K260294 · Surgical Fusion Technologies GmbH · MAI · Mar 27, 2026 · Orthopedic
Device Facts
Record ID
K260294
Device Name
SF Push-in Anchor
Applicant
Surgical Fusion Technologies GmbH
Product Code
MAI · Orthopedic
Decision Date
Mar 27, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The SF Push-in Anchor is intended to be used for suture or tissue fixation in the foot, ankle, hand, wrist, elbow, knee, hip and shoulder. The SF Push-in Anchor is designed only to be inserted with the SupraFuser® Generator.
Device Story
SF Push-in Anchor is a bioresorbable implant for soft tissue-to-bone reattachment. System includes anchors (1.6-3.6mm), SupraFuser ultrasonic generator, handpiece, sonotrode, and surgical instruments (drills, stoppers, reamers). During surgery, surgeon inserts anchor into predrilled bone hole; ultrasonic energy from SupraFuser liquefies Poly-L-lactide-co-D,L-lactide polymer at bone interface. Liquid polymer flows into cancellous bone marrow space, quenching immediately to provide mechanical fixation. Device used in orthopedic procedures; operated by surgeons. Output is secure suture fixation for soft tissue. Benefits include bioresorbable material and stable fixation via ultrasonic welding.
Clinical Evidence
Bench testing only. Data includes gamma sterilization validation, packaging validation, shelf-life validation, and anchor pull-out testing to demonstrate mechanical integrity and substantial equivalence.
Technological Characteristics
Implants made of bioresorbable Poly-L-lactide-co-D,L-lactide (Resomer LR 706 S). Diameters: 1.6, 2.3, 3.0, 3.6mm. Energy source: Ultrasonic (SupraFuser generator). Instruments: Stainless steel (1.4197) drills/reamers, PEEK stoppers/front covers. Sterilization: Gamma radiation (VDmax 25 kGy). Connectivity: Standalone system.
Indications for Use
Indicated for suture or tissue fixation in foot, ankle, hand, wrist, elbow, knee, hip, and shoulder. Specific procedures include ligament reconstruction (scapholunate, carpal, collateral, lateral/medial stabilization, metatarsal), tendon repair/reconstruction (flexor, extensor, biceps, Achilles, patellar, posterior oblique), arthroplasty (carpometacarpal, bunionectomy), and labral/capsular repairs (Bankart, SLAP, acetabular).
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
K143063 — Stryker SonicPin System, Stryker SonicAnchor System · Stryker Trauma GmbH · Jan 27, 2015
K082906 — UNITY BETA ANCHOR SYSTEM · Bonutti Research, Inc. · Oct 30, 2008
K032072 — LACTOSORB L15 SCREW ANCHOR WITH LACTOCARBONATE SUTURE · Arthrotek, Inc. · Oct 1, 2003
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
March 27, 2026
Surgical Fusion Technologies GmbH
% Kelliann Payne
Partner
Hogan Lovells US LLP
1735 Market St.
23rd Floor
Philadelphia, Pennsylvania 19103
Re: K260294
Trade/Device Name: SF Push- in Anchor
Regulation Number: 21 CFR 888.3030
Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories
Regulatory Class: Class II
Product Code: MAI, HTY, GAT
Dated: January 29, 2026
Received: January 29, 2026
Dear Kelliann Payne:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260294 - Kelliann Payne
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K260294 - Kelliann Payne
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
CHRISTOPHER FERREIRA -S
Christopher Ferreira, M.S.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K260294 | ? |
| Please provide the device trade name(s). | | ? |
| SF Push-in Anchor | | |
| Please provide your Indications for Use below. | | ? |
| The SF Push-in Anchor is intended to be used for suture or tissue fixation in the foot, ankle, hand, wrist,
elbow, knee, hip and shoulder. The SF Push-in Anchor is designed only to be inserted with the
SupraFuser® Generator. | | |
| SF Push-in Anchor 1.6mm:
Hand/Wrist: Scapholunate Ligament Reconstruction, Carpal Ligament Reconstruction, Repair/
Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP and MCP
joints for all digits, digital tendon transfers, Carpometacarpal joint arthroplasty (basal thumb joint
arthroplasty)
Elbow: Ulnar or Radial Collateral Ligament Reconstruction,
Shoulder: Bankart Repair, SLAP Lesion Repair, Capsular Shift or Capsulolabral Reconstruction,
Foot/Ankle: Lateral Stabilization, Medial Stabilization, Metatarsal Ligament Repair, Hallux Valgus
reconstruction, Digital Tendon Transfers, Mid-foot Reconstruction,
Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair. | | |
| SF Push-in Anchor 2.3mm:
Hand/Wrist: Scapholunate Ligament Reconstruction, Carpal Ligament Reconstruction, Repair/
Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP and MCP
joints for all digits, digital tendon transfers, Carpometacarpal joint arthroplasty (basal thumb joint
arthroplasty), Triangular Fibrocartilage Complex (TFCC)
Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction, Lateral
epicondylitis repair,
Shoulder: Rotator Cuff Repairs, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-
Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction,
Foot/Ankle: Lateral Stabilization, Medial Stabilization, Metatarsal Ligament Repair, Hallux Valgus
reconstruction, Digital Tendon Transfers, Mid-foot Reconstruction, Achilles Tendon Repair, Bunionectomy,
Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair,
Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis,
Hip: Acetabular labral repair, capsular repair. | | |
| SF Push-in Anchor 3.0mm and 3.6mm:
Hand/Wrist: Scapholunate Ligament Reconstruction, Carpal Ligament Reconstruction, Repair/
Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP and MCP
joints for all digits, digital tendon transfers, Carpometacarpal joint arthroplasty (basal thumb joint
arthroplasty), Triangular Fibrocartilage Complex (TFCC)
Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction, lateral
epicondylitis repair,
Shoulder: Rotator Cuff Repairs, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-
Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction, | | |
{4}
Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Metatarsal Ligament Repair, Hallux Valgus reconstruction, Digital Tendon Transfers, Mid-foot Reconstruction, Metatarsal tendon repair, Bunionectomy,
Knee: Medial collateral ligament repair, lateral collateral ligament repair, patellar tendon repair, posterior oblique ligament repair, and iliotibial band tenodesis,
Hip: Acetabular labral repair, capsular repair.
Please select the types of uses (select one or both, as applicable).
☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
?
{5}
K260294
1 of 10
# Surgical Fusion Technologies GmbH's SF-Push-in Anchors 510(k) Summary
## Submitter's Name, Address, Telephone Number, Contact Person
Surgical Fusion Technologies GmbH
Wagistrasse 6 Schlieren, Switzerland
Phone: +41 44 204 60 18
Facsimile: +41 44 204 61 2820
Contact Person: Joerg Mayer, CTO
Date Prepared: January 29, 2026
## Name of Device and Name
SF Push-in Anchor
## Common or Usual Name
Fastener, fixation, biodegradable, soft tissue (MAI), class II Pin, fixation, smooth (HTY), class II
Suture, nonabsorbable, synthetic, polyethylene (GAT), class II
## Classification Name
Single/multiple component metallic bone fixation appliances and accessories (21 CFR 888.3030)
Smooth or threaded metallic bone fixation fastener (21 CFR 888.3040) Nonabsorbable poly(ethylene terephthalate) surgical suture (21 CFR 878.5000)
## Predicate Devices
K240288 Surgical Fusion SF-Push-in Anchor (Primary Predicate)
K183091 SpineWelding Elaris Pedicle Screw System (Material reference)
## Indications for Use
The SF Push-in Anchor is intended to be used for suture or tissue fixation in the foot, ankle, hand, wrist, elbow, knee, hip and shoulder. The SF Push-in Anchor is designed only to be inserted with the SupraFuser Generator.
## SF Push-in Anchor 1.6mm:
Hand/Wrist: Scapholunate Ligament Reconstruction, Carpal Ligament Reconstruction, Repair/Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the
{6}
K260294
2 of 10
PIP, DIP and MCP joints for all digits, digital tendon transfers, Carpometacarpal joint arthroplasty (basal thumb joint arthroplasty)
Elbow: Ulnar or Radial Collateral Ligament Reconstruction
Shoulder: Bankart Repair, SLAP Lesion Repair, Capsular Shift or Capsulolabral Reconstruction
Foot/Ankle: Lateral Stabilization, Medial Stabilization, Metatarsal Ligament Repair, Hallux Valgus reconstruction, Digital Tendon Transfers, Mid-foot Reconstruction
Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair
## SF Push-in Anchor 2.3mm:
Hand/Wrist: Scapholunate Ligament Reconstruction, Carpal Ligament Reconstruction, Repair/Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP and MCP joints for all digits, digital tendon transfers, Carpometacarpal joint arthroplasty (basal thumb joint arthroplasty), Triangular Fibrocartilage Complex (TFCC)
Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction, Lateral epicondylitis repair
Shoulder: Rotator Cuff Repairs, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction
Foot/Ankle: Lateral Stabilization, Medial Stabilization, Metatarsal Ligament Repair, Hallux Valgus reconstruction, Digital Tendon Transfers, Mid-foot Reconstruction, Achilles Tendon Repair, Bunionectomy
Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis
Hip: Acetabular labral repair, capsular repair
## SF Push-in Anchor 3.0mm and 3.6mm:
Hand/Wrist: Scapholunate Ligament Reconstruction, Carpal Ligament Reconstruction, Repair/Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP and MCP joints for all digits, digital tendon transfers, Carpometacarpal joint arthroplasty (basal thumb joint arthroplasty), Triangular Fibrocartilage Complex (TFCC)
Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction, lateral epicondylitis repair
Shoulder: Rotator Cuff Repairs, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction
{7}
K260294
3 of 10
Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Metatarsal Ligament Repair, Hallux Valgus and Varus reconstruction, Digital Tendon Transfers, Mid-foot Reconstruction, Metatarsal tendon repair, Bunionectomy
Knee: Medial collateral ligament repair, lateral collateral ligament repair, patellar tendon repair, posterior oblique ligament repair, and iliotibial band tenodesis,
Hip: Acetabular labral repair, capsular repair
## Technological Characteristics
The SF Push-in Anchor system consists of implantable anchors with 1.6, 2.3, 3.0 and 3.6mm diameter, an ultrasonic system, a sonotrode, dedicated anchor size specific drills and stoppers, a handpiece front cover, a guide with integrated wrench, and a set of reamers.
The SF Push-in Anchors are delivered in a dedicated Implant Case to facilitate suture loading and suture handling during implantation.
The SF Push-in Anchors are made of biocompatible and fully bioresorbable Poly-L-lactide-co-D,L-lactide. The SF Push-in Anchors are fully bioresorbable implants designed for soft tissue reattachment to bone by means of suture materials. The implantation process employs ultrasonic energy to liquefy the polymeric components of the Push-in Anchors at the interface with bone tissue. The liquid polymer flows into the marrow space of the surrounding cancellous bone, where it is immediately quenched and provides anchorage of the implant.
The ultrasonic energy for the implantation of the Push-in Anchors is produced by the SupraFuser® ultrasonic generator and applied via the attached handpiece. The sonotrode is mounted on the handpiece. It transmits the ultrasonic vibrations to the Push-in Anchor. The drills and the stoppers are dedicated to be used with the Push-in Anchor of the respective size. The Geomax Reamer is used for the SF Push-in Anchors 1.6, 2.3 and 3.0 in cases of thick cortical bone or oblique insertion.
## Performance Data
The following tests were performed to demonstrate the substantial equivalence of the SF Push-in Anchor:
- Gamma sterilization validation
- Packaging validation
- Shelf-life validation
- Anchor pull out testing
{8}
K260294
4 of 10
# Substantial Equivalence
The SF Push-in Anchors are as safe and effective as the primary predicate SF Push-in Anchors. The SF Push-in Anchors have the same intended uses and similar indications, technological characteristics, and principles of operation. The minor technological differences between the subject and predicate SF Push-in Anchors and its predicate devices, summarized in the table below, raise no new issues of safety or effectiveness. Performance data demonstrate that the SF Push-in Anchors are as safe and effective as the predicate. Thus, the SF Push-in Anchors are substantially equivalent.
| Company | Surgical Fusion Technologies | Surgical Fusion Technologies |
| --- | --- | --- |
| Device | SF- Push-in Anchor Portfolio | SF- Push-in Anchor Portfolio |
| FDA 510(K) | Subject submission | K240288 |
| Indications | **SF Push-in Anchor 1.6mm:**
**Hand/Wrist:** Scapholunate Ligament Reconstruction, Carpal Ligament Reconstruction, Repair/Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP and MCP joints for all digits, digital tendon transfers, Carpometacarpal joint arthroplasty (basal thumb joint arthroplasty),
**Elbow:** Ulnar or Radial Collateral Ligament Reconstruction
**Shoulder:** Bankart Repair, SLAP Lesion Repair, Capsular Shift or Capsulolabral Reconstruction
**Foot/Ankle:** Lateral Stabilization, Medial Stabilization, Metatarsal Ligament Repair, Hallux Valgus reconstruction, Digital Tendon Transfers, Mid-foot Reconstruction
**Knee:** Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair | **SF Push-in Anchor 1.6mm:**
**Hand/Wrist:** Scapholunate Ligament Reconstruction, Carpal Ligament Reconstruction, Repair/Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP and MCP joints for all digits, digital tendon transfers
**Elbow:** Ulnar or Radial Collateral Ligament Reconstruction
**Shoulder:** Bankart Repair, SLAP Lesion Repair, Capsular Shift or Capsulolabral Reconstruction
**Foot/Ankle:** Lateral Stabilization, Medial Stabilization, Metatarsal Ligament Repair, Hallux Valgus reconstruction, Digital Tendon Transfers, Mid-foot Reconstruction
**Knee:** Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair |
| | | **SF Push-in Anchor 2.3mm:**
**Hand/Wrist:** Scapholunate Ligament Reconstruction, Carpal Ligament Reconstruction, Repair/Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP and MCP joints for all digits, digital tendon transfers |
{9}
K260294
5 of 10
| Company | Surgical Fusion Technologies | Surgical Fusion Technologies |
| --- | --- | --- |
| | **SF Push-in Anchor 2.3mm:**
**Hand/Wrist:** Scapholunate Ligament Reconstruction, Carpal Ligament Reconstruction, Repair/Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP and MCP joints for all digits, digital tendon transfers, Carpometacarpal joint arthroplasty (basal thumb joint arthroplasty), Triangular Fibrocartilage Complex (TFCC)
**Elbow:** Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction, Lateral epicondylitis repair
**Shoulder:** Rotator Cuff Repairs, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction
**Foot/Ankle:** Lateral Stabilization, Medial Stabilization, Metatarsal Ligament Repair, Hallux Valgus reconstruction, Digital Tendon Transfers, Mid-foot Reconstruction, Achilles Tendon Repair, Bunionectomy
**Knee:** Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis
**Hip:** Acetabular labral repair, capsular repair
**SF Push-in Anchor 3.0mm and 3.6mm:**
**Hand/Wrist:** Scapholunate Ligament Reconstruction, Carpal Ligament Reconstruction, Repair/Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP and MCP joints for all digits, digital tendon transfers, Carpometacarpal joint | **Elbow:** Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction, Lateral epicondylitis repair
**Shoulder:** Rotator Cuff Repairs, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction
**Foot/Ankle:** Lateral Stabilization, Medial Stabilization, Metatarsal Ligament Repair, Hallux Valgus reconstruction, Digital Tendon Transfers, Mid-foot Reconstruction, Achilles Tendon Repair, Bunionectomy
**Knee:** Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis
**Hip:** Acetabular labral repair, capsular repair
**SF Push-in Anchor 3.0mm and 3.6mm:**
**Hand/Wrist:** Scapholunate Ligament Reconstruction, Carpal Ligament Reconstruction, Repair/Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP and MCP joints for all digits, digital tendon transfers
**Elbow:** Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction, lateral epicondylitis repair
**Shoulder:** Rotator Cuff Repairs, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction
**Foot/Ankle:** Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Metatarsal Ligament Repair, Hallux Valgus reconstruction, Digital Tendon Transfers, Mid-foot Reconstruction, Metatarsal tendon repair, Bunionectomy |
{10}
K260294
6 of 10
| Company | Surgical Fusion Technologies | Surgical Fusion Technologies |
| --- | --- | --- |
| | arthroplasty (basal thumb joint arthroplasty), Triangular Fibrocartilage Complex (TFCC)
**Elbow:** Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction, lateral epicondylitis repair
**Shoulder:** Rotator Cuff Repairs, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction
**Foot/Ankle:** Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Metatarsal Ligament Repair, Hallux Valgus and Varus reconstruction, Digital Tendon Transfers, Mid-foot Reconstruction, Metatarsal tendon repair, Bunionectomy
**Knee:** Medial collateral ligament repair, lateral collateral ligament repair, patellar tendon repair, posterior oblique ligament repair, and iliotibial band tenodesis,
**Hip:** Acetabular labral repair, capsular repair | **Knee:** Medial collateral ligament repair, lateral collateral ligament repair, patellar tendon repair, posterior oblique ligament repair, and iliotibial band tenodesis,
**Hip:** Acetabular labral repair, capsular repair |
| **Device Diameter/Length** | 1.6mm/5.2mm
2.3mm/7.2mm
3.0mm/8.7mm
3.6mm/10.2mm | 1.6mm/5.2mm
2.3mm/7.2mm
3.0mm/8.7mm
3.6mm/10.2mm |
| **Material** | PLDLLA: Resomer® LR 706 S | PLDLLA: Resomer® LR 706 S |
{11}
K260294
7 of 10
| Company | Surgical Fusion Technologies | Surgical Fusion Technologies |
| --- | --- | --- |
| Manufacturing | Injection molding at Surgical Fusion Technologies GmbH | Injection molding at Samaplast AG |
| Sterilization | VDmax 25 kGy Gamma radiation | VDmax 17.5 kGy Gamma radiation |
| Acceptance Level for Molecular Weight (Inherent Viscosity) post sterilization | 1.0 cm^{3}/g | 1.2 cm^{3}/g |
| Sutures | Supplied with, or recommended sutures
Non-resorbable polyethylene terephthalate sutures or UHMWPE sutures or a combination of UHMWPE and polyester sutures | Supplied with, or recommended sutures
Non-resorbable polyethylene terephthalate sutures or UHMWPE sutures or a combination of UHMWPE and polyester sutures |
| Insertion Method | Insert anchor into predrilled hole while applying US energy | Insert anchor into predrilled hole while applying US energy |
| Fixation Methods | Ultrasonic melting of the polymer into the porous cancellous bone | Ultrasonic melting of the polymer into the porous cancellous bone |
| Length of ultrasound energy delivery | Not more than 6 seconds | Not more than 6 seconds |
| Packaging | Primary: Implant in Implant Case in Blister
Secondary: Blister in Aluminum Pouch
Tertiary: Product box | Primary: Implant in Aluminium Pouch
Secondary: PE pouch
Tertiary: Product box |
{12}
K260294
8 of 10
| Company | Surgical Fusion Technologies | Surgical Fusion Technologies |
| --- | --- | --- |
| **Shelf-life** | 3 years (test in process for 7 years) | 7 years |
| **SupraFuser Ultrasonic System** | | |
| **Components** | Ultrasound generator
Handpiece
Sonotrode S-PI | Ultrasound generator
Handpiece
Sonotrode S-PI |
| **Power Source** | 100-240V AC ±10%
Max. Power consumption: 50VA | 100-240V AC ±10%
Max. Power consumption: 50VA |
| **Software** | V3.1 | V2.5 |
| **Sterilizable Instruments** | | |
| **Drill Sizes** | Drill S-PI A1.6
Drill S-PI A2.3
Drill S-PI A3.0
Drill S-PI A3.6
Drill S-PI A.1.6 extra short (length 5mm instead of 6.9mm in the Drill S-PI A1.6) | Drill S-PI A1.6
Drill S-PI A2.3
Drill S-PI A3.0
Drill S-PI A3.6 |
| **Drill Material** | Stainless steel (1.4197 medical/X20CrNiMos13-1)
Additional Laser marking on tip | Stainless steel (1.4112) |
| **Manual Reamer** | Geomax Reamer S-PI-A1.6
Geomax Reamer S-PI-A2.3
Geomax Reamer S-PI-A3.0 | Not available |
{13}
K260294
9 of 10
| Company | Surgical Fusion Technologies | Surgical Fusion Technologies |
| --- | --- | --- |
| Reamer Material | Reamer - Stainless steel grade 1.4197
Handle – Medical grade silicone with SS core | Nor available |
| Stopper/Front Cover | Stopper S-PI-A1.6
Stopper S-PI-A2.3
Stopper S-PI-A3.0
Stopper S-PI-A3.6
Front Cover | Stopper S-PI-A1.6
Stopper S-PI-A2.3
Stopper S-PI-A3.0
Stopper S-PI-A3.6
Front Cover |
| Stopper/Front Cover Material | Medical grade PEEK | Medical grade PEEK |
{14}
K260294
10 of 10
## Conclusions
The SF Push-in Anchors are substantially equivalent to the predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.