Phasix ST Mesh

K173143 · C.R. Bard, Inc. · OWT · Apr 25, 2018 · General, Plastic Surgery

Device Facts

Record IDK173143
Device NamePhasix ST Mesh
ApplicantC.R. Bard, Inc.
Product CodeOWT · General, Plastic Surgery
Decision DateApr 25, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.3300
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Phasix™ ST Mesh is indicated for use in the reinforcement of soft tissue, where weakness exists, in procedures involving soft tissue repair, such as for the repair of hernias, including hiatal hernias.

Device Story

Phasix™ ST Mesh is a sterile, fully resorbable surgical mesh prosthesis for soft tissue reinforcement and reconstruction. It consists of a co-knitted mesh of poly-4-hydroxybuterate (P4HB) and polyglycolic acid (PGA) fibers, coated on the visceral side with a resorbable hydrogel (sodium hyaluronate, carboxymethylcellulose, and polyethylene glycol). The mesh is implanted by surgeons during soft tissue repair procedures (e.g., hernia repair) to provide mechanical support during the healing period. The fascial side promotes tissue ingrowth, while the hydrogel coating minimizes visceral tissue attachment. The P4HB fibers undergo slow hydrolytic and enzymatic degradation, maintaining strength throughout the 12-week healing period, with significant degradation occurring between 12-18 months. The hydrogel coating resorbs within 30 days. The device is intended to provide structural support to weakened tissues, potentially reducing recurrence rates in hernia repairs.

Clinical Evidence

Evidence includes two retrospective clinical studies on hiatal hernia repair. Study 1 (n=50) showed one asymptomatic recurrence and no mesh-related complications at 3-18 months. Study 2 (n=180) with median 4-month follow-up (n=84 objective follow-up) reported two recurrent asymptomatic hernias (1-2 cm and 2-3 cm) requiring no reintervention, with no mesh-related complications or erosions observed.

Technological Characteristics

Co-knitted poly-4-hydroxybuterate (P4HB) and polyglycolic acid (PGA) fibers. Visceral coating: resorbable hydrogel (sodium hyaluronate, carboxymethylcellulose, polyethylene glycol). Resorbable. Sterile. No electrical or metal components. Biocompatibility per ISO 10993.

Indications for Use

Indicated for reinforcement of soft tissue where weakness exists, including hernia repair (e.g., hiatal hernias). Applicable to adult patients undergoing soft tissue repair procedures.

Regulatory Classification

Identification

Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. April 25, 2018 C. R. Bard Inc Mr. Tony John Senior Regulatory Affairs Specialist 100 Crossings Boulevard Warwick, Rhode Island 02886 # Re: K173143 Trade/Device Name: Phasix ST Mesh Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: Class II Product Code: OWT, OWZ, OXC, FTL, OOD Dated: March 28, 2018 Received: March 29, 2018 Dear Mr. John: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may: therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # David Krause -S for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K173143 Device Name Phasix™ ST Mesh #### Indications for Use (Describe) The Phasix™ ST Mesh is indicated for use in the reinforcement of soft tissue, where weakness exists, in procedures involving soft tissue repair, such as for the repair of hernias, including hiatal hernias. Type of Use (Select one or both, as applicable) | <span> <b> ☑ </b> Prescription Use (Part 21 CFR 801 Subpart D) </span> | |--------------------------------------------------------------------------------------------------------| | <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ### Phasix™ ST Mesh #### 510(k) Summary 21 CFR 807.92 As required by the Safe Medical Devices Act of 1990, coded under Section 513, Part (()(3)(A) of the Food, Drug and Cosmetic Act, a 510(k) summary upon which substantial equivalence determination is based is as follows: ### Submitter Information: | Company Name: | Davol Inc., Subsidiary of C. R. Bard, Inc. | |---------------------|----------------------------------------------------| | Company Address: | 100 Crossings Boulevard<br>Warwick, RI 02886 | | Telephone: | (401) 825-8692 | | Fax: | (401) 825-8765 | | Contact person: | Tony John, MS<br>Sr. Regulatory Affairs Specialist | | Date of Submission: | September 20, 2017 | #### Subject Device Name: | Name of Device: | Phasix™ ST Mesh | |-----------------------|----------------------------------------------| | Common or Usual Name: | Surgical Mesh | | Classification Name: | Mesh, Surgical, Absorbable, Abdominal Hernia | | Regulatory Class: | Class II | | Regulation Number: | 878.3300 | | Product Code: | OWT, OOD, FTL, OXC, OWZ | ## Predicate Device: | Name of Device: | Gore® BioAbsorbable (Bio-A®) Mesh | | |--------------------------------------------|-----------------------------------|--| | 510(k)#: | K033671 | | | Common or Usual Name: Bioabsorbablel Mesh | | | | Classification Name: | Mesh, Surgical, Polymeric | | | Regulatory Class: | Class II | | | Regulation Number: | 878.3300 | | | Product Code: | OWT, OWZ, OXC | | ### Reference Predicate Device: | Name of Device: | Phasix™ ST Mesh | |-----------------------|----------------------------------------------| | 510(k)#: | K143380 | | Common or Usual Name: | Surgical Mesh | | Classification Name: | Mesh, Surgical, Absorbable, Abdominal Hernia | | Regulatory Class: | Class II | | Regulation Number: | 878.3300 | {4}------------------------------------------------ #### Product Code: OWT, OOD, FTL ### Device Description: The Phasix™ ST Mesh is a fully resorbable mesh with a resorbable hydrogel coating. It is a sterile mesh prosthesis designed for the reinforcement and reconstruction of soft tissue deficiencies. Phasix ™ ST Mesh is co-knitted using poly-4-hydroxybuterate (P4HB) and polyglycolic acid (PGA) fibers. P4HB is produced from a naturally occurring monomer and is processed into monofilament fibers and then knitted into a surgical mesh. P4HB degrades through a process of hydrolysis and a hydrolytic enzymatic digestive process. It has been developed to reinforce areas where weakness exists while minimizing the variability of resorption rate (loss of mass) and strength to provide support throughout the expected healing period. Preclinical implantation studies indicate that resorption of the P4HB fibers is minimal throughout the 12 week expected healing period and up to 26 weeks post implantation. Significant degradation of the mesh fibers observed in preclinical studies within 12 to 18 months indicates loss in mechanical integrity and strength. While fiber segments were observed at 18 months, they continued to degrade. Phasix™ ST Mesh is coated on the PGA surface with a resorbable, chemically modified sodium hyaluronate (HA), carboxymethylcellulose (CMC) and polyethylene glycol (PEG) based hydrogel. The fascial side of the mesh allows for a prompt fibroblastic response through the interstices of the mesh, allowing for complete tissue ingrowth, similar to P4HB mesh alone. The visceral side of the mesh is a resorbable hydrogel coating, separating the mesh from underlying tissues and organ surfaces to help minimize tissue attachment to the mesh. Shortly after hydration, the biopolymer coating becomes a hydrated gel that is resorbed from the site in less than 30 days. #### Indications for Use of Device: Phasix™ ST Mesh is indicated for use in the reinforcement of soft tissue, where weakness exists, in procedures involving soft tissue repair, such as for the repair of hernias, including hiatal hernias. # Technological Comparison to Predicate Devices: The proposed Phasix™ ST Mesh is identical in material and design to the predicate Phasix™ ST Mesh (K143380) and has comparable performance with the Gore Bio-A reference predicate. The proposed and predicate devices have the same intended use and similar indications for use statements. {5}------------------------------------------------ The following technological differences exist between the subject and the primary predicate device: - . They are composed of different resorbable materials. - They have minor difference in the following physical characteristics: . - Mesh Thickness o - Mesh Density o - Mesh Thickness o - There are differences in the following performance characteristics: ● - Ball Burst Strength о - Tear Strength o - Resorption Profile o Where minor technological differences exist between the proposed and reference predicate device, testing demonstrates that the differences do not adversely affect the safety and performance of the proposed device. ### Performance Data: Since the proposed and reference predicate device are identical, the following biocompatibility and animal performance data are being leveraged from the K143380 510(k) submission in support of the substantial equivalence determination. ### Biocompatibility Testing The materials used in the construction of the Phasix™ ST Mesh are were evaluated with complete biocompatibility testing in accordance with FDA's Blue Book Memorandum #G95-1 issued on May 1, 1995, Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing" and FDA Guidance "Use of International Standard ISO-10993, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing Within a Risk Management Process". The following studies were completed: - . Cytotoxicity - Sensitization . - . Intracutaneous Reactivity - Systemic Toxicity (Acute) - Pyrogenicity - . Genotoxicity {6}------------------------------------------------ - Local and Systemic Toxicity (4 and 13 week) . - Local Toxicity (4, 8, 13, and 20 week) . ### Animal Studies In vivo porcine studies were performed to characterize the mechanical strength, tissue response, and resorption profile of the Phasix™ ST Mesh device at 4, 12, and 24 weeks. ### Bench Testing Bench testing was performed to compare the proposed Phasix™ ST Mesh to the cited predicate Gore® Bio-A® device. In accordance with FDA's "Guidance for the Preparation of a Premarket Notification Application for a Surgical Mesh" published March 2, 1999, the following physical and performance characteristics were evaluated: - . Mesh thickness - . Mesh density - Mesh pore size ● - Burst strength - . Mesh knit/weave characteristics - Device stiffness - Tear strength/resistance ● - Suture pullout strength. . ### Electrical safety and electromagnetic compatibility (EMC) There are no electrical or metal components in the Phasix™ Mesh; therefore the proposed device does not require EMC and Electrical Safety evaluation. ### Software Verification and Validation Testing The proposed Phasix™ Mesh does not contain software. ### Clinical Study Data and discussion from 2 separate retrospective clinical studies were included in this submission, demonstrating the safety and performance of the Phasix™ ST Mesh in hiatal hernia repair. {7}------------------------------------------------ - A retrospective review of 50 consecutive patients who had a paraesophageal hernia . repair with Phasix™ ST Mesh reinforcement of the crural closure at a single institution. | Demographics | Total | Female | Male | |--------------|-----------------------|----------|----------| | Patients | 50 | 34 (68%) | 16 (32%) | | Mean Age | 65 years | | | | Mean BMI | 30.24 kg/m2 | | | | BMI Range | 17.1 kg/m2-41.5 kg/m2 | | | The following procedures were performed: - Primary hiatal hernia repair o - o Fundolipication - Partial gastroectomy o - o Collisgastrophy - Relaxing incision o All patients either underwent objective follow-up (EGD & UGI) or non-objective follow-up at time points ranging from 3, 12, and 18 months. There was one small, asymptomatic recurrence and no mesh related complications or erosions. - A retrospective review of 180 patients who underwent laparoscopic hiatal hernia ● repair with Phasix™ ST Mesh to reinforce the crural closure was performed at a single institution with multiple surgeons. This review included any patient who had Phasix™ ST Mesh used to reinforce the crural closure and included other concomitant procedures. | Demographics | Total | Male | Female | |-------------------------------------|-------------|----------|-----------| | Patients | 180 | 60 (33%) | 120 (67%) | | Median Age | 66 years | | | | Preoperative Diagnosis | | | | | Gastrointestinal Reflux<br>(GERD) | 59 (32.8%) | | | | Paraesophageal hernia<br>(PEH) | 120 (66.7%) | | | | Morgagni hernia | 1 (0.5%) | | | | Comorbidity | | | | | Steroid use or Diabetes<br>Mellitus | 22 | | | | Operation | | | | {8}------------------------------------------------ | First time | 142 (78.9%) | |------------------|-------------| | Re-operation | 38 (21.1%) | | Median ASA Score | 2 | The following procedures were performed: - Nissen fundoplication O - o Partial (Toupet or Dor) fundoplication - LINX implantation o - partial or total gastrectomy O - Collis gastroplasty o - Diaphragm relaxing incision o Objective follow-up was conducted in 84 patients with a median follow-up time of 4 months postoperatively. Esophagogastroduodenoscopy (EDG) was performed in 37 patients, barium swallow (BS) in 20 patients and both EGD and BS in 27 patients. There were 2 recurrent asymptomatic hernias identified in two patients. The first was identified in a patient by EDG during routine Collis gastroplasty follow-up and the second identified during percutaneous endoscopic gastrostomy (PEG) tube placement. The sizes of the hernias were 1-2 cm and 2-3 cm respectively and did not require any reintervention. This study did not observe any mesh related complications or mesh erosion. #### Conclusions: All test results provided in this submission demonstrate that the proposed Phasix™ ST Mesh is substantially equivalent to both the Gore® Bio-A® predicate and the Phasix™ ST Mesh reference predicate.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...