Nexus Suture

K151009 · Baylis Medical Company, Inc. · GAT · Aug 18, 2015 · General, Plastic Surgery

Device Facts

Record IDK151009
Device NameNexus Suture
ApplicantBaylis Medical Company, Inc.
Product CodeGAT · General, Plastic Surgery
Decision DateAug 18, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.5000
Device ClassClass 2
AttributesTherapeutic

Intended Use

Nexus™ Suture is indicated for use in approximation and/or ligation of soft tissues, including use of allograft tissue for orthopedic surgeries.

Device Story

Nexus Suture is a nonabsorbable, sterile, surgical suture used for soft tissue approximation and ligation; including orthopedic allograft tissue. Device consists of ultra-high molecular weight polyethylene (UHMWPE) fibers, either undyed or braided with blue polypropylene strands. Provided in sizes 2-0, 0, and 2; available in various cut lengths without needles or on reels. Used by surgeons in clinical settings to secure tissues. Mechanism of action is mechanical tissue approximation/ligation, identical to predicate sutures. Benefits include reliable tissue fixation during surgical procedures.

Clinical Evidence

Bench testing only. Verification testing included biocompatibility, diameter testing, and tensile strength testing to demonstrate performance equivalence to predicate devices.

Technological Characteristics

Nonabsorbable surgical suture composed of ultra-high molecular weight polyethylene (UHMWPE) or UHMWPE braided with blue polypropylene. Blue colorant is [phthalocyaninato(2-)] copper (max 0.5% by weight). No coating. Meets USP requirements for non-absorbable surgical suture (except diameter). Sterile, provided in sizes 2-0, 0, and 2.

Indications for Use

Indicated for approximation and/or ligation of soft tissues, including allograft tissue for orthopedic surgeries.

Regulatory Classification

Identification

Nonabsorbable poly(ethylene terephthalate) surgical suture is a multifilament, nonabsorbable, sterile, flexible thread prepared from fibers of high molecular weight, long-chain, linear polyesters having recurrent aromatic rings as an integral component and is indicated for use in soft tissue approximation. The poly(ethylene terephthalate) surgical suture meets U.S.P. requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be provided uncoated or coated; and it may be undyed or dyed with an appropriate FDA listed color additive. Also, the suture may be provided with or without a standard needle attached.

Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a bird in flight, formed by three stylized human profiles facing to the right. The profiles are stacked on top of each other, creating a sense of depth and movement. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 August 18, 2015 Baylis Medical Company Incorporated Ms. Meghal Khakhar Director of Regulatory & Scientific Affairs 2645 Matheson Boulevard East Mississauga, Ontario L4W 5S4 CANADA Re: K151009 Trade/Device Name: Nexus™ Suture Regulation Number: 21 CFR 878.5000 Regulation Name: Nonabsorbable poly(ethylene terephthalate) surgical suture Regulatory Class: Class II Product Code: GAT Dated: June 30, 2015 Received: July 6, 2015 Dear Ms. Khakhar: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 {1}------------------------------------------------ CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # David Krause -S - for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K151009 Device Name Nexus™ Suture Indications for Use (Describe) Nexus™ Suture is indicated for use in approximation and/or ligation of soft tissues, including use of allograft tissue for orthopedic surgeries. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|---------------------------------------------| | Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 7. 510(k) Summary # Submitter Information - A. Company Name: Baylis Medical Company Inc. - B. Company Address: 2645 Matheson Blvd. East Mississauga, Ontario L4W 5S4 Canada - C. Company Phone: (905) 602-4875 - D. Company Facsimile: (905) 602-5671 - E. Contact Person: Meghal Khakhar, Director of Regulatory & Scientific Affairs - F. Summary Prepared on: 13-Apr-2015 # Device Identification - A. Device Trade Name: Nexus™ Suture - B. Device Common Name: Suture, nonabsorbable, synthetic, polyethylene - C. Classification Name: CFR 878.5000 Nonabsorbable poly(ethylene terephthalate) surgical suture - D. Product Code: GAT - E. Device Class: Class II # Identification of Predicate Devices The predicate devices are provided in Table 7.1. {4}------------------------------------------------ #### Table 7.1: Predicate Devices | Predicate Device | Manufacturer | 510(k) | |------------------------------------------------------------------------------|------------------|---------| | Force Fiber Polyethylene<br>Nonabsorbable<br>Surgical Suture | Teleflex Medical | K033654 | | Force Fiber Blue Co-Braid<br>Polyethylene Non-<br>Absorbable Surgical Suture | Teleflex Medical | K040472 | #### Indications for Use Nexus™ Suture is indicated for use in approximation and/or ligation of soft tissues, including use of allograft tissue for orthopedic surgeries. #### Device Description Nexus Suture is a nonabsorbable, sterile, surgical suture provided in sizes 2-0, 0 and 2. It is composed of either undyed (white) ultra high molecular weight polyethylene (UHMWPE) or white UHMWPE braided with one or two strands of blue polypropylene to add color. The blue colorant is [phthalocyaninato(2-)] copper with a concentration not to exceed 0.5% by weight. It does not have a coating. Nexus Suture will be provided to the user in a variety of cut lengths without needles or on reels that facilitate suture handling. Nexus Suture meets all USP requirements for non-absorbable surgical suture, except for oversized diameter. #### Comparison to Predicate Devices Nexus Suture and its predicate devices share the same intended use and fundamental scientific technology, including principles of operation and mechanism of action. The results of verification testing support the substantial equivalence determination to the predicate devices (Table 7.2). {5}------------------------------------------------ | Characteristic | Proposed White Nexus<br>Suture Compared to<br>Predicate Teleflex<br>Medical Suture<br>(K033654) | Proposed White/Blue<br>Nexus Suture Compared<br>to Predicate Teleflex<br>Medical Suture<br>(K040472) | |--------------------------------------|-------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------| | Intended Use | Identical | Identical | | Indications for Use | Identical | Identical | | Fundamental scientific<br>technology | Identical | Identical | | Operating principles | Identical | Identical | | Mechanism of action | Identical | Identical | | Technological aspects | Identical | Identical | | Available Sizes | Similar* | Similar* | #### Table 7.2: Comparison of Proposed and Predicate Devices *The 2-0, 0 and 2 sizes are identical. Additional sizes available in the predicate are not applicable to the subject device. ## Performance Testing To demonstrate substantial equivalence of the Nexus Suture to the predicate devices, bench top testing was performed. The devices were subject to the following verification tests: biocompatibility, diameter testing and tensile strength testing. ## Conclusions Nexus Suture and its predicate devices share the same intended use and fundamental scientific technology, including principles of operation and mechanism of action.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...