AcuPulse W CO2 Laser Systems, Delivery Devices and Accessories
K201663 · Lumenis, Ltd. · GEX · Jul 16, 2020 · General, Plastic Surgery
Device Facts
Record ID
K201663
Device Name
AcuPulse W CO2 Laser Systems, Delivery Devices and Accessories
Applicant
Lumenis, Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Jul 16, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Lumenis AcuPulse W CO2 Laser System, Delivery Devices and Accessories is indicated for incision, excision, ablation, vaporization and coagulation of body soft tissues in medical specialties including aesthetic (dermatology and plastic surgery), podiatry, otolaryngology (ENT), gynaecology, neurosurgery, orthopaedics, general and thoracic surgery. The use with the scanning unit is indicated for ablative skin resurfacing.
Device Story
AcuPulse W is a scaled-down CO2 laser system for soft tissue surgery. Input: electrical power; Output: CO2 laser energy delivered via articulated arm and handpieces/scanners. Operated by clinicians in clinical settings. System includes laser console (laser tube, power source, cooling, electronics), articulated arm, footswitch, and various delivery accessories. New 90° side-firing handpiece (Ti-6Al-4V) added. Software is identical to predicate AcuPulse (K180597), with features enabled/disabled via hardware key (HASP dongle). Clinicians use the system to perform precise tissue ablation or coagulation; output affects clinical decision-making by providing controlled surgical energy for various specialties. Benefits include precise tissue management and reduced thermal damage compared to traditional methods.
Clinical Evidence
Bench testing only. No clinical data presented. Verification included risk analysis (ISO 14971), electrical/laser safety (IEC 60601-1, IEC 60601-1-2, IEC 60825-1, IEC 60601-2-22), biocompatibility (ISO 10993), cleaning (AAMI TIR 30), and sterilization validation (ISO 17665-1) for the new 90° side-firing handpiece.
Technological Characteristics
CO2 laser system; 10.6 µm wavelength. Components: laser console, articulated arm, footswitch, handpieces. New handpiece material: Titanium (Ti-6Al-4V). Connectivity: hardware key (HASP dongle) for feature configuration. Sterilization: validated for reusable handpieces. Software: embedded, identical to predicate. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60825-1, IEC 60601-2-22, ISO 14971, ISO 10993, ISO 17665-1, AAMI TIR 30.
Indications for Use
Indicated for incision, excision, ablation, vaporization, and coagulation of soft tissues in dermatology, plastic surgery, podiatry, ENT, gynecology, neurosurgery, orthopedics, general, and thoracic surgery. Includes ablative skin resurfacing when used with a scanning unit.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
July 16, 2020
Image /page/0/Picture/1 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue.
Lumenis Ltd. Shlomit Segman Senior Manager RA 6 Hakidma Street PO Box 240 Yokneam, 2069204 Israel
Re: K201663
Trade/Device Name: AcuPulse W CO2 Laser Systems, Delivery Devices and Accessories Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: June 15, 2020 Received: June 19, 2020
Dear Shlomit Segman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Colin Kejing Chen Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
### K201663
#### Device Name
Lumenis AcuPulse W CO2 Laser System, Delivery Devices and Accessories
Indications for Use (Describe)
Lumenis AcuPulse W CO2 Laser System, Delivery Devices and Accessories is indicated for incision, ablation, vaporization and coagulation of body soft tissues in medical specialties including aesthetic (dermatology and plastic surgery), podiatry, otolaryngology (ENT), gynaecology, neurosurgery, orthopaedics, general and thoracic surgery. The use with the scanning unit is indicated for ablative skin resurfacing.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 20px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
*PRAStaff@fda.hhs.gov*
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*
{3}------------------------------------------------
#### K201663
# 510(k) SUMMARY
### The Lumenis AcuPulse W Laser System, Delivery Devices and Accessories
| Applicant Name: | Lumenis Ltd.<br>6 Hakidma Street PO Box 240<br>Yokneam Industrial Park,<br>Yokneam 2069204, Israel<br>Tel: +972-4-9599000<br>Fax: +972-4-9599046 |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Shlomit Segman Lumenis Ltd.<br>6 Hakidma Street PO Box 240,<br>Yokneam Industrial Park,<br>Yokneam 2069204, Israel<br>Tel: +972-4-9599230<br>Fax: +972-4-9599198<br>Email:Shlomit.Segman@lumenis.com |
| Date Prepared: | 15 June 2020 |
| Trade Name: | The Lumenis AcuPulse W Laser System, Delivery Devices and<br>Accessories |
| Classification Name: | Powered laser surgical instrument |
| Product Code: | GEX |
| Device Class: | Class II |
| Regulation Number: | 21 CFR 878.4810 |
| Panel: | General & Plastic Surgery |
| Predicate Devices: | AcuPulse Family of CO2Lasers system (K180597) |
### Intended Use/ Indications for Use:
Lumenis AcuPulse W CO2 Laser System, Delivery Devices and Accessories is indicated for incision, excision, ablation, vaporization and coagulation of body soft tissues in medical specialties including aesthetic (dermatology and plastic surgery), podiatry, otolaryngology (ENT), gynaecology, neurosurgery, orthopaedics, general and thoracic surgery. The use with the scanning unit is indicated for ablative skin resurfacing.
{4}------------------------------------------------
## Device Description:
The subject AcuPulse W system is a new scaled-down version of the recently FDA cleared Lumenis AcuPulse CO2 Laser System, Delivery Devices and Accessories under K180597. It is based on the following hardware components in the AcuPulse:
- . A Laser Console with a Free Beam Port to which an articulated arm attached.
- Identical to the AcuPulse, the Laser Console of the AcuPulse W System houses the laser O tube and optical system, the laser power source, console electronics, laser output ports, cooling system, purge air pump, control components (including touch screen, emergency stop button) and rear connector panel.
- . An articulated arm to guide laser output via a set of folding mirrors to the connected accessory for delivery of laser energy to the tissue.
- . A footswitch to activate the laser treatment beam and allow the selected laser energy to be transmitted via the delivery device to the target location.
- . A variety of Free Beam Delivery Device and accessories.
The AcuPulse and AcuPulse W systems have the same proprietary software, which is embedded in the Main Controller, Peripheral Controller units and PC. The differences between the two are the enabled or disabled software features, through a hardware key (HASP dongle), that reflects the capabilities of each laser system or the particular configuration of a given system. Importantly, both systems use the same Software version, which was already validated, verified and submitted under K180597.
The AcuPulse W is offered with a subset of the previously cleared AcuPulse Handpieces/Tips that connect to the articulated arm or/and scanners for controlled delivery of laser energy to the target tissue. In addition, Lumenis is adding a Titanium (Ti-6Al-4V) 90° Side-firing Handpiece design. Like the cleared handpieces, it has multiple uses. The 90° Side-firing handpiece connects to the cleared AcuScan 120 Microscanner through two adaptors (a Handpiece Adapter and Third Lens Adapter) for delivery of laser energy. The 90° Side-firing handpiece is comprised of the following components:
- Handpiece Adapter
- Tip (Standard or Fine)
- . Limiter ring (Standard or Fine)
This handpiece is based on the same technology and principles of operation of the cleared CO2 delivery devices and its introduction does not raise any new questions of safety and/or efficacy.
## Substantial Equivalence
The intended use of the AcuPulse W System is a shortened version of the cleared AcuPulse, with the changes to reflect the reduced options proposed for the AcuPulse W System and to follow the format that we have noted for the reference, competing device DEKA SmartXide2 (K133895) and other cleared CO2 laser systems (see also K101555, K063001, K160312 and K172096).
{5}------------------------------------------------
In addition, the same technological characteristics and principles of operation apply for all laser systems, Delivery Devices and accessories.
Design verification processes were performed as a result of this risk analysis assessment to verify that no different questions of safety and effectiveness have been raised due to the modifications introduced. The test methods are essentially the same as those used to support to the clearance of the AcuPulse CO2 Laser System, Delivery Devices and Accessories (K082809, K100415, K180597).
The following activities were performed:
- . Risk analysis (ISO 14971)
- Electrical and laser safety and electromagnetic compatibility testing as required to conform . to performance standards as follows:
- IEC 60601-1: Medical electrical equipment Part 1: General requirements for o basic safety and essential performance
- IEC 60601-1-2: Medical electrical equipment Part 1-2: General requirements for o basic safety and essential performance- Collateral standard: Electromagnetic compatibility- requirements and tests
- IEC 60825-1: Safety of laser Products-Part 1: Equipment classification and o requirements
- IEC 60601-2-22: Medical electrical equipment- part 2: particular requirements for o the safety of diagnostic and therapeutic laser equipment.
- . 90° Side-firing Handpiece testing demonstrating demonstrate that the accessory together with the AcuPulse W system performs in compliance with their specifications and requirements
- 90° Side-firing Handpiece Cleaning (AAMI TIR 30) and sterilization validation activities (ISO ● 17665-1)
- 90° Side-firing Handpiece Biocompatibility of materials (ISO 10993) .
Test results indicated that the new configuration AcuPulse W CO2 Laser System, Delivery Devices and accessories perform in accordance with its requirements and specifications similarly to its predicate device. Consequently, Lumenis Ltd. believes that AcuPulse W CO2 Laser System, Delivery Devices and Accessories is substantially equivalent to the cleared predicate and it does not raise any different questions of safety and/or effectiveness.