The NeoFlex™ Ultrasound Transducer Cover is a protective cover or sheath placed over diagnostic ultrasound transducer / probe / scanhead instruments. The cover allows use of the transducer in scanning and needle guided procedures for body surface, endocavity, and intra-operative diagnostic ultrasound, while helping to prevent transfer of microorganisms, body fluids, and material to the patient and healthcare worker during reuse of the transducer (both sterile and non-sterile covers). The cover also provides a means for maintenance of a sterile field (sterile covers only). NeoFlex™ is latex-free and therefore beneficial when treating a patient with known type I hypersensitivity, or for the healthcare worker who has become type I sensitized. Transducer covers are furnished sterile & non-sterile; single use patient / procedure, disposable.
Device Story
NeoFlex™ is a thin, conformal, one-piece polychloroprene (neoprene) synthetic rubber sheath; designed to fit various ultrasound transducer geometries. Device acts as a physical barrier; prevents transmission of pathogens, body fluids, and particulate material between patient and transducer during reuse. Used in clinical settings for body surface, endocavity, and intra-operative ultrasound procedures; including needle-guided interventions. Sterile versions maintain sterile fields. Clinicians apply cover to transducer before scanning; disposable after single use. Benefits include latex-free composition; eliminates risk of type I natural latex protein allergen exposure for sensitized patients or staff. Provides strength and elasticity comparable to latex; does not impair ultrasound imaging.
Clinical Evidence
Bench testing only. Biocompatibility testing (ISO 10993-1) confirmed non-toxic, non-sensitizing, non-irritating, non-hemolytic, and non-pyrogenic properties. Barrier effectiveness validated using a protocol adapted from surgical glove/endoscope sheath testing, demonstrating resistance to bloodborne pathogen penetration. Strength and elasticity testing confirmed performance equivalent to latex predicate during application, scanning, and needle guide attachment.
Technological Characteristics
Material: Polychloroprene (neoprene) synthetic rubber. Form factor: One-piece, closed-end, rolled sheath in various geometries. Thickness: 0.0065". Sterilization: 100% EtO (where applicable) per ANSI/AAMI/ISO 11135. Biocompatibility: ISO 10993-1. Manufacturing: Dip-molding/vulcanization in Class 10,000 cleanroom.
Indications for Use
Indicated for use as a protective barrier over diagnostic ultrasound transducers during body surface, endocavity, and intra-operative scanning and needle-guided procedures. Suitable for patients and healthcare workers with type I natural latex hypersensitivity. Single-use, disposable.
Regulatory Classification
Identification
A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.
{1}------------------------------------------------
JUN 2 1999
CIVCO
MEDICAL INSTRUMENTS
K991236 (5 pages)
# 510(k) SUMMARY OF SAFETY & EFFECTIVENESS
(As required by 21 CFR 807.92)
# NeoFlex™ Ultrasound Transducer Cover
## A. General Information
| Submitter's Name: | CIVCO Medical Instruments Company, Inc. |
|-------------------|------------------------------------------------|
| Address: | 102 First Street South, Kalona, IA 52247 |
| Telephone No.: | phone (319) 656-4447 fax: (319) 656-4451 |
| Contact Person: | J. William Jones, Manager - Regulatory Affairs |
1937223 Establishment Reqistration Number: CIVCO Medical Instruments is registered as a medical device manufacturer.
| Device Trade: | NeoFlex™ Ultrasound Transducer Cover |
|-----------------------------|----------------------------------------------|
| Device Common: | Ultrasound Transducer Cover / Sheath / Drape |
| Device Classification Name: | Ultrasonic Diagnostic Transducer Accessories |
Classification: Classification Panel: Classification Procode: Class II under 21 CFR 892.1570 Radiology 90 ITX
Performance Standards: No applicable performance standards have been issued under Section 514 of the Food, Drug and Cosmetic Act.
#### B. Device Description
The NeoFlex™ Ultrasound Transducer Cover device provides a thin, conformal covering to fit various & specific ultrasound transducer geometries. Device is manufactured as a one-piece design that provides a covering that helps prevent the transmission of pathogens as the ultrasound transducer is reused from one patient to another.
Cover material is polychloroprene (neoprene) synthetic rubber similar to that of nonlatex medical examination / surgical gloves. Type I natural latex allergy does not occur in response to polychloroprene synthetic rubber since the synthetic rubber does not contain the natural protein allergen residuals present in latex.
Page II - 1
#### CIVCO North America
02 First Street South Kalona, IA 52247-9589 USA Phone: 319.656.4447 Fax: 319.656.4451 www.civcomedical.com
Image /page/0/Picture/18 description: The image contains two logos. The logo on the left is a circle with the text "QUALITY ASSURED FIRM" around the top half of the circle. Inside the circle is the text "ISO 9001" with a checkmark below it. The text "SGS" is below the checkmark. The logo on the right is a square with a crown on top. Inside the square is a checkmark and the text "NATIONAL ACCREDITATION OF CERTIFICATION BODIES".
Image /page/0/Picture/19 description: The image is a black and white logo for a company that is quality assured. The logo is circular in shape, with the words "QUALITY ASSURED FIRM" written around the top half of the circle. The letters "EN" are written above the number "46001" in the center of the circle. A check mark is below the number, and the letters "SGS" are at the bottom of the circle.
#### IV CO Europe
Avenue Louise 65, box 11 1050 Brussels Belgium Phone: +32(02)535.7881 Fax: +32(02)535.7700 www.civcomedical.com
{2}------------------------------------------------
Various sizes and shapes of covers are offered in order to customize the fit to specific transducer geometries. Product categories / models include:
General Purpose NeoFlex™ Transducer Covers (sterile and non-sterile)
Endocavity NeoFlex™ Transducer Covers (sterile and non-sterile)
Surgi-Tip™ Intraoperative* Transducer Covers (sterile) *polyethylene cord cover w/ NeoFlex ™ tip
Covers are packaged in both sterile "procedure kit" form for single patient / procedure, disposable use. Cover kits are supplied with fasteners, and with or without coupling gel packet. Transducer covers are also combined with disposable needle guide devices into custom kits that CIVCO builds for ultrasound OEMs and end users.
### C. Intended Use / Indications for Use
The NeoFlex™ Ultrasound Transducer Cover is a protective cover or sheath placed over diagnostic ultrasound transducer / probe / scanhead instruments. The cover allows use of the transducer in scanning and needle guided procedures for body surface, endocavity, and intra-operative diagnostic ultrasound, while helping to prevent transfer of microorganisms, body fluids, and material to the patient and healthcare worker during reuse of the transducer (both sterile and non-sterile covers). The cover also provides a means for maintenance of a sterile field (sterile covers only). NeoFlex™ is latex-free and therefore beneficial when treating a patient with known type I hypersensitivity, or for the healthcare worker who has become type I sensitized. Transducer covers are furnished sterile & non-sterile; single use patient / procedure, disposable.
The intended use and indications for use place NeoFlex™ Ultrasound Transducer Covers in device body contact categories as follows:
- a) surface devices, intact skin / mucosal membranes / breached surfaces, limited contact duration (< 24 hours)
- external communicating devices, tissue communicating, b) limited contact duration (< 24 hours)
#### D. Predicate Device
The NeoFlex™ Ultrasound Transducer Cover device is identified as substantially equivalent to CIVCO Medical's currently, legally marketed Latex Ultrasound Transducer Covers:
| Predicate Device(s) | 510(k) Reference | Manufacturer | Parameter | NeoFlex™ Ultrasound Transducer<br>Cover | Predicate Device<br>CIVCO Latex Ultrasound<br>Transducer Cover (K970515) |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| and and the program and the management of the commended to the first of the first of the first of the first of the first of the first of the first of the first and | and and and the property of the commended to the controlled in the contrôleant of the complex of the complex of | and and the first for the mannel of the program and any first and and on | Intended Use /<br>Indications for<br>Use | Same.<br>Additionally,<br>NeoFlex™ is non-latex and<br>therefore beneficial when treating a<br>patient with known type I<br>hypersensitivity, or for the healthcare<br>worker who has become type I<br>sensitized. | Provides a thin, conformal protective<br>cover system for diagnostic ultrasound<br>transducer usage in body surface,<br>endocavity, and intra-operative patient<br>environments; helps to prevent transfer<br>of microorganisms, body fluids, and<br>particulate material to the patient and<br>healthcare worker during reuse of the<br>transducer, and helps to maintain the<br>sterile field where applicable;<br>disposable device - for single patient /<br>procedure use. |
| Latex Ultrasound Transducer Cover | K970515 | CIVCO Medical | Design | Same. | One-piece, closed end, rolled (condom<br>style) with various dimensional<br>configurations necessary to<br>accommodate differences in ultrasound<br>transducer geometries. |
| Material | Polychloroprene, synthetic rubber materials used in compounding meet the recommended safe levels as specified in the US Food and Drugs Administration CFR, Title 21, Section 177.2600 and 182.5991. USP Absorbable Dusting Powder. synthetic rubber does not contain the natural protein allergen residuals present in latex. | Latex, natural rubber materials used in compounding meet the recommended safe levels as specified in the US Food and Drugs Administration CFR, Title 21, Section 177.2600 and 182.5991. USP Absorbable Dusting Powder. | | | |
| Manufacturing | Same. | dip-molding / rubber vulcanization. packaged in class 10,000 cleanroom per Federal Std 209e. packaging system per ANSI / AAMI / ISO 11607. | | | |
### E. Substantial Equivalence Summary
The NeoFlex™ Ultrasound Transducer Cover is substantially equivalent in safety and effectiveness to the CIVCO Latex Ultrasound Transducer Cover. The comparison table on the following page demonstrates this substantial equivalence.
{3}------------------------------------------------
# Comparison of Device to Substantially Equivalent, Legally Marketed Device
{4}------------------------------------------------
## Comparison of Device to Substantially Equivalent, Legally Marketed Device cont.
| Parameter | NeoFlex™ Ultrasound Transducer Cover | Predicate Device<br>CIVCO Latex Ultrasound Transducer Cover (K970515) |
|------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Quality Systems | Same. | FDA/QSR cGMP 21CFR Part 820. ISO 9001 / EN46001 / ISO 13485. |
| Sterility | Same. | sterilization (when applicable) by 100% EtO method; validated ANSI / AAMI / ISO 11135. SAL 10-6. |
| Device Body Contact Category | Same. | surface devices, intact skin / mucosal membranes / breached surfaces; limited contact duration (< 24 hours) External communicating devices, tissue communicating; limited contact duration (< 24 hours) |
| Safety | Biocompatibility tests for acute systemic toxicity, irritation, sensitization, hemolysis, material mediated pyrogen, and ethylene oxide sterilization residuals have demonstrated the NeoFlex™ polychloroprene material / cover device is: non-toxic. non-sensitizing. non-irritating. non-hemolytic. non-pyrogenic. Testing is in accordance with - ISO 10993-Part 1 Biological Evaluation of Medical Devices, FDA Blue Book Memorandum #G95-1, and FDA-Good Laboratory Practices (GLP).<br>Type I natural latex allergy does not occur in response to polychloroprene synthetic rubber. | Biocompatibility tests for acute systemic toxicity, irritation, sensitization, hemolysis, material mediated pyrogen, and ethylene oxide sterilization residuals have demonstrated the latex material / cover device is: non-toxic. non-sensitizing. non-irritating. non-hemolytic. non-pyrogenic. Testing is in accordance with - ISO 10993-Part 1 Biological Evaluation of Medical Devices, FDA Blue Book Memorandum #G95-1, and FDA-Good Laboratory Practices (GLP). |
| Parameter | NeoFlex™ Ultrasound Transducer<br>Cover | Predicate Device<br>CIVCO Latex Ultrasound<br>Transducer Cover (K970515) |
| Effectiveness | Testing for NeoFlex™ polychloroprene<br>covers has shown that the material is<br>adequate for the intended use:<br>strength and elastic characteristics<br>are effectively similar to that of latex<br>and allows use without tearing or<br>pinholing the cover - a) during<br>application and removal of cover<br>from transducer, b) during scanning<br>under intended uses, and c)<br>attaching / removing a disposable<br>needle guide to the transducer<br>bracket over the cover. same nominal thickness of .0065". ultrasound imaging is not impaired. NeoFlex™ polychloroprene<br>transducer cover provides an<br>effective barrier to the prevention of<br>microbial migration - tested under<br>protocol adapted from that used to<br>evaluate the barrier properties /<br>resistance of surgical gloves and<br>endoscope sheaths to penetration by<br>bloodborne pathogens using viral<br>penetration as a test system. polychloroprene (neoprene)<br>material is used for medical<br>examination / surgical gloves. | Experience and testing has shown that<br>latex natural rubber covers:<br>latex has sufficient strength and<br>elasticity for the intended<br>application. nominal thickness is .0065". do not impair ultrasound imaging. are an effective barrier to the<br>prevention of microbial migration. |
{5}------------------------------------------------
#### Comparison of Device to Substantially Equivalent, Legally Marketed Device
cont.
### F. Conclusions
This promarket submission for the NeoFlex™ Ultrasound Transducer Cover has demonstrated Substantial Equivalence as defined and understood in the Federal Food, Drug and Cosmetic act and various guidance documents issued by the Center for Devices and Radiological Health.
{6}------------------------------------------------
Public Health Service
Image /page/5/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its body and wings. The bird is facing to the right. Encircling the bird is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
J. William Jones Regulatory Affairs Manager Civco Medical Instruments Company, Inc. 102 Fist Street South Kalona, Iowa 55247
RE:
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
K991236 NeoFlex Ultrasound Transducer Cover Dated: April 8, 1999 Received: April 12, 1999 Regulatory Class: II 21 CFR 892.892.1570/Procode: 90 ITX
Dear Mr. Jones:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been redassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive. Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
#### INDICATIONS FOR USE STATEMENT
Page 1 of 1
K991236 510(k) Number (if known):
NeoFlex™ Ultrasound Transducer Cover Device Name:
Indications For Use:
Protective cover or sheath placed over diagnostic ultrasound transducer / probe / scanhead instruments. The cover allows use of the transducer in scanning and needle quided procedures for body surface, endocavity, and intra-operative diagnostic ultrasound, while helping to prevent transfer of microorganisms, body fluids, and particulate material to the patient and healthcare worker during reuse of the transducer (both sterile and non-sterile covers). The cover also provides a means for maintenance of a sterile field (sterile covers only). CIVCO NeoFlex™ Ultrasound Transducer Covers are furnished sterile & non-sterile: single use patient / procedure, disposable.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | |
|--------------------------------------------------------------------|---------|
| Division of Reproductive, Abdominal, ENT, and Radiological Devices | |
| 510(k) Number | K991236 |
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use (Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.