Monkey Bars Pin to Bar External Fixation System

K242452 · Paragon 28, Inc. · KTT · Nov 1, 2024 · Orthopedic

Device Facts

Record IDK242452
Device NameMonkey Bars Pin to Bar External Fixation System
ApplicantParagon 28, Inc.
Product CodeKTT · Orthopedic
Decision DateNov 1, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The Monkey Bars Pin to Bar External Fixation System is intended to be used in adult and pediatric patients (specifically, children and adolescents aged 2 through 21 years) for provisional fixation of open and/or unstable fractures in the lower and upper extremities and pelvis. It may also be used for temporary fixation of peri-aticular fractures. Additionally, the device can be used on fractures where soft tissue injury or an infected fracture site may preclude the use of other fracture fixation treatments.

Device Story

External fixation system for stabilization of bone fractures; consists of carbon fiber composite bars, titanium combination clamps, aluminum multi-pin clamps, aluminum posts, and stainless steel pins. Pins implanted into bone; connected via clamps, rods, and posts to form rigid construct holding bone fragments in place. Used by surgeons in clinical settings for fracture management. Benefits patient by providing rigid stabilization for unstable or infected fracture sites where internal fixation is contraindicated.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including Minimal Construct Static Compression Testing.

Technological Characteristics

Materials: Carbon fiber composite (bars), titanium (combination clamp), aluminum (multi-pin clamps, posts), stainless steel (pins, instruments). Mechanical fixation system; non-powered. Components include 11mm diameter bars (150-500mm), 3-5mm diameter pins, and transfixing pins. Sterilization method not specified.

Indications for Use

Indicated for adult and pediatric patients (ages 2-21) for provisional fixation of open/unstable fractures of lower/upper extremities and pelvis; temporary fixation of peri-articular or intra-articular fractures; and fractures where soft tissue injury or infection precludes other fixation methods.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. November 1, 2024 Paragon 28, Inc. Edward Wells-Spicer Regulatory Affairs Specialist II 14445 Grasslands Drive Englewood, Colorado 80112 Re: K242452 Trade/Device Name: Monkey Bars Pin to Bar External Fixation System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: KTT Dated: August 16, 2024 Received: August 19, 2024 Dear Edward Wells-Spicer: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" {1}------------------------------------------------ (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {2}------------------------------------------------ Sincerely, Lixin Liu -S Lixin Liu, Ph.D Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K242452 Device Name Monkey Bars Pin to Bar External Fixation System ### Indications for Use (Describe) The Monkey Bars Pin to Bar External Fixation System is intended to be used in adult and pediatric patients (specifically, children and adolescents aged 2 through 21 years) for provisional fixation of open and/or unstable fractures in the lower and upper extremities and pelvis. It may also be used for temporary fixation of peri-aticular fractures. Additionally, the device can be used on fractures where soft tissue injury or an infected fracture site may preclude the use of other fracture fixation treatments. | Type of Use (Select one or both, as applicable) | <table><tr><td><div> <span> <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div></td></tr><tr><td><div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div></td></tr></table> | <div> <span> <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | |------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------| | <div> <span> <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | | | | | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | | | | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # 510(K) SUMMARY | Manufacturer: | Paragon 28, Inc.<br>14445 Grasslands Dr.<br>Englewood, CO 80112 | |-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact: | Edward Wells-Spicer<br>Regulatory Affairs Specialist II<br>Paragon 28, Inc.<br>14445 Grasslands Dr.<br>Englewood, CO 80112<br>Phone: 720-994-5481<br>ewspicer@paragon28.com | | Date Prepared: | October 31, 2024 | | Device Trade Name: | Monkey Bars Pin to Bar External Fixation System | | Device Class and Common Name: | Class II, External Fixation System | | Classification: | 21 CFR 888.3030: Single/multiple component metallic bone fixation appliances and accessories. | | Product Code: | KTT | | Indications for Use: | The Monkey Bars Pin to Bar External Fixation system is intended to be used in adult and pediatric patients (specifically, children and adolescents aged 2 through 21 years) for provisional fixation of open and/or unstable fractures in the lower and upper extremities and pelvis. It may also be used for temporary fixation of peri-articular or intra-articular fractures. Additionally, the device can be used on fractures where soft tissue injury or an infected fracture site may preclude the use of other fracture fixation treatments. | | Device Description: | The device is used for the external stabilization of bone fractures. It consists of: Carbon fiber composite bars, 11mm diameter by 150mm to 500mm lengths, titanium combination clamp, aluminum 5 & 8 hole multi-pin clamps, aluminum straight and 30 degree angled posts, stainless steel 3mm, 4mm | {5}------------------------------------------------ and 5mm thread diameter pins with 5mm diameter shanks that come in blunt tip and self tapping with 125mm through 230mm overall lengths and 20mm through 65mm thread lengths, stainless steel transfixing pin with 5mm shank, 7.2mm thread diameter, and 325mm overall length, and stainless steel instruments for implantation. The pins are implanted into bone and then they are connected using the clamps, rods and posts to form a rigid construct which holds the bone fragments rigidly in place. The pins are offered in various lengths and thicknesses. - OIC External Fixation System (K211112) Primary Predicate: Additional Predicate: Orthofix Galaxy Fixation System (K113770) - Reference Device: Monkey Rings External Fixation System (K212895) - The Monkey Bars Pin to Bar External Fixation System is Substantial substantially equivalent to the legally marketed predicate Equivalence: device systems with respect to intended use and design. The subject Monkey Bars Pin to Bar External Fixation System is similar to the cleared OIC External Fixation System in terms of material, shape, and dimensions. The primary difference between the subject Monkey Bars Pin to Bar External Fixation System and the primary predicate device is that the jaw can rock in both directions as it closes on either the bar or pin. The subject device has a shorter bar length range, a larger combination clamp design and size, and a longer kickstand (with one predicate size being wider). The subject device pins have a shorter length range and have the same diameter as the primary predicate pins. The differences in technological characteristics between the subject device and predicate device were shown not to introduce different questions of safety and effectiveness. - Performance Testing: All necessary testing has been performed on representative Monkey Bars components to assure substantial equivalence to its predicate and demonstrate the subject device performs as intended. Clinical data are not needed to support the safety and effectiveness of the subject device. Minimal Construct Static Compression Testing was performed. {6}------------------------------------------------ Conclusions: The Monkey Bars Pin to Bar External Fixation System subject to this submission possesses the same intended use and has similar technological characteristics as the predicate device system components. All performance testing conducted for the Monkey Bars Pin to Bar External Fixation System predetermined acceptance criteria or were otherwise considered acceptable. As such, the Monkey Bars Pin to Bar External Fixation System are substantially equivalent to the predicate devices.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...