Intended for use by a qualified/trained physician or technician on both adult and pediatric subjects for taking diagnostic x-rays. Not for mammography.
Device Story
Mobile digital x-ray system; captures diagnostic images using portable high-frequency x-ray generators and digital receptor panels. System components include MinXray portable generators (HF120/60H, HF100H+, or HF1202H), digital flat panel detectors (Toshiba FDX3543RP/RPW or PerkinElmer XRpad 4336), and dicomPACS software on a laptop or industrial computer. Operated by physicians or technicians in clinical or bedside settings. X-ray generator produces ionizing radiation; digital panel converts x-rays to digital signals; software processes, displays, and manages images. Output used by clinicians for diagnostic assessment of patient anatomy. Benefits include streamlined digital radiography, portability for trauma/bedside exams, and high-quality diagnostic imaging.
Clinical Evidence
No clinical data. Substantial equivalence established via bench testing. System performance verified using i.b.a. Test Device DIGI-13 for quality tests (DIN V 6868-58, DIN 6868-13). Images evaluated for diagnostic quality across all generator/panel combinations. Compliance with electrical safety (UL 60601-1) and electromagnetic compatibility (IEC 60601-1-2) confirmed.
Technological Characteristics
Mobile x-ray system; high-frequency inverter generators (120 VAC); digital flat panel detectors (Toshiba/PerkinElmer); laptop/industrial PC. Connectivity via Ethernet or Wi-Fi. Standards: 21 CFR 1020.30, UL 60601-1 (2005), IEC 60601-1-2 (2007), IEC 60601-2-54. Software: dicomPACS.
Indications for Use
Indicated for use by qualified/trained physicians or technicians on adult and pediatric patients for diagnostic x-ray imaging. Contraindicated for mammography.
Regulatory Classification
Identification
A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
CMDR-2ST & CMDR-2SLWT Digital Portable X-ray System (K141885)
Reference Devices
Perkinelmer, XRpad 4336 MED Flat Panel Detector (K140551)
Oehm Und Rehbein Gmbh, DicomPACS® DX-R with flat panel (K141440)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract symbol that resembles three human profiles stacked on top of each other.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 7, 2017
MinXray, Inc. % Daniel Kamm, P.E. Principal Engineer Kamm & Associates 8870 Ravello Ct. NAPLES FL 34114
Re: K171353
Trade/Device Name: CMDR 2ST (Multiple Models), CMDR 2SPE (Multiple Models), Integris Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: II Product Code: IZL, MQB, LLZ Dated: May 3, 2017 Received: May 9, 2017
Dear Mr. Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{1}------------------------------------------------
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Sincerely,
Michael D'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K171353
Device Name
CMDR 2ST (Multiple Models), CMDR 2SPE (Multiple Models), Integris
Indications for Use (Describe)
Intended for use by a qualified trained physician on both adult and pediatric subjects for taking diagnostic x-rays. Not for mammography.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary: 510(k) Number K171353 MinXray, Inc. 3611 Commercial Avenue Northbrook, Illinois 60062, USA Toll Free 1-800-221-2245 (USA & Canada) Date Prepared: May 27, 2017 Contact: Keith Kretchmer, President
- 1) Identification of the Device: Trade/Device Name: CMDR 2ST (Multiple Models), CMDR 2SPE (Multiple Models), Integris Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: II Product Codes: IZL, MQB, and LLZ. Common/Usual Name: Digital Mobile Diagnostic X-Ray System
- 2) Equivalent legally marketed device: K141885 Trade/Device Name: CMDR-2ST & CMDR-2SLWT Digital Portable X-ray System Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: II Product Codes: IZL, MQB, and LLZ. Common/Usual Name: Digital Mobile Diagnostic X-Ray System
- 3) Reference devices (We employ these cleared devices without modification):
- a) Trade/Device Name: K140551 Perkinelmer, XRpad 4336 MED Flat Panel Detector Regulation Number: 21 CFR 892. 1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB
- b) Trade/Device Name: K141440; Oehm Und Rehbein Gmbh, DicomPACS® DX-R with flat panel Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary X-ray System Regulatory Class: Class II Product Code: MQB
- 4) Indications for Use (intended use): These digital radiographic systems are intended for use by a qualified/trained physician or technician on both adult and pediatric subjects for taking diagnostic x-rays. Not for mammographic use.
5) Description of the Device: This represents the straightforward combination of three devices: One of three cleared MinXray Portable HF X-ray generators: HF120/60H PowerPlus cleared in K040046, (and in K141885) OR HF100H+ cleared in K052721 OR HF1202 PowerPlus cleared in K153059. One of three cleared digital X-ray receptor panels: Toshiba FDX3543RP OR the Toshiba FDX3543RPW cleared in K162687 (and others) OR the PerkinElmer Solid State Imager, (K140551) PLUS: the dicomPACS® software package (K141885) (Same as our predicate). The X-ray generators are portable units which operate from 120 V 50-60° AC. The generator unit utilizes a high frequency inverter and can be mounted to a tripod or support arm. The usual safety precautions regarding the use of x-rays must be observed by the operator. The digital panel features the formerly used Toshiba panels or PerkinElmer flat panel technology in a sleek and compact unit. The portable panels provide digital X-ray image capture for a wide range of applications. The lightweight design, generous imaging area, and fast processing times of the
detector make it easy to capture high quality diagnostic images for routine diagnosis, as well as challenging trauma and bedside exams. It's a portable solution for a faster, more streamlined approach to digital radiography. The only difference between this modified device and our predicate device is the digital x-ray receptor panel. .
{4}------------------------------------------------
## 6) Substantial Equivalence Chart
| Characteristic | Predicate:<br>CMDR-2ST & CMDR-2SLWT<br>Digital X-Ray System<br>K141885 | CMDR 2ST (Multiple Models) | CMDR 2SPE (Multiple Models) | Integris |
|--------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use: | Intended for use by a<br>qualified/trained physician or<br>technician on both adult and<br>pediatric subjects for taking<br>diagnostic x-rays. Not for<br>mammography | SAME | SAME | SAME |
| Configuration | Mobile System with digital x-<br>ray panel and image<br>acquisition computer | SAME | SAME | SAME |
| Computer | Dell laptop | Dell Precision M3510 or Dell<br>Latitude 14 (ruggedized) | Dell Precision M3510 or Dell<br>Latitude 14 (ruggedized) | Beckhoff C6920/C6925 |
| X-ray<br>Generator(s)<br>(All made by<br>Mikasa X-Ray) | HF120/60H PowerPlus<br>(K040046) | SAME OR<br>HF100H+ (K052721)<br>HF1202H PowerPlus (K153059) | SAME OR<br>HF100H+ (K052721)<br>HF1202H PowerPlus (K153059) | HF1202H PowerPlus (K153059) |
| Basic<br>Generator<br>Character-<br>istics | 120 VAC Line operated<br>HF120/60H PowerPlus™<br>produces up to 120 kVP and<br>2.4 kW peak | SAME OR<br>HF100H+: up to 100 kVp and 2<br>kW peak<br>HF1202H PowerPlus: up to 120<br>kVp up to 3.0 kW peak | SAME OR<br>HF100H+: up to 100 kVp and 2 kW<br>peak<br>HF1202H PowerPlus: up to 120 kVp<br>up to 3.0 kW peak | HF1202H PowerPlus: up to 120<br>kVp up to 3.0 kW peak |
| Collimator | Collimare LED Collimator | Collimare LED Collimator | Collimare LED Collimator | Collimare LED Collimator |
| Characteristic | Predicate:<br>CMDR-2ST & CMDR-2SLWT<br>Digital X-Ray System<br>K141885 | CMDR 2ST (Multiple Models) | CMDR 2SPE (Multiple Models) | Integris |
| Generator<br>Photo(s) | HF120/60HPPWV PowerPlus<br>Image: HF120/60HPPWV PowerPlus generator | HF120/60HPPWV PowerPlus<br>Image: HF120/60HPPWV PowerPlus generator<br>OR: HF100H+<br>Image: HF100H+ generator<br>OR: HF1202H PowerPlus<br>Image: HF1202H PowerPlus generator | CMDR 2SPE 120/60 PP<br>Image: CMDR 2SPE 120/60 PP generator<br>OR: HF100H+<br>Image: HF100H+ generator<br>OR: HF1202H PowerPlus<br>Image: HF1202H PowerPlus generator | HF1202H PowerPlus<br>Image: HF1202H PowerPlus generator |
| Digital X-ray<br>Panel<br>Supplied | Toshiba FDX3543RP | Toshiba FDX3543RPW. Note:<br>this panel was cleared in<br>K143257 | PerkinElmer XRpad 4336 (Cleared in<br>K140551) | PerkinElmer XRpad 4336 (Cleared<br>in K140551) |
| Characteristic | Predicate:<br>CMDR-2ST & CMDR-2SLWT<br>Digital X-Ray System<br>K141885 | CMDR 2ST (Multiple Models) | CMDR 2SPE (Multiple Models) | Integris |
| Panel<br>Performance | Pixel size 143 x 143μm<br>Matrix size 2448x2984<br>Size 14" x 17" | Pixel Pitch 140 μ m<br>Pixel Matrix 2466 (H) x 3040 (V)<br>Size 14" x 17" | Pixel size 100 x 100μm<br>Matrix size 3556 × 4320<br>Size 14" x 17" (Cleared in K140551) | Pixel size 100 x 100μm<br>Matrix size 3556 × 4320<br>Size 14" x 17" (Cleared in<br>K140551) |
| Panel Power<br>Source | DC Adapter | DC Adapter or Lithium Ion<br>rechargeable battery | DC Adapter or Lithium Ion<br>rechargeable battery | DC Adapter or Lithium Ion<br>rechargeable battery |
| Panel<br>Interface | Ethernet | Ethernet or Wi-Fi wireless, panel<br>cleared in | Ethernet or Wi-Fi wireless, panel<br>cleared in K140551 | Ethernet or Wi-Fi wireless, panel<br>cleared in K140551 |
| Performance<br>Standard | 21 CFR 1020.30 | SAME | SAME | SAME |
| PACS<br>software | dicomPACS® (Cleared in<br>K141440) | dicomPACS® (Cleared with the<br>Toshiba panels in K141440) | dicomPACS® (Cleared with the<br>PerkinElmer panel in K141440) | dicomPACS® (Cleared with the<br>PerkinElmer panel in K141440) |
| Power Source | 120 V 50/60 Hz AC 20 amp | SAME | SAME | SAME |
| Digital Panel<br>Power Source | 120 V 50/60 Hz AC | 120 V 50/60 Hz AC or Lithium<br>Ion Rechargeable Battery. | 120 V 50/60 Hz AC or Lithium Ion<br>Rechargeable Battery. | 120 V 50/60 Hz AC or Lithium Ion<br>Rechargeable Battery. |
| Digital Panel | Image: Digital Panel | Image: Digital Panel | Image: Digital Panel | Image: Digital Panel |
| Characteristic | Predicate:<br>CMDR-2ST & CMDR-2SLWT<br>Digital X-Ray System<br>K141885 | CMDR 2ST (Multiple Models) | CMDR 2SPE (Multiple Models) | Integris |
| Model Details | HF120/60HPPWV Generator<br>with Toshiba FDX3543RP and<br>O&R Imaging Software | CMDR 2ST 100+/ DP for Dell<br>Precision or a /DL for Dell<br>Latitude Computer<br>Uses HF100H+ Generator with<br>Toshiba FDX3543RP and O&R<br>Imaging Software<br>CMDR 2STW 100H+/DP for Dell<br>Precision or a /DL for Dell<br>Latitude Computer<br>Uses HF100H+ Generator with<br>Toshiba FDX3543RPW and O&R<br>Imaging Software<br>CMDR 2STW 1202/DP for Dell<br>Precision or a /DL for Dell<br>Latitude Computer<br>Uses HF1202 Generator with<br>Toshiba FDX3543RPW and O&R<br>Imaging Software<br>CMDR 2STW-MIL (Uses Dell<br>Latitude 14 with the<br>HF120/60HPPWV Generator<br>with Toshiba FDX3543RPW and<br>O&R Imaging Software | CMDR 2SPE 120/60 PP<br>CMDR 2SPE 100H+<br>CMDR 2SPE 1202<br>(The suffix is the generator model<br>number)<br>Add a /DP for Dell Precision<br>Computer or a /DL for Dell Latitude<br>Computer<br>All use the PerkinElmer XRpad 4336<br>and O&R Imaging Software | Integris.<br>Panel: Perkin Elmer Wireless<br>XRpad 4336 and O&R Imaging<br>Software<br>Computer: Beckhoff<br>C6920/C6925<br>Generator: HF1202H PowerPlus |
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
- 7) The technological characteristics, including design, materials, composition, and energy source, are substantially the same, so there are no issues impacting safety and effectiveness.
Safety and Effectiveness, comparison to predicate device. The results of bench testing indicates that the new devices are as safe and effective as the predicate devices. Proper system operation is fully verified upon installation. We verified that the modified combination of components worked properly and produced diagnostic quality images as good as our predicate generator/panel combination. NO HARDWARE OR SOFTWARE MODIFICATIONS TO ALREADY CLEARED DEVICES WERE REQUIRED TO CREATE THESE NEW MODELS.
- 8) Summary of non-clinical testing: Prototype systems covering all generator/panel combinations were assembled and tested. First the dicomPACS® software was installed on the Dell Inspiron laptop computer. The proper installation was verified by running the software. A Wi-Fi connection was confirmed between the PerkinElmer panel and the Dell laptop. The panel was confirmed to be charged and turned on. Then the MinXray HF portable generator was turned on. The MinXray portable X-ray generator was set to generate an exposure. The generator was aimed at the PerkinElmer panel, and a radiographic phantom was placed on the panel. Several test exposures showed that the system was operating properly. No modifications were necessary to any of the hardware or software other than changing the digital panel. The completed system complies with DHHS radiation safety standards currently in effect, and has undergone testing for compliance with UL 60601-1 (2005) (Electrical medical device safety), IEC 60601-1-2 (2007) (Electromagnetic Compatibility). Additionally, the HF1202H PowerPlus generator meets IEC 60601-2-54: Medical electrical equipment - Part 2-54: Particular requirements for the basic safety and essential performance of X-ray equipment for radiography and radioscopy. The other generators were designed and cleared prior to the isuance of this standard. The software supplier Oehm Und Rehbein GmbH verified compatibility with the new PerkinElmer digital panel and supplied us with a test report. The risks and hazardous impacts of the device modification were analyzed by FMEA methodology. The specific risk control and protective measures to mitigate the risks from the modification were reviewed and implemented as part of product design. The overall assessment concluded that all identified risks and hazardous conditions were successfully mitigated and accepted.
We employed the i.b.a. Test Device DIGI-13, a device for quality tests at CR and DR systems (e.g. for acceptance tests according to DIN V 6868-58 and constancy tests according to DIN 6868-13) to obtain images from both the predicate and the new digital panel. All panel/generator combinations were tested. The images were evaluated and found to be of diagnostic quality.
- 9) Summary of clinical testing: Clinical testing was not required to establish substantial equivalence.
- 10) Conclusion: After analyzing bench and clinical tests, it is the conclusion of MinXray Inc. that the modified Digital Diagnostic X-Ray Systems are as safe and effective as the predicate device, have few technological differences, and has the same indications for use, thus rendering it substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.