K171228 · Sportwelding GmbH · MAI · Jan 19, 2018 · Orthopedic
Device Facts
Record ID
K171228
Device Name
Fiji Anchor
Applicant
Sportwelding GmbH
Product Code
MAI · Orthopedic
Decision Date
Jan 19, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Fiji Anchor is intended to be used for suture or tissue fixation in the foot, ankle, hand, wrist, elbow, and shoulder. The Fiji Anchor is designed only to be inserted with the BoneWelder 150-W1 system. Specific indications are listed below: Elbow: Ulnar or Radial Collateral Ligament Reconstruction Shoulder: Bankart Repair, Capsular Shift or Capsulolabral Reconstruction Hand/Wrist: Scapholunate Ligament Reconstruction. Carpal Ligament Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP and MCP joints for all digits, digital tendon transfers Foot/Ankle: Lateral Stabilization, Metatarsal Ligament Repair, Hallux Valqus reconstruction, digital tendon transfers. Mid-foot reconstruction
Device Story
Fiji Anchor is a bioresorbable implant for soft tissue-to-bone reattachment; used in orthopedic surgeries (foot, ankle, hand, wrist, elbow, shoulder). System includes implantable anchor, BoneWelder 150-W1 ultrasonic generator, handpiece with sonotrode, and twist drill. Surgeon drills hole; inserts anchor; applies ultrasonic energy via handpiece for <6 seconds. Energy liquefies Poly-L-lactide-co-D,L-lactide (PLDLLA) polymer at bone interface; liquid polymer flows into cancellous bone marrow space; immediate quenching provides mechanical anchorage. Suture fixation achieved via manual knotting. Device benefits patient through bioresorbable material that metabolizes into H2O and CO2, eliminating need for permanent hardware removal. Used in clinical/OR settings by orthopedic surgeons.
Clinical Evidence
Bench testing only. Performance data includes: static and dynamic pullout testing (pre/post-aging), creep testing, material characterization (GPC, inherent viscosity, GCMS, ICP), cytotoxicity (ISO 10993-5), extraction/analysis (ISO 10993-18), temperature rise in porcine bone, and lot-to-lot endotoxin testing (USP <85>). Software validation and electrical safety (IEC 60601-1, 60601-1-2) were performed.
Technological Characteristics
Implant material: Poly-L-lactide-co-D,L-lactide (PLDLLA). Dimensions: 2.3 x 7.2 mm. Fixation: Ultrasonic energy-induced polymer liquefaction. Energy source: BoneWelder 150-W1 ultrasonic generator. Connectivity: Handpiece-to-generator. Sterilization: Sterile. Software: Embedded firmware for ultrasonic generator control.
Indications for Use
Indicated for patients requiring suture or tissue fixation in the foot, ankle, hand, wrist, elbow, and shoulder for ligament reconstruction, tendon repair, and stabilization procedures.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
January 19, 2018
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
SportWelding GmbH % Janice M. Hogan Regulatory Counsel Hogan Lovells US LLP 1835 Market Street, 29th Floor Philadelphia, Pennsylvania 19103
Re: K171228
Trade/Device Name: Fiji Anchor Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: MAI, HTY, GAT Dated: December 13, 2017 Received: December 13, 2017
Dear Ms. Hogan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Katherine D. Kavlock -S
for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known)
K171228 Device Name
Fiji Anchor
Indications for Use (Describe)
The Fiji Anchor is intended to be used for suture or tissue fixation in the foot, ankle, hand, wrist, elbow, and shoulder. The Fiji Anchor is designed only to be inserted with the BoneWelder 150-W1 system. Specific indications are listed below:
Elbow: Ulnar or Radial Collateral Ligament Reconstruction
Shoulder: Bankart Repair, Capsular Shift or Capsulolabral Reconstruction
Hand/Wrist: Scapholunate Ligament Reconstruction. Carpal Ligament Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP and MCP joints for all digits, digital tendon transfers
Foot/Ankle: Lateral Stabilization, Metatarsal Ligament Repair, Hallux Valqus reconstruction, digital tendon transfers. Mid-foot reconstruction
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
□ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.
{3}------------------------------------------------
## K171228
### 510(k) SUMMARY
### SportWelding's Fiji Anchor
# Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
Sport Welding GmbH Wagistrasse 6 Schlieren, Switzerland
Phone: +41 44 204 61 2821 Facsimile: +41 44 204 61 2820 Contact Person: Joerg Mayer, Managing Director, CTO
Date Prepared: January 17, 2018
### Name of Device and Name
Fiji Anchor
### Common or Usual Name
Fastener, fixation, biodegradable, soft tissue (MAI), class II
Pin, fixation, smooth (HTY), class II
Suture, nonabsorbable, synthetic, polyethylene (GAT), class II
#### Classification Name
Single/multiple component metallic bone fixation appliances and accessories (21 CFR 888.3030)
Smooth or threaded metallic bone fixation fastener (21 CFR 888.3040)
Nonabsorbable poly(ethylene terephthalate) surgical suture (21 CFR 878.5000)
#### Predicate Devices
- K063479 Arthrex's Bio-PushLock (Primary)
- . K143063 Stryker's SonicAnchor System (Reference)
### Indications for Use
The Fiji Anchor is intended to be used for suture or tissue fixation in the foot, ankle, hand, wrist, elbow, and shoulder. The Fiji Anchor is designed only to be inserted with the BoneWelder 150-W1 system. Specific indications are listed below:
{4}------------------------------------------------
Elbow: Ulnar or Radial Collateral Ligament Reconstruction
Shoulder: Bankart Repair, Capsular Shift or Capsulolabral Reconstruction
Hand/Wrist: Scapholunate Ligament Reconstruction, Carpal Ligament Reconstruction, Repair/Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP and MCP joints for all digits, digital tendon transfers
Foot/Ankle: Lateral Stabilization, Medial Stabilization, Metatarsal Liqament Repair, Hallux Valgus reconstruction, digital tendon transfers, Mid-foot reconstruction
# Technological Characteristics
The Fiji anchor system consists of an implantable anchor, an ultrasound system, and handpiece tip, and a twist drill.
The Fiji Anchor® is made of biocompatible and fully bioresorbable Poly-L-lactide-co-D,Llactide. The in vivo degradation of the Fiji Anchor® is based on the natural physiologic process of hydrolysis, which results in the complete metabolization of the polymer into H₂O and CO2.
The Fiji Anchor® is a fully bioresorbable implant designed for soft tissue reattachment to bone by means of suture material. The BoneWelding® technology employs ultrasonic energy to liquefy the polymeric components of the Fiji Anchor® at the interface with bone tissue. The liquid polymer flows into the marrow space of the surrounding cancellous bone where it is immediately quenched and provides anchorage of the implant.
The ultrasonic energy for the implantation of the Fiji Anchor® is produced by the BoneWelder® ultrasonic generator and applied via the attached handpiece. The handpiece tip (sonotrode) is mounted on the handpiece. It transmits the ultrasonic energy to the Fiji Anchor®. The twist drill is dedicated to be used with the Fiji Anchor®.
#### Performance Data
The following tests were performed to demonstrate the substantial equivalence of the Fiji Anchor:
- Software documentation and validation per FDA guidance
- . IEC 60601-1 and 60601-1-2 testing
- . Material characterization using GPC, inherent viscosity, GCMS, and ICP
- . Static pullout testing before and after aging
- . Dynamic pullout testing before and after aging
- Creep testing
- . Extraction and analysis per ISO 10993-18
{5}------------------------------------------------
- Cytotoxicity testing per ISO 10993-5 .
- . Temperature rise in surrounding bone (porcine bone)
- . Lot-to-lot endotoxin testing is also performed per USP <85>
# Substantial Equivalence
The Fiji Anchor is as safe and effective as the Arthrex's Bio-PushLock. The Fiji Anchor has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate device. The minor technological differences between the Fiji Anchor and its predicate devices, summarized in the table below, raise no new issues of safety or effectiveness. Performance data demonstrate that the Fiji Anchor is as safe and effective as Arthrex's Bio-PushLock. Thus, the Fiji Anchor is substantially equivalent.
| Property | SportWelding Fiji | Arthrex 2.5 mm Bio- | Stryker Sonic Anchor |
|--------------------------------------------|-------------------------------------------------------------------------|------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| | Anchor | PushLock | |
| 510(k) | Subject | K063497 | K143063 |
| Components | Anchor, suture, | PEEK eyelet, suture, | Anchor, suture, |
| | sonotrode, US generator | anchor | sonotrode, US generator |
| Size | 2.3 x 7.2 mm | 2.5 x 8 mm | 2.5 x 12 mm |
| Suture Sizes | #4-0, #3-0, #2-0, #0, #2 | #2-0, #0 | #2-0, #0, #2 |
| Insertion Method | Insert anchor into<br>predrilled hole while<br>applying US energy | Insert eyelet into<br>predrilled hole, then<br>secure with<br>bioresorbable anchor | Partially insert anchor<br>into predrilled hole, then<br>fully insert while<br>applying US energy |
| Fixation Method | Ultrasonic melting of the<br>polymer into the porous<br>cancellous bone | Barbs in a mechanical<br>press-fit | Ultrasonic melting of the<br>polymer into the porous<br>cancellous bone |
| Length of<br>ultrasound<br>energy delivery | Not more than 6<br>seconds | N/A | Not more than 12<br>seconds |
| Suture fixation | Manual knotting | Knotless | Manual knotting |
| Material | PLDLLA | PLLA and PEEK | PLDLLA |
| Resorbable | Yes | Partly | Yes |
| Sterilization | Sterile | Sterile | Sterile |
# Conclusions
The Fiji Anchor is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.