Bencox M Stem (Bencox Hip System) and Bencox ID Cemented Stem (with Centralizer) & Bencox Bone Plug
Applicant
Corentec Co., Ltd.
Product Code
LPH · Orthopedic
Decision Date
Apr 29, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3358
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Bencox M Stem (Bencox Hip System) is intended for Cementless use in partial or total hip arthroplasty in primary or revision surgery for the following conditions: a. Non-inflammatory degenerative joint disease, such as avascular necrosis, osteoarthritis, traumatic arthritis b. Inflammatory degenerative joint disease, such as rheumatoid arthritis c. Treatment of non-union, femoral neck fracture and trochantric fractures of the proximal femur with head involvement, unmanageable using other techniques d. Patients with failed previous surgery where pain, deformity, or dysfunction persists e. Revision of previously failed total hip arthroplasty The Bencox ID Cemented Stem (with Centralizer) & Bencox Bone Plug is intended for cemented use in total or partial hip arthroplasty and in primary or revision surgery for the following conditions, a) Non-inflammatory. degenerative joint disease, such as avascular necrosis, osteoarthritis, traumatic arthritis b) Inflammatory degenerative joint disease, such as rheumatoid arthritis c) Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques d) Patients with failed previous surgery where pain, deformity, or dysfunction persists e) Revision of previously failed total hip arthroplasty The Centralizer is intended to be used with Bencox ID Cemented Stem to centralize the femoral stem within the intramedullary canal. The Bencox Bone Plug is intended to be used with cement procedures to control and restrict the flow of cement within the intramedullary canal.
Device Story
Bencox M Stem and Bencox ID Cemented Stem are femoral hip prostheses used in partial or total hip arthroplasty. Bencox M Stem is a double-tapered, press-fit, cementless stem; Bencox ID is a collarless, cemented stem. Accessories include a PMMA centralizer for stem alignment and a UHMWPE bone plug to restrict cement flow. Used in OR by orthopedic surgeons. Implants replace damaged femoral head/neck; restore joint function; reduce pain. Bencox M Stem uses Ti6Al4V alloy with plasma-sprayed titanium coating. Bencox ID uses stainless steel. Instrumentation set includes rasps and impactors. Devices are supplied sterile (gamma radiation).
Clinical Evidence
Bench testing only. Performance documented via FEA, mechanical fatigue testing, static pull-off testing, rotational torque testing, dynamic compression testing, corrosion testing, and range of motion testing. Testing performed per ISO 7206-4, ISO 7206-6, ISO 21535, ISO 7206-9, ISO 7206-10, ASTM F2009, and ASTM F1875. Sterilization validated per ISO 11137-1 & 2; shelf-life validated per ASTM F1980.
Technological Characteristics
Bencox M Stem: Ti6Al4V alloy (ASTM F136), plasma-sprayed pure titanium coating (ASTM F1580). Bencox ID Stem: Stainless steel. Centralizer: PMMA. Bone Plug: UHMWPE. Design: Tapered rectangular/double-tapered geometry, 12/14 conical taper for head coupling. Sterilization: Gamma radiation (SAL 10^-6). Shelf life: 5 years.
Indications for Use
Indicated for patients requiring partial or total hip arthroplasty (primary or revision) due to non-inflammatory degenerative joint disease (avascular necrosis, osteoarthritis, traumatic arthritis), inflammatory degenerative joint disease (rheumatoid arthritis), non-union, femoral neck/trochanteric fractures, or failed previous surgery.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. In the center of the circle is an abstract image of a human face in profile, with three overlapping faces.
April 29, 2016
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Corentec Company, Limited J.S. Daniel Senior Manager/Engineer - RA & OA 8F Chungho Tower, 483, Gangnam-daero Seocho Gu, Seoul, 137-040 KOREA
Re: K152084
Trade/Device Name: BENCOX M Stem and Bencox ID Cemented Stem (with Centralizer) & Bone Plug Regulation Number: 21 CFR 888.3358 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: LPH, JDI, KWY, LZO Dated: March 28, 2016 Received: March 29, 2016
Dear J.S. Daniel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing
{1}------------------------------------------------
(21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Pq.1/1
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
#### 510(k) Number (if known) K152084
#### Device Name
Bencox M Stem (Bencox Hip System) and Bencox ID Cemented Stem (with Centralizer) & Bencox Bone Plug
#### Indications for Use (Describe)
Bencox M Stem (Bencox Hip System) is intended for Cementless use in partial or total hip arthroplasty in primary or revision surgery for the following conditions:
- a. Non-inflammatory degenerative joint disease, such as avascular necrosis, osteoarthritis, traumatic arthritis
- b. Inflammatory degenerative joint disease, such as rheumatoid arthritis
c. Treatment of non-union, femoral neck fracture and trochantric fractures of the proximal femur with head involvement, unmanageable using other techniques
d. Patients with failed previous surgery where pain, deformity, or dysfunction persists
e. Revision of previously failed total hip arthroplasty
The Bencox ID Cemented Stem (with Centralizer) & Bencox Bone Plug is intended for cemented use in total or partial hip arthroplasty and in primary or revision surgery for the following conditions,
a) Non-inflammatory. degenerative joint disease, such as avascular necrosis, osteoarthritis, traumatic arthritis
b) Inflammatory degenerative joint disease, such as rheumatoid arthritis
c) Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques
d) Patients with failed previous surgery where pain, deformity, or dysfunction persists
e) Revision of previously failed total hip arthroplasty
The Centralizer is intended to be used with Bencox ID Cemented Stem to centralize the femoral stem within the intramedullary canal.
The Bencox Bone Plug is intended to be used with cement procedures to control and restrict the flow of cement within the intramedullary canal.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(K) SUMMARY
Corentec Co., Ltd. Bencox M Stem (Bencox Hip System) and Bencox ID Cemented Stem (with Centralizer) & Bone Plug 25th April 2016
### ADMINISTRATIVE INFORMATION
| Manufacturer | Corentec Co., Ltd. |
|--------------|---------------------------------------------------|
| | 12, Yeongsanhong 1-gil, Ipjang-Myeon, Seobuk-Gu |
| | Cheonan-si, Chungchongnam-do, 331-822 South Korea |
| | Telephone: +82-41-585-7114 |
| | Fax: +82-41-585-7113 |
| Official Contact J.S. Daniel |
|--------------------------------------|
| Senior Manager/Engineer - RA/QA |
| Corentec Co., Ltd |
| 8F Chungho Tower, 483, Gangnam-daero |
| Seocho Gu, Seoul, Korea 137-040 |
| Ph: +82 70 4393 3819 |
| Fax: +82 2 3445 5467 |
| Email: jsdaniel@corentec.com |
#### DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name: | Bencox M Stem (Bencox Hip System) & Bencox ID Cemented Stem (with Centralizer) and Bencox Bone Plug |
|-----------------------------|-----------------------------------------------------------------------------------------------------|
| Common Name: | Hip Prosthesis |
| Classification Regulations: | 21 CFR 888.3358, 888.3390, 888.3350, 888.3353 |
| Class: | Class II |
| Product Codes: | LPH, JDI, KWY, LZO |
| Classification Panel: | Orthopedic Products Panel |
| Reviewing Branch: | Orthopedic Devices Branch |
{4}------------------------------------------------
### INTENDED USE
#### Bencox M Stem (Bencox Hip System)
Bencox M Stem (Bencox Hip System) is intended for Cementless use in partial or total hip arthroplasty in primary or revision surgery for the following conditions:
a. Non-inflammatory degenerative joint disease, such as avascular necrosis, osteoarthritis, traumatic arthritis
b. Inflammatory degenerative joint disease, such as rheumatoid arthritis
c. Treatment of non-union, femoral neck fracture and trochantric fractures of the proximal femur with head involvement, unmanageable using other techniques
d. Patients with failed previous surgery where pain, deformity, or dysfunction persists
e. Revision of previously failed total hip arthroplasty
#### BENCOX ID Cemented Stem (with Centralizer) & Bone Plug
The Bencox ID Cemented Stem (with Centralizer) & Bone Plug is intended for cemented use in total or partial hip arthroplasty and in primary or revision surgery for the following conditions.
- a) Non-inflammatory. degenerative joint disease, such as avascular necrosis, osteoarthritis, traumatic arthritis
- b) Inflammatory degenerative joint disease, such as rheumatoid arthritis
- c) Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques
- d) Patients with failed previous surgery where pain, deformity, or dysfunction persists
- e) Revision of previously failed total hip arthroplasty
The Centralizer is intended to be used with Bencox ID Cemented Stem to centralize the femoral stem within the intramedullary canal.
The Bencox Bone Plug is intended to be used with cemented hip replacement procedures to control and restrict the flow of cement within the intramedullary canal.
### DEVICE DESCRIPTION
Bencox M Stem (Bencox Hip System) - Sterile
The Bencox M Stem is a double tapered wedge shaped stem intended for press-fit partial or total hip arthroplasty. When used in total hip arthroplasty the stem is coupled to a Femoral Head and used with Bencox Acetabular Cups cleared under K103431, K112019,
{5}------------------------------------------------
K120924. K121665 & K150007. When used in partial hip arthroplasty Bencox M Stem is coupled to a Femoral Head and used with Bencox Bipolar Cup cleared under K112019.
Bencox M Stem is made of Ti6AI4V alloy conforming to ASTM F136. The proximal half of the stem is plasma sprayed coating of pure titanium as per ASTM F1580. The materials used in subject device, Bencox M Stem, have been cleared for use in K112019 & K150007 for Bencox ID Stem & Bencox ID Stem (offset). The stem has a tapered rectangular section. The stem has a neck with a standard 12/14 conical taper to couple to Bencox Femoral Heads which have been cleared earlier. The Bencox M Stem is available 13 sizes with CCD angle of 132°. The stem is available in 13 sizes.
All the above described technological characteristics of subject device is similar to the identified predicate devices.
### BENCOX ID Cemented Stem (with Centralizer) & Bone Plug - Sterile
Bencox ID Cemented Stem has bone plug and centralizer as accessories. The Collarless Bencox ID Cemented Stem consists of a polished double tapered distal geometry which generates the radial compressive load in distal body. The neck shape is similar to cleared Corentec's hip femoral stems in K103431, K122099 and K150007. The standard Morse taper (12/14) is used for femoral head assembly similar to cleared devices to couple to Bencox Femoral Heads and their compatible acetabular components which have been cleared in previous premarket notifications. The Bencox ID Cemented Stem is available 06 sizes with CCD angle of 132°. Bencox ID Cemented Stem is made of SS allov.
The deformable centralizer is made of PMMA with flexible wing like projections that extend outwards from the base of the centralizer and fits over the outside diameter of the stem tip. These projections contact the walls of the femoral canal, centralizing the hip stem within the canal. The hollow centralizer is designed to allow stem to engage distally within the cement mantle, subjecting the cement to compressive loading, and reducing end bearing of the stem directly onto the cement. The centralizer is a molded acrylic device manufactured from PMMA.
Bencox bone plug has a double flanged geometry very similar to the predicate devices. The plug blocks the propagation of the doughy bone cement downward in the medullary cavity. The proximal flange provides a proximal barrier, while the more flexible distal flange forms a secondary seal to prevent cement leakage. Bencox bone plug is manufactured from UHMWPE.
All the above described technological characteristics of subject device is similar to the identified predicate devices.
{6}------------------------------------------------
#### Accessory BENCOX Hip Instrumentation: (Non Sterile)
The Bencox Hip Instrumentation is a set of accessories to be used with Bencox Hip Implants. The instruments are designed to be simple, conventional, and accurate and all parts of which are used for their respective procedures by qualified orthopedic surgeons.
The parts of the instruments are made of stainless steel and/or polymers and/or silicon which are cleared for use in previous premarket notifications of Corentec.
Bencox Hip System Instrumentation to be used Bencox M Stem (Bencox Hip System) & Bencox ID Cemented Stem is same with Instrumentation cleared under K103431, K112019, K120924 & K122099 along with the inclusion of instruments specific to Bencox M Stem such as Rasp, Impactor etc.
#### SUBSTANTIAL EQUIVALENCE
The subject devices are substantially equivalent in indications, design principles and technological characteristics to the following predicate devices, each of which has been determined by FDA to be substantially equivalent, as below,
| Subject<br>Devices | Predicate Device/s | | | |
|--------------------------------------------------------|-----------------------|------------------------|----------------------------------|----------------------------------|
| | Predicate<br>Category | Manufacturer | Trade Name | 510(k) |
| BENCOX M<br>STEM | Primary | DePuy | TriLock | K010367 |
| | Additional | Stryker | Allocade II | K103479 &<br>K120578 |
| BENCOX<br>ID<br>Cemented<br>STEM (with<br>Centralizer) | Primary | Stryker<br>(Howmedica) | Exeter/Exeter V40 Stem | K803126;<br>K011623; K110290 |
| | | | Exeter Distal Centralizer | K974054 |
| | Additional | DePuy | C Stem AMT Stem | K042959 |
| | | | C Stem AMT Distal<br>Centralizer | K042959 &<br>K013350 |
| BENCOX<br>BONE<br>PLUG | Primary | Smith &<br>Nephew | Buck Cement Restrictor | K791125 (Referred<br>in K023680) |
| | Additional | Zimmer | Allen Medullary Bone Plug | K001733 |
As described in the Device Description section, the technological characteristics, such as design, material, indications, and operating principles are similar to the identified predicate devices. The minor differences do not affect the safety and effectiveness of the device when used for the said indications, based on non clinical testing.
{7}------------------------------------------------
### PERFORMANCE DATA
The Bencox M Stem (Bencox Hip System) and BENCOX ID Cemented Stem was subjected to a series of testing protocols to document the performance of the components and to demonstrate substantial equivalence. Non-clinical testing and analysis included FEA, mechanical fatigue testing, and static pull off testing, rotational torque testing, dynamic compression testing, corrosion and range of motion testing.
The testing methods are described in the standards, ISO 7206-4. ISO 7206-6, ISO 21535, ISO 7206-9, ISO 7206-10, ASTM F2009, and ASTM F1875. The results of this testing showed that the subject devices are expected to be safe and effective for the proposed indications and are substantially equivalent to the predicate devices.
### STERILIZATION
For all sterile implants, following to gamma sterilization, the packaging was subjected to sterile barrier testing to validate a shelf life of 5 years as per ISO & ASTM standards which confirms the stability and effectiveness of packaging of the sterilized product during the shelf-life, by evaluating changes by accelerated aging, as per ASTM F1980. Sterilization validation as per ISO 11137-1 & 2 Sterilization of health care products -Radiation ensures sterility of the components for an SAL of 10°.
All non sterile instruments used in the surgery must be sterilized by the hospital, prior to use, as mentioned in the instrument IFU.
Both Bencox M Stem (Bencox Hip System) and BENCOX ID Cemented Stem (with Centralizer) & Bone Plug have same shelf life of 5 years, similar to other cleared sterile devices of Corentec Co., Ltd.
#### CONCLUSION
Overall, the Bencox M Stem (Bencox Hip System) and BENCOX ID Cemented Stem (with Centralizer) & Bone Plug has similarities to the predicate devices with the same intended use, same fundamental scientific technology, similar design features, same operating principles, same materials and are supplied Sterile. The safety and effectiveness of the subject devices are adequately supported by the substantial equivalence information, material information and analysis of data in this PMN. Any differences in technological characteristic between the subject and predicate devices do not raise new issues of safety or efficacy.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.