K151486 · Siemens Medical Solutions USA, Inc. · KPS · Jul 17, 2015 · Radiology
Device Facts
Record ID
K151486
Device Name
Biograph mCT and mCT Flow PET/CT Scanners
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
KPS · Radiology
Decision Date
Jul 17, 2015
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1200
Device Class
Class 2
Intended Use
The Siemens Biograph mCT systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information. The CT component produces cross-sectional images of the body by computer reconstruction of X-Ray transmission data from either the same axial plane taken at different angles or spiral planes taken at different angles. The PET subsystem images and measures the distribution of PET radiopharmaceuticals in humans for the purpose of determining various metabolic (molecular) and physiologic functions within the human body and utilizes the CT for fast attenuation correction maps for PET studies and precise anatomical reference for the fused PET and CT images. The system maintains independent functionality of the CT and PET devices, allowing for single modality CT and / or PET diagnostic imaging. These systems are intended to be utilized by appropriately trained health care professionals to aid in detecting, localizing, diagnosing, staging and restaging of lesions, tumors, disease and organ function for the evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The images produced by the system can also be used by the physician to aid in radiotherapy treatment planning and interventional radiology procedures.
Device Story
Biograph mCT and mCT Flow are combined PET/CT scanners for whole-body oncology, neurology, and cardiology imaging. Systems acquire X-ray transmission data (CT) and radiopharmaceutical distribution data (PET) to provide registered, fused physiologic and anatomic images. Operated by trained healthcare professionals in clinical settings, the system includes patient handling, acquisition, and processing workstations. Software manages patient data, scan control, image reconstruction, and archival. Physicians use fused images to detect/stage disease, plan radiotherapy, and guide interventional procedures. Benefits include improved diagnostic accuracy through combined metabolic and anatomic visualization. Modifications include software upgrades (Syngo, Somaris, PETsyngo), new features like 4D lung scanning, iterative metal artifact reduction (iMAR), temporal intensity projections (tMIP), and enhanced iterative reconstruction (ADMIRE).
Clinical Evidence
Bench testing only. PET subsystem performance evaluated per NEMA NU2:2012 standards on 3-ring and 4-ring configurations. Metrics included transverse/axial resolution (FWHM), sensitivity, peak NECR, peak trues, count rate bias, and scatter fraction. All results met predetermined acceptance criteria. No clinical data presented.
Technological Characteristics
Combined PET/CT scanner. CT component: multi-slice X-ray transmission. PET component: radiopharmaceutical distribution imaging. Software: Syngo, Somaris, and PETsyngo platforms. Features: iterative reconstruction (ADMIRE), iterative metal artifact reduction (iMAR), 4D lung scanning, and temporal intensity projections. Connectivity: DICOM compliant. Standards: IEC 60601-1, 21 CFR 1020.30, 21 CFR 1020.33, NEMA NU2:2012.
Indications for Use
Indicated for use by trained healthcare professionals to aid in detecting, localizing, diagnosing, staging, and restaging of lesions, tumors, disease, and organ function in patients for evaluation of conditions including cardiovascular disease, neurological disorders, and cancer. Images may also be used for radiotherapy treatment planning and interventional radiology procedures.
Regulatory Classification
Identification
An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.
Predicate Devices
Biograph mCT and mCT Flow PET/CT Scanners (K123737)
K251671 — Biograph Vision PET/CT Systems; Biograph mCT PET/CT Systems · Siemens Medical Solutions USA, Inc. · Jul 3, 2025
K113448 — BIOGRAPH MCT FAMILY · Siemens Medical Solutions USA, Inc. · Dec 17, 2011
K101550 — BIOGRAPH MCT FAMILY · Siemens Medical Solutions USA, Inc. · Sep 3, 2010
K193248 — Biograph Vision, Biograph mCT Family of PET/CTs · Siemens Medical Solutions USA, Inc. · Feb 14, 2020
K170904 — Biograph Horizon PET/CT · Siemens Medical Solutions USA, Inc. · Jun 1, 2017
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized design featuring three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 3, 2015
Siemens Medical Solutions USA, Inc. % Alaine Medio, RAC PET and PCS Regulatory Projects Manager 810 Innovation Drive KNOXVILLE TN 37932
Re: K151486
Trade/Device Name: Biograph mCT and mCT Flow PET/CT Scanners Regulation Number: 21 CFR 892.1200 Regulation Name: Emission computed tomography system Regulatory Class: II Product Code: KPS, JAK Dated: June 1, 2015 Received: June 2, 2015
Dear Ms. Medio:
This letter corrects our substantially equivalent letter of July 17, 2015.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Michael D'Hara For
Robert Ochs, Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K151486
Device Name Biograph mCT and mCT Flow PET/CT Scanners
Indications for Use (Describe)
The Siemens Biograph mCT systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.
The CT component produces cross-sectional images of the body by computer reconstruction of X-Ray transmission data from either the same axial plane taken at different angles or spiral planes taken at different angles. The PET subsystem images and measures the distribution of PET radiopharmaceuticals in humans for the purpose of determining various metabolic (molecular) and physiologic functions within the human body and utilizes the CT for fast attenuation correction maps for PET studies and precise anatomical reference for the fused PET and CT images.
The system maintains independent functionality of the CT and PET devices, allowing for single modality CT and / or PET diagnostic imaging.
These systems are intended to be utilized by appropriately trained health care professionals to aid in detecting, localizing, diagnosing, staging and restaging of lesions, tumors, disease and organ function for the evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The images produced by the system can also be used by the physician to aid in radiotherapy treatment planning and interventional radiology procedures.
Type of Use (Select one or both, as applicable)
| <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------|
| <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff @fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
as required by 21 CFR Part 807.87(h)
| Identification of the Submitter | |
|---------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | M. Alaine Medio, RAC<br>PET and PCS Regulatory Projects Manager<br>Siemens Medical Solutions USA, Inc.<br>Molecular Imaging<br>810 Innovation Drive<br>Knoxville, TN 37932 |
| Manufacturers: | Siemens Medical Solutions USA, Inc.<br>Molecular Imaging<br>810 Innovation Drive<br>Knoxville, TN 37932<br>Siemens Medical Solutions USA, Inc.<br>Molecular Imaging<br>2501 North Barrington Road<br>Hoffman Estates, IL 60192 |
| Telephone Number: | (865)218-2703 |
| Fax Number: | (865)218-3019 |
| Date of Submission: | May 29th, 2015 |
| Identification of the product | |
| Device Proprietary Name: | Biograph mCT and mCT Flow PET/CT Scanners |
| Common Name: | Positron Emission Tomography (PET) System<br>Computed Tomography (CT) System |
| Classification Name: | Emission Computed Tomography System per 21 CFR<br>892.1200<br>Computed Tomography X-Ray System per 21 CFR<br>892.1750 |
| Product Code: | 90 KPS and 90 JAK |
| Classification Panel: | Radiology |
| Device Class: | Class II |
| Device Proprietary Name: | Biograph mCT and mCT Flow PET/CT Scanners |
| Manufacturer: | Siemens Medical Solutions USA, Inc. |
| Product Code: | 90 KPS and 90 JAK |
| Device Class: | Class II |
| 510(k) Number: | K123737 |
| Reference Devices: | |
| Device Name: | Somatom Definition AS/AS+ |
| | Somatom Definition Edge |
| 510(k) Numbers: | K143400 (Somatom Definition AS/AS+) |
| | K143401 (Somatom Definition Edge) |
{4}------------------------------------------------
## Marketed Devices to which Equivalence is claimed
Predicate:
#### Device Description:
The Biograph mCT and mCT Flow PET/CT Scanners are combined multi-slice X-Ray Computed Tomography and Positron Emission Tomography scanners. These systems are designed for whole body oncology, neurology and cardiology examinations. The Biograph mCT systems provide registration and fusion of high-resolution metabolic and anatomic information from the two major components of each system (PET and CT). Additional components of the system include a patient handling system and acquisition and processing workstations with associated software.
Biograph mCT software is a command based program used for patient management, data manaqement, scan control, image reconstruction and image archival and evaluation. All images conform to DICOM imaging format requirements.
The software for the Biograph mCT systems which is the subject of this application is substantially equivalent to the commercially available Biograph mCT software. Modifications include:
- Syngo software upgrade
- o Update of the synqo software to the latest revision (drivers, etc...)
- . Somaris Software (cleared in K143400, K143401))
- Upgrade to the latest revision of Somaris Software O
- Addition of new software features o
- Scan Time, Respiratory Rate and Pitch adjustments to provide for Full 4D lung scans
{5}------------------------------------------------
- . Twin beam scanning feature for Biograph mCT-X and Biograph mCT Edge
- . HD Field of View Pro
- Iterative Metal Artifact Reduction (iMAR)
- Temporal minimum intensity projection and temporal maximum intensity projections (tMIP)
- FAST 3D align / 3D Reconstruction ■
- FAST DE Results
- Repiratory Online mode
- Enhancements to the Iterative Reconstruction software (ADMIRE) o
- PETsyngo software ●
- Update of the software to add stitched whole body and stitched HD-Chest O feature
- Update to add Multi-series CT for Attenuation Correction O
- Summed whole body dynamic O
- Update of the software corrections O
- Update of the TrueD software component o
- = Address anomalies related to loading of PET images
- Update to improve usability
- Improvements to workflow o
- Improvements to performance (faster reconstruction speeds) O
# Intended Use:
The Siemens Biograph mCT systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.
The CT component produces cross-sectional images of the body by computer reconstruction of X-Ray transmission data from either the same axial plane taken at different angles or spiral planes taken at different angles. The PET subsystem images and measures the distribution of PET radiopharmaceuticals in humans for the purpose of determining various metabolic (molecular) and physiologic functions within the human body and utilizes the CT for fast attenuation correction maps for PET studies and precise anatomical reference for the fused PET and CT images.
The system maintains independent functionality of the CT and PET devices, allowing for single modality CT and / or PET diagnostic imaging.
These systems are intended to be utilized by appropriately trained health care professionals to aid in detecting, localizing, diagnosinq, staging and restaging of lesions, tumors, disease and organ function for the evaluation of diseases and disorders such as, but not limited to,
{6}------------------------------------------------
cardiovascular disease, neurological disorders and cancer. The images produced by the system can also be used by the physician to aid in radiotherapy treatment planning and interventional radiology procedures.
# Performance Testing / Safety and Effectiveness:
Performance testing for the CT subsystem was included in the original premarket notification for the CT subsystems and there have been no changes affecting this testing.
PET Testing in accordance with NEMA NU2:2012 was conducted on two different confiqurations of the Biograph mCT systems, a 3 rinq version and a 4 ring version.
| Performance Criteria (Standard) | Results | Acceptance |
|------------------------------------|---------|--------------------------------------------------------------|
| Transverse Resolution FWHM @ 1 cm | Pass | <= 7.3 mm |
| Transverse Resolution FWHM @ 10 cm | Pass | <= 7.5 mm |
| Axial Resolution FWHM @ 1 cm | Pass | <= 6.0 mm |
| Axial Resolution FWHM @ 10 cm | Pass | <= 6.6 mm |
| Sensitivity @ 435 keV LLD | Pass | >= 5.0 cps/kBq (3R)<br>>= 9.4 cps/kBq (4R) |
| Count Rate peak NECR | Pass | >= 95 kcps @ 30 kBq/cc (3R)<br>>= 165 kcps @ 28 kBq/cc (4R) |
| Count Rate peak trues | Pass | >= 350 kcps @ 46 kBq/cc (3R)<br>>= 575 kcps @ 40 kBq/cc (4R) |
| Count Rate bias (mean) | Pass | <=+/- 5% |
| Scatter Fraction | Pass | <= 37% (3R)<br><= 40% (4R) |
All Performance testing met the predetermined acceptance values.
The device labeling contains instructions for use and any necessary cautions and warnings to provide for safe and effective use of the device.
Risk Management is ensured via a risk analysis in compliance with ISO 14971 to identify and provide mitiqation to potential hazards beginning early in the design cycle and continuing throughout the development of the product. Siemens Medical Solutions, USA Inc. adheres to recognized and established industry standards such as IEC 60601-1 series and 21 CFR 1020.30 and 21 CFR 1020.33 to minimize electrical, mechanical and radiation hazards.
Verification and validation of Siemens systems is performed in accordance with documented procedures, design and code reviews, test plans and specifications. Traceability of the requirements specified in the requirement specifications and functional specifications is ensured during component integration, software validation and system testing.
{7}------------------------------------------------
## Statement regarding Substantial Equivalence:
There have been no changes implemented in the modifications to the Biograph mCT that impact either the fundamental technology or the indications for use. The Biograph mCT with the modifications outlined in this Premarket Notification is substantially equivalent to the currently commercially available predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.