Synthetic polyisoprene probe covers are indicated for use during diagnostic ultrasound procedures in cases where patient sensitivity to natural rubber latex is speculative.
Device Story
SheathesISO Ultrasound Transducer Covers are thin, conformal, cylindrical synthetic polyisoprene sheaths designed to fit various ultrasound transducer shapes. The device acts as a protective barrier to minimize cross-contamination between the patient and the transducer during diagnostic ultrasound procedures. It is intended for single-patient, disposable use and may be supplied in procedure kits containing elastic fasteners, coupling gel, or needle guides. The device is used in clinical settings by healthcare professionals. By providing a latex-free barrier, it allows for safe ultrasound imaging in patients with suspected natural rubber latex sensitivity.
Clinical Evidence
No clinical data. Evidence consists of bench testing and biocompatibility assessments. Bench testing included viral penetration, stretch testing, pinhole testing, and gel compatibility. Biocompatibility testing included material-mediated pyrogen, cytotoxicity, systemic toxicity, hemolysis, irritation, and sensitization, confirming compliance with ISO 10993-1:2003 for surface-contacting devices in contact with the body for less than 24 hours.
Technological Characteristics
Material: Cis-1,4-polyisoprene (synthetic, latex-free). Form factor: Cylindrical, thin, conformal dip-molded sheath. Energy source: None (passive accessory). Connectivity: None. Sterilization: Sterile and non-sterile options provided. Biocompatibility: Compliant with ISO 10993-1:2003.
Indications for Use
Indicated for patients undergoing diagnostic ultrasound procedures where natural rubber latex sensitivity is a concern.
Regulatory Classification
Identification
A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.
K013721 — SYNTHETIC POLYSIOPRENE ULTRASOUND TRANSDUCER COVER · CIVCO Medical Instruments Co., Inc. · Dec 28, 2001
K970515 — GENERAL PURPOSE LATEX TRANSDUCER COVERS · CIVCO Medical Instruments Co., Inc. · Jun 20, 1997
K050320 — FREESCAN LATEX-FREE ULTRASOUND TRANSDUCER COVER KIT · Ascendia Medtech AB · Aug 15, 2005
K981069 — ULTRA-COVER PU-TRANSDUCER COVER · International Medical Products B.V. · Sep 23, 1998
K970513 — GENERAL PURPOSE TRANDUCER COVER · CIVCO Medical Instruments Co., Inc. · Jun 20, 1997
Submission Summary (Full Text)
{0}------------------------------------------------
10091828
### Section 5: 510(K) Summarv
(As required by 21 CFR 807.92)
# Synthetic Polyisoprene Ultrasound Transducer Cover
September 25, 2009 (updated)
SEP 2 8 2009
Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (WO66-0609) 10903 New Hampshire Avenue Silver Spring, MD 20993-0002
To Whom it may Concern:
This letter, along with the attached materials is to notify your office of the intention of Sheathing Technologies to market the following revice starting on or after (90) days from this date.
Device/Specification Developer:
Sheathing Technologies, Inc. 18431 Technology Drive Morgan Hill, CA 95037
Establishment Registration No.: 2950776
Contact Persons:
Richard Stevens Director of Product Development 1-408-782-2720
Jennifer Downing Manager of Quality & Research 1-408-782-2720
Trade Name:
Common Name:
Ultrasound Transducer Cover/Sheath/Drape
SheathesISO Ultrasound Transducer Covers
Classification Name:
Ultrasonic Diagnostic Transducer Accessories
5-1
{1}------------------------------------------------
Equivalence:
510(K) K013721, Synthetic Polvisoprene Ultrasound Transducer Cover, Civco Medical Instruments
510(K) K991236, NeoFlex™ Ultrasound Transducer Cover, Civco Medical Instruments
### Labeling and Usage:
The following information will be found on each box/bag . (See Attachment E):
- Proprietary name 1 ..
- Quantity of sheathes packed 2.
- Name and Location of Manufacturer 3.
- 4. Sterile/NS (if applicable)
- 5. Expiration date
- 6. Size of probe cover
- 7. Caution statements: Follow manufacturer's instructions for cleaning and disinfecting the transducer.
- Prescription Statement: "Caution: 8. Federal law restricts this device to sale by or on the order of a physician or a practitioner trained in its use."
The Sheathing Technologies, Inc CIS-Isoprene Ultrasound Transducer Cover provides a thin, conformal covering to fit various and specific ultrasound transducer shapes. The cover is made by dipping to form a general cylindrical shape. It is closed at the proximal end and open at the distal end for insertion of the ultrasound transducer. This device is an accessory used on diagnostic probes to help minimize cross contamination.
The cover material is a Cis-1,4-polyisoprene, a synthetic version of natural rubber latex made without using natural rubber latex in the formulation. Various sizes and shapes of covers are offered to address the variations in Ultrasound Transducer configurations.
Product categories/models include
Device Description:
5-2
{2}------------------------------------------------
- 1. General Purpose SheathesISO Ultrasound Transducer Covers (sterile and non-sterile)
Covers are packaged in both sterile and nonsterile, individually wrapped or in bag/box quantities, as well as in "procedure kit" form for single patient/procedure, disposable use. The "procedure kit" may include elastic fasteners, with or without coupling gel packets. Transducer covers are also combined with disposable needle guide devices into kits.
Substantial Equivalence:
Non-Clinical Tests:
The Sheathing Technologies, Inc. CIS-Polyisoprene Ultrasound transducer cover is identified as substantially equivalent to CIVCO's current, legally marketed NeoFlex™ Ultrasound Transducer Covers and CIVCO's Synthetic Polyisoprene Ultrasound Transducer Cover.
- 1. Bench Testing:
- a. Viral Penetration Testing
- b. Stretch testing
- c. Pinhole testina
- d. Gel Compatibility
- 2. Biocompatibility
- a. Material Mediated Pyrogen
- b. Cytotoxicity
- c. Systemic Toxicity
- d. Hemolysis
- e. Irritation
- Sensitization f.
Conclusions from Non-Clinical Tests: Sheathing Technologies's CIS-Polyisoprene ultrasound transducer cover is compatible with ultrasound gel, has sufficient strength and elasticity for the intended application, and is biocompatible according to
the ISO 10993-1:2003 biocompatibility standard for a surface-contacting device that is in contact with the body for less than 24 hours.
ર-3
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features an abstract design with three upward-pointing lines that resemble stylized flames or ribbons. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the design.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Ms. Jennifer Downing Manager of Quality & Research Sheathing Technologies, Inc. 18431 Technology Drive MORGAN HILL CA 95037
SEP 2 8 2009
Re: K091828
Trade/Device Name: SheathesISO Ultrasound Transducer Covers Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasound transducer Regulatory Class: II Product Code: ITX Dated: August 28, 2009 Received: September 1, 2009
Dear Ms. Downing:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{4}------------------------------------------------
Page 2 -
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR. Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Janine M. Morris
Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Section 4: Indications for Use Statement
510(k) Number (if known):
K091828
Device Name: SheathesISO Ultrasound Transducer Covers
Indication For Use: Synthetic polyisoprene probe covers are indicated for use during diagnostic ultrasound procedures in cases where patient sensitivity to natural rubber latex is speculative.
Prescription Use X (21 CFR Part 801 Subpart D) Subpart C)
And/Or
Over the Counter Use (21 CFR Part 801
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K0911828
4-1
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.