OIC External Fixation System

K171211 · Orthopaedic Implant Company · KTT · Sep 27, 2017 · Orthopedic

Device Facts

Record IDK171211
Device NameOIC External Fixation System
ApplicantOrthopaedic Implant Company
Product CodeKTT · Orthopedic
Decision DateSep 27, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The OIC External Fixation system is intended to be used in adult and pediatic patients for provisional fixation of open and or unstable fractures in the lower and upper extremities and pelvis. It may also be used for temporary fixation of periarticular or intra-articular fractures. Additionally, the device can be used on fractures where soft tissue injury or an infected fracture site may preclude the use of other fracture fixation treatments.

Device Story

External fixation system for stabilization of bone fractures; consists of carbon fiber composite bars, titanium/aluminum clamps, and stainless steel pins/posts. Pins implanted into bone; connected via clamps, rods, and posts to form rigid construct holding bone fragments in place. Used by surgeons in clinical settings for provisional or temporary fracture fixation. Benefits include stabilization of fractures where soft tissue injury or infection precludes internal fixation.

Clinical Evidence

Bench testing only. FEA simulations and mechanical static axial load testing were performed in accordance with ASTM F1541-02 to evaluate clamping ability and structural integrity compared to predicate devices.

Technological Characteristics

Materials: carbon fiber composite (bars), titanium and aluminum (clamps), stainless steel (pins/posts). Components: 11mm diameter bars (100-600mm), 4mm/5mm pins, clamps, and posts. Mechanical fixation principle. Non-powered, manual assembly. Sterilization not specified.

Indications for Use

Indicated for adult and pediatric patients requiring provisional fixation of open/unstable fractures of lower/upper extremities and pelvis, or temporary fixation of peri-articular/intra-articular fractures, including cases with soft tissue injury or infected fracture sites.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 September 27, 2017 Orthopaedic Implant Company Douglas Fulton Ouality Assurance Manager 770 Smithridge Drive #400 Reno, Nevada 89502 Re: K171211 Trade/Device Name: OIC External Fixation System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: KTT Dated: August 2, 2017 Received: August 3, 2017 Dear Douglas Fulton: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR {1}------------------------------------------------ Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely. # Mark N. Melkerson -S Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use 510(k) Number (if known) K171211 Device Name OIC External Fixation System #### Indications for Use (Describe) The OIC External Fixation system is intended to be used in adult and pediatic patients for provisional fixation of open and or unstable fractures in the lower and upper extremities and pelvis. It may also be used for temporary fixation of periarticular or intra-articular fractures. Additionally, the device can be used on fractures where soft tissue injury or an infected fracture site may preclude the use of other fracture fixation treatments. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below. {3}------------------------------------------------ #### 510(k) Summary Prepared 8/1/2017 Name and Address of Manufacturer: The Orthopaedic Implant Company (OIC) 770 Smithridge Drive. Suite 400 Reno, NV 89502 Contact: Douglas Fulton Quality Assurance Manager Telephone: 775-636-8281 Fax: 775-636-8284 Email: doug@orthoimplantcompany.com Device Identification: Trade Name: OIC External Fixation System Classification Name: Single/multiple component metallic bone fixation appliances and accessories Classification: Class II, 21 CFR 888.3030 Panel: Orthopedic Product Code: KTT Indications for Use: The OIC External Fixation system is intended to be used in adult and pediatric patients for provisional fixation of open and/or unstable fractures in the lower and upper extremities and pelvis. It may also be used for temporary fixation of peri-articular or intra-articular fractures. Additionally, the device can be used on fractures where soft tissue injury or an infected fracture site may preclude the use of other fracture fixation treatments. Device Description: The device is used for the external stabilization of bone fractures. It consists of: Carbon fiber composite bars, 11mm diameter, 100mm to 600mm lengths Titanium and aluminum combination clamp. 5 & 8 hole pin clamp Aluminum straight and angled posts Stainless steel 4mm and 5mm pins, blunt tip and threaded 150mm lengths, 15mm through 120mm thread lengths. Stainless steel 5mm transfixing pin, 250mm length Stainless steel instruments for implantation The pins are implanted into bone and then they are connected using the clamps, rods and posts to form a rigid construct which holds the bone fragments rigidly in place. The pins are offered in various lengths and thicknesses. Comparison of Technological Characteristics (Substantial Equivalence): Predicate devices: K110965 Renovis T 710 Large External Fixation System K072212 Jet-X Bar System Clamps, Bars and Posts - MR Conditional K952730 Hoffman II External Fixation System K061493 Apex Pins K031428 Synthes Large External Fixation Clamps The OIC External Fixation System has the following similarities to those which previously received 510(k) concurrence: has the same indicated use, uses the same operating principle, incorporates the same design and incorporates the same or similar materials Performance Testing: FEA simulations were conducted using methods outlined in ASTM F1541-02 to ascertain that the clamping ability of the OIC clamps have acceptable mechanical characteristics for the intended uses. Mechanical static axial load testing performed per ASTM F1541-02 determined that the device will withstand greater loads than the predicate device. Conclusion: The OIC External Fixation System described in this substantially equivalent to the predicate devices.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...