CMDR 2C (Multiple Models)

K201575 · Minxray, Inc. · IZL · Jul 8, 2020 · Radiology

Device Facts

Record IDK201575
Device NameCMDR 2C (Multiple Models)
ApplicantMinxray, Inc.
Product CodeIZL · Radiology
Decision DateJul 8, 2020
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 892.1720
Device ClassClass 2
AttributesPediatric

Indications for Use

Intended for use by a qualified/trained physician or technician on both adult and pediatric subjects for taking diagnostic x-rays. Not for mammography.

Device Story

Mobile digital x-ray system; combines portable high-frequency x-ray generator (HF120/60H PowerPlus, HF100H+, or HF1202 PowerPlus), CareView 1500C flat-panel detector, and dicomPACS software on ruggedized laptop. Operator (physician/technician) positions generator and detector; captures x-ray images; processes via software for display. Used in clinical/bedside/trauma settings. Output provides diagnostic-quality images for routine/trauma diagnosis. Benefits include streamlined digital radiography workflow, portability, and high-quality imaging.

Clinical Evidence

No clinical data. Substantial equivalence established via bench testing. System performance verified using i.b.a. Test Device DIGI-13 for quality tests (DIN V 6868-58/DIN 6868-13). Compliance with UL 60601-1, IEC 60601-1-2, and IEC 60601-2-54 verified. FMEA methodology used for risk analysis.

Technological Characteristics

Mobile x-ray system; high-frequency inverter generator; Csl scintillator flat-panel detector (154 μm pixel pitch, 16-bit A/D resolution). Connectivity: Gigabit Ethernet. Power: 120/240V AC or Li-ion battery. Software: dicomPACS. Standards: UL 60601-1, IEC 60601-1-2, IEC 60601-2-54.

Indications for Use

Indicated for use by qualified/trained physicians or technicians on adult and pediatric subjects for diagnostic x-ray imaging. Contraindicated for mammography.

Regulatory Classification

Identification

A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ July 8, 2020 Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" encircling a symbol. To the right of the seal, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue. MinXray, Inc. % Mr. Daniel Kamm Principal Engineer Kamm & Associates 8870 Ravello Ct NAPLES FL 34114 Re: K201575 Trade/Device Name: CMDR 2C (Multiple Models) Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: Class II Product Code: IZL, MQB, LLZ Dated: June 10, 2020 Received: June 11, 2020 Dear Mr. Kamm: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for {1}------------------------------------------------ devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. For Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K201575 Device Name CMDR 2C (Multiple Models) Indications for Use (Describe) Intended for use by a qualified/trained physician or technician on both adult and pediatric subjects for taking diagnostic x-rays. Not for mammography. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ 510(k) Summary: 510(k) Number K201575 MinXray, Inc. 3611 Commercial Avenue Northbrook, Illinois 60062, USA Toll Free 1-800-221-2245 (USA & Canada) Date Prepared: June22, 2020 Contact: Keith Kretchmer, President - 1) Identification of the Device: Trade/Device Name: CMDR 2C (Multiple Models) Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: II Product Codes: IZL, MQB, and LLZ. Common/Usual Name: Digital Mobile Diagnostic X-Ray System - 2) Equivalent legally marketed device: K191451, MinXray, Inc. Trade/Device Name: CMDR 2CW (Multiple Models) Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: II Product Codes: IZL, MQB, and LLZ. Common/Usual Name: Digital Mobile Diagnostic X-Ray System - 3) Reference device (We employ these cleared devices without modification): Trade/Device Name: K153058 CareView 1500C/CareView 1500L X-ray Flat Panel Detectors; Manufacturer: CareRay Digital Medical Systems Co., Ltd Regulation Number: 21 CFR 892. 1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB - Indications for Use (intended use): These digital radiographic systems are intended for use by a qualified/trained 4) physician or technician on both adult and pediatric subjects for taking diagnostic x-rays. Not for mammographic use. - 5) Description of the Device: This represents the straightforward combination of three devices: One of three cleared MinXray Portable HF X-ray generators: - a) HF120/60H PowerPlus cleared in K040046, (and in K141885) OR - HF100H+ cleared in K052721 OR b) - C) HF1202 PowerPlus cleared in K153059. - d) Plus: A 510(k) cleared (K153058) Digital X-Ray Receptor Panel CareView 1500C X-ray Flat Panel Detector. - PLUS: the dicomPACS® software package (Same as our predicate). e) The x-ray generators are portable units which operate from 120/240V 50-60° AC. The generator unit utilizes a high frequency inverter and can be mounted to a tripod or support arm. The usual safety precautions regarding the use of x-rays must be observed by the operator. The digital panel features the Careray flat panel technology in a sleek and compact unit. The portable panel provides digital X-ray image capture for a wide range of applications. The lightweight design, generous imaging area, and fast processing times of the detector make it easy to capture high quality diagnostic images for routine diagnosis, as well as challenging trauma and bedside exams. It's a portable solution for a faster, more streamlined approach to digital radiography. The only difference between this modified device and our predicate devices is the model number of the digital x-ray receptor panel. The predicate panel can communicate either by wireless or wired connection. The subject device communicates by Ethernet only. {4}------------------------------------------------ ## 6) Substantial Equivalence Chart | Characteristic | Predicate: K191451<br>CMDR 2CW (Multiple Models) | CMDR 2C (Multiple Models) | |--------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use: | Intended for use by a qualified/trained<br>physician or technician on both adult and<br>pediatric subjects for taking diagnostic x-<br>rays. Not for mammography | SAME | | Configuration | Mobile System with digital x-ray panel and<br>image acquisition computer | SAME | | Computer | Dell Precision M3530 or Dell Latitude 7424<br>(ruggedized) | Dell Precision M3530 or Dell Latitude 7424<br>(ruggedized) | | X-ray<br>Generator(s)<br>(All made by<br>Mikasa X-Ray) | HF120/60H PowerPlus (K040046) or<br>HF100H+ (K052721) or<br>HF1202H PowerPlus (K153059) | SAME | | Basic<br>Generator<br>Characteristics | 120 VAC Line operated:<br>HF120/60H PowerPlus™ produces up to 120<br>kVP and 2.4 kW peak<br>HF100H+: up to 100 kVp and 2 kW peak<br>HF1202H PowerPlus: up to 120 kVp up to<br>3.0 kW peak | SAME | | Collimator | Collimare LED Collimator | Collimare LED Collimator | | Generator<br>Photo(s)<br>The same<br>generators will<br>be used for the<br>modified device | Image: HF120/60HPPWV | Image: HF100H+<br>Image: HF1202 H | | Digital X-ray<br>Panel Supplied | K150929 CareView 1500Cw X-ray Flat Panel<br>Detector manufactured by CareRay | K153058 CareView 1500C X-ray Flat Panel<br>Detector manufactured by CareRay | | Panel<br>Performance | Pixel Pitch 154 μm<br>2304 × 2816 pixels<br>Size 14" x 17"<br>(Cleared in K150929)<br>Csl Scintillator<br>Spatial Resolution Min. 3.3 line pair/mm<br>A/D Resolution: 16 bit, 65536 grayscale<br>MTF - 65% @ 1 lp/mm<br>DQE Performance: > 65% @ 0 lp/mm | Pixel Pitch 154 μm<br>2304 × 2816 pixels<br>Size 14" x 17"<br>(Cleared in K153058)<br>Csl Scintillator<br>Spatial Resolution Min. 3.3 line pair/mm<br>A/D Resolution: 16 bit, 65536 grayscale<br>MTF - 65% @ 1 lp/mm<br>DQE Performance: > 65% @ 0 lp/mm<br>Note: This panel was cited as the predicate in<br>K150929. The panels are identical except for the<br>method of image communication to the host. | | Panel<br>Communication | IEEE 802.11a/b/g/n (2.4 GHz /5 GHz) | Gigabit Ethernet | | Characteristic | Predicate: K191451<br>CMDR 2CW (Multiple Models) | CMDR 2C (Multiple Models) | | Panel Power<br>Source | DC Adapter or Lithium Ion rechargeable<br>battery | DC Adapter or Lithium Ion rechargeable<br>battery | | Panel Interface | Ethernet or Wi-Fi wireless | Ethernet only | | Meets US<br>Performance<br>Standard | YES | YES | | PACS software | dicomPACS® (Cleared with the Toshiba<br>panels in K141440) | SAME | | Power Source | 120 V 50/60 Hz AC 20 amp | SAME | | Digital Panel<br>Power Source | 120 V 50/60 Hz AC or Lithium Ion<br>Rechargeable Battery. | SAME | | Digital Panel | Image: Digital Panel | Image: Digital Panel | | Model Details | (All models use O&R Imaging Software with the<br>Careray Digital Panel and a Dell Laptop) (Five<br>Models) Model Details: | (All models use O&R Imaging Software with the<br>Careray Digital Panel and a Dell Laptop) (Five<br>Models) Model Details: | | | CMDR.CW.120.60.S for Dell Precision or<br>CMDR.CW.120.60.R for Dell Latitude Laptop or<br>equivalent Uses HF120/60HPPWV Generator<br>with CareView 1500Cw and O&R Imaging<br>Software | CMDR.C.120.60.S for Dell Precision or<br>CMDR.C.120.60.R for Dell Latitude Laptop or<br>equivalent Uses HF120/60HPPWV Generator<br>with CareView 1500Cw and O&R Imaging<br>Software | | | CMDR.CW.100.S for Dell Precision or<br>CMDR.CW.100.Rfor Dell Latitude Laptop or<br>equivalent Uses HF100H+ Generator with<br>CareView 1500Cw and O&R Imaging Software. | CMDR.C.100.S for Dell Precision or<br>CMDR.C.100.Rfor Dell Latitude Laptop or<br>equivalent Uses HF100H+ Generator with<br>CareView 1500Cw and O&R Imaging Software. | | | CMDR.CW.1202.S for Dell Precision Laptop or<br>equivalent or<br>CMDR.CW.1202.R for Dell Latitude Laptop or<br>equivalent Uses HF1202 Generator with<br>CareView 1500Cw and O&R Imaging Software | CMDR.C.1202.S for Dell Precision Laptop or<br>equivalent or<br>CMDR.C.1202.R for Dell Latitude Laptop or<br>equivalent Uses HF1202 Generator with<br>CareView 1500Cw and O&R Imaging Software | | | CMDR 2CW-MIL (Uses Dell Latitude 7424 or<br>equivalent with the HF120/60HPPWV Generator<br>with CareView 1500Cw and O&R Imaging<br>Software | CMDR 2C-MIL (Uses Dell Latitude 7424 or<br>equivalent with the HF120/60HPPWV Generator<br>with CareView 1500Cw and O&R Imaging<br>Software | {5}------------------------------------------------ {6}------------------------------------------------ - 7) The technological characteristics, including design, materials, composition, and energy source, are substantially the same, so there are no issues impacting safety and effectiveness. Safety and Effectiveness, comparison to predicate device. The results of bench testing indicate that the new devices are as safe and effective as the predicate devices. Proper system operation is fully verified upon installation. We verified that the modified combination of components worked properly and produced diagnostic quality images as good as our predicate generator/panel combination. NO HARDWARE OR SOFTWARE MODIFICATIONS TO ALREADY CLEARED DEVICES WERE REQUIRED TO CREATE THESE NEW MODELS. The main difference from the predicate device is the addition of the CareView 1500 CW flat-panel detector, that replaces the Toshiba and Perkin-Elmer digital imagers used with the predicate system. - 8) Summary of non-clinical testing: Prototype systems covering all generator/panel combinations were assembled and tested. First the dicomPACS® software was installed on the Dell Inspiron laptop computer. The proper installation was verified by running the software. A Wi-Fi connection was confirmed between the PerkinElmer panel and the Dell laptop. The panel was confirmed to be charged and turned on. Then the MinXray HF portable generator was turned on. The MinXray portable X-ray generator was set to generate an exposure. The generator was aimed at the PerkinElmer panel, and a radiographic phantom was placed on the panel. Several test exposures showed that the system was operating properly. No modifications were necessary to any of the hardware or software other than changing the digital panel. The completed system complies with DHHS radiation safety standards currently in effect, and has undergone testing for compliance with UL 60601-1 (2005) (Electrical medical device safety), IEC 60601-1-2 (2007) (Electromagnetic Compatibility). Additionally, the HF1202H PowerPlus generator meets IEC 60601-2-54: Medical electrical equipment - Part 2-54: Particular requirements for the basic safety and essential performance of X-ray equipment for radiography and radioscopy. The other generators were designed and cleared prior to the isuance of this standard. We verified software compatibility with the new CareView Cw digital panel. The risks and hazardous impacts of the device modification were analyzed by FMEA methodology. The specific risk control and protective measures to mitigate the risks from the modification were reviewed and implemented as part of product design. The overall assessment concluded that all identified risks and hazardous conditions were successfully mitigated and accepted. We employed the i.b.a. Test Device DIGI-13, a device for quality tests at CR and DR systems (e.g. for acceptance tests according to DIN V 6868-58 and constancy tests according to DIN 6868-13) to obtain images from both the predicate and the new digital panel. All panel/generator combinations were tested. The images were evaluated and found to be of diagnostic quality. In recognition of the FDA Guidance on Cybersecurity, we added Cybersecurity precautions to our labeling and obtained Cybersecurity information from our DICOM software supplier Oehm Und Rehbein. - 9) Summary of clinical testing: Clinical testing was not required to establish substantial equivalence. - 10) Conclusion: After analyzing bench and clinical tests, it is the conclusion of MinXray Inc. that the modified Digital Diagnostic X-Ray Systems are as safe and effective as the predicate device, have few technological differences, and has the same indications for use, thus rendering it substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...