K211112 · Orthopaedic Implant Company · KTT · May 5, 2021 · Orthopedic
Device Facts
Record ID
K211112
Device Name
OIC External Fixation System
Applicant
Orthopaedic Implant Company
Product Code
KTT · Orthopedic
Decision Date
May 5, 2021
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Intended Use
The OIC External Fixation system is intended to be used in adult and pediatic patients for provisional fixation of open and/or unstable fractures in the lower and upper extremities and pelvis. It may also be used for temporary fixation of periarticular or intra-articular fractures. Additionally, the device can be used on fractures where soft tissue injury or an infected fracture site may preclude the use of other fracture fixation treatments.
Device Story
External fixation system for stabilization of bone fractures; consists of carbon fiber composite bars, titanium/aluminum clamps, and stainless steel pins. Pins implanted into bone; connected via clamps, rods, and posts to form rigid construct holding bone fragments in place. Used in clinical settings by surgeons for fracture management. Provides mechanical stability to bone fragments; facilitates healing in complex fracture cases where internal fixation is contraindicated due to soft tissue or infection. System includes various pin sizes and lengths to accommodate different patient anatomies.
Clinical Evidence
Bench testing only. Geometric comparison performed on 3mm pins against secondary predicate. CAD modeling and analysis conducted to evaluate MRI safety, supporting MR Conditional labeling.
Indicated for adult and pediatric patients requiring provisional or temporary fixation of open, unstable, periarticular, or intra-articular fractures of the lower/upper extremities and pelvis, including cases with soft tissue injury or infection.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
K183682 — OIC External Fixation System · Orthopaedic Implant Company · Apr 8, 2019
K171211 — OIC External Fixation System · Orthopaedic Implant Company · Sep 27, 2017
K242452 — Monkey Bars Pin to Bar External Fixation System · Paragon 28, Inc. · Nov 1, 2024
K141912 — AOS SMALL BONE EXTERNAL FIXATION, THREADED HALF PINS, AOS SMALL BONE EXTERNAL FIXATION, CARBON FIBER RODS, AOS SMALL BON · Advanced Orthopaedic Solutions, Inc. · Oct 29, 2014
K960014 — PRACTIFIX EXTERNAL FIXATION SYSTEM · Orthopaedic Innovations, Inc. · Mar 18, 1996
Submission Summary (Full Text)
{0}------------------------------------------------
May 5, 2021
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
Orthopaedic Implant Company Douglas Fulton Quality Assurance Manager 770 Smithridge Dr. #400 Reno. Nevada 89502
## Re: K211112
Trade/Device Name: OIC External Fixation System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: KTT Dated: April 12, 2021 Received: April 14, 2021
# Dear Douglas Fulton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Ting Song, Ph.D., R.A.C. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K21112
Device Name
OIC External Fixation System
#### Indications for Use (Describe)
The OIC External Fixation system is intended to be used in adult and pediatic patients for provisional fixation of open and/or unstable fractures in the lower and upper extremities and pelvis. It may also be used for temporary fixation of periarticular or intra-articular fractures. Additionally, the device can be used on fractures where soft tissue injury or an infected fracture site may preclude the use of other fracture fixation treatments.
Type of Use (Select one or both, as applicable)
× Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
{3}------------------------------------------------
#### 510(k) Summary
Prepared 4/12/2021
Name and Address of Manufacturer: The Orthopaedic Implant Company (OIC) 770 Smithridge Drive. Suite 400 Reno, NV 89502
Contact: Douglas Fulton Quality Assurance Manager Telephone: 775-636-8281 Fax: 775-636-8284 Email: doug@orthoimplantcompany.com
Device Identification: Trade Name: OIC External Fixation System Common Name: External fixation components Classification Name: Single/multiple component metallic bone fixation appliances and accessories Classification: Class II, 21 CFR 888.3030 Panel: Orthopedic Product Code: KTT
Indications for Use:
The OIC External Fixation system is intended to be used in adult and pediatric patients for provisional fixation of open and/or unstable fractures in the lower and upper extremities and pelvis. It may also be used for temporary fixation of peri-articular or intra-articular fractures. Additionally, the device can be used on fractures where soft tissue injury or an infected fracture site may preclude the use of other fracture fixation treatments.
Device Description:
The device is used for the external stabilization of bone fractures. It consists of:
Carbon fiber composite bars, 11mm diameter, 100mm to 650mm lengths
Titanium and aluminum combination clamp, 5 & 8 hole pin clamp
Aluminum straight and angled posts
Stainless steel 3mm, 4mm and 5mm pins, blunt tip and threaded 85mm through 250mm lengths, 10mm through 120mm thread lengths.
Stainless steel 5mm transfixing pin, 300mm length
Stainless steel instruments for implantation
The pins are implanted into bone and then they are connected using the clamps, rods and posts to form a rigid construct which holds the bone fragments rigidly in place. The pins are offered in various lengths and thicknesses.
Comparison of Technological Characteristics (Substantial Equivalence):
Predicate devices: K183682 OIC External Fixation System,
Secondary predicate device: K061493 Apex Pins.
The OIC External Fixation System has the following similarities to those which previously received 510(k) concurrence:
- · has the same indicated use,
- · uses the same operating principle,
- · incorporates the same design, and
- · incorporates the same or similar materials
#### Performance Testing:
A geometric comparison was performed on the OIC External Fixation System 3mm pins as compared to the secondary predicate. The comparison shows that the system has acceptable characteristics for the intended uses. CAD modeling and analysis were performed on the 3mm pins to evaluate the MRI safety of the new pins. The analysis indicates that these devices should be labeled MR Conditional using the same recommended MRI safety labeling as the OIC predicate.
#### Conclusion:
The OIC External Fixation System described in this substantially equivalent to the predicate devices.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.