Intended for use by a qualified/trained physician or technician on both adult and pediatric subjects for taking diagnostic x-rays. Not for mammography.
Device Story
Mobile digital x-ray system; combines portable high-frequency x-ray generators (HF120/60H PowerPlus, HF100H+, or HF1202 PowerPlus) with CareView 1500Cw flat-panel detector and dicomPACS software. Used in clinical settings for routine, trauma, and bedside diagnostic imaging. Operator (physician/technician) positions generator and detector; system captures x-ray signals, converts to digital images via flat-panel detector, and processes/displays images on a laptop (Dell Precision or Latitude). Output assists clinicians in diagnostic decision-making. System provides portable, streamlined digital radiography workflow.
Clinical Evidence
No clinical data. Bench testing only. System performance verified using i.b.a. Test Device DIGI-13 for quality tests (DIN V 6868-58 and DIN 6868-13). Images evaluated for diagnostic quality. Compliance with UL 60601-1, IEC 60601-1-2, and IEC 60601-2-54 verified.
Technological Characteristics
Mobile x-ray system; high-frequency inverter generators (120/240V AC); CareView 1500Cw flat-panel detector (154 μm pixel pitch, 2304 x 2816 matrix); Dell laptop workstation. Connectivity via Wi-Fi/wired. Standards: UL 60601-1, IEC 60601-1-2, IEC 60601-2-54. Software: dicomPACS.
Indications for Use
Indicated for use by qualified physicians or technicians on adult and pediatric patients for diagnostic x-ray imaging. Contraindicated for mammography.
Regulatory Classification
Identification
A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 3, 2019
MinXray, Inc. % Daniel Kamm, P.E. Principal Engineer Kamm & Associates 8870 Ravello Ct NAPLES FL 34114
Re: K191451
Trade/Device Name: CMDR 2CW (Multiple Models) Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: Class II Product Code: IZL, MQB, LLZ Dated: May 28, 2019 Received: May 31, 2019
Dear Mr. Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K191451
Device Name CMDR 2CW (Multiple Models)
Indications for Use (Describe)
Intended for use by a qualified/trained physician or technician on both adult and pediatric subjects for taking diagnostic x-rays. Not for mammography.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary: 510(k) Number K191451 MinXray, Inc. 3611 Commercial Avenue Northbrook. Illinois 60062. USA Toll Free 1-800-221-2245 (USA & Canada) Date Prepared: June 17, 2019 Contact: Keith Kretchmer, President
- Identification of the Device: 1) Trade/Device Name: CMDR 2CW (Multiple Models) Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: II Product Codes: IZL, MQB, and LLZ. Common/Usual Name: Digital Mobile Diagnostic X-Ray System
- 2) Equivalent legally marketed device: K171353 Trade/Device Name: CMDR 2ST (Multiple Models), CMDR 2SPE (Multiple Models), Integris Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: II Product Codes: IZL, MQB, and LLZ. Common/Usual Name: Digital Mobile Diagnostic X-Ray System
- 3) Reference device (We employ these cleared devices without modification): Trade/Device Name: K150929 CareView 1500Cw X-ray Flat Panel Detector made by CareRay. Regulation Number: 21 CFR 892. 1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB
- Indications for Use (intended use): These digital radiographic systems are intended for use by a qualified/trained 4) physician or technician on both adult and pediatric subjects for taking diagnostic x-rays. Not for mammographic use.
- 5) Description of the Device: This represents the straightforward combination of three devices: One of three cleared MinXray Portable HF X-ray generators:
- a) HF120/60H PowerPlus cleared in K040046, (and in K141885) OR
- b) HF100H+ cleared in K052721 OR
- c) HF1202 PowerPlus cleared in K153059.
- Plus: A 510(k) cleared (K150929) Digital X-Ray Receptor Panel CareView 1500Cw X-ray Flat Panel Detector. d)
- e) PLUS: the dicomPACS® software package (Same as our predicate).
The x-ray generators are portable units which operate from 120/240V 50-60° AC. The generator unit utilizes a high frequency inverter and can be mounted to a tripod or support arm. The usual safety precautions regarding the use of x-rays must be observed by the operator. The digital panel features the Careray flat panel technology in a sleek and compact unit. The portable panel provides digital X-ray image capture for a wide range of applications. The lightweight design, generous imaging area, and fast processing times of the detector make it easy to capture high quality diagnostic images for routine diagnosis, as well as challenging trauma and bedside exams. It's a portable solution for a faster, more streamlined approach to digital radiography. The only difference between this modified device and our predicate devices is the supplier of the digital x-ray receptor panel.
{4}------------------------------------------------
## 6) Substantial Equivalence Chart
| Characteristic | Predicate: K171353 CMDR 2ST (Multiple<br>Models) CMDR 2SPE (Multiple Models) | CMDR 2CW (Multiple Models) |
|--------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use: | Intended for use by a qualified/trained<br>physician or technician on both adult and<br>pediatric subjects for taking diagnostic x-rays.<br>Not for mammography | SAME |
| Configuration | Mobile System with digital x-ray panel and<br>image acquisition computer | SAME |
| Computer | Dell Precision M3510 or Dell Latitude 14 | Dell Precision M3530 or Dell Latitude 7424<br>(ruggedized) |
| X-ray<br>Generator(s)<br>(All made by<br>Mikasa X-Ray) | HF120/60H PowerPlus (K040046) or<br>HF100H+ (K052721) or<br>HF1202H PowerPlus (K153059) | SAME |
| Basic<br>Generator<br>Characteristics | 120 VAC Line operated:<br>HF120/60H PowerPlus™ produces up to 120<br>kVP and 2.4 kW peak<br>HF100H+: up to 100 kVp and 2 kW peak<br>HF1202H PowerPlus: up to 120 kVp up to 3.0<br>kW peak | SAME |
| Collimator | Collimare LED Collimator | Collimare LED Collimator |
| Generator<br>Photo(s)<br>The same<br>generators will<br>be used for the<br>modified device | Image: HF120/60HPPWV | HF100H+<br>HF1202 H |
| Digital X-ray<br>Panel<br>Supplied | Toshiba FDX3543RPW. Note: this panel was<br>cleared in K143257 OR PerkinElmer XRpad<br>4336 (Cleared in K140551) | K150929 CareView 1500Cw X-ray Flat Panel<br>Detector manufactured by CareRay |
| Panel<br>Performance | Pixel Pitch 140 μm<br>Pixel Matrix 2466 (H) x 3040 (V)<br>Size 14" x 17" OR<br>Pixel Pitch 100 x 100μm<br>Matrix size 3556 × 4320<br>Size 14" x 17" (Cleared in K140551) | Pixel Pitch 154 μm<br>2304 × 2816 pixels<br>Size 14" x 17"<br>(Cleared in K150929) |
| Panel Power<br>Source | DC Adapter or Lithium Ion rechargeable<br>battery | DC Adapter or Lithium Ion rechargeable<br>battery |
| Characteristic | Predicate: K171353 CMDR 2ST (Multiple<br>Models) CMDR 2SPE (Multiple Models) | CMDR 2CW (Multiple Models) |
| Meets US<br>Performance<br>Standard | YES | YES |
| PACS software | dicomPACS® (Cleared with the Toshiba panels<br>in K141440) | SAME |
| Power Source | 120 V 50/60 Hz AC 20 amp | SAME |
| Digital Panel<br>Power Source | 120 V 50/60 Hz AC or Lithium Ion Rechargeable<br>Battery. | SAME |
| Digital Panel | Image: Digital Panel | Image: Digital Panel |
| Model Details | (All models use O&R Imaging Software, and<br>either a wired or wireless Toshiba panel) (Five<br>Models) | (All models use O&R Imaging Software with the<br>Careray Digital Panel and a Dell Laptop) (Five<br>Models) Model Details: |
| | HF120/60HPPWV Generator with Toshiba<br>FDX3543RP<br>CMDR 2ST 100+/ DP for Dell Precision or a /DL for<br>Dell Latitude Computer: Uses HF100H+ Generator<br>with Toshiba FDX3543RP<br>CMDR 2STW 100H+/DP for Dell Precision or a /DL<br>for Dell Latitude Computer: Uses HF100H+<br>Generator with Toshiba FDX3543RPW<br>CMDR 2STW 1202/DP for Dell Precision or a /DL for<br>Dell Latitude Computer: Uses HF1202 Generator<br>with Toshiba FDX3543RPW<br>CMDR 2STW-MIL (Uses Dell Latitude 14 with the<br>HF120/60HPPWV Generator with Toshiba<br>FDX3543RPW | CMDR.CW.120.60.S for Dell Precision or<br>CMDR.CW.120.60.R for Dell Latitude Laptop or<br>equivalent Uses HF120/60HPPWV Generator with<br>CareView 1500Cw and O&R Imaging Software<br>CMDR.CW.100.S for Dell Precision or<br>CMDR.CW.100.Rfor Dell Latitude Laptop or<br>equivalent Uses HF100H+ Generator with CareView<br>1500Cw and O&R Imaging Software.<br>CMDR.CW.1202.S for Dell Precision Laptop or<br>equivalent or<br>CMDR.CW.1202.R for Dell Latitude Laptop or<br>equivalent Uses HF1202 Generator with CareView<br>1500Cw and O&R Imaging Software<br>CMDR 2CW-MIL (Uses Dell Latitude 7424 or<br>equivalent with the HF120/60HPPWV Generator |
{5}------------------------------------------------
{6}------------------------------------------------
- 7) The technological characteristics, including design, materials, composition, and energy source, are substantially the same, so there are no issues impacting safety and effectiveness. Safety and Effectiveness, comparison to predicate device. The results of bench testing indicate that the new devices are as safe and effective as the predicate devices. Proper system operation is fully verified upon installation. We verified that the modified combination of components worked properly and produced diagnostic quality images as good as our predicate generator/panel combination. NO HARDWARE OR SOFTWARE MODIFICATIONS TO ALREADY CLEARED DEVICES WERE REQUIRED TO CREATE THESE NEW MODELS. The main difference from the predicate device is the addition of the CareView 1500 CW flat-panel detector, that replaces the Toshiba and Perkin-Elmer digital imagers used with the predicate system.
- 8) Summary of non-clinical testing: Prototype systems covering all generator/panel combinations were assembled and tested. First the dicomPACS® software was installed on the Dell Inspiron laptop computer. The proper installation was verified by running the software. A Wi-Fi connection was confirmed between the PerkinElmer panel and the Dell laptop. The panel was confirmed to be charged and turned on. Then the MinXray HF portable generator was turned on. The MinXray portable X-ray generator was set to generate an exposure. The generator was aimed at the PerkinElmer panel, and a radiographic phantom was placed on the panel. Several test exposures showed that the system was operating properly. No modifications were necessary to any of the hardware or software other than changing the digital panel. The comples with DHHS radiation safety standards currently in effect, and has undergone testing for compliance with UL 60601-1 (2005) [Electrical medical device safety), IEC 60601-1-2 (2007) (Electromagnetic Compatibility). Additionally, the HF1202H PowerPlus generator meets IEC 60601-2-54: Medical electrical equipment - Part 2-54: Particular requirements for the basic safety and essential performance of X-ray equipment for radiography and radioscopy. The other generators were designed and cleared prior to the issuance of this standard. We verified software compatibility with the new CareView Cw digital panel. The risks and hazardous impacts of the device modification were analyzed by FMEA methodology. The specific risk control and protective measures to mitigate the risks from the modification were reviewed and implemented as part of product design. The overall assessment concluded that all identified risks and hazardous conditions were successfully mitigated and accepted.
We employed the i.b.a. Test Device DIGI-13, a device for quality tests at CR and DR systems (e.g. for acceptance tests according to DIN V 6868-58 and constancy tests according to DIN 6868-13) to obtain images from both the predicate and the new digital panel. All panel/generator combinations were tested. The images were evaluated and found to be of diagnostic quality.
In recognition of the FDA Guidance on Cybersecurity, we added Cybersecurity precautions to our labeling and obtained Cybersecurity information from our DICOM software supplier Oehm Und Rehbein.
- 9) Summary of clinical testing: Clinical testing was not required to establish substantial equivalence.
- 10) Conclusion: After analyzing bench and clinical tests, it is the conclusion of MinXray Inc. that the modified Digital Diagnostic X-Ray Systems are as safe and effective as the predicate device, have few technological differences, and has the same indications for use, thus rendering it substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.