K060165 · Axya Medical, Inc. · GAT · Sep 1, 2006 · General, Plastic Surgery
Device Facts
Record ID
K060165
Device Name
AXYAFLEX POLYESTER SUTURE
Applicant
Axya Medical, Inc.
Product Code
GAT · General, Plastic Surgery
Decision Date
Sep 1, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5000
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
AxyaFlex™ polyester suture is indicated for use in general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic and neurological procedures.
Device Story
AxyaFlex™ is a nonabsorbable, braided, sterile surgical suture composed of poly(ethylene terephthalate co-isophthalate). Designed for general soft tissue approximation and ligation; used in cardiovascular, ophthalmic, and neurological procedures. Braided structure provides optimal handling properties. Device is used by surgeons in clinical settings. Performance is verified through USP-compliant testing of diameter, length, and knot tensile strength. Biocompatibility and chemical analyses confirm equivalence to established polyester suture materials. Sterilized using industry-standard processes.
Clinical Evidence
Bench testing only. No clinical data. Studies evaluated suture diameter, length, and knot tensile strength per USP requirements and FDA Class II Special Controls Guidance. Biocompatibility and chemical analyses performed to demonstrate material equivalence to predicate polyester sutures.
Technological Characteristics
Nonabsorbable, braided, sterile surgical suture. Material: poly(ethylene terephthalate co-isophthalate). High molecular weight, long-chain, linear polyester with recurrent aromatic rings. Physical properties (diameter, length, knot tensile strength) meet USP standards. Sterilization via process equivalent to predicate manufacturers.
Indications for Use
Indicated for general soft tissue approximation and/or ligation in patients requiring cardiovascular, ophthalmic, or neurological surgical procedures.
Regulatory Classification
Identification
Nonabsorbable poly(ethylene terephthalate) surgical suture is a multifilament, nonabsorbable, sterile, flexible thread prepared from fibers of high molecular weight, long-chain, linear polyesters having recurrent aromatic rings as an integral component and is indicated for use in soft tissue approximation. The poly(ethylene terephthalate) surgical suture meets U.S.P. requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be provided uncoated or coated; and it may be undyed or dyed with an appropriate FDA listed color additive. Also, the suture may be provided with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
Genzyme Surgical Products "Cottony" II Dacron Polyester Nonabsorbable Suture (K001434)
Reference Devices
Ethicon, Inc. Mersilene® Polyester Fiber Suture
Submission Summary (Full Text)
{0}------------------------------------------------
K060165
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS (Per 21 CFR 807.92)
# General Company Information
SEP - 1 2006
- Name: Axya Medical, Inc. Contact: Howard Schraver Regulatory Affairs Consultant
- 100 Cummings Center Address: Suite 444C Beverly, MA 01915
- Telephone: (978) 232 - 9997 Fax: (978) 232 - 9998
- Date Prepared July 20, 2006
## General Device Information
- Product Name: AxyaFlex™ polyester suture
- "Nonabsorbable poly(ethylene terephthalate) surgical suture" Classification: Product code: GAT - Class II 21 CFR 878.5000
### Predicate Devices
Aesculap, Inc. Synthofil™ PET Sutures [510 (k) Number K990088]
Genzyme Surgical Nextstitch™ PET Sutures [510(k) Number K001440]
United States Surgical Corp. Nonabsorbable Dacron® sutures [510(k) Number K902873]
Sharpoint, Inc. Dacron® Sutures [510(k) Number K904263]
{1}------------------------------------------------
K060165
Page 2 of 3
Ethicon, Inc. Mersilene® Polyester Fiber Suture A pre-enactment device
Genzyme Surgical Products "Cottony" II Dacron Polyester Nonabsorbable Suture [501(k) Number K001434]
### Description
AxyaFlex™ polyester suture is a nonabsorbable, braided, sterile, surgical suture composed of poly(ethylene terephthalate co-isophthalate). It is prepared from fibers of high molecular weight, long-chain, linear polyesters having recurrent aromatic rings as an integral component. AxyaFlex™ sutures are braided for optimal handling properties.
## Intended Use (Indications)
AxyaFlex™ suture is indicated for use in general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic and neurological procedures.
## Substantial Equivalence
This submission supports the position that the AxyaFlex™ polyester suture is substantially equivalent to a number of previously cleared devices, including the Ethicon, Inc. Mersilene® polyester suture [pre-enactment] and the Genzyme Surgical Products "Cottony" II Dacron Polyester Nonabsorbable Suture.
The 510(k) Notice contains summaries of studies that were conducted to evaluate suture diameter, length, and knot tensile strength as specified in the USP and in the FDA Guidance Document entitled "Class II Special Controls Guidance Document: Surgical Sutures, Guidance for Industry and FDA" (June 23, 2003). In addition, biocompatibility testing and chemical analyses were conducted to demonstrate that the suture material was equivalent to predicate polyester material.
The data presented demonstrate that the AxyaFlex™ polyester suture satisfied the USP requirements and was equivalent to the predicate materials.
The sutures are provided sterile. The suture material is sterilized using a process equivalent to the process used by the predicate suture manufacturers.
{2}------------------------------------------------
K060165
# Conclusions
Axya Medical, Inc. believes that the information provided establishes that similar legally marketed sutures have been used for the same clinical applications as the indications as the indications for the AxyaFlex™ polyester suture. The material from which the Axya device is fabricated has an established history of use in medical applications; and devices produced by Axya have been tested in accordance with applications, and devices
{3}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP - 1 2006
Axya Medical, Inc. % Mr. Howard L. Schrayer 100 Cummings Center, Suite 444C Beverly, Massachusetts 01915
Re: K060165
Trade/Device Name: AxyaFlex™ polyester suture Regulation Number: 21 CFR 878.5000 Regulation Name: Nonabsorbable poly(ethylene terephthalate) surgical suture Regulatory Class: II Product Code: GAT Dated: July 21, 2006 Received: July 25, 2006
Dear Mr. Schrayer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Mr. Howard L. Schrayer
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark N. Mekerson
Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Page 1 of 1
### INDICATIONS FOR USE
KO60165 510(k) Number (if known):
Device Name: AxyaFlex™ polyester suture
Indications For Use:
AxyaFlex™ polyester suture is indicated for use in general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic and neurological procedures.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
```
(Division Sign-Off)
Division of General, Restorative, and Neurological Devices
510(k) Number 4 k060163
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.