K113021 · CSA Medical, Inc. · GEH · Feb 7, 2012 · General, Plastic Surgery
Device Facts
Record ID
K113021
Device Name
TRUFREEZE SYSTEM, TRUFREEZE SPRAY KIT
Applicant
CSA Medical, Inc.
Product Code
GEH · General, Plastic Surgery
Decision Date
Feb 7, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4350
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The truFreeze System is intended for use as a cryosurgical tool for the destruction of unwanted tissue in the fields of dermatology, gynecology and general surgery.
Device Story
Cryosurgical tool applying medical-grade liquid nitrogen to ablation sites via low-pressure, open-tipped, flexible catheters. System comprises mobile console with touch panel computer (TPC), cryogen/suction/electronics modules, and disposable spray kit. Operator (physician) controls cryogen level, filling, pressure, cooling, defrost, suction, and timing via touch panel or dual foot pedal; wireless remote provides auxiliary timer control. On-board suction system manages cryogen decompression via CDT tubes. Console includes safety features: tank pressure relief, isolated low-voltage power, emergency shut-off. Used in clinical settings to destroy unwanted tissue; provides controlled cryo-ablation therapy.
Clinical Evidence
Bench testing included electrical safety, electromagnetic compatibility, software testing, biocompatibility, and sterilization testing per applicable standards. User validation was conducted using animal models. No clinical data on human subjects reported.
Technological Characteristics
Cryosurgical unit; mobile cart form factor; touch panel computer interface; liquid nitrogen delivery; low-pressure open-tipped flexible catheters; on-board suction system; wireless remote control; isolated low-voltage power system. Software manages cryogen sensing, pressure, and timing.
Indications for Use
Indicated for the destruction of unwanted tissue in dermatology, gynecology, and general surgery patients.
Regulatory Classification
Identification
(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.
{0}------------------------------------------------
## 510(k) Summary
### truFreeze® System
Image /page/0/Picture/2 description: The image shows a logo for CSA medical. The logo consists of the letters "CSA" in a bold, sans-serif font, with the word "medical" in a smaller font below. Above the letters is a semi-circle with a series of dots above it, arranged in a radiating pattern. The dots appear to be emanating from a central point.
K113021
Applicant Establishment Registration Number Contact Person
Summary Date Proprietary Name Classification Classification Name Regulation Number Classification Product Code Predicate Devices
CSA Medical 3004534508 Colleen A. Kistler, RAC tel. 443-524-9688 fax. 866-300-5183 ckistler@csamedical.com
FEB - 7 2012
January 23, 2012 truFreeze® System Class II Cryosurgical Unit, Cryogenic Surgical Device 21 CFR 878.4350 GEH CryMed Cryo-Ablator (K040809) CryoSpray Ablation System (K060555, K0702651) Mobile Care Monitor (K090138) Smart Wand-DTX System (K093557)
### Device Description
The truFreeze® System is a cryosurgical tool that applies medical-grade liquid nitrogen to the ablation area via a small, low pressure, open tipped catheter. The truFreeze System consists of (1) a console and (2) a disposable spray kit.
The console is the central interface of the system and is comprised of a touch panel computer (TPC) and cryogen, suction and electronics modules packaged in a mobile cart. Users interact with the console through a dual foot pedal and the touch panel. A controller and associated software manage the cryogen level sensing, filling, pressure, cooling, defrost, suction, timing and data management functions. A wireless remote control provides alternative timer control from a distance in the procedure room. A fill kit, stored on the rear of the console, allows for liquid nitrogen transfer from the source tank to the console. Safety features include indicators, tank pressure relief valves, an isolated low voltage power system, and an emergency shut-off button.
{1}------------------------------------------------
# 510(k) Summary
### truFreeze® System
Image /page/1/Picture/2 description: The image shows a logo for CSA medical. The logo consists of the letters "CSA" in a bold, sans-serif font, with the word "medical" in a smaller font below. Above the letters is a semi-circular shape with a series of dots arranged in a conical pattern, suggesting a beam or spray. The overall design is simple and professional, likely representing a company in the healthcare or medical field.
The truFreeze spray kit consists of sterile single-use catheters and cryogen decompression tubes (CDTs). The catheter is flexible and capable of retroflex in a scope. The CDT and accessory tubes are included for use with the on-board suction system.
### Intended Use/Indications for Use
The truFreeze System is intended for use as a cryosurgical tool for the destruction of unwanted tissue in the fields of dermatology, gynecology and general surgery.
### Technical and Operational Characteristics
The truFreeze System is similar in design and has the same operational and technological characteristics as the predicate devices. The minor differences and additional features improve the overall ease of use, robustness and serviceability and do not raise any new issues of safety and effectiveness.
### Summary of Testing
The truFreeze System was subjected to a comprehensive test program to evaluate conformance to product specifications and demonstrate substantial equivalence to the predicate device(s). Electrical safety and electromagnetic compatibility testing, software testing, biocompatibility and sterilization testing were conducted in accordance with applicable standards. In addition to the design verification and validation testing, user validations using animal models were also conducted. Test results demonstrated that the truFreeze System meets its specifications and does not raise any new safety and/or effectiveness issues.
### Substantial Equivalence
The truFreeze System has the same technological characteristics and principles of operation as the predicate device(s). The minor differences between the truFreeze System and its predicate device do not raise any new issues of safety and effectiveness and therefore is substantially equivalent.
The truFreeze System has the same intended use and indication for use as other legally marketed predicate liquid nitrogen cryosurgical devices and therefore is substantially equivalent.
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, enclosed within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged around the circumference of the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
CSA Medical, Incorporated % Ms. Colleen A. Kistler, RAC Director, Regulatory & Quality 1101 East 33td Street, Suite E305 Baltimore, Maryland 21218
Re: K113021
FEB - 7 2012
Trade/Device Name: truFreeze® System Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical unit and accessories Regulatory Class: Class II Product Code: GEH Dated: January 5, 2012 Received: January 6, 2012
Dear Ms. Kistler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it ' may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
## Page 2 - Ms. Colleen A. Kistler, RAC
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please rr you aw.l/www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
for Pete Dimemat
Mark N. Melkerson
Asp LLin Dr.
Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### INDICATION FOR USE
510(k) Number (if known): K113021
Device Name: truFreeze® System
Indications for Use:
The truFreeze® System is intended for use as a cryosurgical tool for the destruction of unwanted tissue in the fields of dermatology, gynecology, and general surgery.
Prescription Use V (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Surgical, Orthopedic,
and Restorative Devices
510(k) Number k113021
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.