High Strength Suture

K192637 · Yunyi (Beijing) Medical Device Co., Ltd. · GAT · Mar 2, 2020 · General, Plastic Surgery

Device Facts

Record IDK192637
Device NameHigh Strength Suture
ApplicantYunyi (Beijing) Medical Device Co., Ltd.
Product CodeGAT · General, Plastic Surgery
Decision DateMar 2, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.5000
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The device is indicated for use in general soft tissue approximation and/or ligation, including use in allograft tissues for orthopedic procedures.

Device Story

Braided, non-absorbable surgical suture composed of ultra-high molecular weight polyethylene (UHMWPE) with medical adhesive at both ends. Device is undyed, uncoated, and provided sterile. Used by surgeons for soft tissue approximation and ligation, including orthopedic allograft procedures. Functions as a mechanical fixation device. Benefits include high strength for tissue approximation.

Clinical Evidence

No clinical data. Bench testing only. Compliance verified via USP 41-NF 36:2018 standards for tensile strength and diameter, ASTM standards for packaging, and ISO 10993 series for biocompatibility (cytotoxicity, irritation, sensitization, systemic toxicity, pyrogenicity, and implantation).

Technological Characteristics

Braided ultra-high molecular weight polyethylene (UHMWPE) suture; medical adhesive at ends; undyed; uncoated. Sterilized via Ethylene Oxide (EO). Complies with USP 41-NF 36:2018 (Sutures, Tensile Strength, Diameter), ASTM F88/F88M-15, ASTM F1929-15, and ISO 10993 series.

Indications for Use

Indicated for general soft tissue approximation and/or ligation, including use in allograft tissues for orthopedic procedures.

Regulatory Classification

Identification

Nonabsorbable poly(ethylene terephthalate) surgical suture is a multifilament, nonabsorbable, sterile, flexible thread prepared from fibers of high molecular weight, long-chain, linear polyesters having recurrent aromatic rings as an integral component and is indicated for use in soft tissue approximation. The poly(ethylene terephthalate) surgical suture meets U.S.P. requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be provided uncoated or coated; and it may be undyed or dyed with an appropriate FDA listed color additive. Also, the suture may be provided with or without a standard needle attached.

Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. March 2, 2020 Yunyi (Beijing) Medical Device Co., LTD % Diana Hong General Manager Mid-Link Consulting Co. Ltd P.O Box 120-119 Shanghai, 200120 Cn Re: K192637 Trade/Device Name: High Strength Suture Regulation Number: 21 CFR 878.5000 Regulation Name: Nonabsorbable Poly(Ethylene Terephthalate) Surgical Suture Regulatory Class: Class II Product Code: GAT Dated: September 20, 2019 Received: September 24, 2019 Dear Ms. Hong: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, For Cindy Chowdhury, Ph.D., M.B.A. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K192637 Device Name High Strength Suture Indications for Use (Describe) The device is indicated for use in general soft tissue approximation and/or ligation, including use in allograft tissues for orthopedic procedures X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # Tab #6 510(k) Summary This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92. The assigned 510(k) Number: - 1. Date of Preparation: 08/06/2019 - 2. Sponsor Identification ### Yunyi (Beijing) Medical Device Co., LTD. Floor 1-4, Building 4, No. 9 Tianfu Street, Biomedical Base, Daxing District, Beijing 102600, People's Republic of China Establishment Registration Number: Not registered Contact Person: Xiaowei Liu Position: General Manager Tel: +86-10-57234533 Fax: +86-10-57234533 Email:liuxiaowei@yunyimedical.com - 3. Designated Submission Correspondent Ms. Diana Hong (Primary Contact Person) Mr. Ying Xu (Alternative Contact Person) Mid-Link Consulting Co., Ltd P.O. Box 120-119, Shanghai, 200120, China Tel: +86-21-22815850 Fax: 360-925-3199 Email: info@mid-link.net {4}------------------------------------------------ #### Identification of Proposed Device 4. Trade Name: High Strength Suture Common Name: Polyblend Suture, Non-absorbable Surgical Sutures Size: UP-(2-0) W, UP-0W, UP-2W, UP-5W and UP-7W ### Regulatory Information Classification Name: Nonabsorbable Poly (Ethylene Terephthalate) Surgical; Classification: II; Product Code: GAT; Regulation Number: 21 CRFR878.5000 Review Panel: General& Plastic Surgery; ### Indication for Use: The device is indicated for use in general soft tissue approximation and/or ligation, including use in allograft tissues for orthopedic procedures. ### Device Description The proposed device is a braided synthetic, non-absorbable surgical suture made of ultra-high molecular weight polyethylene and medical adhesive at both ends of the suture. The proposed device is undyed and uncoated. The sutures are available in a range of sizes. The device does not contain needle. And the device is provided in sterile. - ડ. Identification of Predicate Device 510(k) Number: K190817 Product Name: HS Fiber® Suture Manufacturer: Riverpoint Medical - 6. Non-Clinical Test Conclusion Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards: - > USP 41-NF 36:2018 Non-absorbable Surgical Suture - > USP 41-NF 36:2018 <881> Tensile Strength {5}------------------------------------------------ - > USP 41-NF 36:2018 <861> Sutures – Diameter - > ASTM F88/F88M-15 Standard test method for seal strength of flexible barrier materials - > ASTM F1929-15 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration - > ISO 10993-5:2009 Biological Evaluation Of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity - > ISO 10993-10:2010 Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization - > ISO 10993-7:2008 Biological Evaluation Of Medical Devices - Part 7: Ethylene Oxide Sterilization Residuals - > ISO 10993-11:2017 Biological Evaluation Of Medical Devices - Part 11: Tests For Systemic Toxicity - > USP 41 NF 36 <151> Pyrogen Test (USP Rabbit Test) - > ISO 10993-3:2014 Biological Evaluation Of Medical Devices - Part 11: Tests For Systemic Toxicity - > ISO 10993-6:2016 Biological Evaluation Of Medical Devices -- Part 6: Tests For Local Effects After Implantation - USP 41 NF 36 <85> Bacterial Endotoxins Test > - 7. Clinical Test Conclusion No clinical study is included in this submission. {6}------------------------------------------------ #### 8. Substantially Equivalent (SE) Comparison | Table 1 Comparison of Technology Characteristics | | | |--------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Item | Proposed Device | Predicate Device | | | | K190817 | | Product Code | GAT | GAT | | Regulation Number | 21 CRFR878.5000 | 21 CRFR878.5000 | | Indication for Use | The device is indicated for use in general soft tissue approximation and/or ligation, including use in allograft tissues for orthopedic procedures. | HS Fiber sutures are indicated for use in general soft tissue approximation and/or ligation, including use in cardiovascular surgery, and the use of allograft tissues for orthopedic procedures. | | Material | ultra-high molecular weight polyethylene | ultra-high molecular weight polyethylene | | Color | Undyed | Dyeing | | Absorbable /<br>Non-absorbable | Non-absorbable | Non-absorbable | | Braided /<br>Monofilament | Braided | Braided | | Sterile | EO Sterilized | Unknown | | Suture Size | UP-(2-0) W, UP-0W, UP-2W, UP-5W ,UP-7W | Unknown | | Single Use | Yes | Yes | | Performance | Comply with:<br>USP 41 <861><br>USP 41 <881> | Comply with:<br>USP<861><br>USP<881> | | Biocompatibility | | | | Cytotoxicity | No cytotoxicity | Comply with ISO 10993 | | Intracutaneous Study | No irritation | | | Sensitization | No sensitization | | | Pyrogenicity Test | No pyrogenicity | | | Acute Systemic Toxicity | No systemic toxicity | | | Bacterial Reverse Mutation | Not induce backward mutation | | | Gene Mutation Test | Non-mutagenic | | | Muscle Implantation Test | Non-irritant to the muscle tissue | | | Subchronic Toxicity Test | No subchronic toxicity | | 9. Substantially Equivalent (SE) Conclusion {7}------------------------------------------------ Based on the comparison and analysis above, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate devices.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...