
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Nobel Biocare TiUnite® Implants are root-form endosseous implants intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as an artificial tooth, in order to restore patient esthetics and chewing function. Nobel Biocare TiUnite® Implants are indicated for single or multiple unit restorations in splinted applications. Nobel Biocare TiUnite® Implants may be placed immediately and put into immediate function providing that the initial stability requirements detailed in the surgical manuals are satisfied. TiUnite implants are indicated for use in soft bone whenever immediate or early loading is applied. The TiUnite implants are preferred in these soft bone indications because bone formation is more rapid and greater than on machined surface implants resulting in better maintenance of initial implant stability, faster and stronger osseointegration, and higher success rates.
Device Story
TiUnite Implants are threaded, root-form dental implants; machined from titanium with a titanium oxide surface layer. Placed in upper or lower jaw bone via single or two-stage surgical procedures; support prosthetic devices like artificial teeth. Surface treatment promotes rapid bone-to-implant contact; enhances osseointegration; maintains initial stability in soft bone. Used by dental surgeons in clinical settings. Output is a stable foundation for dental prosthetics; enables immediate or early loading protocols; improves success rates in soft bone compared to machined implants.
Clinical Evidence
No clinical data provided in the summary. Substantial equivalence is based on technological characteristics and performance comparisons to the predicate device.
Technological Characteristics
Root-form threaded dental implants; material: titanium with titanium oxide (TiUnite) surface layer. Available in straight or tapered contours. Designed for endosseous placement in jaw bone.
Indications for Use
Indicated for patients requiring single or multiple unit dental restorations in the upper or lower jaw. Suitable for immediate or early loading, particularly in soft bone, to restore esthetics and chewing function.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
K050705
#### 510(k) Summary of Safety and Effectiveness 1.4
MAY 1 9 2005
| Submitted by: | Herbert Crane<br>Manager, Regulatory Affairs |
|----------------------------------------|----------------------------------------------------------------------------|
| Address: | Nobel Biocare USA LLC<br>22715 Savi Ranch Parkway<br>Yorba Linda, CA 92887 |
| Telephone: | (714) 282-4800, ext. 5074 |
| Facsimile: | (714) 282-9023 |
| Date of Submission: | March 18, 2005 |
| Classification Name: | Endosseous Implant (21 CFR 872.3640) |
| Trade or Proprietary<br>or Model Name: | TiUnite™ Implants |
| Legally Marketed Device(s): | Nobel Biocare Endosseous Implants (K041661) |
#### Device Description:
Nobel Biocare TiUnite" Implants are threaded, root-form dental implants intended for use in the upper and/or lower jaw to support prosthetic devices, such as artificial teeth, in order to restore patient esthetics and chewing function to partially or fully edentulous patients.
Til hite Implants machined from titanium and have a surface treatment consisting of a thin, uniform titanium oxide (TiUnite") layer. Nobel Biocare TiUnite Implants are available with a straight or tapered contour.
Nobel Biocare TiUnite" Implants may be placed in the oral cavity using either a single stage surgical procedure or a two stage surgical procedure. If a single stage procedure is used, the implants may be immediately loaded following insertion where good initial stability can be obtained.
Due to a more rapid formation and greater amount of bone to implant contact during healing. Nobel Biocare's TiUnites surfaced implants result in faster and stronger ossecuntegration and better maintenance of the initial implant stability than machined titanium implants. When piaced in soft bone and immediately loaded, the faster and stronger osseointegration of Nobel Biocare's TiUnites surfaced implant results in a higher success rate compared to machined implants
### Indications for Use:
Nobel Biocare TiUnite® Implants are root-form endosseous implants intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as an artificial tooth, in order to restore patient esthetics and chewing function. Nobel Biocare TilUnite® Implants are indicated for single or multiple unit restorations in splinted applications. Nobel Biocare TiUnite® Implants may be placed immediately and put into immediate function providing that the initial stability requirements detailed in the surgical manuals are satisfied.
TiUnite implants are indicated for use in soft bone whenever immediate or early loading is applied. The TiUnite implants are preferred in these soft bone indications because bone formation is more rapid and greater than on machined surface implants resulting in better maintenance of initial implant stability. faster and stronger osseointegration, and higher success rates
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines representing its body and wings. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter.
MAY 1 9 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Nobel Biocare AB C/O Mr. Herbert Crane Manager Regulatory Affairs Nobel Biocare USA, LLC 22715 Savi Ranch Parkway Yorba Linda, California 92887
Re: K050705
Trade/Device Name: TiUnite® Implants Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Implant Regulatory Class: II Product Code: DZE Dated: March 17, 2005 Received: March 18, 2005
Dear Mr. Crane:
We have reviewed your Section 510(k) premarket notification of intent to market the device WE have reviewed your becares researce is substantially equivalent (for the referenced above and have determined are to tegally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device interstate comments, or to that bay been reclassified in accordance with the forcement Amendinents, or to de vices that have obct (Act) that do not require approval of a prematice the Federal Food, Drug, and Cosment 1 10 ( 16) - 16) - 16, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 1 approval application (1 Mr ). - Four my - Four - San - San - Sat include controls provisions of the rich - 110 gisting of devices, good manufacturing practice, requirements for animons against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III If your device is classified (see ass rolling controls. Existing major regulations affecting (PMA), it may of subject to Such adaments. Title 21, Parts 800 to 898. In the Cl., Parts 800 to 898. In your device can be found in the Oode of receined in the Federal Register.
{2}------------------------------------------------
## Page 2 - Mr. Crane
Please be advised that FDA's issuance of a substantial equivalence determination does not Please be advised that IDA S Issualice of a babound. Vice complies with other requirements mean that FDA has made a decemmanent and regulations administered by other Federal agencies. of the Act of ally rederal statues and regulations, but not limited to: registration it You must comply with an the Het brequired (21 CFR Part 801); good manufacturing practice and ifsing (21 CFR Part 607), laceming (21 CFR Pat 820), regulation (21 CFR Part 820); and if requirements as set form in the quality brovisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 5 10(k) This letter will anow you to begin mailing of substantial equivalence of your device to a premarket notification. The PDA midness of bassification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Patt 801), If you desire specific advice for your do records on the Also, please note the regulation prease contact the Office of Ochiphants and entification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the may obtain other general miormation on and Consumer Assistance at its toll-free Division of Binas 2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Chiu Lin, Ph.D.
Chiu Lin. Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
-
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Device Name: TiUnite® Implants
Indications For Use:
dications I of Or Obe.
Nobel Biocare TiUnite® Implants are root-form endosseous implants intended to the surgical tooth, i Nobel Biocare Titlnite® Implants are foot-omr enobsedds in includes, such as an atificial toth, in
bone of the upper or lower jaw arches to proxtins . Nohol Rincere , such as bone of the upper or lower jaw arches to provide suppor Til bith Implants are indicated for order to restore patient esthetics and chewing function. Nobel Biocare Titlenter Intent malants
single or multiple unit restorations in splinted or non-splinted the intilal single or multiple unit restorations in splinted of nor-splined applications working the initial stability requirements
may be placed immediately and put into immediate funct may be place -----------------------------------------------------------------------------------------------------------------------------------------------------------------
TiUnite® implants are indicated for use in soft bone whenever immediate or early bading is applied. The TiUnite® implants are indicated for use in Sol bone indications because bone formation is more rapid and Tillnite® implants are preferred in these soft bother maintenance of initial implant stability, faster greater than on midonix.
and stronger osseointegration, and higher success rates.
Prescription Use (Part 21 CFR 801 Subpart D) ANDIOR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Susan Penn
Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Device
Page 1 of 1
510(k) Number.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.