GENERAL PURPOSE LATEX TRANSDUCER COVERS

K970515 · CIVCO Medical Instruments Co., Inc. · ITX · Jun 20, 1997 · Radiology

Device Facts

Record IDK970515
Device NameGENERAL PURPOSE LATEX TRANSDUCER COVERS
ApplicantCIVCO Medical Instruments Co., Inc.
Product CodeITX · Radiology
Decision DateJun 20, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1570
Device ClassClass 2

Indications for Use

Protective cover or sheath placed over diagnostic ultrasound transducer / probe / scanhead instruments. The cover allows use of the transducer in scanning and needle guided procedures for body surface, endocavity, and intra-operative diagnostic ultrasound, while helping to prevent transfer of microorganisms, body fluids, and particulate material to the patient and healthcare worker during reuse of the transducer (both sterile and non-sterile covers). The cover also provides a means for maintenance of a sterile field (sterile covers only). CIVCO Latex Transducer Covers are furnished sterile & non-sterile; single use patient / procedure, disposable.

Device Story

Device is a thin, conformal, dip-molded natural rubber latex sheath designed to fit various ultrasound transducer geometries. It acts as a physical barrier to prevent transmission of pathogens, body fluids, and particulates between patients and healthcare workers during diagnostic ultrasound procedures. Applied by unrolling over the transducer; secured with bands if necessary. Used in clinical settings for body surface, endocavity, and intra-operative imaging. Facilitates maintenance of a sterile field when sterile versions are used. Does not impair ultrasound imaging. Disposable, single-use. Benefits include infection control and protection of both patient and clinician.

Clinical Evidence

No clinical trials performed. Evidence based on bench testing, including biocompatibility (acute systemic toxicity, irritation, sensitization, hemolysis, material-mediated pyrogen) per ISO 10993-1 and FDA Blue Book #G95-1. Barrier effectiveness validated via independent laboratory testing using a viral penetration protocol adapted from surgeon's glove standards to demonstrate resistance to bloodborne pathogens.

Technological Characteristics

Material: Natural Rubber Latex (conforming to 21 CFR 177.2600). Form factor: One-piece, dip-molded, beaded/rolled sheath. Sterilization: Ethylene Oxide (EtO) for sterile models. Manufacturing: Class 10,000 cleanroom. Biocompatibility: Non-toxic, non-sensitizing, non-irritating, non-hemolytic, non-pyrogenic.

Indications for Use

Indicated for use as a protective barrier over diagnostic ultrasound transducers during body surface, endocavity, and intra-operative scanning and needle-guided procedures. Intended for single-patient use to prevent cross-contamination between patients and healthcare workers. Suitable for use on intact skin, mucosal membranes, and breached surfaces.

Regulatory Classification

Identification

A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.

Predicate Devices

Submission Summary (Full Text)

{1} K970515 CIVCO MEDICAL INSTRUMENTS Solutions for Ultrasound B2 of B4 JUN 20 1997 510 (k) SUMMARY CIVCO Latex Ultrasound Transducer Cover Date Summary Prepared: 30 January 1997 This summary of the safety and effectiveness information upon which the substantial equivalence determination is based is being submitted in accordance with the requirements of SMDA 1990. Submitter's Name: CIVCO Medical Instruments Company, Inc. Address: 102 First Street South, Kalona, IA 52247 Telephone No.: (319) 656-4447 fax: (319) 656-4451 Contact Person: J. William Jones, Manager - Regulatory Affairs Establishment Registration Number: 1937223 CIVCO Medical Instruments is registered as a medical device manufacturer. Device Trade / Proprietary Name: CIVCO Latex Ultrasound Transducer Cover Device Common / Usual Name: Ultrasound Transducer Cover / Sheath / Drape Device Classification Name: Ultrasonic Diagnostic Transducer Accessories Classification: Class II under 21 CFR 892.1570 Classification Panel: 90 Radiology Classification Procode: ITX Description of Predicate Device(s): The CIVCO Latex Ultrasound Transducer Cover is equivalent to CIVCO's currently, legally marketed Latex Ultrasound Sheath, 510(k) reference number K895614. ORIGINALS ARE CODED RED sheet 1 of 3 102 First Street South • Kalona, IA 52247 USA Phone: 319/656-4447 • Fax: 319/656-4451 E-mail: info@civcomedical.com ![img-0.jpeg](img-0.jpeg) {2} 510 (k) SUMMARY con't. CIVCO Latex Ultrasound Transducer Cover B3 of B4 Description of Subject Device Submitted for Premarket Notification: Latex Ultrasound Transducer Cover - this device provides an efficient, thin, conformal covering to fit various & specific ultrasound transducer geometries. Devices are manufactured as a one-piece design that provides a covering that helps prevent the transmission of pathogens from one patient to another. Covers are beaded and rolled for easy application to the transducer. Cover materials are dip-molded Natural Rubber Latex similar to that of patient / surgical gloves. Ultrasound imaging is not impaired by use of the cover as it is intended. Adequate coupling between the cover and the transducer is required - the sterile or non-sterile Latex Ultrasound Transducer Cover is utilized by applying sterile transmission, coupling, or lubricating gel onto the transducer face or into closed end of cover, inserting ultrasound transducer into closed end of cover and unrolling cover over length of the transducer as desired, and securing open end of cover with bands as necessary. The removal process is accomplished by pulling or unrolling the cover off the transducer in a reverse method from the application, resulting in the cover being turned inside out. Various sizes and shapes of covers are offered in order to customize the fit to specific transducer geometries. Product categories / models include: - General Purpose Latex Transducer Covers (sterile and non-sterile) - PROcovers™ Endocavity Latex Transducer Covers (sterile and non-sterile) - Endocavity Latex Transducer Covers (sterile and non-sterile) - Bulk Packaged Endocavity Latex Transducer Covers (non-sterile) - Endocavity Accessories - non-sterile latex fingercots - Transesophageal Echocardiography (TEE) Covers (sterile) Covers are packaged in both sterile and non-sterile "procedure kit" form for single patient / procedure, disposable use. Cover kits are supplied with fasteners, and with or without coupling gel packet. Latex transducer covers are also combined with disposable needle guide devices that CIVCO custom kits for ultrasound OEMs. Intended Uses / Indications for Use: Protective cover or sheath placed over diagnostic ultrasound transducer / probe / scanhead instruments. The cover allows use of the transducer in scanning and needle guided procedures for body surface, endocavity, and intra-operative diagnostic ultrasound, while helping to prevent transfer of microorganisms, body fluids, and particulate material to the patient and healthcare worker during reuse of the transducer (both sterile and non-sterile covers). The cover also provides a means for maintenance of a sterile field (sterile covers only). CIVCO Latex Transducer Covers are furnished sterile & non-sterile; single use patient / procedure, disposable. The intended use and indications for use place CIVCO Latex Ultrasound Transducer Covers in device body contact categories as follows: a) surface devices, intact skin / mucosal membranes / breached surfaces, limited contact duration (< 24 hours) b) external communicating devices, blood path indirect / tissue communicating, limited contact duration (< 24 hours) sheet 2 of 3 ORIENTAL CARE CODED SED {3} 510 (k) SUMMARY con't. CIVCO Latex Ultrasound Transducer Cover B4 of B4 # Comparison of Device to Substantially Equivalent, Legally Marketed Device(s): **Intended Use:** both provide a thin, conformal protective cover system for ultrasound transducer usage in body surface, endocavity, and intra-operative patient environments; both help to prevent transfer of microorganisms, body fluids, and particulate material to the patient and healthcare worker during reuse of the transducer, and help maintain the sterile field where applicable; both are disposable devices - for single patient / procedure use. **Design:** both are one-piece, open on one end, closed on other end with various dimensional configurations necessary to accommodate differences in ultrasound transducer geometries. Covers are externally applied to ultrasound transducer. **Material:** both are Natural Rubber Latex; material formulations conform to US Food and Drugs Administration CFR, Title 21, Section 177.2600, and are manufactured under FDA - CGMP under 21 CFR, Part 820. **Manufacturing:** manufacturing processes - dip-molding, packaging (in class 10,000 cleanroom), and sterilization (when applicable) are same as for predicate device. **Safety:** materials and manufacturing processing (including EtO sterilization) affects to the healthcare worker and patient via intended use / indications for use contact of this device have been biologically evaluated using biocompatibility tests for acute systemic toxicity, irritation, sensitization, hemolysis, and material mediated pyrogen. Testing is in accordance with ISO 10993-Part 1 Biological Evaluation of Medical Devices, FDA Blue Book Memorandum #G95-1, and FDA-Good Laboratory Practices (GLP). CIVCO Latex Ultrasound Transducer Cover products / materials have been evaluated for safe use under device categories of limited contact duration and body contact for surface devices (skin / mucosal membranes / breached surfaces) and body contact for external communicating devices (blood path indirect / tissue communicating). Biocompatibility testing was conducted using sterilized (where applicable), finished devices. Testing has demonstrated subject materials / devices to be non-toxic, non-sensitizing, non-irritating, non-hemolytic, and non-pyrogenic. **Effectiveness:** physical / mechanical properties of the finished device are the same as for that of the predicate device; material strength and elasticity is adequate to allow use without tearing or pinholing the cover - a) during application and removal of cover from transducer, b) during scanning under intended uses, and c) attaching / removing a disposable needle guide to the transducer bracket over the cover; natural rubber latex materials have been effectively used in CIVCO ultrasound transducer covers applications for the past seven (7) years; CIVCO latex transducer covers have been tested by an independent laboratory under protocol adapted from that used to evaluate the barrier properties / resistance of surgeons glove materials to penetration by bloodborne pathogens using viral penetration as a test system. This testing has demonstrated that the latex transducer cover is an effective barrier to the prevention of microbial migration. sheet 3 of 3 ORIGINALS ARE CODED RED {4} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUN 20 1997 J. William Jones Manager, Regulatory Affairs CIVCO Medical Instruments Company, Inc. Medical Instruments 102 First Street South Kalona, IA 52247 Re: K970515 CIVCO Latex Ultrasound Transducer Cover Dated: June 2, 1997 Received: June 3, 1997 Regulatory class: II 21 CFR 892.1570/Procode: 90 ITX Dear Mr. Jones: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4591 for Radiology devices, or 594-4613 for Ear, Nose and Throat devices. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, ![img-1.jpeg](img-1.jpeg) Enclosure {5} 510(k) Notification - CIVCO Latex Ultrasound Transducer Cover page C3 # INDICATIONS FOR USE STATEMENT Page 1 of 1 510(k) Number (if known): K970515 Device Name: CIVCO Latex Ultrasound Transducer Cover Indications For Use: Protective cover or sheath placed over diagnostic ultrasound transducer / probe / scanhead instruments. The cover allows use of the transducer in scanning and needle guided procedures for body surface, endocavity, and intra-operative diagnostic ultrasound, while helping to prevent transfer of microorganisms, body fluids, and particulate material to the patient and healthcare worker during reuse of the transducer (both sterile and non-sterile covers). The cover also provides a means for maintenance of a sterile field (sterile covers only). CIVCO Latex Transducer Covers are furnished sterile & non-sterile; single use patient / procedure, disposable. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) ![img-2.jpeg](img-2.jpeg) (Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices 510(k) Number K970515 Prescription Use ☑ (Per 21 CFR 801.109) OR Over-The-Counter Use ☐ (Optional Format 1-2-96) 0000000000000000000000000000000000000000000000000000000000
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...