The Bencox Hip System is intended for cementless use in total or partial hip arthroplasty in primary or revision surgery for the following conditions: a. Non-inflammatory degenerative joint disease, such as avascular necrosis, osteoarthritis, traumatic arthritis b. Inflammatory degenerative joint disease, such as rheumatoid arthritis c. Treatment of non-union, femoral neck fracture and trochantric fractures of the proximal femur with head involvement, unmanageable using other techniques d. Patients with failed previous surgery where pain, deformity, or dysfunction persists e. Revision of previously failed total hip arthroplasty
Device Story
Bencox II Stem and Bencox Metal Head are hip prosthesis components for cementless total or partial hip arthroplasty. Modification includes proximal neck region geometry change for the stem and addition of an extra-large (XL) neck length for the metal head. Components are implanted by orthopedic surgeons during hip replacement surgery to restore joint function and alleviate pain. Device functions as a mechanical replacement for the femoral component of the hip joint. Performance validated via fatigue strength and range of motion testing per ISO 7206-4, ISO 7206-6, and ISO 21535. Sterilized via radiation (ISO 11137-1 & 2) with 5-year shelf life.
Clinical Evidence
Bench testing only. Fatigue strength and range of motion testing performed per ISO 7206-4, ISO 7206-6, and ISO 21535. Results demonstrate performance similar to or better than predicate devices.
Technological Characteristics
Materials: Ti-6Al-4V alloy (ASTM F136) and Co-Cr-Mo alloy (ASTM F1537). Components: Hip prosthesis stem and metal head. Sterilization: Radiation (ISO 11137-1 & 2). Shelf life: 5 years. Connectivity: None (mechanical implant).
Indications for Use
Indicated for patients requiring total or partial hip arthroplasty (primary or revision) due to non-inflammatory degenerative joint disease (e.g., avascular necrosis, osteoarthritis, traumatic arthritis), inflammatory degenerative joint disease (e.g., rheumatoid arthritis), non-union, femoral neck/trochanteric fractures with head involvement, or failed previous surgery.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
{0}------------------------------------------------
Special 510(k): Device Modification
K12209
# 510(K) SUMMARY
Corentec Co., Ltd.
ОСТ 5 2012
r
# Bencox II Stem & Bencox Metal Head
July 10th, 2012
#### ADMINISTRATIVE INFORMATION
Manufacturer Corentec Co., Ltd. 247 Giro-ri, Ipjang-myeon, Seobuk-Gu Cheonan-si, Chungchongnam-do, 331-822 South Korea Telephone: +82-41-585-7114 Fax: +82-41-585-7113
J.S. Daniel Official Contact Project Manager - RA/QA Corentec Co., Ltd 8th Chungho Tower, 748-1 Banpo I Dong Seocho Gu, Seoul, Korea 137-040 Ph: +82 70 4393 3819 Fax: +82 2 3445 5467 Email: jsdaniel@corentec.com
#### DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name: | Bencox II Stem & Bencox Metal Head |
|-----------------------------|------------------------------------|
| Common Name: | Hip Prosthesis |
| Classification Regulations: | 21 CFR 888.3358; 888.3390 |
| Class: | II |
| Product Codes: | LPH; KWY |
| Classification Panel: | Orthopedic Products Panel |
| Reviewing Branch: | Orthopedic Devices Branch |
{1}------------------------------------------------
Special 510(k): Device Modification
Bencox II Stem & Bencox Metal Head
## INTENDED USE
The Bencox Hip System is intended for cementless use in total or partial hip arthroplasty in primary or revision surgery for the following conditions:
a. Non-inflammatory degenerative joint disease, such as avascular necrosis, osteoarthritis, traumatic arthritis
b. Inflammatory degenerative joint disease, such as rheumatoid arthritis
c. Treatment of non-union, femoral neck fracture and trochantric fractures of the proximal femur with head involvement, unmanageable using other techniques d. Patients with failed previous surgery where pain, deformity, or dysfunction persists e. Revision of previously failed total hip arthroplasty
## DEVICE DESCRIPTION
Bencox II Stem cleared under K103431 is modified at the proximal neck region. The other aspects of design / material / specification / manufacturing & sterilization / packaging are not altered and remain the same as discussed in K103431. The component is manufactured from Ti-6Al-4V alloy conforming to ASTM F136 Standard Specification for Wrought Titanium-6 Aluminum-4 Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications (UNS R56401); Co-Cr-Mo alloy conforming to ASTM F1537 Standard Specification for Wrought Cobalt-28-Chromium-6-Molybdenum Alloys for Surgical Implants.
Bencox Metal Head is cleared under K103431 & K112019 with the neck lengths of small, medium and large. This submission includes extra large (XL) neck length as an inclusion to the existing specification. The other aspects of design / material / specification / manufacturing & sterilization/packaging are not altered and remain the same as discussed in K103431 & K112019. The component is manufactured from Co-Cr-Mo alloy conforming to ASTM F1537 Standard Specification for Wrought Cobalt-28-Chromium-6-Molybdenum Alloys for Surgical Implants.
## SUBSTANTIAL EQUIVALENCE
The Bencox II Stem & Bencox Metal Head are substantially equivalent in indications and design principles to the following predicate devices, each of which has been determined by FDA to be substantially equivalent, as below,
- . Corentec Co., Ltd., Coren Total Hip System (now renamed as Bencox Total Hip System) cleared under K103431
- Corentec Co., Ltd., Bencox Bipolar Cup System cleared under K112019 .
{2}------------------------------------------------
Bencox II Stem & Bencox Metal Head
## PERFORMANCE DATA
The Bencox II Stem neck change & Bencox Metal Head components were subjected to a series of testing protocols to document the performance of the components and to demonstrate substantial equivalence for fatigue strength and range of motion and the included methods are described in the standards, ISO 7206-4, ISO 7206-6 & ISO 21535.
The Bencox Metal Head-XL was also tested with Bencox ID Stem (machined) & Bencox ID Stem (forged). Bencox 11 Stem neck change & Bencox Metal Head-XL components performed either similar or better than the predicate devices cleared under K103431 & K112019.
## STERILIZATION & PACKAGING
Sterilization validation as per ISO 11137 - 1 & 2 Sterilization of health care products - Radiation ensures sterility of the components for an SAL of 106. Both Bencox Il Stem & Bencox Metal Head and the predicate devices cleared under K103431 & K112019 have same shelf life. Following sterilization, packaging was subjected to sterile barrier testing to validate a shelf life of 5 years for both components.
Any differences in technological characteristic between the subject and predicate devices do not raise new issues of safety or efficacy. Overall, the Bencox II Stem neck change & Bencox Metal Head-XL has similarities to the predicate devices with the same intended use, same fundamental scientific technology, same operating principles, same materials and are supplied Sterile.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes representing the three levels of government: federal, state, and local. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Corentec Company, Limited % Mr. J.S. Daniel Project Manager - QA/RA 11th Chungho Tower, 748-1 Banpo 1 Dong Seocho Gu, Seoul, Korea 137-040
OCT 5 2012
Re: K122099
Trade/Device Name: Bencox II Stem & Bencox Metal Head Regulation Number: 21 CFR 888.3358 Regulation Name: Hip Joint metal/polymer/meta1 semi-constrained porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: LPH; KWY Dated: September 4, 2012 Received: September 6, 2012
Dear Mr. Daniel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{4}------------------------------------------------
Page 2 - Mr. J.S. Daniel
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Erine l. Keith
-> Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## INDICATIONS FOR USE
# 510(k) Number (if known): K122099
Device Name: Bencox II Stem & Bencox Metal Head
The Bencox Hip System is intended for cementless use in total or partial hip arthroplasty in primary or revision surgery for the following conditions:
a. Non-inflammatory degenerative joint disease, such as avascular necrosis, osteoarthritis, traumatic arthritis
b. Inflammatory degenerative joint disease, such as rheumatoid arthritis
c. Treatment of non-union, femoral neck fracture and trochantric fractures of the proximal femur with head involvement, unmanageable using other techniques d. Patients with failed previous surgery where pain, deformity, or dysfunction persists e. Revision of previously failed total hip arthroplasty
Prescription Use: X (Per 21 CFR 801 Subpart D)
AND / OR
Over-The Counter Use :-(Per 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
AS~b
(Division Sign-Oft) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K122099
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.