Stryker SonicPin System, Stryker SonicAnchor System
Applicant
Stryker Trauma GmbH
Product Code
HTY · Orthopedic
Decision Date
Jan 27, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Stryker SonicPin System is intended to maintain alignment and fixation of bone fractures, osteotomies or bone grafts in hallux valgus applications in the presence of appropriate immobilization (e.g. rigid fixation implants, cast and brace). The Stryker SonicPin is designed only to be inserted with the SonicFusion equipment. The Stryker SonicAnchor System is intended to be used for suture or tissue fixation in open procedures in the foot, ankle, knee, hand, wrist, elbow and shoulder. The Stryker SonicAnchor is designed only to be inserted with the SonicFusion equipment.
Device Story
System comprises bioresorbable PLDLLA pin or anchor implants and SonicFusion equipment; SonicFusion uses ultrasonic energy to liquefy polymer implant during insertion, creating fixed interface with host bone. SonicAnchor includes Teleflex Force Fiber suture (UHMWPE/PP). Used in open orthopedic procedures; operated by surgeons. SonicFusion equipment includes ultrasonic generator, handpiece, footswitch, drill, and depth gauge. Output is mechanical fixation of bone or soft tissue. Benefits include secure fixation via ultrasonic welding process. MR safety evaluated.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including mechanical performance (tensile strength, dynamic fatigue), biocompatibility (ISO 10993), electrical safety/EMC (IEC 60601-1/1-2), and software verification/validation.
Indicated for bone fracture/osteotomy/graft fixation in hallux valgus (SonicPin) and suture/tissue fixation to cancellous bone in foot, ankle, knee, hand, wrist, elbow, and shoulder (SonicAnchor). Includes Achilles tendon repair, lateral/medial stabilization, hallux valgus/midfoot reconstruction, metatarsal ligament repair, digital tendon transfer, acromio-clavicular separation repair, proximal deltoid repair, ulnar/radial collateral ligament reconstruction, patellar tendon repair, scapholunate ligament reconstruction, carpal ligament reconstruction, and collateral ligament repair/reconstruction.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 27, 2015
Stryker Trauma GmbH Ms. Estela Celi Senior Regulatory Affairs Specialist 325 Corporate Mahwah, New Jersey 07430
Re: K143063 Trade/Device Name: Stryker SonicPin System SonicAnchor System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: HTY, MAI, GAT Dated: October 22, 2014 Received: October 29, 2014
Dear Ms. Celi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing
{1}------------------------------------------------
Page 2 - Ms. Estela Celi
(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K143063
Device Name Stryker SonicPin System
Indications for Use (Describe)
The Stryker SonicPin System is intended to maintain alignment and fixation of bone fractures, osteotomies or bone grafts in hallux valgus applications in the presence of appropriate immobilization (e.g. rigid fixation implants, cast and brace). The Stryker SonicPin is designed only to be inserted with the SonicFusion equipment.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW .*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
#### K143063
Device Name Stryker Sonic Anchor System
Indications for Use (Describe)
The Stryker SonicAnchor System is intended to be used for suture or tissue fixation in open procedures in the foot, ankle, knee, hand, wrist, elbow and shoulder. The Stryker SonicAnchor is designed only to be inserted with the SonicFusion equipment.
Indications Include:
| Foot/Ankle | |
|------------------------------------------------------|--|
| • Achilles Tendon Repair | |
| • Lateral Stabilization | |
| • Medial Stabilization | |
| • Hallux Valgus Reconstruction | |
| • Midfoot Reconstruction | |
| • Metatarsal Ligament Repair | |
| • Digital Tendon Transfer | |
| Shoulder | |
| • Acromio-Clavicular Separation Repair | |
| • Proximal Deltoid Repair | |
| Elbow | |
| • Ulnar or Radial Collateral Ligament Reconstruction | |
| Knee | |
| • Patellar Tendon Repair | |
| Hand/Wrist | |
| • Scapholunate Ligament Reconstruction | |
| • Carpal Ligament Reconstruction | |
| • Repair/Reconstruction of Collateral Ligaments | |
Type of Use (Select one or both, as applicable)
| <span style="text-decoration: underline;"></span> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
|----------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{4}------------------------------------------------
# 510(k) Summary
| Submitter: | Stryker Trauma GmbH<br>Prof. Kuetscher Str. 1-5<br>24232 Schönkirchen<br>Germany<br>Phone: (201) 831-6461<br>Fax: (201) 831-3461 |
|------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Estela Celi<br>Senior Regulatory Affairs Specialist |
| Device<br>Name of Device: | Stryker SonicPin System/SonicAnchor System |
| Common Name: | Stryker SonicPin System- Smooth Fixation Pin |
| | Stryker SonicAnchor System-Smooth Fixation Pin/Fastener<br>& Suture |
| Classification Name: | Stryker SonicPin System<br>Smooth or threaded metallic bone fixation fastener<br>21 CFR §888.3040 |
| | Stryker SonicAnchor System<br>Single/multiple component metallic bone fixation<br>appliances and accessories.<br>21 CFR §888.3030<br>Nonabsorbable poly(ethylene terephthalate) surgical<br>suture<br>21 CFR §878.5000 |
| Regulatory Class:<br>Product Code: | II<br>Stryker SonicPin System: HTY<br>Stryker SonicAnchor System: MAI, GAT |
| Predicate Devices: | Stryker SonicPin System<br>Primary Predicate: Stryker SonicPin System-K091955 |
| | Stryker SonicAnchor System<br>Primary Predicate-Arthrex 2.5 mm PushLock -K063479<br>Reference Device- Stryker SonicPin System-K091955 |
| Date Prepared: | October 22, 2014 |
{5}------------------------------------------------
## Description:
## Stryker SonicPin System
This Traditional 510(k) submission is being supplied to the U.S. FDA to seek clearance for modifications made to the Stryker SonicPin System, with regard to the SonicFusion equipment and included in the previously cleared Stryker SonicPin System under K091995 and the addition of accessories. The Stryker SonicPin System consists of a sterile bioresorbable pin implant, the SonicFusion equipment and associated accessories intended for the internal fixation of bone fractures and fragments. The pin is made of PLDLLA (Poly(L-lactide-co-D.L-lactide) which has been evaluated for magnetic resonance safety and is implanted using SonicFusion technology which is a process that employs ultrasonic energy to liquefy a polymer and facilitate a fixed interface between the implant and host bone.
The associated accessories include:
- SonicFusion equipment
- . Drill
- Handpiece Tip
- TipTool ●
- Direct Depth Gauge ●
## Stryker SonicAnchor System
This Traditional 510(k) submission is being supplied to the U.S. FDA to provide authorization to market the new Stryker SonicAnchor System. The Stryker SonicAnchor System consists of a bioresorbable anchor with suture implant, the SonicFusion equipment and associated accessories intended for the fixation of suture or soft tissue to cancellous bone. The implant is comprised of the anchor, made of PLDLLA (Poly(L-lactide-co-D,L-lactide) which has been evaluated for magnetic resonance safety and a Teleflex Force Fiber suture, made of braided ultrahigh molecular weight polyethylene (UHMWPE) and polypropylene (PP) strands.
- The associated accessories include: ● SonicFusion equipment
- Drill ●
- Handpiece Tip
- TipTool ●
## Intended Use:
## Stryker SonicPin System
The Stryker SonicPin System is intended to maintain alignment and fixation of bone fractures, osteotomies or bone grafts in hallux valgus applications in the presence of appropriate immobilization (e.g. rigid fixation implants, cast and brace). The Stryker SonicPin is designed only to be inserted with the SonicFusion equipment.
Both the subject and predicate have the same intended use for the internal fixation of bone fractures and fragments.
{6}------------------------------------------------
#### Stryker SonicAnchor System
The Stryker SonicAnchor System is intended to be used for suture or tissue fixation in open procedures in the foot, ankle, knee, hand, wrist, elbow and shoulder. The Stryker SonicAnchor is designed only to be inserted with the SonicFusion equipment.
Both the subject and predicate have the same intended use for the fixation of suture or soft tissue to cancellous bone.
#### Indications for Use:
#### Stryker SonicPin System
The Stryker SonicPin System is intended to maintain alignment and fixation of bone fractures, osteotomies or bone grafts in hallux valgus applications in the presence of appropriate immobilization (e.g. rigid fixation implants, cast and brace). The Stryker SonicPin is designed only to be inserted with the SonicFusion equipment.
#### Stryker SonicAnchor System
The Stryker SonicAnchor System is intended to be used for suture or tissue fixation in open procedures in the foot, ankle, knee, hand, wrist, elbow and shoulder. The Stryker SonicAnchor is designed only to be inserted with the SonicFusion equipment. Indications Include:
#### Foot/Ankle
- . Achilles Tendon Repair
- Lateral Stabilization ●
- Medial Stabilization ●
- Hallux Valgus Reconstruction ●
- Midfoot Reconstruction ●
- Metatarsal Ligament Repair ●
- Digital Tendon Transfer
## Shoulder
- Acromio-Clavicular Separation Repair ●
- Proximal Deltoid Repair
#### Elbow
- Ulnar or Radial Collateral Ligament Reconstruction ●
## Knee
- . Patellar Tendon Repair
## Hand/Wrist
- Scapholunate Ligament Reconstruction
- Carpal Ligament Reconstruction ●
- Repair/Reconstruction of Collateral Ligaments ●
{7}------------------------------------------------
# Comparison of Technological Characteristics with the Predicate Device:
## Stryker SonicPin System
SonicFusion is the technological principle for both the subject Stryker SonicPin System and the predicate Stryker SonicPin System. SonicFusion technology is based on the application of controlled ultrasonic energy which liquefies the polymer implant as it is placed in the bone. At a high level, the subject and predicate devices are based on the following same technological elements:
- . Use of a polymer pin implant
- SonicFusion technology used for implant application ●
The following differences exist between the subject and predicate devices:
- Sonic Pin implant material has been evaluated for magnetic resonance safety ●
- New SonicFusion ultrasonic generator, handpiece and footswitch
- Use of interchangeable tips on the SonicFusion handpiece
- Ultrasound delivered has a decreased amplitude
- Added footswitch containing power switch to SonicFusion equipment ●
# Stryker SonicAnchor System
The Stryker SonicAnchor System and the predicate device are both intended to be used for suture or tissue fixation to bone.
At a high level, the subject and predicate devices require the following same technological elements:
- Surgical site preparation
- . Use of a drill for surgical site preparation
- Use of a suture ●
The following differences exist between the subject and predicate devices:
- Use of different fixation methods
# Performance Data:
## Mechanical Testing
# Stryker SonicPin System
Mechanical testing was performed to evaluate the implantation process with the modifications made to the SonicFusion equipment.
- . Tensile Strength Testing
Additionally the magnetic resonance was evaluated as per FDA Guidance for Industry and FDA Staff - Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment, 2008.
## Stryker SonicAnchor System
Testing was completed to evaluate mechanical performance of the subject device compared against the predicate.
- Dynamic Fatigue Strength Testing
{8}------------------------------------------------
Additionally the magnetic resonance safety and ultrasound heat safety were evaluated as per FDA Guidance for Industry and FDA Staff - Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment, 2008.
## Biocompatibility Testing
The biocompatibility evaluation for the Stryker SonicPin System and Stryker SonicAnchor System was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing,"" May 1, 1995, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The battery of testing included the following tests:
# Stryker SonicPin System
- Rationales for Irritation and Sensitization testing
- Cytotoxicity ●
- . Biological Compatibility Risk Assessment
## Stryker SonicAnchor System
- Rationales for Irritation and Sensitization testing ●
- Cytotoxicity
- Genotoxicity ●
- Biodegradation ●
- . Biological Compatibility Risk Assessment
# Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the SonicFusion equipment used in both the Stryker SonicPin System and the Stryker SonicAnchor System, consisting of the ultrasonic generator, handpiece and footswitch. The SonicFusion equipment complies with the IEC 60601-1 standards for safety and the IEC 60601-1-2 Edition 3: 2007-03 standards for EMC.
## Software Verification and Validation Testing
Software verification and validation testing was completed on the SonicFusion equipment, used in both the Stryker SonicPin System and the Stryker SonicAnchor System, was conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern, since a failure or latent flaw in the software could directly result in serious injury or death to the patient or operator.
# Clinical Testing:
## Stryker SonicPin System
Clinical testing was not required to support substantial equivalence of the proposed Stryker SonicPin System to the predicate SonicPin System.
{9}------------------------------------------------
#### Stryker SonicAnchor System
Clinical testing was not required to support substantial equivalence of the Stryker SonicAnchor System in comparison to the predicate device.
#### Conclusion:
#### Stryker SonicPin System
Modifications made to the Stryker SonicPin System have been evaluated in comparison to the predicate device and non-clinical data have been used to support substantial equivalence. The hardware and software verification and validation for the SonicFusion equipment demonstrate that the subject Stryker SonicPin System should perform as intended in the specified use conditions.
#### Stryker SonicAnchor System
The non-clinical data for the Stryker SonicAnchor System including the SonicFusion equipment support the safety of the device and demonstrates that the Stryker SonicAnchor System performs comparably to the predicate device that is currently marketed for the same intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.