K101555 · Advanced Technology Laser Co., Ltd. · GEX · Sep 17, 2010 · General, Plastic Surgery
Device Facts
Record ID
K101555
Device Name
CO2 LASER SYSTEM
Applicant
Advanced Technology Laser Co., Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Sep 17, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
eBeam CO2 Laser System is indicated for use in surgical applications requiring the ablation, vaporization, excision, incision, and coagulation of soft tissue in medical specialties including: aesthetic (dermatology and plastic surgery), general surgery ,gynecology,podiary,dental and otorhinolaryngology. Dermatology, Plastic Surgery and General Surgery procedures including but not limited to: Laser skin resurfacing Treatment of furrows and wrinkles Removal of skin tags, actinic keratosis, acne scars, keloids, tattoos, telangiectasia, squamous and basel cell carcinoma, warts and uneven pigmentation. Treatment of cysts, abscesses, hemorrhoids and other soft tissue applications. Blepharoplasty Site preparation for hair transplants. Dental procedure including but not limited to - Periodontal procedures such as - gingivectomy, removal of hyperplasias gingivoplasty (incision and excision); Oral Surgery procedures such as - aphous ulcer excision, frenectomy, benign /malignant lesion ablation, operculectomy and homeostasis. Podiatry procedures including but not limited to - Ablation, vaporization and excision of soft tissue lesions such as ingrown nail, fungal nail, porokeratoma, matrixectomy and verrucae vulgares. Otorhinolaryngology (ENT) procedures including but not limited to - Treatment of leukoplakia of larynx, nasal obstruction, rhinophyma, verrucrea vulgares, choanal atresia, rhinophyma, LAUP and papillomatosis polyps. Gynecology Treatment of condyloma acuminata, cervical intraepithelial neoplasia, leukoplakia and vulvar/vaginal intraepithelial neoplasia, cervical dysplasia. This device is for prescription use only.
Device Story
eBeam CO2 Laser System is a sealed-off CO2 laser (10.6 µm wavelength) for soft tissue surgery. System comprises control unit, user interface, power supply, cooling system, and 7-joint spring-balanced articulated arm with handpieces (scanning/normal). Laser energy delivered via mirrors in articulated arm to handpiece for ablation, vaporization, excision, incision, and coagulation. Operated by physicians in clinical settings. Output is laser beam; provider directs beam to target tissue to achieve surgical effect. Benefits include precise soft tissue management across multiple specialties. Note: safety/effectiveness not evaluated for fractionated scanning.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by bench testing and adherence to mandatory/voluntary safety standards (IEC 60601 series, 21 CFR 1040.10/11).
Technological Characteristics
Sealed-off CO2 laser; 10.6 µm wavelength. 7-joint spring-balanced articulated arm delivery system. Includes scanning and normal handpieces. Class IV laser. Complies with IEC 60825-1, IEC 60601-2-22, IEC 60601-1, and IEC 60601-1-2. Quality system certified to ISO 9001 and ISO 13485.
Indications for Use
Indicated for patients requiring soft tissue ablation, vaporization, excision, incision, and coagulation across dermatology, plastic surgery, general surgery, gynecology, podiatry, dental, and ENT specialties. Specific conditions include skin lesions, dental/periodontal pathologies, podiatric lesions, and gynecological/ENT soft tissue conditions. Prescription use only.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
SLIM Evolution Family of CO2 Laser and Delivery Device Accessories (K063001)
Submission Summary (Full Text)
{0}------------------------------------------------
K101555
SEP 1:7 2010
ADVANCED TECHNOLOGY LASER CO., LTD. 510(k) Submission Report-510(k) summary
# Section III. 510(k) summary
eBeam CO2 Laser System
Advanced Technology Laser Co., Ltd
# (As required by 21 CFR 807.92)
# 1. Date Prepared: March 22, 2010
# 2. Sponsor Information
Advanced Technology Laser Co., Ltd 920 Jian-Chuan Road, Bldg. A2, Level 5 Shanghai, 200240, China
Contact Person: Jing Wang, Registration specialist Phone: +86-21-54712151(ext:837) Fax: +86-21-54712152
# 3. Submission Correspondent
Ms. Diana Hong Mr. Lee Fu Shanghai Mid-Link Business Consulting Co., Ltd Suite 5D, No.19, Lane 999, Zhongshan No.2 Road(S) Shanghai,200030, China Tel: +86-21-64264467 Fax: +86-21-64264467-809 Email: Diana.hong@mid-link.net
{1}------------------------------------------------
ADVANCED TECHNOLOGY LASER CO., LTD. 510(k) Submission Report -510(k) summary
# 4. Proposed Device Information
Device Common or Usual Name: Carbon dioxide laser; Device Trade or Proprietary Name: eBeam CO2 Laser System Model: eBeam Classification Name: Laser instrument, Surgical, Powered Regulation Number: 21 CFR 878.4810 Product Code: GEX Panel: 878 General and Plastic Surgery
### 5. Predicate Device
SLIM Evolution Family of CO2 Laser and Delivery Device Accessories (K063001)
## 6. Device Description
eBeam CO2 Laser System is an intelligent sealed-off CO2 laser. eBeam CO2 Laser System delivers laser energy to the treatment site via a 7-joint spring-balanced articulated arm in which mirrors reflect the laser beam along the axis of arms to enter the output device- the specially designed handpiece.
eBeam CO2 Laser System can produce a laser beam with a wavelength of 10.6 m which is effective on all soft tissues, thus it can be used in surgical applications requiring the ablation, vaporization, excision, incision, and coagulation of soft tissue in medical specialties including: aesthetic(dermatology and plastic surgery), general surgery, gynecology, podiatry, dental and otorhinolaryngology.
eBeam CO2 Laser System consists of control system, user interface, power supply, laser emission and delivery system, cooling system, safety features and handpieces (including scanning handpiece and normal handpiece).
#### 7. Intended use
{2}------------------------------------------------
#### ADVANCED TECHNOLOGY LASER CO., LTD. 510(k) Submission Report-510(k) summary
eBeam CO2 Laser System is indicated for use in surgical applications requiring the ablation, vaporization, excision, incision, and coagulation of soft tissue in medical specialties including: aesthetic (dermatology and plastic surgery), general surgery, gynecology, podiatry, dental and otorhinolaryngology.
#### 8. Substantial Equivalence
eBeam CO2 Laser System shares the same indications for use and safety compliance, similar design features, functional features, and therefore are substantially equivalent to the predicate device, SLIM Evolution Family of CO2 Lasers and Delivery Device Accessories (K063001). In addition, a review of the predicate device demonstrate that eBeam CO2 Laser System is safe and effective as the predicate device as they share equivalent wavelengths, and are used to perform the same indicated surgical procedures. Therefore the proposed device is substantially equivalent (SE) to the predicate device.
### 9. Testing
eBeam CO2 Laser System is designed, tested and will be manufactured in accordance with both mandatory and voluntary standards, including:
- IEC 60825-1: 2007, Safety of laser products Part 1: Equipment classification, . requirements and user's guide.
- IEC 60601-2-22: 2007, Medical Electrical Equipment Part 2: Particular . requirements for the safety of diagnostic and therapeutic laser equipment.
- IEC 60601-1:1988+A1:1991+A2:1995, Medical Electrical Equipment Partl: ● General requirements for safety.
- IEC60601-1-2:2001+A1:2004, Medical Electrical Equipment Part 1: General ● requirements for safety-2, Collateral Standard: Electromagnetic compatibility -Requirements and tests.
The devices also comply with European Medical Directive 93/42/EEC and the US Federal Performance Standards 21 CFR 1002.10 Requirements (21CFR 1040.10 and 21CFR 1040.11 for Class IV Laser Products), Part 820 - Quality System Regulation, and have passed ISO9001 and ISO13485 System Certification.
{3}------------------------------------------------
ADVANCED TECHNOLOGY LASER CO., LTD. 510(k) Submission Report -510(k) summary
#### Non-Clinical Conclusion:
Laboratory testing was conducted to validate and verify that the proposed device, eBeam CO2 Laser System met all design specifications and was substantially equivalent to the predicate device.
Clinical Conclusion: No Clinical Information is required
{4}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. In the center of the seal is an abstract image of an eagle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Advanced Technology Laser Co., Ltd. % Underwriters Laboratories, Inc. Laboratory and Testing Mr. Marc M. Mouser 2600 N.W. Lake Road Camas, Washington 98607-9526
SEP 1 7 2010
Re: K101555
Trade/Device Name: eBeam CO2 Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: September 08, 2010 Received: September 09, 2010
Dear Mr. Mouser:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
#### Page 2 - Mr. Marc M. Mouser
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
For
Dep D.R
Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
×101555 SEP 1 7 2010
DVANCED TECHNOLOGY LASER CO., LTD. 510(k) Submission Report -Indications for Use Statement
# Attachment IV Indications for Use Statement
510(k) Number: K101555 Device Name:
eBeam CO2 Laser System
#### Indications for Use:
eBeam CO2 Laser System are indicated for use in surgical applications requiring the ablation, vaporization, excision, incision, and coagulation of soft tissue in medical specialties including: aesthetic (dermatology and plastic surgery), general surgery ,gynecology,podiary,dental and otorhinolaryngology
Dermatology, Plastic Surgery and General Surgery procedures including but not limited to:
Laser skin resurfacing
Treatment of furrows and wrinkles
Removal of skin tags, actinic keratosis, acne scars, keloids, tattoos, telangiectasia, squamous and basel cell carcinoma, warts and uneven pigmentation.
Treatment of cysts, abscesses, hemorrhoids and other soft tissue applications.
Blepharoplasty
Site preparation for hair transplants
#### Dental procedure including but not limited to -
Periodontal procedures such as - gingivectomy, removal of hyperplasias gingivoplasty (incision and excision);
Oral Surgery procedures such as - aphous ulcer excision, frenectomy, benign /malignant lesion ablation, operculectomy and homeostasis
Over-The-Counter Use Prescription Use _ AND/OR X (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THE LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
CDRH. Office of Device Evaluation (ODE) Concurrence of
Neil R. Oyler Co.
Division of Surgical, Orthopedic, and Restorative Devices
Page 1 of 2
510(k) Number. K101555
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Advanced Technology Laser Co., LTD. The logo includes the letters ATL inside of a circle. Below the logo, the text "510(k) Submission Report -Indications for Use Statement" is visible. The text is left-aligned and in a smaller font size than the company name.
#### Continued from Previous Page
#### Podiatry procedures including but not limited to -
Ablation, vaporization and excision of soft tissue lesions such as ingrown nail, fungal nail, porokeratoma, matrixectomy and verrucae vulgares.
#### Otorhinolaryngology (ENT) procedures including but not limited to -
Treatment of leukoplakia of larynx, nasal obstruction, rhinophyma, verrucrea vulgares, choanal atresia, rhinophyma, LAUP and papillomatosis polyps.
#### Gynecology
Treatment of condyloma acuminata, cervical intraepithelial neoplasia, leukoplakia and vulvar/vaginal intraepithelial neoplasia, cervical dysplasia.
This device is for prescription use only.
The safety and effectiveness of this scanner/device has not been evaluated as a fractionated scanner/device.
AND/OR Over-The-Counter Use Prescription Use _____________________________________________________________________________________________________________________________________________________________ × (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THE LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 2 of 2
NielReOgle for mxm
- Orvision of Surgical, Orthopedic, Restorative Devic
510(k) Number K101555
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.