The Siemens Biograph Vision and Biograph mCT systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information. The CT component produces cross-sectional images of the body by computer reconstruction of X-Ray transmission data from either the same axial plane taken at different angles or spiral planes taken at different angles. The PET subsystem images and measures the distribution of PET radiopharmaceuticals in humans for the purpose of determining various metabolic (molecular) and physiologic functions within the human body and utilizes the CT for fast attenuation correction maps for PET studies and precise anatomical reference for the fused PET and CT images. The system maintains independent functionality of the CT and PET devices, allowing for single modality CT and / or PET diagnostic imaging. These systems are intended to be utilized by appropriately trained health care professionals to aid in detecting, localizing, diagnosing, staging and restaging of lesions, tumors, disease and organ function for the evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The images produced by the system can also be used by the physician to aid in radiotherapy treatment planning and interventional radiology procedures. This CT system can be used for low dose lung cancer screening in high risk populations.
Device Story
Combined PET/CT scanner; inputs X-ray transmission data and PET radiopharmaceutical distribution; outputs fused high-resolution physiologic and anatomic images. Used in clinical settings by trained healthcare professionals for oncology, neurology, and cardiology diagnostics; radiotherapy planning; interventional radiology. System includes patient handling, acquisition/processing workstations, and software for scan control, reconstruction, and archival. New features include AI-based data-driven gating (OncoFreeze AI), PET planning (FlowMotion AI), and whole-body scatter correction. Physician uses fused images to localize lesions/tumors and assess organ function. Benefits include improved diagnostic accuracy, precise anatomical reference for PET, and optimized workflow.
Clinical Evidence
No clinical trials; bench testing only. PET performance validated per NEMA NU2-2018 (resolution, sensitivity, count rate, image quality, co-registration). AI features validated via comparative analysis: OncoFreeze AI compared to Anzai-based gating (SUV/volume metrics); FlowMotion AI evaluated for bed range identification accuracy (87.1-100%); Whole Body Scatter Correction validated via phantom simulation study comparing single-bed vs. whole-body scatter correction against ground truth.
Technological Characteristics
Combined PET/CT scanner; multi-slice CT and PET subsystems. Connectivity via DICOM. Software-based image reconstruction and processing. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-44, NEMA XR 25/28/29, NEMA PS3.1-3.20. Cybersecurity controls implemented. Software developed per IEC 62304.
Indications for Use
Indicated for patients requiring PET/CT imaging for detection, localization, diagnosis, staging, and restaging of lesions, tumors, disease, and organ function evaluation (cardiovascular, neurological, cancer). CT component indicated for low-dose lung cancer screening in high-risk populations.
Regulatory Classification
Identification
An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.
{0}------------------------------------------------
February 14, 2020
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Siemens Medical Solutions USA, Inc % Ms. Tabitha Estes Regulatory Affairs Professional 810 Innovation Drive KNOXVILLE TN 37932
## Re: K193248
Trade/Device Name: Biograph Vision PET/CT and Biograph mCT PET/CT Regulation Number: 21 CFR 892.1200 Regulation Name: Emission computed tomography system Regulatory Class: Class II Product Code: KPS, JAK Dated: January 16, 2020 Received: January 17, 2020
Dear Ms. Estes:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
#### K193248
Device Name Biograph Vision PET/CT and Biograph mCT PET/CT
#### Indications for Use (Describe)
The Siemens Biograph Vision and Biograph mCT PET/CT systems are combined X-Ray Computed Tomography (CT) and Position Emission Tomography (PET) scanners that provide registration of high resolution physiologic and anatomic information.
The CT component produces cross-sectional images of the body by computer reconstruction of X-Ray transmission data from either the same axial plane taken at different angles or spiral planes taken at different angles. The PET subsystem images and measures the distribution of PET radiopharmaceuticals in humans for the purpose of determining various metabolic (molecular) and physiologic functions within the human body and utilizes the CT for fast attenuation correction maps for PET studies and precise anatomical reference for the fused PET and CT images.
The system maintains independent functionality of the CT and PET devices, allowing for single modality CT and / or PET diagnostic imaging. These systems are intended to be utilized by appropriately trained health care professionals to aid in detecting, localizing, diagnosing, staging of lesions, tumors, disease and organ function for the evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The images produced by the system can also be used by the physician to aid in radiotherapy treatment planning and interventional radiology procedures.
This CT system can be used for low dose lung cancer screening in high risk populations *
* As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature. for further information.
| Type of Use ( <i>Select one or both, as applicable</i> ) | |
|-----------------------------------------------------------------------------------------------------------|---------------------------------------------------------|
| <div><span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div> Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary - K193248
as required by 21 CFR Part 807.87(h)
### Identification of the Submitter
| Submitter: | Tabitha Estes<br>Regulatory Affairs<br>Siemens Medical Solutions USA, Inc.<br>Molecular Imaging<br>810 Innovation Drive<br>Knoxville, TN 37932 |
|-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| Alternative Contact: | M. Alaine Medio<br>Regulatory Affairs |
| Manufacturer: | Siemens Medical Solutions USA, Inc.<br>Molecular Imaging<br>2501 North Barrington Road<br>Hoffman Estates, IL 60192 |
| Telephone Number: | (865)804-4553 |
| Fax Number: | (865)218-3019 |
| Date of Submission: | November 22, 2019 |
| Identification of the product | |
| Device Proprietary<br>Name: | Biograph Vision PET/CT and Biograph mCT PET/CT |
| Common Name: | Positron Emission Tomography (PET) System<br>Computed Tomography (CT) System |
| Classification Name: | Emission Computed Tomography System per 21 CFR<br>892.1200<br>Computed Tomography X-Ray System per 21 CFR 892.1750 |
| Product Code: | KPS and JAK |
| Classification Panel: | Radiology |
| Device Class: | Class II |
{4}------------------------------------------------
### Marketed Devices to which Equivalence is claimed
Primary Predicate Device:
| Device Proprietary<br>Name: | Biograph Vision PET/CT Systems |
|-----------------------------|-------------------------------------|
| Manufacturer: | Siemens Medical Solutions USA, Inc. |
| Product Code: | KPS and JAK |
| Device Class: | Class II |
| 510(k) Number: | K190900 |
Predicate Device:
| Device Proprietary<br>Name: | Biograph mCT PET/CT Systems |
|-----------------------------|-------------------------------------|
| Manufacturer: | Siemens Medical Solutions USA, Inc. |
| Product Code: | KPS and JAK |
| Device Class: | Class II |
| 510(k) Number: | K173578 |
#### Reference Devices:
| Device Name and 510(k) numbers: | SOMATOM Definition CT | K190578 |
|---------------------------------|-----------------------|---------|
|---------------------------------|-----------------------|---------|
#### Device Description:
The Biograph Vision and Biograph mCT PET/CT systems are combined multi-slice X-Ray Computed Tomography and Positron Emission Tomography scanners. These systems are designed for whole body oncology, neurology and cardiology examinations. The Biograph Vision and Biograph mCT systems provide registration and fusion of high-resolution metabolic and anatomic information from the two major components of each system (PET
{5}------------------------------------------------
and CT). Additional components of the system include a patient handling system and acquisition and processing workstations with associated software.
Biograph Vision and Biograph mCT software is a command-based program used for patient management, data management, scan control, image reconstruction and image archival and evaluation. All images conform to DICOM imaging format requirements.
The software for the Biograph Vision and Biograph mCT systems which are the subject of this application is substantially equivalent to the commercially available Biograph Vision and Biograph mCT software. Modifications include, corrections to software anomalies and addition of new software features, including:
- OncoFreeze AI (Data Driven Gating)
- FlowMotion Al (PET FAST Planning)
- FAST PET Workflow
- Updates to HD FoV
- . Updates to PET DICOM dose Report
- Whole Body Scatter Correction (for Biograph Vision)
Additionally, minor modifications have been made to the computers due to obsolescence issue. These changes do not affect system performance characteristics and have no impact on safety or effectiveness.
#### Intended Use:
The Siemens Biograph Vision and Biograph mCT systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.
The CT component produces cross-sectional images of the body by computer reconstruction of X-Ray transmission data from either the same axial plane taken at different angles or spiral planes taken at different angles. The PET subsystem images and measures the distribution of PET radiopharmaceuticals in humans for the purpose of determining various metabolic (molecular) and physiologic functions within the human body and utilizes the CT for fast attenuation correction maps for PET studies and precise anatomical reference for the fused PET and CT images.
The system maintains independent functionality of the CT and PET devices, allowing for single modality CT and / or PET diagnostic imaging.
These systems are intended to be utilized by appropriately trained health care professionals to aid in detecting, localizing, diagnosing, staging and restaging of lesions, tumors, disease and organ function for the evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The images produced by the
{6}------------------------------------------------
system can also be used by the physician to aid in radiotherapy treatment planning and interventional radiology procedures.
This CT system can be used for low dose lung cancer screening in high risk populations.*
*As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
{7}------------------------------------------------
## Performance Testing / Safety and Effectiveness:
Performance testing for the CT subsystem was included in the original premarket notification for the CT subsystems and there have been no changes affecting this testing.
PET Testing in accordance with NEMA NU2-2018 was conducted on two different configurations of the Biograph Vision systems, a 6-ring version and an 8-ring version and on two configurations of the Biograph mCT, a 3-ring version and a 4-ring version.
| Performance Criteria | Results | Biograph Vision 6-ring<br>Acceptance | Biograph Vision 8-ring<br>Acceptance |
|----------------------------------------------------------------|---------|--------------------------------------|--------------------------------------|
| Resolution - Full Size | | | |
| Transverse Resolution<br>FWHM @ 1 cm | Pass | ≤ 4.0 mm | ≤ 4.0 mm |
| Transverse Resolution<br>FWHM @ 10 cm | Pass | ≤ 4.8 mm | ≤ 4.8 mm |
| Transverse Resolution<br>FWHM @ 20 cm | Pass | ≤ 5.2 mm | ≤ 5.2 mm |
| Axial Resolution FWHM<br>@ 1 cm | Pass | ≤ 4.3 mm | ≤ 4.3 mm |
| Axial Resolution FWHM<br>@ 10 cm | Pass | ≤ 5.4 mm | ≤ 5.4 mm |
| Axial Resolution FWHM<br>@ 20 cm | Pass | ≤ 5.4 mm | ≤ 5.4 mm |
| Count Rate / Scatter / Sensitivity | | | |
| Sensitivity @435 keV LLD | Pass | ≥ 8.0 cps/kBq | ≥ 15.0 cps/kBq |
| Count Rate peak NECR | Pass | ≥ 140 kcps @ ≤ 32 kBq/cc | ≥ 250 kcps @ ≤ 32 kBq/cc |
| Count Rate peak trues | Pass | ≥600 kcps @ ≤ 56 kBq/cc | ≥1100 kcps @ ≤ 56 kBq/cc |
| Scatter Fraction at peak<br>NECR | Pass | ≤ 43% | ≤ 43% |
| Mean bias (%) at peak<br>NEC | Pass | ≤ +/- 6% | ≤ +/- 6% |
| Image Quality (4 to 1) - (% Contrast / Background Variability) | | | |
| 10mm sphere | Pass | ≥ 55% / ≤ 10% | ≥ 55% / ≤ 10% |
| 13mm sphere | Pass | ≥ 60% / ≤ 9% | ≥ 60% / ≤ 9% |
| 17mm sphere | Pass | ≥ 65% / ≤ 8% | ≥ 65% / ≤ 8% |
| 22mm sphere | Pass | ≥ 70% / ≤ 7% | ≥ 70% / ≤ 7% |
| 28mm sphere | Pass | ≥ 75% / ≤ 6% | ≥ 75% / ≤ 6% |
| 37mm sphere | Pass | ≥ 80% / ≤ 5% | ≥ 80% / ≤ 5% |
| Co-Registration Accuracy | | | |
| Max Error | Pass | ≤ 5 mm | ≤ 5 mm |
| Performance Criteria | Results | Biograph mCT 3-ring<br>Acceptance | Biograph mCT 4-ring<br>Acceptance |
| Resolution - Full Size | | | |
| Transverse Resolution<br>FWHM @ 1 cm | Pass | ≤ 4.7 mm | ≤ 4.0 mm |
| Transverse Resolution<br>FWHM @ 10 cm | Pass | ≤ 5.4 mm | ≤ 4.8 mm |
| Transverse Resolution<br>FWHM @ 20 cm | Pass | ≤ 6.3 mm | ≤ 5.2 mm |
| Axial Resolution FWHM<br>@ 1 cm | Pass | ≤ 4.9 mm | ≤ 4.3 mm |
| Axial Resolution FWHM<br>@ 10 cm | Pass | ≤ 6.5 mm | ≤ 5.4 mm |
| Axial Resolution FWHM<br>@ 20 cm | Pass | ≤ 8.8 mm | ≤ 5.4 mm |
| Count Rate / Scatter / Sensitivity | | | |
| Sensitivity @435 keV LLD | Pass | ≥ 5.0 cps/kBq | ≥ 9.4 cps/kBq |
| Count Rate peak NECR | Pass | ≥ 95 kcps @ ≤ 30 kBq/cc | ≥ 165 kcps @ ≤ 28 kBq/cc |
| Count Rate peak trues | Pass | ≥350 kcps @ ≤ 46 kBq/cc | ≥575 kcps @ ≤ 40 kBq/cc |
| Scatter Fraction at peak<br>NECR | Pass | ≤ 40% | ≤ 40% |
| Mean bias (%) at peak<br>NEC | Pass | ≤ +/- 6% | ≤ +/- 6% |
| Image Quality (4 to 1) - (% Contrast / Background Variability) | | | |
| 10mm sphere | Pass | ≥ 10% / ≤ 10% | ≥ 10% / ≤ 10% |
| 13mm sphere | Pass | ≥ 25% / ≤ 10% | ≥ 25% / ≤ 10% |
| 17mm sphere | Pass | ≥ 40% / ≤ 10% | ≥ 40% / ≤ 10% |
| 22mm sphere | Pass | ≥ 55% / ≤ 10% | ≥ 55% / ≤ 10% |
| 28mm sphere | Pass | ≥ 60% / ≤ 10% | ≥ 60% / ≤ 10% |
| 37mm sphere | Pass | ≥ 65% / ≤ 10% | ≥ 65% / ≤ 10% |
| Co-Registration Accuracy | | | |
| Max Error | Pass | ≤ 5 mm | ≤ 5 mm |
{8}------------------------------------------------
All features (including the new features listed in the device description) were tested during verification and validation testing and met the predetermined acceptance criteria.
Additionally, the changes below had additional scientific evaluations performed. A brief description of testing activities is provided.
- OncoFreeze Al (Data Driven Gating) .
An evaluation of change in SUVmax, SUVmean and volume measurement has been performed comparing Anzai based OncoFreeze and OncoFreeze Al.
{9}------------------------------------------------
| | Anzai-based OncoFreeze | OncoFreeze Al |
|-------------------------------|------------------------|---------------|
| ΔSUVmax (relative to static) | +29% ± 22% | +27% ± 22% |
| ΔSUVmean (relative to static) | +27% ± 22% | +26% ± 22% |
| ΔVolume (relative to static) | -34% ± 23% | -31% ± 19% |
- FlowMotion AI (PET FAST Planning)
Testing was performed to evaluate the accuracy with which FlowMotion AI correctly defines the bed ranges. Successful identification of all ranges in a FlowMotion Al configuration ranged from 87.1% to 100%, dependent on the configuration set.
- Whole Body Scatter Correction (for Biograph Vision)
An evaluation was performed to evaluate ROI's using single bed scatter compared to whole body scatter.
| Difference from ground truth<br>In simulation study of phantom | Single Bed<br>Scatter<br>Correction | Whole Body<br>Scatter correction |
|------------------------------------------------------------------------------|-------------------------------------|----------------------------------|
| Representative region of interest close to<br>phantom exhibiting high signal | +87% | -2% |
| Representative region of interest close to<br>phantom exhibiting low signal | -42% | -3% |
| Representative region of interest inside<br>phantom | +0.5% | -0.4% |
The device labeling contains instructions for use and any necessary cautions and warnings to provide for safe and effective use of the device.
Risk Management is ensured via a risk analysis in compliance with ISO 14971 to identify and provide mitigation to potential hazards beginning early in the design cycle and continuing throughout the development of the product. Siemens Medical Solutions, USA Inc. adheres to recognized and established industry standards such as IEC 60601-1 series and 21 CFR 1020.30 and 21 CFR 1020.33 to minimize electrical, mechanical and radiation hazards.
Siemens claims compliance with the following product standards for the Biograph Vision and Biograph mCT systems:
- IEC 60601-1: 2005+ A1:2012
- IEC 60601-1-2: 2014
- IEC 60601-1-3: 2013
- IEC 60601-1-6:2010 +A1:2013
{10}------------------------------------------------
- IEC 60601-2-28:2010 ●
- IEC 60601-2-44: 2009 + A1:2012
- IEC 62366-1:2015
- . IEC 61223-2-6:2006
- IEC 61223-3-5:2004
- NEMA XR 25: 2010
- . NEMA XR 28: 2013
- NEMA XR 29: 2013
- NEMA PS3.1-3.20
Additionally, the Biograph Vision and Biograph mCT systems have been developed in accordance with the requirements of the following standards:
- IEC 62304:2006 +A1:2015
- ISO 14971:2012 (ISO 14971:2007)
The device labeling contains instructions for use and any necessary cautions and warnings to provide for safe and effective use of the device.
Cybersecurity information in accordance with FDA Guidance documents issued October 2, 2014 has been provided. The Biograph Vision and Biograph mCT software has specific cybersecurity controls to prevent unauthorized access, modifications, misuse or denial of use. Additionally, controls are enabled to prevent the unauthorized use of information that is stored, accessed or transferred between the Biograph Vision, Biograph mCT and external devices.
Verification and validation of Siemens systems is performed in accordance with documented procedures, design and code reviews, test plans and specifications. Traceability of the requirements specified in the requirement specifications and functional specifications is ensured during component integration, software validation and system testing.
#### Statement regarding Substantial Equivalence:
There have been no changes implemented in the modifications to the Biograph Vision and Biograph mCT systems that impact either the fundamental technology or the indications for use. The Biograph Vision and Biograph mCT systems with the modifications outlined in this Premarket Notification are substantially equivalent to the currently commercially available predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.