K152668 · CSA Medical, Inc. · GEH · Oct 14, 2015 · General, Plastic Surgery
Device Facts
Record ID
K152668
Device Name
truFreeze System
Applicant
CSA Medical, Inc.
Product Code
GEH · General, Plastic Surgery
Decision Date
Oct 14, 2015
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4350
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The truFreeze® System is intended for cryogenic destruction of tissue requiring either active or passive venting during surgical procedures. The truFreeze® System is Indicated for use as a cryosurgical tool in the fields of dermatology, gynecology, and general surgery, to ablate benign and malignant lesions.
Device Story
The truFreeze System is a cryosurgical device that delivers medical-grade liquid nitrogen to a treatment area via a small, low-pressure, open-tipped catheter. The system comprises a mobile console and a disposable spray kit. The console includes a touch panel computer, cryogen, suction, and electronics modules. Users operate the system via a dual foot pedal, touch panel, or wireless remote control. An off-the-shelf controller manages cryogen level sensing, filling, pressure, cooling, defrost, suction, timing, and data management. The device is used in clinical settings by healthcare providers to ablate tissue. The system provides controlled cryo-ablation, which can benefit patients by removing benign or malignant lesions. Safety features include tank pressure relief valves, an isolated low-voltage power system, and an emergency stop button.
Clinical Evidence
No clinical data. Verification testing was performed to confirm that the modified surface finish and counterbore depth specifications prevented cryogen leaks during the filling process. Testing included multiple fill cycles on transfer blocks at the low and high ends of the modified depth specifications. All acceptance criteria were met with no leaking observed.
Technological Characteristics
Cryosurgical unit consisting of a mobile console and disposable sterile, single-use catheters. Console features a touch panel computer, cryogen/suction/electronics modules, and wireless remote. Operates via liquid nitrogen delivery. Modifications include improved surface finish and tightened counterbore depth dimensions for the seal receptacle. Software manages cryogen level, pressure, cooling, defrost, and timing. Electrical safety and EMC compliance maintained.
Indications for Use
Indicated for use as a cryosurgical tool in dermatology, gynecology, and general surgery to ablate benign and malignant lesions requiring active or passive venting.
Regulatory Classification
Identification
(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, with flowing lines above them that resemble a bird's wing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
CSA Medical Incorporated Ms. Sherrie Coval-Goldsmith Vice President Regulatory Affairs/Quality Assurance 91 Harwell Avenue Lexington, Massachusetts 02421
October 14, 2015
Re: K152668 Trade/Device Name: truFreeze® System Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical unit and accessories Regulatory Class: Class II Product Code: GEH Dated: September 16, 2015 Received: September 17, 2015
Dear Ms. Coval-Goldsmith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR [SELECT ONE: Part 801 [or, for IVDs only] Parts 801 and 809]); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS)
{2}------------------------------------------------
regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Joshua C. Nipper -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration<br><b>Indications for Use</b> | Form Approved: OMB No. 0910-0120<br>Expiration Date: January 31, 2017<br>See PRA Statement below. |
|-------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number (if known) | K152668 |
| Device Name | truFreeze® System |
| Indications for Use (Describe) | The truFreeze® System is intended for cryogenic destruction of tissue requiring either active or passive venting during surgical procedures.<br><br>The truFreeze® System is Indicated for use as a cryosurgical tool in the fields of dermatology, gynecology, and general surgery, to ablate benign and malignant lesions. |
| Type of Use (Select one or both, as applicable) | |
| | <div> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </div> <div> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
| FORM FDA 3881 (1/14) | Page 1 of 1 |
|----------------------|----------------------------------------|
| | PSC Publishing Services (301) 443-6740 |
1-9
{4}------------------------------------------------
Page 1 of 3
# Attachment D - 510(k) Summary (truFreeze® System)
Applicant Establishment Registration Number Contact Person
- Summary Date Proprietary Name Classification Classification Name Regulation Number Classification Predicate Device
CSA Medical 3004534508 Sherrie Coval-Goldsmith Vice President RA/QA CSA Medical 91 Hartwell Ave Lexington, Ma 02421 Phone: 781-538-7447 Fax: 781-538-4730 sgoldsmith@csamedical.com September 16, 2015 truFreeze® System Class II Cryosurgical Unit, Cryogenic Surgical Device 21 CFR 878.4350 Product Code GEH K143625 (truFreeze System)
## Device Description
The truFreeze system is a cryosurgical tool that applies medical-grade liquid nitrogen to the treatment area via a small, low pressure, open tipped catheter. The truFreeze System consists of a console and a disposable spray kit.
### Console:
The console is the central interface of the system and is comprised of a touch panel computer (TPC) and cryogen, suction, and electronics modules packaged in a mobile cart. Users interact with the console through a dual foot pedal and a touch panel. An off-the-shelf controller and associated software manage the cryogen level sensing, filling, pressure, cooling, defrost, suction, timing and data management functions. A wireless remote control provides alternative timer control from a distance in the treatment room. A fill kit, stored on the rear of the console, allows for liquid nitrogen transfer from the source tank to the console. Safety features include indicators, tank pressure relief valves, an isolated low voltage power system, and an emergency button to be used in the event of user or technical malfunction.
### Disposable spray kit:
There are no proposed changes to the disposable spray kit. There are 2 types of spray kits available. One kit is available for active venting procedures and one is available for passive venting procedures. Both active and passive venting kits include are provided in a carton of five (5) individually packaged sterile, single-use catheters with introducers in individual pouches. Additionally, the active venting kit is provided with includes a carton of five (5) individually packaged sterile, single-use CDTs with associated tubing in individual pouches. Each carton within a spray kit contains the instructions for use.
## Labeling (Intended Use/Indications for Use and Instructions for Use Document)
The truFreeze System is identical in its Intended Use as the predicate device (K150920 truFreeze system). Both devices describe the ablation of benign and malignant tissue in general terms and the requirement to use either active or passive venting during surgical procedures. Both devices are indicated for use as a cryosurgical tool in the fields of dermatology, gynecology, and general surgery, to ablate benign and malignant lesions. This is the same indication for use as previously cleared for the previous version of the truFreeze System (K150920).
{5}------------------------------------------------
# Technical and Operational Characteristics
This 510(k) implements a design modification to the surface finish and depth of the bore where the seal is seated in the receptacle body. These modifications consisted of (1) improving the finish of the surface that contacts the seal OD from a lower to a smoother finish and (2) tightening the counterbore depth dimension to ensure compression of the cryo seal flange by the mating collar. The proposed modifications are tighter than the current specifications, but still fall within the current range of these two specifications. Other than this specification modification, the technical and operational characteristics of the truFreeze System are unchanged.
## Testing
The truFreeze System was previously subjected to a comprehensive test program that included electrical safety and electromagnetic compatibility test, software tests, animal testing, biocompatibility and sterilization testing. The proposed modifications were tested in CSAM's verification testing.
Components were built to the smoother surface finish as well as at both ends of the proposed modified depth specification. Two parts at each end of the specification were tested. Testing at the low and high end modified depth specification confirmed that performance was acceptable across the entire tolerance range. Fill testing was performed to subject the transfer blocks to multiple fill cycles, to confirm that leaking would not result with the proposed changes to the specifications. Each transfer block was subjected to 6 fills, or 3 times the number of fills when the leaks occurred. No leaks were observed across multiple fills, nor did they introduce any new safety concerns as the receptacle body performed as intended. All acceptance criteria were met. Therefore, the proposed specifications are considered verified.
| Transfer<br>Interface<br>Assembly<br>Sample # | Leaking Result | Acceptance Criteria | Pass/Fail |
|-----------------------------------------------|---------------------|---------------------|-----------|
| B-1 (Low end of<br>depth spec) | No leaking observed | No leaking observed | PASS |
| | No leaking observed | | PASS |
| | No leaking observed | | PASS |
| | No leaking observed | | PASS |
| | No leaking observed | | PASS |
| | No leaking observed | | PASS |
| | No leaking observed | | PASS |
| B-2 (Low end of<br>depth spec) | No leaking observed | | PASS |
| | No leaking observed | | PASS |
| | No leaking observed | | PASS |
| | No leaking observed | | PASS |
| | No leaking observed | | PASS |
| | No leaking observed | | PASS |
| C-1(High end of<br>depth spec) | No leaking observed | | PASS |
| | No leaking observed | | PASS |
| | No leaking observed | PASS | |
| | No leaking observed | PASS | |
| Table 1. Summary of Test Results | |
|----------------------------------|--|
{6}------------------------------------------------
Page 3 of 3
| | No leaking observed | PASS |
|--------------------------------|---------------------|------|
| | No leaking observed | PASS |
| C-2(High end of<br>depth spec) | No leaking observed | PASS |
| | No leaking observed | PASS |
| | No leaking observed | PASS |
| | No leaking observed | PASS |
| | No leaking observed | PASS |
| | No leaking observed | PASS |
# Rationale For Substantial Equivalence
The labeling as well as the technological characteristics of the proposed truFreeze System and the predicate device (K150920 truFreeze system) were compared. The Intended Use/Indications for Use statement of the two devices had identical general claims and do not raise new questions of safety and performance. The proposed changes to the technology were compared and are identical. The specification changes fall within the current specifications and are considered a tightening of the specifications. The tightening of the specifications will address complaints of cryogen leak occurring during the filling of the truFreeze system prior to system use and do not raise new questions of safety and performance.
# Conclusion
Based on the comparison of labeling, technology and verification testing comparisons, the truFreeze device is substantially equivalent to the predicate device listed above.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.