K962644 · Deutschland Surgical, Inc. · JDO · Sep 6, 1996 · Orthopedic
Device Facts
Record ID
K962644
Device Name
VIGOR COMPRESSION HIP SCREW SYSTEM
Applicant
Deutschland Surgical, Inc.
Product Code
JDO · Orthopedic
Decision Date
Sep 6, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Vigor Compression Hip Screw is indicated as a means of rigid internal fixation of stable and unstable fractures of the proximal femur in which a stable medial buttress can be reconstructed and is intended for the management of intertrochanteric, subtrochanteric and basilar neck fractures and selected trochanteric non-unions.
Device Story
System provides rigid internal fixation for proximal femur fractures. Components include bone screws, compression screws, lag screws, and side plates. Lag screws feature flattened square base for rotational stability. Implants available in various lengths and plate configurations. Used by surgeons in clinical settings to stabilize fractures, facilitating bone healing. Instrumentation set included for surgical procedure. Implants are single-use.
Clinical Evidence
Bench testing only; no clinical data provided.
Technological Characteristics
Material: 316L Stainless Steel. Components: Bone screws (22-60mm), lag screws (50-150mm), and side plates (4, 5, 6-hole; 135°, 140°, 145° angles). Non-sterile; requires user sterilization to 10^-6 SAL.
Indications for Use
Indicated for patients with stable or unstable proximal femur fractures, including intertrochanteric, subtrochanteric, basilar neck fractures, and selected trochanteric non-unions, where a stable medial buttress can be reconstructed.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
K140018 — TSP HIP FRACTURE PLATING SYSTEM · Biomet, Inc. · Apr 4, 2014
K971654 — OSTEO COMPRESSION HIP SCREW SYSTEM · Osteonics Corp. · Jul 9, 1997
K233134 — I.T.S. INS Proximal Femur Nail · I.T.S. GmbH · Feb 12, 2024
K972629 — DEPUY AE TK2 HIP SCREW SYSTEM · Depuy, Inc. · Oct 9, 1997
K152757 — SIGN Hip Construct (SHC) · Sign Fracture Care International · Nov 13, 2015
Submission Summary (Full Text)
{0}
SEP - 6 1996
K962644
# SUMMARY OF SAFETY AND EFFECTIVENESS
## VIGOR COMPRESSION HIP SCREW SYSTEM
### I. General Information
Classification Name: Device, Fixation, Proximal Femoral Implant
Common Name: Compression Hip Screw
Device Trade Name: Vigor Compression Hip Screw System
Classification Code: 87JDO
Submitter's Name & Address: Deutschland Surgical Incorporated
1222 Commerce Street, Suite #415
Dallas, Texas 75202-4306
Establishment Registration No: 1646853
Contact Person: Thomas Wolter
Summary Preparation Date: July 5, 1996
### II. Predicate Device
The Vigor Compression Hip Screw is claimed to be substantially equivalent in material, design, and function to the Synthes Dynamic Hip Screw cleared by FDA under 510(k) K791619 on August 28, 1979.
### III. Device Description
The Vigor Compression Hip Screw System is intended as a means of rigid internal fixation for fractures of the proximal femur. The Vigor System implants consist of Bone Screws; Compression Screws, Lag Screws and Side Plates, and are offered in a wide, incremental range of sizes to ensure optimum selection for a proper fixation. The Vigor full-length threaded Bone Screws and Compression Bone Screws are available in a range of sizes from 22.0 mm to 60.0 mm in length. Vigor Lag Screws and Super Lag Screws feature a flattened, square base that engages the inner surface of the implant barrel for increased rotational stability. These Lag Screws are available in a range of sizes from 50.0 mm to 150.0 mm in length, in 5.0 mm increments. And, finally, a trio of low-profile Side Plates are included in the system. Vigor offers the standard 4-hole, 5-hole and 6-hole plates, each available in a 135°, 140° or 145° configuration.
The Vigor System instrumentation set consists of the standard instrumentation required for each indicated procedure.
Premarket Notification:
Vigor Compression Hip Screw
Deutschland Surgical Incorporated:
Submitted: July 5, 1996
Page 83
{1}
# IV. Sterilization
Vigor Implants and Instrumentation are provided non-sterile. Implants are for single use only. Both implants and instruments must be sterilized prior to use in accordance with the recommended sterilization parameters described in the package insert in order to achieve a sterility assurance level of $10^{-6}$.
# V. Indications for Use
The Vigor Compression Hip Screw is indicated as a means of rigid internal fixation of stable and unstable fractures of the proximal femur in which a stable medial buttress can be reconstructed and is intended for the management of intertrochanteric, subtrochanteric and basilar neck fractures and selected trochanteric non-unions.
# VII. Substantial Equivalence
The Vigor Compression Hip Screw System is considered to be substantially equivalent to the Synthes Dynamic Hip Screw[K791619] distributed by Synthes (U.S.A) at 1690 Russell Road, Paoli, PA 19301.
| Feature | Vigor Compression Hip Screw | Synthes Dynamic Hip Screw |
| --- | --- | --- |
| Material | 316L Stainless Steel | 316L Stainless Steel |
| Design | Bone, Compression and Lag Screws, low-profile, angled side plates and associated instrumentation. | Bone, Compression and Lag Screws, low-profile, angled side plates and associated instrumentation |
| Function | Rigid internal fixation to fractures of the proximal femur. | Rigid internal fixation to fractures of the proximal femur. |
# VIII. Conclusion
The Vigor Compression Hip Screw System is considered to be substantially equivalent in design, material and function to the Synthes Dynamic Hip Screw and is believed to perform as well as the Synthes System.
Premarket Notification:
Vigor Compression Hip Screw
Deutschland Surgical Incorporated:
Submitted: July 5, 1996
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.