The Bencox Forte and Bencox Delta ceramic heads is intended for use in total or partial hip arthroplasty in primary or revision surgery for the following conditions: a. Non-inflammatory degenerative joint disease, such as avascular necrosis, osteoarthritis, traumatic arthritis b. Inflammatory degenerative joint disease, such as rheumatoid arthritis c. Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques d. Patients with failed previous surgery where pain, deformity, or dysfunction persists e. Revision of previously failed hip arthroplasty
Device Story
Bencox Forte and Bencox Delta are ceramic femoral head prostheses designed for total or partial hip arthroplasty. The device acts as the articulating component of a hip system, mechanically fixed to a mating hip stem via a standard 12/14 taper. It articulates against UHMWPE (highly cross-linked or conventional) liners. Used in clinical settings by orthopedic surgeons during primary or revision hip surgery. The device restores joint function and reduces pain in patients with degenerative joint disease or fractures. It is manufactured from bio-inert alumina ceramic (Al2O3).
Clinical Evidence
Bench testing only. Testing included component burst strength testing of BIOLOX forte ball heads on titanium test tapers per ISO 7206-10 and influence of diameter/neck length on burst strength. Sterilization validation performed per ISO 11137-1 and 11137-2 to a SAL of 10^-6.
Technological Characteristics
Material: Alumina ceramic (Al2O3) per ASTM F603 / ISO 6474. Form factor: Femoral ball heads, 28-40 mm diameter. Interface: 12/14 taper. Sterilization: Gamma radiation (Co-60).
Indications for Use
Indicated for patients requiring total or partial hip arthroplasty (primary or revision) due to non-inflammatory degenerative joint disease (avascular necrosis, osteoarthritis, traumatic arthritis), inflammatory degenerative joint disease (rheumatoid arthritis), femoral neck/trochanteric fractures with head involvement, or failed previous hip surgery.
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
Coren Total Hip System & Bencox Bipolar Cup System (K103431 & K112019)
Submission Summary (Full Text)
{0}------------------------------------------------
K121665 (pg 1/3)
# 510(K) SUMMARY
EP 25 2012
### Corentec Co., Ltd.
# BENCOX® Forte & BENCOX® Delta
4th June, 2012
#### ADMINISTRATIVE INFORMATION
| Manufacturer | Corentec Co., Ltd. |
|------------------|---------------------------------------------------|
| | 247 Giro-ri, Ipjang-myeon, Seobuk-Gu |
| | Cheonan-si, Chungchongnam-do, 331-822 South Korea |
| | Telephone: +82-41-585-7114 |
| | Fax: +82-41-585-7113 |
| Official Contact | J.S. Daniel |
| | Project Manager - RA/QA |
| | Corentec Co., Ltd |
| | 11th Chungho Tower, 748-1 Banpo 1 Dong |
| | Seocho Gu, Seoul, Korea 137-040 |
| | Ph: +82 2 3445 5492 |
| | Fax: +82 2 3445 5497 |
Email: jsdaniel@corentec.com
## DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name: | Bencox Forte & Bencox Delta |
|-----------------------------|-----------------------------|
| Common Name: | Femoral Head Prosthesis |
| Classification Regulations: | 21 CFR 888.3353 |
| Class | II |
| Product Codes: | LZO |
| Classification Panel: | Orthopedic Products Panel |
| Reviewing Branch: | Orthopedic Devices Branch |
#### INTENDED USE
The Bencox Forte and Bencox Delta ceramic heads is intended for use in total or partial hip arthroplasty in primary or revision surgery for the following conditions: a. Non-inflammatory degenerative joint disease, such as avascular necrosis, osteoarthritis, traumatic arthritis
b. Inflammatory degenerative joint disease, such as rheumatoid arthritis
{1}------------------------------------------------
c. Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques d. Patients with failed previous surgery where pain, deformity, or dysfunction persists e. Revision of previously failed hip arthroplasty
### DEVICE DESCRIPTION
The Bencox Biolox Ceramic Head is a component of a Hip Prosthesis System and intended for mechanical fixation to a mating hip stem, with a standard 12/14 taper. It is intended to articulate between both Highly Cross Linked UHMWPE as well as Conventional UHMWPE.
The Bencox Ceramic Head is available as standard type and is manufactured from bio-inert alumina ceramic, Al2O3 as per ASTM F603 / ISO 6474. The ceramic femoral balls heads range in diameter from 28 mm to 40 mm with small, medium & large versions.
Radiation (gamma) sterilization is used for Bencox Forte and Bencox Delta Biolox ceramic heads. Sterilization by Co-60 gamma irradiation has been validated to a sterility assurance level (SAL) of 10-6 by the bioburden method, according to ISO 11137-1 & ISO 11137-2. Following sterilization packaging was validated for a shelf life of 5 years for BENCOX Forte & Delta ceramic femoral heads.
### SUBSTANTIAL EQUIVALENCE TO MARKETED DEVICES
BENCOX Forte & BENCOX Delta Biolox components are substantially equivalent in indications and design principles to the predicate devices, each of which has been determined by FDA to be substantially equivalent:
- MEDACTA International SA; MectaCer BIOLOX forte Femoral Heads cleared . under K073337,
- DePuy Orthopedics, Inc., DePuy Delta Ceramic Femoral Head cleared under . K062748,
- Corentec Co., Ltd., Coren Total Hip System & Bencox Bipolar Cup System . cleared under K103431 & K112019 (Femoral Heads).
### PERFORMANCE TESTING
Performance testing of the Bencox Forte/Delta Femoral Heads was conducted in accordance with the international standards. All required testing per "Guidance Document for the Preparation of Premarket Notifications of Ceramic Ball Hip Systems" were performed.
{2}------------------------------------------------
- Component testing of BIOLOX forte ball head 28-12/14 L on titanium test tapers . per CeramTec AG test procedure VA 02 04 4129, ISO 7206-10.
- Influence of diameter and neck length on burst strength of BIOLOX forte and . BIOLOX delta ball heads with taper type 12/14. Burst test setup as per ISO 7206-10.
- Sterilization validation as per ISO 11137 1 & 2 Sterilization of health care . products - Radiation.
Any differences in technological characteristic between the subject and predicate devices do not raise new issues of safety or efficacy of the BENCOX Forte & BENCOX Delta Biolox components for its intended use.
Overail, the BENCOX Forte & BENCOX Delta Biolox components have the following similarities to the predicate devices:
- has the same intended use, .
- uses the same operating principles, .
- incorporates the same basic designs, .
- incorporates the same or similar materials, and ●
- has similar packaging and sterilized using the same materials and processes. .
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
SEP 25 2012
Corentec Co., Ltd. Seoul Office % J.S. Daniel· Project Manager - RA/QA 11th Chungho Tower 748-1 Banpo 1 Dong Seocho Gu, Seoul Republic of Korea 137-040
#### Re: K121665
Trade/Device Name: BENCOX® Forte & BENCOX® Delta Regulation Number: 21 CFR 888.3353 Regulation Name: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis Regulatory Class: Class II
Product Code: LZO Dated: August 28, 2012 Received: August 30, 2012
#### Dear Mr. Daniel
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. J.S. Daniel
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## INDICATIONS FOR USE
510(k) Number (if known):
Device Name: BENCOX Forte & BENCOX Delta
The Bencox Forte and Bencox Delta ceramic heads are intended for use in total or partial hip arthroplasty in primary or revision surgery for the following conditions:
- a) Non-inflammatory degenerative joint disease, such as avascular necrosis, osteoarthritis, traumatic arthritis
- b) Inflammatory degenerative joint disease, such as rheumatoid arthritis
- c) Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques
- d) Patients with failed previous surgery where pain, deformity, or dysfunction persists
- e) Revision of previously failed hip arthroplasty.
Prescription Use: X _ (Per 21 CFR 801 Subpart D) AND / OR
Over-The Counter Use: (Per 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
A. Suls.
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
×121665 510(k) Number
Corentec Co. Ltd
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.