K172096 · Ilooda Co,., Ltd. · GEX · Aug 9, 2017 · General, Plastic Surgery
Device Facts
Record ID
K172096
Device Name
Fraxis
Applicant
Ilooda Co,., Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Aug 9, 2017
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Fractional mode is indicated only for ablative skin resurfacing. Non-fractional mode is indicated for incision, excision, ablation, vaporization and coagulation of body soft tissues including intraoral tissues, in medical specialties including aesthetic (dermatology and plastic surgery), otorhinolaryngology (ENT), gynecology, neurosurgery, dental and oral surgery and genitourinary surgery.
Device Story
Ilooda Fraxis CO2 Laser is a surgical laser system for soft tissue procedures and skin resurfacing. It utilizes a 10.6µm CO2 laser source delivered via an articulated arm to a handpiece. The device operates in fractional mode for skin resurfacing or non-fractional mode for surgical applications (incision, excision, ablation, vaporization, coagulation). The system includes a main console, delivery handpieces, and a footswitch. It is operated by clinicians in clinical settings. The user interface is an LCD touch screen. The device provides precise laser energy to target tissues, enabling controlled tissue removal or coagulation. It benefits patients by facilitating surgical procedures and dermatological treatments. The device is non-sterile and reusable, requiring cleaning between uses.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including software verification and validation, biocompatibility, and adherence to IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, and IEC 60825-1 standards.
Technological Characteristics
CO2 laser, 10.6µm wavelength, 30W output power. Pulse duration 20-5,000µs. Fractional pulse energy max 150mJ. Repetition rate 1,000Hz. Air cooling. Articulated arm delivery. LCD touch screen interface. Electrical requirements 100-240VAC, 50-60Hz. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, and IEC 60825-1.
Indications for Use
Indicated for patients requiring ablative skin resurfacing (fractional mode) or soft tissue incision, excision, ablation, vaporization, and coagulation (non-fractional mode) across dermatology, plastic surgery, ENT, gynecology, neurosurgery, and dental/oral surgery.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{1}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes entwined around it. The caduceus is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the logo.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 9, 2017
Ilooda Co. LTD % Ms. Kathy Maynor Emvera Technologies, LLC 641 10th St Cedartown, Georgia 30125
Re: K172096
Trade/Device Name: Ilooda Fraxis CO2 Laser Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX, ONG Dated: July 10, 2017 Received: July 11, 2017
Dear Ms. Maynor:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Actinclude requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{2}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
# Jennifer R. Stevenson -S3
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K172096
Device Name Ilooda Fraxis CO2 Laser
Indications for Use (Describe)
### - C02 LASER Part:
Fractional mode is indicated only for ablative skin resurfacing.
Non-fractional mode is indicated for incision, ablation, vaporization and coagulation of body soft tissues including intraoral tissues, in medical specialties including aesthetic (dermatology and plastic surgery), otorhinolaryngology (ENT), gynecology, neurosurgery, dental and oral surgery and genitourinary surgery.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## Section 8 – 510(k) Summary or 510(k) Statement
#### I. General Information [21 CFR 807.92(a) (1)]
| Submitter: | Ilooda<br>120, Jangan-Ro 458 Beon-Gil<br>Jangan-Gu<br>Suwon-Si Gyeonggido, KOREA, REPUBLIC OF<br>16200 | |
|---------------------------|--------------------------------------------------------------------------------------------------------|--|
| Contact Person: | Kathy Maynor<br>Consultant<br>352-586-3113 (cell)<br>Kmaynor77@gmail.com | |
| Summary Preparation Date: | August 9, 2017 | |
#### II. Names [21 CFR 807.92 (a) (2)]
| Device Names: | The Ilooda Fraxis CO2 Laser |
|-------------------------------|----------------------------------------------------------------------------------------|
| Primary Classification Names: | Surgical Powered Light Instrument, |
| Regulation No.: | 21 CFR 878.4810 |
| Device Regulation Name: | Laser surgical instrument for use in general and plastic<br>surgery and in dermatology |
| Product Code: | GEX, ONG |
| Device Class: | II |
#### Predicate Devices [21 CFR 807.92(a) (3)] III.
| K # | Predicate Device |
|---------|------------------|
| K160312 | Fraxis Duo |
#### IV. Product Description [21 CFR 807.92(a) (4)]
The Ilooda Fraxis CO2 Laser is comprised of the following major components:
- The main console unit 1.
- 2. Delivery handpieces (2)
- 3. Footswitch.
- 4. Accessories
{5}------------------------------------------------
The device is re-usable and non-sterile; instructions for cleaning its components between uses are provided in the labeling.
#### V. Indications for Use [21 CFR 807.92(a) (5)]
- CO2 LASER Part: Fractional mode is indicated only for ablative skin resurfacing. Non-fractional mode is indicated for incision, excision, ablation, vaporization and coagulation of body soft tissues including intraoral tissues, in medical specialties including aesthetic (dermatology and plastic surgery). otorhinolaryngology (ENT), gynecology, neurosurgery, dental and oral surgery and genitourinary surgery.
#### VI. Summary of Technical Characteristics Compared to Predicate [21 CFR 807.92(a)(6)]
Table 1: Salient Characteristics of the Ilooda Fraxis CO2 Laser and the Predicate Device
| Characteristic | K172096<br>Ilooda Fraxis CO2 Laser | K160312<br>Fraxis Duo |
|----------------------------------------------------------|----------------------------------------|----------------------------------------|
| Product Code & | GEX, ONG | GEX |
| Regulation No. | 21 CFR 878.4810 | 21 CFR 878.4810 |
| Laser Type | CO2 | CO2 |
| Laser wavelength | 10.6µm | 10.6µm |
| Output power | 30W | 30W |
| Pulse Duration | 20-5,000µs | 20-5,000µs |
| Fractional Pulse energy | Max 150mJ | Max 150mJ |
| Repetition rate | 1,000Hz | 1,000Hz |
| Scan area | 20x20mm | 20 x 20mm |
| Spot size | 100-200µm<br>Non-fractional: Max 1.3mm | 100-200µm<br>Non-fractional: Max 1.3mm |
| Number of microbeams<br>per surface area<br>(fractional) | Max 289 spot/cm2 | Max 289 spot/cm2 |
| Energy per microbeam<br>(fractional) | 150mJ | 150mJ |
| Total power per surfaced<br>area (fractional) | Max 30W | Max 30W |
| Treatment Time | 10-15 min | 10-15 min |
| Pulse rate (non-<br>fractional) | 1Hz - 1,000Hz | 1Hz - 1,000Hz |
{6}------------------------------------------------
| Pulse width (non-<br>fractional) | 20μs – 5000μs | 20μs – 5000μs |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Operational mode | Fractional mode, normal mode<br>(CW, Pulse, Single Pulse) | Fractional mode, surgical<br>mode (CW, pulsed, single<br>pulse) |
| Aiming beam | Diode laser(Red), Max 4mW | Diode laser(Red), Max 4mW |
| Cooling | Air cooling | Air cooling |
| User Interface | LCD touch screen | LCD touch screen |
| Optical guide | Articulated arm | Articulated arm |
| Electrical Requirements | 100-240VAC,<br>50-60 Hz,<br>6.3 A, | 100-240VAC<br>50-60 Hz |
| Indications for Use | - CO2 LASER Part:<br>Fractional mode is indicated<br>only for ablative skin<br>resurfacing.<br>Non-fractional mode is<br>indicated for incision, excision,<br>ablation, vaporization and<br>coagulation of body soft tissues<br>including intraoral tissues, in<br>medical specialties including<br>aesthetic (dermatology and<br>plastic surgery),<br>otorhinolaryngology (ENT),<br>gynecology, neurosurgery,<br>dental and oral surgery and<br>genitourinary surgery. | - CO2 LASER Part:<br>Fractional mode is indicated<br>only for ablative skin<br>resurfacing.<br>Non-fractional mode is<br>indicated for incision,<br>excision, ablation,<br>vaporization and coagulation<br>of body soft tissues<br>including intraoral tissues, in<br>medical specialties including<br>aesthetic (dermatology and<br>plastic surgery),<br>otorhinolaryngology (ENT),<br>gynecology, neurosurgery,<br>dental and oral surgery and<br>genitourinary surgery.<br>- HF electrosurgical Part: The<br>FRAXIS DUO is intended for<br>use in dermatologic and<br>general surgical procedures for<br>electro-coagulation and<br>hemostasis |
#### VII. Differences Between the Fraxis and Fraxis Duo
The Fraxis does not have the RF micro-needling handpiece named GynoLase, so they are not included in the equivalence chart above. These handpieces are included in the Fraxis Duo, the predicate device.
#### Performance Testing [21 CFR 807.92(b)(1)] VIII.
IEC 60601-1 Medical Electrical Equipment-Part 1: General Requirements for safety IEC 60601-1-2 Medical Electrical Equipment 1-2 General Requirements for basic safety and essential performance
{7}------------------------------------------------
IEC 60601-2-22 Medical Electrical Equipment-Part 2-22: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment IEC 60825-1Safety of laser products-Part 1: Equipment Classification, requirements and user's guide
In addition, software verification and validation testing was performed and biocompatibility was established.
#### IX.. Clinical Data [21 CFR 807.92(b) (2)]
Based on the similarities in the safety and effectiveness profiles of the subject, primary predicate and reference devices, no clinical studies were deemed needed to support this submission.
#### X. Conclusion
The Ilooda Fraxis CO2 laser was found to be substantially equivalent to the predicate devices.
The Ilooda Fraxis CO2 laser shares the same or similar indications for use, similar design features, and functional features with, and thus is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.