K082809 · Lumenis, Inc. · GEX · Dec 18, 2008 · General, Plastic Surgery
Device Facts
Record ID
K082809
Device Name
ACUPULSE 30 AND 40 CARBON DIOXIDE LASER SYSTEM
Applicant
Lumenis, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Dec 18, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The intended use of the AcuPulse series is for the vaporization, excision, ablation, or coagulation of soft tissue in the surgical specialties of: ENT, Gynecology, Laparoscopic Surgery including GYN Laparoscopy, Aesthetic Surgery, Dental and Oral Surgery, Neurosurgery, Orthopedics. General Surgery and Podiatry. The intended use of the AcuPulse series is for the performance of specific surgical applications in the surgical specialties of: ENT, Gynecology, Laparoscopic Surgery including GYN Laparoscopy, Aesthetic Surgery, Dental and Oral Surgery, Orthopedics, General Surgery and Podiatry as follows:
Device Story
AcuPulse 30 and 40 are computer-controlled CO2 laser systems; utilize sealed-off CO2 laser tube (mid-infrared wavelength) to deliver 30-40 Watts of energy. Energy absorbed by water in soft tissue; enables excision, incision, ablation, vaporization, and coagulation. System comprises main cabinet, articulated arm beam delivery, and attachable accessories; activated via footswitch. Used in clinical settings (OR, clinic) by physicians. Integrates GUI and SurgiTouch scanner for precise tissue interaction. Output allows surgeons to perform precise soft tissue procedures; benefits patients through minimally invasive surgical options, reduced bleeding, and targeted tissue removal.
Indicated for soft tissue vaporization, excision, ablation, or coagulation in ENT, Gynecology, Laparoscopic Surgery, Aesthetic Surgery, Dental/Oral Surgery, Neurosurgery, Orthopedics, General Surgery, and Podiatry. Specific applications include skin resurfacing, lesion removal (benign/malignant), tumor resection, ulcer debridement, and various endoscopic/laparoscopic procedures across listed specialties.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Sharplan Models 20C, 30C and 40C CO2 Laser (K963229)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains a series of handwritten digits and characters. The sequence starts with the characters 'KC', followed by the numbers '82809'. The handwriting style appears somewhat cursive, with some characters connected.
# DEC 1 8 2008
## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990.
#### SUBMITTER INFORMATION: A.
| Submitter's Name: | Lumenis, Inc. |
|----------------------|------------------------------------------------|
| Address: | 5302 Betsy Ross Drive<br>Santa Clara, CA 95054 |
| Contact: | Mark H. Smith |
| Phone: | (408) 764-3615 |
| Fax: | (408) 764-3934 |
| Date of Preparation: | June 13, 2008 |
#### B. DEVICE NAME:
| Trade Name(s): | AcuPulse 30 and 40 Carbon Dioxide Laser System |
|-----------------------|---------------------------------------------------------------------------------------------------------|
| Common/Usual Name: | Laser instrument, surgical, powered |
| Classification Names: | 79 GEX, Laser Powered Surgical Instrument |
| CFR Reference: | 21 CFR 878.4810, Laser surgical instrument for use in general<br>and plastic surgery and in dermatology |
#### PREDICATE DEVICES NAMES: C.
| Trade Name(s): | 1) | Sharplan Models 20C, 30C and 40C CO2 Laser (K963229 |
|----------------|----|-----------------------------------------------------|
| | 2) | UtraPulse SurgiTouch CO2 Laser Systems (K030147) |
#### D. DEVICE DESCRIPTION:
The AcuPulse 30 and 40 series is the next generation of Lumenis CO2 lasers. The new product will retain the same console, articulated arm, delivery devices and other basic features of the current Compact 20C, 30C & 40C series, while integrating improved versions of the GUI and SurgiTouch scanner currently used with the UltraPulse SurgiTouch.
The AcuPulse 30 and 40 Carbon Dioxide Laser is an advanced computer-controlled, user friendly CO2 laser system based on a sealed-off CO2 laser tube providing up to 30 or
40 Watts (model dependent) on tissue. The system incorporates the CO2 laser tube within the main cabinet, an articulated arm beam delivery system and attachable laser accessories. The system is activated for laser emission by a footswitch.
This system incorporates a DC-excited CO2 laser tube. The CO2 laser wavelength falls in the mid-infrared region of the electromagnetic spectrum. This wavelength is invisible to the human eye. CO2 laser energy is readily absorbed by water in tissue. Since soft tissue is comprised primarily of water, CO2 laser energy can be used effectively for the excision, incision, ablation, vaporization and coagulation of soft tissue
{2}------------------------------------------------
The principles of operation and fundamental scientific technology are the same for the subject devices and predicate devices.
#### INTENDED USE: E.
The intended use of the AcuPulse series is for the vaporization, excision, ablation, or coagulation of soft tissue in the surgical specialties of: ENT, Gynecology, Laparoscopic Surgery including GYN Laparoscopy, Aesthetic Surgery, Dental and Oral Surgery, Neurosurgery, Orthopedics. General Surgery and Podiatry.
The intended use of the AcuPulse series is for the performance of specific surgical applications in the surgical specialties of: ENT, Gynecology, Laparoscopic Surgery including GYN Laparoscopy, Aesthetic Surgery, Dental and Oral Surgery, Orthopedics, General Surgery and Podiatry as follows:
## Dermatology
The AcuPulse laser is indicated for use in dermatology and plastic surgery for the following applications.
- · Ablation, vaporization, excision, incision, and coagulation of soft tissue in the performance of:
- => Laser skin resurfacing
- => Laser dermabrasion
- > Laser burn debridement
- · Laser skin resurfacing (ablation and/or vaporization) for the treatment of:
- > Wrinkles, rhytids, and furrows (including fine lines and texture irregularities)
- · Laser skin resurfacing (ablation, and/or vaporization) of soft tissue for the reduction, removal, and/or treatment of:
- > Keratoses. including actinic and sebortheic keratosis, seborrhoecae vulgares, seborrheic wart and verruca seborrheica
- > Vermillionectomy of the lip
- > Cutaneous horns
- => Solar/actinic elastosis
- > Cheilitis, including actinic cheilitis
- > Lentigines, including lentigo maligna or Hutchinson's malignant freckle
- > Uneven pigmentation/dyschromia
- > Acne scars
- => Surgical scars
- > Keloids including acne keloidalis nuchae
↑
{3}------------------------------------------------
## Dermatology, continued
- => Hemangiomas (including Buccal, port wine and pyogenic granulomas/granuloma pyogenicum/granuloma telangiectaticum)
- ⇒ Tattoos
- > Telangiectasia
- > Removal of small skin tumors, including periungual (Koenen) and subungual fibromas
- > Superficial pigmented lesions
- > Adenosebaceous hypertrophy or sebaceous hyperplasia
- => Rhinophyma reduction
- > Cutaneous papilloma (skin tags)
- ⇒ Milia
- => Debridement of eczematous or infected skin
- => Basal and squamous cell carcinoma, including keratoacanthomas, Bowen's disease (Erythroplasia of Queyrat), and Bowenoid Papulosis (BP) lesions
- > Nevi, including spider, epidermal and protruding
- > Neurofibromas
- > Laser de-epithelialization
- > Tricoepitheliomas
- ⇒ Xanthelasma palpebrarum
- > Syringoma
- · Laser ablation, vaporization, and/or excision for complete and partial nail matrixectomy
- · Vaporization/coagulation of:
- > Benign/malignant vascular/avascular skin lesions
- > Moh's surgery
- > Lipectomy
- > Verrucae and seborrhoecae vulgares, including paronychial, periungal, and subungual warts
- · Laser incision and/or excision of soft tissue for the performance of upper and lower eyelid blepharoplasty
- · Laser incision and/or excision of soft tissue for the creation of recipient sites for hair transplantation
{4}------------------------------------------------
## Podiatry
The AcuPulse laser is indicated for use in podiatry for the following applications:
- · Laser ablation, vaporization and/or excision of soft tissue for the reduction, removal, and/or treatment of:
- → Verrucae vulgares/plantar (warts), including paronychial, periungal, and subungual warts
- ⇒ Fungal nail treatment
- ⇒ Porokeratoma ablation
- => Ingrown nail treatment
- ⇒ Neuromas/fibromas, including Morton's neuroma
- ⇒ Debridement of ulcers
- ⇒ Other soft tissue lesions
- Laser ablation, vaporization, and/or excision for complete and partial (nail) matrixectomy
{5}------------------------------------------------
## Otolaryngology (ENT)
The AcuPulse laser is indicated for laser incision, ablation, and/or vaporization of soft tissue in otolaryngology for the treatment of:
- · Choanal atresia
- · Leukoplakia, including oral, larynx, uvula, palatal, and upper lateral pharyngeal tissue
- · Nasal obstruction
- · Adult and juvenile papillomatosis polyps
- · Polypectomy of nose and nasal passages
- · Lymphangioma removal
- · Removal of vocal cord/fold nodules, polyps and cysts
- · Removal of recurrent papillomas in the oral cavity, nasal cavity, larynx, pharynx and trachea, including the uvula, palatal, upper lateral pharyngeal tissue, tongue and vocal cords
- · Laser/tumor surgery in the larynx, pharynx, nasal, ear and oral structures and tissue
- · Zenker's Diverticulum/ pharyngoesophageal diverticulum [endoscopic laser-assisted esophagodiverticulostomy (ELAED)]
- · Stenosis, including subglottic stenosis
- · Tonsillectomy (including tonsilar cryptolysis and neoplasma) and tonsil ablation/tonsillotomy
- · Pulmonary bronchial and tracheal lesion removal
- · Benign and malignant nodules, tumors and fibromas (larynx, pharynx, trachea, tracheobronchial/endobronchial)
- · Benign and malignant lesions and fibromas (nose and nasal passages)
- · Benign and malignant tumors and fibromas (oral)
- · Acoustic neuroma in the ear
- · Superficial lesions of the ear, including chondrodermatitis nondularis chronica helices/Winkler's disease
- · Telangiectasia/hemangioma of larynx, pharynx, and trachea (includes uvula, palatal, or upper lateral pharyngeal tissue)
- · Cordectomy, cordotomy (for the treatment of vocal fold motion impairment), and cordal lesions of larynx, pharynx, and trachea
- · Myringotomy/tympanostomy (tympanic membrane fenestration)
- · Uvulopalatoplasty (LAUP, laser UPP)
- · Turbinectomy and turbinate reduction/ablation
- · Septal spur ablation/reduction and septoplasty
- · Partial glossectomy
- · Tumor resection of oral, subfacial and neck tissues
- · Rhinophyma
{6}------------------------------------------------
- · Verrucae vulgares (warts)
- · Gingivoplasty/gingivectomy
### Gynecology and GYN Laparoscopy Indications
The AcuPulse laser is indicated for use in gynecology for the following applications:
- · Laser incision, excision, ablation and/or vaporization of soft tissue in gynecology for the treatment of:
- > Conization of the cervix, including cervical intraepithelial neoplasia (CIN), and vulvar and vaginal intraepithelial neoplasia (VIN, VAIN)
- > Condyloma acuminata, including cervical, genital, vulvar, perineal, and Bowenoid papulosa (BP) lesions
- > Leukoplakia (vulvar dystrophies)
- -> Incision and drainage (I&D) of Bartholin's and nabothian cysts
- > Herpes vaporization
- > Urethral caruncle vaporization
- => Cervical dysplasia
- => Benign and malignant tumors
- > Hemangiomas
- · Vaporization, incision, excision, ablation, or photocoagulation of soft tissue in endoscopic and laparoscopic surgery, including gynecological laparoscopy, for the treatment of:
- > Endometrial lesions, including ablation of endometriosis
- > Excision/lysis of adhesions
- > Salpingostomy
- => Oophorectomy
- = Fimbrioplasty
- > Metroplasty
- > Microsurgery (tubal)
- > Uterine myomas and fibroids
- > Ovarian fibromas and follicle cysts
- => Uterosacral ligament ablation
- > Hysterectomy
{7}------------------------------------------------
### Neurosurgery Indications
The AcuPulse laser is indicated for laser incision, ablation and/or vaporization of soft tissue in neurosurgery for the treatment of:
- · Cranial
- > Posterior fossa tumors
- > Peripheral neurectomy
- > Benign and malignant tumors and cysts, for example, gliomas, meningiomas (including basal tumors), acoustic neuromas, lipomas, and large tumors
- > Arteriovenous malformation
- > Pituitary gland tumors (transphenoidal approach)
- · Spinal Cord
- > Incision/excision and vaporization of benign and malignant tumors and cysts
- => Intra- and extradural lesions
- > Laminectomy/laminotomy/microdiscectomy
## Orthopedic Indications
The AcuPulse laser is indicated for incision, excision, and vaporization of soft tissue in orthopedic surgery, including the following applications:
- · Arthroscopy
- > Meniscectomy
- ⇒ Chondromalacia
- => Chondroplasty
- > Ligament release (lateral and other)
- => Excision of plica
- > Partial synovectomy
- · General
- > Debridement of traumatic wounds
- => Debridement of decubitus and diabetic ulcers
- > Microsurgery
- => Artificial joint revision
- > PMMA removal
{8}------------------------------------------------
## General and Thoracic Surgery
The AcuPulse laser is indicated for incision, excision, and vaporization of soft tissue in general and thoracic surgery, including endoscopic and open procedures. Applications include:
- · Debridement of decubitus ulcers, stasis, diabetic, and other ulcers
- · Mastectomy
- Debridement of burns
- · Rectal and anal hemorrhoidectomy
- · Breast biopsy
- · Reduction mammoplasty
- · Cytoreduction for metastatic disease
- Laparotomy and laparoscopic applications
- · Mediastinal and thoracic lesions and abnormalities
- · Skin tag vaporization
- · Atheroma
- · Cysts, including sebaceous cysts, pilar cysts, and mucous cysts of the lips
- · Pilonidal cyst removal and repair
- Abscesses
- · Other soft tissue applications
{9}------------------------------------------------
## Dental and Oral Surgery
The AcuPulse laser is indicated for incision, excision, and vaporization of soft tissue in dentistry and oral surgery. Applications include:
- · Gingivectomy/removal of hyperplasias
- · Gingivoplasty
- · Incisional and excisional biopsy
- · Treatment of ulcerous lesions, including aphthous ulcers
- · Incision of infection when used with antibiotic therapy
- · Frenectomy (frenum release)
- · Excision and ablation of benign and malignant lesions
- · Homeostasis
- · Operculectomy
- · Crown lengthening
- · Removal of soft tissue, cysts, and tumors
- · Oral cavity tumors and hemangiomas
- · Abscesses
- · Extraction site hemostasis
- · Salivary gland pathologies
- · Preprosthetic gum preparation
- Leukoplakia
- · Partial glossectomy
- · Periodontal gum resection
Both the subject and predicate devices have nearly identical indications for use and intended use:
Both the subject and predicate devices have the same indications for use and intended use.
The intended use has not changed from the predicate devices (K963229, K030147).
#### TECHNOLOGICAL CHARACTERISTICS SUMMARY & SUBSTANTIAL EQUIVALENCE F. STATEMENT:
The subject device, the AcuPulse 30 and 40 Carbon Dioxide Laser System, has the same intended use, general design and fundamental scientific technology as the predicate devices (K963229, K030147).
The AcuPulse 30 and 40 Carbon Dioxide Laser System uses technology substantially equivalent to the Sharplan Models 20C, 30C and 40C CO2 Laser (K963229), and the
{10}------------------------------------------------
UltraPulse SurgiTouch CO2 Laser Systems (K030147). There are no new hazards introduced by the AcuPulse 30 and 40 Carbon Dioxide Laser System as compared with the predicate devices.
#### G. PERFORMANCE DATA SUMMARY:
The appropriate testing including safety, performance and functional testing to determine substantial equivalence of the AcuPulse 30 and 40 Carbon Dioxide Laser System has been conducted.
{11}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/10/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its body and wing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## DEC 1 8 2008
Lumenis, Inc. % Frances E. Harrison, RAC Global VP, Regulatory Affairs and Quality Compliance 3959 West 1820 South Salt Lake City, Utah 84104
Re: K082809
Trade/Device Name: AcuPulse 30 and 40 Carbon Dioxide Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Lascr surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: September 22, 2008 Received: September 24, 2008
Dear Frances E. Harrison:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{12}------------------------------------------------
Page 2 - Frances E. Harrison
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 100-1100.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a logally marketed predicate device results in a classification for your device and thus, permits your levice to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance averse at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Mark M. Wilkerson
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{13}------------------------------------------------
#### IFU Statement
510(k) Number (if known): K082809
Device Name: AcuPulse 30 and 40 Carbon Dioxide Laser System
Indications for Use:
The intended use of the AcuPulse series is for the vaporization, incision, excision, ablation, or coagulation of soft tissue in the surgical specialties of: ENT, Gynecology, Laparoscopic Surgery including GYN Laparoscopy, Aesthetic Surgery, Dental and Oral Surgery, Neurosurgery, Orthopedics, General Surgery and Podiatry,
The intended use of the AcuPulse series is for the performance of specific surgical applications in the surgical specialties of: ENT, Gynecology, Laparoscopic Surgery including GYN Laparoscopy, Aesthetic Surgery, Dental and Oral Surgery, Neurosurgery, Orthopedics, General Surgery and Podiatry as follows:
#### Dermatology
The AcuPulse laser is indicated for use in dermatology and plastic surgery for the following applications.
- · Ablation, vaporization, excision, incision, and coagulation of soft tissue in the performance of:
- => Laser skin resurfacing
- => Laser dermabrasion
- => Laser burn debridement
- · Laser skin resurfacing (ablation and/or vaporization) for the treatment of:
- > Wrinkles, rhytids, and furrows (including fine lines and texture irregularities)
- · Laser skin resurfacing (ablation, and/or vaporization) of soft tissue for the reduction, removal, and/or treatment of:
- > Keratoses, including actinic and seborrheic keratosis, seborrhoecae vulgares, sebortheic wart and verruca seborrheica
- => Vermillionectomy of the lip
- -> Cutaneous horns
- => Solar/actinic elastosis
- > Cheilitis, including actinic cheilitis
- > Lentigines, including lentigo maligna or Hutchinson's malignant free
- > Uneven pigmentation/dvschromia
- Acne scars
- > Surgical scars
- > Keloids including acne keloidalis nuchae
NatarOphth for msm
Division Sign-Off) Division of General, Restorative. and Neurological Devices
510(k) Number
{14}------------------------------------------------
#### Dermatology, continued
- > Hemangiomas (including Buccal, port wine and pyogenic granulomas/granuloma pyogenicum/granuloma telangiectaticum)
- > Tattoos
- > Telangiectasia
- => Removal of small skin tumors, including periungual (Koenen) and subungual fibromas
- > Superficial pigmented lesions
- > Adenosebaceous hypertrophy or sebaceous hyperplasia
- => Rhinophyma reduction
- => Cutaneous papilloma (skin tags)
- ⇒ Milia
- > Debridement of eczematous or infected skin
- > Basal and squamous cell carcinoma, including keratoacanthomas, Bowen's disease (Erythroplasia of Queyrat), and Bowenoid Papulosis (BP) lesions
- => Nevi, including spider, epidermal and protruding
- => Neurofibromas
- -> Laser de-epithelialization
- => Tricoepitheliomas
- > Xanthelasma palpebrarum
- > Syringoma
- · Laser ablation, vaporization, and/or excision for complete and partial nail matrixectomy
- · Vaporization/coagulation of:
- > Benign/malignant vascular/avascular skin lesions
- > Moh's surgery
- > Lipectomy
- => Verrucae and seborrhoecae vulgares, including paronychial, periungal, and subungual warts
- · Laser incision and/or excision of soft tissue for the performance of upper and lower eyelid blepharoplasty
- · Laser incision and/or excision of soft tissue for the creation of recipient sites for hair transplantation
Neil R. Ogden for mx
Division of General, Restorative. and Neurological Devices
**510(k) Number** K082809
{15}------------------------------------------------
## Podiatry
The AcuPulse laser is indicated for use in podiatry for the following applications:
- · Laser ablation, vaporization and/or excision of soft tissue for the reduction, removal, and/or treatment of:
- > Verrucae vulgares/plantar (warts), including paronychial, periungal, and subungual warts
- => Fungal nail treatment
- > Porokeratoma ablation
- => Ingrown nail treatment
- => Neuromas/fibromas, including Morton's neuroma
- => Debridement of ulcers
- > Other soft tissue lesions
- · Laser ablation, vaporization, and/or excision for complete and partial (nail) matrixectomy
Neil RP Ogden-Forman
(Division Sign-Off Division of General, Restorative, and Neurological Devices
**510(k) Number** K082809
{16}------------------------------------------------
### Otolaryngology (ENT)
The AcuPulse laser is indicated for laser incision, excision, ablation, and/or vaporization of soft tissue in otolaryngology for the treatment of:
- · Choanal atresia
- · Leukoplakia, including oral, larynx, uvula, palatal, and upper lateral pharyngeal tissue
- · Nasal obstruction
- · Adult and juvenile papillomatosis polyps
- · Polypectomy of nose and nasal passages
- · Lymphangioma removal
- · Removal of vocal cord/fold nodules, polyps and cysts
- · Removal of recurrent papillomas in the oral cavity, nasal cavity, larynx, pharynx and trachea, including the uvula, palatal, upper lateral pharyngeal tissue. tongue and vocal cords
- · Laser/tumor surgery in the larynx, pharynx, nasal, car and oral structures and tissue
- · Zenker's Diverticulum/ pharyngoesophageal diverticulum fendoscopic laser-assisted esophagodiverticulostomy (ELAED)]
- · Stenosis, including subglottic stenosis
- · Tonsillectomy (including tonsilar cryptolysis and neoplasma) and tonsil ablation/tonsillotomy
- · Pulmonary bronchial and tracheal lesion removal
- · Benign and malignant nodules, tumors and fibromas (larynx, pharynx, trachea, tracheobronchial/endobronchial)
- · Benign and malignant lesions and fibromas (nose and nasal passages)
- · Benign and malignant tumors and fibromas (oral)
- · Acoustic neuroma in the ear
- · Superficial lesions of the ear, including chondrodermatitis nondularis chronica helices/Winkler's disease
- · Telangiectasia/hemangioma of larynx, pharynx, and trachea (includes uvula, palatal, or upper lateral pharyngeal tissue)
- · Cordectomy, cordotomy (for the treatment of vocal fold paralysis/vocal fold motion impairment), and cordal lesions of larynx, pharynx, and trachea
- · Myringotomy/tympanostomy (tympanic membrane fenestration)
- · Uvulopalatoplasty (LAUP, laser UPP)
- · Turbinectomy and turbinate reduction/ablation
- · Septal spur ablation/reduction and septoplasty
- · Partial glossectomy
- · Tumor resection of oral, subfacial and neck tissues
- · Rhinophyma
- · Verrucae vulgares (warts)
- · Gingivoplasty/gingivectomy
Neil R.P. Ogden forman
Division of General. Restorative. and Neurological Devices
**510(k) Number** K082809
{17}------------------------------------------------
### Gynecology and GYN Laparoscopy Indications
The AcuPulse laser is indicated for use in gynecology for the following applications:
- · Laser incision, excision, ablation and/or vaporization of soft tissue in gynecology for the treatment of:
- > Conization of the cervix, including cervical intraepithelial neoplasia (CIN), and vulvar and vaginal intraepithelial neoplasia (VIN, VAIN)
- > Condyloma acuminata, including cervical, genital, vulvar, perineal, and Bowenoid papulosa (BP) lesions
- => Leukoplakia (vulvar dystrophies)
- Incision and drainage (I&D) of Bartholin's and nabothian cysts
- > Herpes vaporization
- > Urethral caruncle vaporization
- > Cervical dysplasia
- => Benign and malignant tumors
- => Hemangiomas
- · Vaporization, incision, excision, ablation, or photocoagulation of soft tissue in endoscopic and laparoscopic surgery, including gynecological laparoscopy, for the treatment of:
- > Endometrial lesions, including ablation of endometriosis
- > Excision/lysis of adhesions
- > Salpingostomy
- > Oophorectomy
- > Fimbrioplasty
- > Metroplasty
- > Microsurgery (tubal)
- > Uterine myomas and fibroids
- > Ovarian fibromas and follicle cysts
- > Uterosacral ligament ablation
- > Hysterectomy
Neil R. Ogden forman
(Division Sign-Off Division of General. Restorative. and Neurological Devices
510(k) Number K082809
{18}------------------------------------------------
#### Neurosurgery Indications
The AcuPulse laser is indicated for laser incision, excision, ablation and/or vaporization of soft tissue in neurosurgery for the treatment of:
- · Cranial
- => Posterior fossa tumors
- => Peripheral neurectomy
- > Benign and malignant tumors and cysts, for example, gliomas, meningiomas (including basal tumors), acoustic neuromas, lipomas, and large tumors
- => Arteriovenous malformation
- > Pituitary gland tumors (transphenoidal approach)
- · Spinal Cord
- > Incision/excision and vaporization of benign and malignant tumors and cysts
- => Intra- and extradural lesions
- > Laminectomy/laminotomy/microdiscectomy
#### Orthopedic Indications
The AcuPulse laser is indicated for incision, excision, and vaporization of soft tissue in orthopedic surgery, including the following applications:
- · Arthroscopy
- > Meniscectomy
- => Chondromalacia
- > Chondroplasty
- => Ligament release (lateral and other)
- > Excision of plica
- -> Partial synovectomy
- · General
- > Debridement of traumatic wounds
- > Debridement of decubitus and diabetic ulcers
- > Microsurgery
- > Artificial joint revision
- > PMMA removal
NeulReDada for mom
Division of General, Restorative, and Neurological Devices
510(k) Number K082809
{19}------------------------------------------------
#### General and Thoracic Surgery
The AcuPulse laser is indicated for incision, excision, and vaporization of soft tissue in general and thoracic surgery, including endoscopic and open procedures. Applications include:
- · Debridement of decubitus ulcers, stasis, diabetic, and other ulcers
- · Mastectomy
- Debridement of burns
- · Rectal and anal hemorrhoidectomy
- · Breast biopsy
- · Reduction mammoplasty
- · Cytoreduction for metastatic disease
- · Laparotomy and laparoscopic applications
- · Mediastinal and thoracic fesions and abnormalities
- · Skin tag vaporization
- · Atheroma
- · Cysts, including sebaceous cysts, pilar cysts, and mucous cysts of the lips
- · Pilonidal cyst removal and repair
- Abscesses
- · Other soft tissue applications
Neitke Oglin for mkn
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
510(k) Number K082809
{20}------------------------------------------------
## Dental and Oral Surgery
The AcuPulse laser is indicated for incision, excision, and vaporization of soft tissue in dentistry and oral surgery. Applications include:
- · Gingivectomy/removal of hyperplasias
- · Gingivoplasty
- · Incisional and excisional biopsy
- · Treatment of ulcerous lesions, including aphthous ulcers
- · Incision of infection when used with antibiotic therapy
- · Frenectomy (frenum release)
- · Excision and ablation of benign and malignant lesions
- · Homeostasis
- · Operculectomy
- · Crown lengthening
- · Removal of soft tissue, cysts, and tumors
- · Oral cavity tumors and hemangiomas
- · Abscesses
- · Extraction site hemostasis
- · Salivary gland pathologies
- · Preprosthetic gum preparation
- · Leukoplakia
- · Partial glossectomy
- · Periodontal gum resection
(Part 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)
## CONCURRENCE OF CDRH, OFFICE OF DEVICE EVALUATION (ODE)
AND/OR
Neufredson
(Division Sign-Off)
Division of General, Restorative
Over-The-Counter Use
and Neurological Devices
(Recommended Format 11/13/2003)
Prescription Use _____________________________________________________________________________________________________________________________________________________________
X
510(k) Number
{21}------------------------------------------------
and the comments of the comments of
and the comments of the comments of the comments of the comments of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.