K190738 · Medacta International S.A. · PHX · Dec 13, 2019 · Orthopedic
Device Facts
Record ID
K190738
Device Name
MyShoulder Placement Guides
Applicant
Medacta International S.A.
Product Code
PHX · Orthopedic
Decision Date
Dec 13, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3660
Device Class
Class 2
Indications for Use
The MyShoulder guides are intended to be used as guides specifically designed for the anatomy of a single patient. The humeral guide is used to define the level and orientation of the humeral head resection cut. The glenoidal guide is used to position and orient the K-wire that will subsequently guide the glenoidal reaming. Both anatomical and reverse total joint configurations are acceptable. Both humeral and glenoid guides are suitable for a delto-pectoral approach only. MyShoulder Patient-Matched Guides are intended for use with Medacta Shoulder System and its cleared indications for use. MyShoulder Patient-Matched Guides are intended for single use only.
Device Story
Patient-specific surgical instruments; input: patient CT scan data; process: 3D surgical planning to create anatomical humeral and glenoidal guides; output: physical guides for humeral resection and K-wire positioning for glenoid reaming; used in clinical setting by surgeons; guides facilitate accurate implant placement for Medacta Shoulder System; benefits: improved surgical precision and alignment; single-use; non-sterile.
Clinical Evidence
No clinical studies were conducted. Evidence consists of non-clinical bench testing, including cadaver lab evaluations for accuracy and positioning variability, surgical evaluation, sterilization effect analysis on geometry, and wear testing.
Technological Characteristics
Material: Medical grade nylon for sintering (Polyamide PA 12). Form factor: Patient-specific anatomical guides. Connectivity: None. Sterilization: Provided non-sterile. Software: Used for 3D planning based on CT data.
Indications for Use
Indicated for patients undergoing shoulder joint replacement surgery (anatomical or reverse total joint configurations) requiring patient-specific humeral resection and glenoid reaming guidance via a delto-pectoral approach.
Regulatory Classification
Identification
A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
Predicate Devices
BLUEPRINT Patient Specific Instrumentation (K143374)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 13, 2019
Medacta International SA % Chris Lussier Director, Quality and Regulatory Medacta USA 3973 Delp Street MEMPHIS TN 38118
Re: K190738
Trade/Device Name: MyShoulder Placement Guides Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder Joint Metal/Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: PHX, KWS, HSD, MBF Dated: November 13, 2019 Received: November 14, 2019
Dear Chris Lussier:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Michael C. Owens Acting Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
### K190738
Device Name MyShoulder™ Placement Guides
### Indications for Use (Describe)
The MyShoulder guides are intended to be used as guides specifically designed for the anatomy of a single patient. The humeral guide is used to define the level and orientation of the humeral head resection cut. The glenoidal guide is used to position and orient the K-wire that will subsequently guide the glenoidal reaming.
Both anatomical and reverse total joint configurations are acceptable.
Both humeral and glenoid guides are suitable for a delto-pectoral approach only.
MyShoulder Patient-Matched Guides are intended for use with Medacta Shoulder System and its cleared indications for use.
MyShoulder Patient-Matched Guides are intended for single use only.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 2.0 510(k) Summary
# K190738
#### I. Submitter
Medacta International SA Strada Regina 6874 Castel San Pietro (CH) Switzerland Phone (+41) 91 696 60 60 Fax (+41) 91 696 60 66
Contact Person: Stefano Baj, Regulatory and Compliance Director, Medacta International SA Date Prepared: Mar 21, 2019 Date Revised: November 13, 2019
#### II. Device
| Device Proprietary Name: | MyShoulder™ Placement Guides |
|--------------------------|-------------------------------------------------------------------|
| Common or Usual Name: | Shoulder Prosthesis System |
| Classification Name: | Shoulder joint metal/polymer semi-constrained cemented prosthesis |
| Primary Product Code: | PHX |
| Secondary Product Code: | KWS, HSD, MBF |
| Regulation Number: | 21 CFR 888.3660, 21 CFR 888.3690, and 21 CFR 888.3670 |
| Device Classification | II |
#### III. Predicate Device
Substantial equivalence is therefore claimed to the following devices:
- > Primary Predicate: BLUEPRINT Patient Specific Instrumentation (K143374), Wright-Tornier
- Reference Device: A
- Anatomic Shoulder Prosthesis (K170910), Medacta International SA
- Shoulder Reverse System (K170452), Medacta International SA ●
- MySpine Pedicle Screw Placement Guides (132788), Medacta International SA ●
#### IV. Device Description
The MyShoulder" Placement Guides are patient specific surgical instruments that allow for guided shoulder implants placement of both Medacta Shoulder Reverse System (K170452) and Medacta Anatomic Shoulder Prosthesis (K170910) and the additional products cleared under K181826 (Glenoid Polyaxial Non-locking screws), K180089 (Short Humeral Stem) and K171058 (Threaded Glenoid Baseplate).
{5}------------------------------------------------
The MyShoulder™ Placement guides uses the patient's CT Scan to perform a 3D planning of the glenohumeral joint replacement surgery. Then, following this planning the humeral and glenoidal guides are created. The MyShoulder™ Placement guides are anatomical guides for a single patient anatomy.
The MyShoulder™ Placement guides subject of this submission are comprised of the following products:
- 1. MyShoulder™ humeral guide: it will allow to perform an accurate humeral resection
- 2. MyShoulder™ glenoidal guide: it will provide a precise alignment reference that will guide the glenoidal reaming and final implant positioning.
The MyShoulder™ Placement guides are manufactured from medical grade nylon for sintering (Polyamide PA 12) which is identical to Medacta reference device MySpine Pedicle Screw Placement Guides (132788). The MyShoulder™ Placement guides are single use, external communicating devices with limited (<24 hours) contact duration and are provided in non-sterile version only.
The MyShoulder™ Placement guides are substantially equivalent to BLUEPRINT Patient Specific K143374), Wright-Tornier, Medacta reference devices Shoulder Reverse System (K170452) Anatomic Shoulder Prosthesis (K170910), and MySpine Pedicle Screw Placement Guides (132788).
#### V. Indications for Use
The MyShoulder guides are intended to be used as guides specifically designed for the anatomy of a single patient. The humeral guide is used to define the level and orientation of the humeral head resection cut. The glenoidal guide is used to position and orient the K-wire that will subsequently guide the glenoidal reaming.
Both anatomical and reverse total joint configurations are acceptable.
Both humeral and glenoid guides are suitable for a delto-pectoral approach only.
MyShoulder Patient-Matched Guides are intended for use with Medacta Shoulder System and its cleared indications for use.
MyShoulder Patient-Matched Guides are intended for single use only.
#### VI. Comparison of Technological Characteristics
MyShoulder glenoid guides are substantially equivalent to Wright-Tornier BLUEPRINT Patient Specific Instrumentation (K143374) for what concerns indications for use, contraindications, design, material and manufacturing process.
MyShoulder guides have substantially equivalent indications for use and contraindications of the reference predicate devices Medacta Shoulder Reverse System (K170452) and Medacta Anatomic Shoulder Prosthesis (K170910).
The technological characteristics including materials, contact duration, software, manufacturing process, device usage, packaging, labeling, and shelf-life of the MyShoulder™ Placement guides are similar to those of the reference device MySpine Pedicle Screw Placement Guides (132788).
{6}------------------------------------------------
The safety and effectiveness of the MyShoulder™ Placement guides are adequately supported by the substantial equivalence information, materials information, and analysis data provided within this Premarket Notification.
#### Performance Data VII.
Risks were identified based on the proposed design and testing was conducted to mitigate those risks. Based on the risk analysis, testing was conducted according to written protocols with acceptance criteria. The following performance data was provided in support of the substantial equivalence determination:
Non-Clinical Studies:
- . Characterization Tests
- Cadaver Lab Test Report 1; o
- Surgical Evaluation Test Report 2; and o
- MyShoulder accuracy and positioning variability cadaver test Test Report 4 o
- MyShoulder accuracy test Test Report 5 o
- Performance Tests ●
- o Effects of the Sterilization on the geometry of the MyShoulder Glenoidal Guide -Test Report 3
- Wear Test, according to RPO-01.028.009 o
Clinical Studies:
- No clinical studies were conducted. .
## VIII. Conclusion
Based on the above information, the MyShoulder™ Placement guides are substantially equivalent to the identified predicate and reference devices.
Substantial equivalence has been demonstrated through a comparison of indication for use, design, and technological characteristics, as well as performance evaluations. The MyShoulder™ Placement guides are as safe and effective as the predicate device BLUEPRINT Patient Specific Instrumentation (K143374), Wright-Tornier and the reference devices Medacta Shoulder Reverse System (K170452), Medacta Anatomic Shoulder Prosthesis (K170910) and MySpine Pedicle Screw Placement Guides (K132788).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.