Biograph Vision PET/CT Systems; Biograph mCT PET/CT Systems
K251671 · Siemens Medical Solutions USA, Inc. · KPS · Jul 3, 2025 · Radiology
Device Facts
Record ID
K251671
Device Name
Biograph Vision PET/CT Systems; Biograph mCT PET/CT Systems
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
KPS · Radiology
Decision Date
Jul 3, 2025
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1200
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Siemens Biograph systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information. The CT component produces cross-sectional images of the body by computer reconstruction of X-Ray transmission data from either the same axial plane taken at different angles or spiral planes taken at different angles. The PET subsystem images and measures the distribution of PET radiopharmaceuticals in humans for the purpose of determining various metabolic (molecular) and physiologic functions within the human body and utilizes the CT for fast attenuation correction maps for PET studies and precise anatomical reference for the fused PET and CT images. The system maintains independent functionality of the CT and PET devices, allowing for single modality CT and/or PET diagnostic imaging. These systems are intended to be utilized by appropriately trained health care professionals to aid in detecting, localizing, diagnosing, staging, and restaging of lesions, tumors, disease, and organ function for the evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders, and cancer. The images produced by the system can also be used by the physician to aid in radiotherapy treatment planning and interventional radiology procedures. This CT system can be used for low dose lung cancer screening in high risk populations.
Device Story
Biograph Vision and Biograph mCT are combined PET/CT scanners; provide registered, fused metabolic and anatomic images. PET subsystem measures radiopharmaceutical distribution; CT subsystem provides attenuation correction and anatomical reference. Systems support independent CT or PET diagnostic imaging. Operated by trained healthcare professionals in clinical settings. Software (Somaris/7 syngo CT VB30) manages patient data, scan control, image reconstruction, and archival. Output images used by physicians for diagnosis, staging, radiotherapy planning, and interventional procedures. Benefits include improved lesion detection and disease evaluation. Updated hardware and software features (FAST Bolus, FAST 4D, FAST Applications) enhance workflow and functionality.
Clinical Evidence
No clinical testing was conducted for this submission. Substantial equivalence is supported by non-clinical verification and validation, including phantom testing in accordance with NEMA NU2-2018 standards, which demonstrated performance comparable to the predicate device.
Technological Characteristics
Combined PET/CT scanner. PET subsystem uses scintillation detectors; CT subsystem uses X-ray tube/detector array. Software: Somaris/7 syngo CT VB30. Connectivity: DICOM compliant. Standards: IEC 60601-1, 21 CFR 1020.30, 21 CFR 1020.33. Hardware updated for obsolescence. Features include FAST Bolus, FAST 4D, FAST Spine, FAST Planning, and Tin Filtration.
Indications for Use
Indicated for use by trained healthcare professionals to detect, localize, diagnose, stage, and restage lesions, tumors, and diseases, including cardiovascular, neurological, and oncological conditions. Includes low-dose lung cancer screening for high-risk populations. Applicable to adult and pediatric patients requiring PET/CT imaging.
Regulatory Classification
Identification
An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.
Predicate Devices
Biograph Vision and Biograph mCT PET/CT Systems (K193248)
Reference Devices
Biograph Vision.X and Biograph Vision.X Edge (K231833)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 3, 2025
Siemens Medical Solutions USA, Inc.
Clayton Ginn
Regulatory Affairs Professional
2501 North Barrington Road
Hoffman Estates, Illinois 60192
Re: K251671
Trade/Device Name: Biograph Vision PET/CT Systems; Biograph mCT PET/CT Systems
Regulation Number: 21 CFR 892.1200
Regulation Name: Emission Computed Tomography System
Regulatory Class: Class II
Product Code: KPS, JAK
Dated: June 27, 2025
Received: June 27, 2025
Dear Clayton Ginn:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251671 - Clayton Ginn
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K251671 - Clayton Ginn
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Daniel M. Krainak, Ph.D.
Assistant Director
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K251671 | ? |
| Please provide the device trade name(s). | | ? |
| Biograph Vision PET/CT Systems;
Biograph mCT PET/CT Systems | | |
| Please provide your Indications for Use below. | | ? |
| The Siemens Biograph systems are combined X-Ray Computed Tomography (CT) and Positron Emission
Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and
anatomic information.
The CT component produces cross-sectional images of the body by computer reconstruction of X-Ray
transmission data from either the same axial plane taken at different angles or spiral
planes taken at different angles. The PET subsystem images and measures the distribution of PET
radiopharmaceuticals in humans for the purpose of determining various metabolic (molecular) and
physiologic functions within the human body and utilizes the CT for fast attenuation correction maps for PET
studies and precise anatomical reference for the fused PET and CT images.
The system maintains independent functionality of the CT and PET devices, allowing for single modality CT
and/or PET diagnostic imaging.
These systems are intended to be utilized by appropriately trained health care professionals to aid in
detecting, localizing, diagnosing, staging, and restaging of lesions, tumors, disease, and organ function for
the evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological
disorders, and cancer. The images produced by the system can also be used by the physician to aid in
radiotherapy treatment planning and interventional radiology procedures.
This CT system can be used for low dose lung cancer screening in high risk populations. * | | |
| * As defined by professional medical societies. Please refer to clinical literature, including the results of the
National Lung Screening Trial (N Engl J Med 2011; 365; 395-409) and subsequent literature, for further
information. | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
SIEMENS Healthineers
# 510(k) Summary Biograph Vision and Biograph mCT PET/CT Systems
In accordance with 21 CFR §807.92 Content and format of a 510(k) summary.
K251671
# 1. Identification of the Submitter
| Submitter / Primary Contact Person | Clayton Ginn
Regulatory Affairs
clayton.ginn@siemens-healthineers.com
+1 (865) 898-2692 |
| --- | --- |
| Secondary Contact Person | Alaine Medio
Regulatory Affairs Manager
alaine.medio@siemens-healthineers.com
+1 (865) 206-0337 |
| Applicant Name and Address | Siemens Medical Solutions, Inc. USA
2501 North Barrington Road
Hoffman Estates IL, 60192, USA
Establishment Registration Number: 1423253 |
# 2. Device Name and Classification
| Product Trade Name: | Biograph Vision and Biograph mCT PET/CT Systems |
| --- | --- |
| Common Name: | Emission computed tomography system (PET)
Computed tomography x-ray system (CT) |
| Classification Name: | System, Tomography, Computed, Emission (PET)
System, X-Ray, Tomography, Computed (CT) |
| Classification Panel: | Radiology |
| CFR Section: | 21 CFR §892.1200 |
| Device Class: | Class II |
| Product Code: | KPS (PET) and
JAK (CT) |
Siemens Medical Solutions USA, Inc.
{5}
SIEMENS Healthineers
# 3. Predicate Devices
Primary Predicate Device:
| Product Trade Name: | Biograph Vision and Biograph mCT PET/CT Systems |
| --- | --- |
| 510(k) Number | K193248 |
| Clearance Date | 02/14/2020 |
| Common Name: | Emission computed tomography system (PET)
Computed tomography x-ray system (CT) |
| Classification Name: | System, Tomography, Computed, Emission (PET)
System, X-Ray, Tomography, Computed (CT) |
| Classification Panel: | Radiology |
| CFR Section: | 21 CFR §892.1200 |
| Device Class: | Class II |
| Product Code: | KPS (PET) and subsequent product code
JAK (CT) |
Reference Devices:
1)
| Product Trade Name: | Biograph Vision.X and Biograph Vision.X Edge |
| --- | --- |
| 510(k) Number | K231833 |
| Clearance Date | 07/13/2023 |
| Common Name: | Emission computed tomography system (PET)
Computed tomography x-ray system (CT) |
| Classification Name: | System, Tomography, Computed, Emission (PET)
System, X-Ray, Tomography, Computed (CT) |
| Classification Panel: | Radiology |
| CFR Section: | 21 CFR §892.1200 |
| Device Class: | Class II |
| Product Code: | KPS (PET) and subsequent product code
JAK (CT) |
Siemens Medical Solutions USA, Inc.
{6}
SIEMENS Healthineers
2)
| Product Trade Name: | Biograph Vision Quadra PET/CT System |
| --- | --- |
| 510(k) Number | K223547 |
| Clearance Date | 12/22/2022 |
| Common Name: | Emission computed tomography system (PET)
Computed tomography x-ray system (CT) |
| Classification Name: | System, Tomography, Computed, Emission (PET)
System, X-Ray, Tomography, Computed (CT) |
| Classification Panel: | Radiology |
| CFR Section: | 21 CFR §892.1200 |
| Device Class: | Class II |
| Product Code: | KPS (PET) and subsequent product code
JAK (CT) |
3)
| Product Trade Name: | SOMATOM Definition Edge, SOMATOM Definition AS/AS+ |
| --- | --- |
| 510(k) Number | K230421 |
| Clearance Date | 06/16/2023 |
| Common Name: | Computed tomography x-ray system |
| Classification Name: | System, X-Ray, Tomography, Computed |
| Classification Panel: | Radiology |
| CFR Section: | 21 CFR §892.1750 |
| Device Class: | Class II |
| Product Code: | JAK |
Siemens Medical Solutions USA, Inc.
{7}
SIEMENS Healthineers
# 4. Device Description
The Biograph Vision and Biograph mCT PET/CT systems are combined multi-slice X-Ray Computed Tomography and Positron Emission Tomography scanners. These systems are designed for whole-body oncology, neurology and cardiology examinations. The Biograph Vision and Biograph mCT systems provide registration and fusion of high-resolution metabolic and anatomic information from the two major components of each system (PET and CT). Additional components of the system include a patient handling system and acquisition and processing workstations with associated software.
Biograph Vision and Biograph mCT software is a command-based program used for patient management, data management, scan control, image reconstruction and image archival and evaluation. All images conform to DICOM imaging format requirements.
The software for the Biograph Vision and Biograph mCT systems, which are the subject of this application, is substantially equivalent to the commercially available Biograph Vision and Biograph mCT software.
- Somaris Software (cleared in K230421)
- Upgrade to the latest revision of Somaris Software (Somaris/7 syngo CT VB30) with modified software features:
- FAST Bolus
- FAST 4D
- FAST Applications (FAST Spine, FAST Planning)
- Automatic Patient Instructions
- Additional default exam protocols
- Additional kV setting for Tin Filtration
- PETsyngo software
- SMART Image Framer (available for Vision 600 and X models only – cleared in K223547)
- Updated computer hardware due to obsolescence issues (cleared in K230421). These changes do not affect system performance characteristics and have no impact on safety or effectiveness.
The Biograph Vision may also use the names Biograph Vision Quantum and Peak for marketing purposes.
# 5. Indications for Use
The Siemens Biograph systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.
The CT component produces cross-sectional images of the body by computer reconstruction of X-Ray transmission data from either the same axial plane taken at different angles or spiral planes taken at different angles. The PET subsystem images and measures the distribution of PET radiopharmaceuticals in humans for the purpose of determining various metabolic (molecular) and physiologic functions within the human body and utilizes the CT for fast attenuation correction maps for PET studies and precise anatomical reference for the fused PET and CT images.
Siemens Medical Solutions USA, Inc.
{8}
SIEMENS Healthineers
The system maintains independent functionality of the CT and PET devices, allowing for single-modality CT and/or PET diagnostic imaging.
These systems are intended to be utilized by appropriately trained health care professionals to aid in detecting, localizing, diagnosing, staging, and restaging of lesions, tumors, disease, and organ function for the evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders, and cancer. The images produced by the system can also be used by the physician to aid in radiotherapy treatment planning and interventional radiology procedures.
This CT system can be used for low-dose lung cancer screening in high-risk populations. *
* As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365; 395-409) and subsequent literature, for further information.
## 6. Indications for Use Comparison to the Predicate Device
The indications for use are the same between the subject device and the primary predicate device.
## 7. Technological Comparison
Biograph Vision and Biograph mCT PET/CT Systems with software version VG85 provide the same technological characteristics in terms of design, materials, chemical composition, energy source, and control mechanisms when compared to the legally marketed predicate devices Biograph Vision and Biograph mCT Systems with software version VG80.
The intended use, indications for use, and fundamental scientific technology for the subject device remain unchanged from the predicate device. No features present from the predicate device have been de-scoped. The software features have been modified in comparison to the predicate device to support enhanced device functionality and computer hardware has been updated due to obsolescence issues.
Differences in technological characteristics described in section 4 above (software feature updates for scanning and reconstruction and computer hardware updates) do not raise different questions of safety and effectiveness as they are cleared in the primary and reference predicate devices. Testing and validation are completed. Test results show that the subject devices are comparable to the predicate devices in terms of technological characteristics, safety, and effectiveness and therefore, are substantially equivalent to the predicate devices.
Siemens Medical Solutions USA, Inc.
{9}
SIEMENS
Healthineers
❤
# 8. Non-Clinical and/or Clinical Test Summary & Conclusions
## Non-Clinical Testing
Performance testing for the CT subsystem was included in the original premarket notification for the CT subsystems, and there have been no changes to this testing.
PET Testing in accordance with NEMA NU2-2018 was conducted on the Biograph Vision and Biograph mCT systems. All Performance testing met the predetermined acceptance values.
| Performance Criteria | Results | Subject Device: Biograph Vision and Biograph mCT (VG85) | Predicate: Biograph Vision Biograph mCT (K193248) |
| --- | --- | --- | --- |
| Resolution – Full Size | | | |
| Transverse Resolution
FWHM @ 1 cm | Pass | Same | ≤ 4.0 mm (Vision)
≤ 4.7 mm (mCT) |
| Transverse Resolution
FWHM @ 10 cm | Pass | Same | ≤ 4.8 mm (Vision)
≤ 5.4 mm (mCT) |
| Transverse Resolution
FWHM @ 20 cm | Pass | Same | ≤ 5.2 mm (Vision)
≤ 6.3 mm (mCT) |
| Axial Resolution
FWHM @ 1 cm | Pass | Same | ≤ 4.3 mm (Vision)
≤ 4.9 mm (mCT) |
| Axial Resolution
FWHM @ 10 cm | Pass | Same | ≤ 5.4 mm (Vision)
≤ 6.5 mm (mCT) |
| Axial Resolution
FWHM @ 20 cm | Pass | Same | ≤ 5.4 mm (Vision)
≤ 8.8 mm (mCT) |
| Count Rate / Scatter / Sensitivity | | | |
| Sensitivity @435 keV LLD | Pass | Same | ≥ 8.0 cps/kBq (Vision 450)
≥ 15.0 cps/kBq (Vision 600)
≥ 5.0 cps/kBq – (mCT 3R)
≥ 9.4 cps/kBq – (mCT 4R) |
| Count Rate peak NECR | Pass | Same | ≥140 kcps @ ≤ 32 kBq/cc (Vision 450)
≥250 kcps @ ≤ 32 kBq/cc (Vision 600 and X)
≥95 kcps @ ≤ 30 kBq/cc (mCT 3R)
≥165 kcps @ ≤ 40 kBq/cc (mCT 4R) |
| Count Rate peak trues | Pass | Same | ≥600 kcps @ ≤ 56 kBq/cc (Vision 450)
≥1100 kcps @ ≤ 56 kBq/cc (Vision 600 and X)
≥350 kcps @ ≤ 46 kBq/cc (mCT 3R)
≥575 kcps @ ≤ 40 kBq/cc (mCT 4R) |
| Scatter Fraction (435 keV LLD) | Pass | Same | ≤43% @ Peak
*≤40% @ low activity (Vision)
≤40% @ Peak
*≤37% @ low activity (mCT) |
Siemens Medical Solutions USA, Inc.
{10}
SIEMENS Healthineers
| Performance Criteria | Results | Subject Device: Biograph Vision and Biograph mCT (VG85) | Predicate: Biograph Vision Biograph mCT (K193248) |
| --- | --- | --- | --- |
| Resolution – Full Size | | | |
| Co-Registration Accuracy | Pass | Same | ≤ 5 mm |
| Time of Flight Resolution at 5.3kBq/cc | Pass | Same | ≤ 214 ps (Vision.X) ≤ 249 (Vision 450 and 600) ≤600 (mCT) |
| Mean bias [%] at peak NEC | Pass | Same | [-6, 6] |
| Image Quality – (% Contrast / Background Variability) | | | |
| 10mm sphere | Pass | Same | ≥ 55% / ≤ 10% (Vision) ≥ 10% / ≤ 10% (mCT) |
| 13mm sphere | Pass | Same | ≥ 60% / ≤ 9% (Vision) ≥ 25% / ≤ 10% (mCT) |
| 17mm sphere | Pass | Same | ≥ 65% / ≤ 8% (Vision) ≥ 40% / ≤ 10% (mCT) |
| 22mm sphere | Pass | Same | ≥ 70% / ≤ 7% (Vision) ≥ 55% / ≤ 10% (mCT) |
| 28mm sphere | Pass | Same | ≥ 75% / ≤ 6% (Vision) ≥ 60% / ≤ 10% (mCT) |
| 37mm sphere | Pass | Same | ≥ 80% / ≤ 5% (Vision) ≥ 65% / ≤ 10% (mCT) |
Verification and validation testing substantiates and is traceable to all requirement and functional specifications, including specifications related to device hazards, and supports the claim of substantial equivalence.
The device labeling contains instructions for use and any necessary cautions and warnings to provide for safe and effective use of the device.
Risk Management is ensured via a risk analysis in compliance with ISO 14971 to identify and provide mitigation to potential hazards beginning early in the design cycle and continuing throughout the development of the product. Siemens Medical Solutions, USA Inc. adheres to recognized and established industry standards such as IEC 60601-1 series and 21 CFR 1020.30 and 21 CFR 1020.33 to minimize electrical, mechanical and radiation hazards.
## Clinical Testing
Clinical testing was not conducted for this submission.
## Conclusion
There are no differences in the Indications for Use, Intended Use, or Fundamental Technological Characteristics of the Biograph Vision and Biograph mCT VG85 software as compared to the currently commercially available Biograph Vision (including Vision.X - K231833) and Biograph mCT software (K193248).
The new features implemented within this release do not raise any new issues of safety and effectiveness as compared to the predicate device. The predicate devices were cleared based on the
Siemens Medical Solutions USA, Inc.
{11}
SIEMENS Healthineers
results of non-clinical testing, which included verification and validation, phantom tests, and supportive literature. The subject device is also validated using the same methods as used for the predicate devices. The non-clinical verification and validation demonstrate that the subject devices should perform as intended in the specified use conditions.
The Biograph Vision and Biograph mCT systems, with the modifications outlined in this Premarket Notification, are substantially equivalent to the commercially available predicate device.
Siemens Medical Solutions USA, Inc.
8
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.