SPINEVU ENDOSCOPIC SPINE SYSTEM (SESS)

K121548 · Spine View, Inc. · HRX · Aug 16, 2012 · Orthopedic

Device Facts

Record IDK121548
Device NameSPINEVU ENDOSCOPIC SPINE SYSTEM (SESS)
ApplicantSpine View, Inc.
Product CodeHRX · Orthopedic
Decision DateAug 16, 2012
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.1100
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Spine View Minimally Invasive Surgical (MIS) Decompression System is indicated for use to facilitate access and visualization in the surgical area of the cervical, thoracic, or lumbar spine and is accessorized with surgical tools for interventional spinal procedures such as discectomy, nucleotomy and foraminotomy.

Device Story

System consists of sterile, single-use arthroscopic surgical accessories; includes enCise Bone Cutter and enVue Cannula. Used in OR by surgeons to facilitate access and visualization during spinal procedures. EnCise Bone Cutter features tubular sheath with mechanical chisel tip for cutting soft and hard tissue (bone/cartilage) under direct endoscopic visualization. EnVue Cannula provides working channel for endoscope and instruments; includes articulating jaw for tissue retraction. Surgeon operates instruments manually; visual feedback provided via endoscope. System enables minimally invasive removal of osteophytes and calcifying tissues; benefits include reduced tissue trauma compared to open surgery.

Clinical Evidence

No clinical data. Substantial equivalence supported by non-clinical bench testing, including device-to-device compatibility, jaw cycle integrity, joint integrity, jaw activation, tip rotation, tip robustness, handle/joint integrity, jaw/trigger force, biocompatibility, design validation, packaging, and sterility testing.

Technological Characteristics

Materials: Stainless steel, Nylon-12. Principle: Mechanical cutting/retraction via manual actuation. Energy: Manual. Form factor: Tubular cannulas and chisel-tipped cutters. Connectivity: None. Sterilization: Irradiated (sterile, single-use).

Indications for Use

Indicated for patients requiring cervical, thoracic, or lumbar spine surgery, specifically for interventional procedures including discectomy, nucleotomy, and foraminotomy.

Regulatory Classification

Identification

An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 121548 Spine View, Inc. Special 510(k) Notification Minimally Invasive Surgical (MIS) Decompression System # 510(k) Summary AUG 1 6 2012 This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. - Name and Address of Applicant A. Spine View, Inc. 48810 Kato Road, Suite 100E Fremont, CA 94538 Phone: (510) 743-5090 Fax: (510) 490-1753 ### B. Contact Person Diana DeGregorio Lincé Consulting Regulatory Affairs Consultant Phone: (925) 980-8047 - C. Date Prepared July 13, 2012 - D. Device Name Minimally Invasive Surgical (MIS) Decompression System Trade Name: Common Name: Arthroscope Classification Name: · Arthroscope ### ய் Device Classification | Classification: | 21 CFR §888.1100 | |-----------------|------------------| | Product Code: | HRX | | Device Class: | Class II | ### F. Predicate Device The modified Minimally Invasive Surgical (MIS) Decompression System is substantially equivalent to the Spine View SpineVu Endoscopic Spine System (SESS) (K113362 and K08151) and predicate cutting instruments offered as part of the Richard Wolf Medical Instruments Corp. Minimally Invasive Spinal Surgery Set (K994363) and several Class I orthopedic manual surgical instruments, including the Joimax EndoReamer and EndoChisel Assembly. {1}------------------------------------------------ KIZISY8 Spine View, Inc. ### Special 510(k) Notification Minimally Invasive Surgical (MIS) Decompression System ### G. Device Description The Minimally Invasive Surgical (MIS) Decompression System is a collection of arthroscopic surgical accessories provided sterile (irradiated) and intended for single-use only. As a group, the accessories are provided to facilitate delivery of an endoscope and other instruments to the targeted treatment site. The addition of a new accessory, the enCise Bone Cutter, as an additional surgical tool to facilitate removal of soft and hard (bone) tissue during interventional spinal procedures. The new component will be offered in two configurations: one compatible with flexible endoscopes, and the second compatible with rigid endoscopes. Two additional enVue Cannula models are also being introduced for compatibility with rigid endoscopes: Long and Standard Jaw. Introduction of the new enCise Bone Cutter and additional configurations of the already cleared enVue Cannula do not affect the intended use or alter the fundamental scientific technology of the system. The additional accessories being proposed in the modified MIS Decompression System are listed below: - . enCise Bone Cutter: a tubular sheath with a mechanical "chisel" tip feature designed to cut soft and hard (cartilaginous, bone) tissue during spinal procedures. The enCise Bone Cutter is offered in two configurations (one compatible with flexible endoscopes, and the second compatible with rigid endoscopes). - enVue Cannula, Standard and Long Jaw: configurations for use with rigid . endoscopes. The working end of the accessories is identical to the previously cleared enVue Cannulas. The only difference is in the handle, which is modified to accommodate rigid endoscopes (up to 5.9 mm OD; 205 mm working length). #### H. Intended Use The Spine View Minimally Invasive Surgical (MIS) Decompression System is indicated for use to facilitate access and visualization in the surgical area of the cervical, thoracic, or lumbar spine and is accessorized with surgical tools for interventional spinal procedures such as discectomy, nucleotomy and foraminotomy. ### l. Technological Comparison The Spine View Minimally Invasive Surgical (MIS) Decompression System has similar features as compared to the predicate devices as shown in the tables below: {2}------------------------------------------------ K121548 f-3.0+5 Spine View, Inc. enCise Bone Cutter | | Predicate | Predicate | Subject | |------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------| | Company, K# | Richard Wolf Minimally<br>Invasive Spinal Surgery<br>Set,<br>K994363 | Joimax, Inc., TESSYS<br>Surgical Instruments, N/A<br>- Class I Exempt | Spine View MIS<br>Decompression System,<br>K121548 | | Device Name | Punches | EndoReamer, EndoChisel | enCise Bone Cutter | | Intended Use | endoscopically controlled<br>dissection, exploration, and<br>manipulation of tissue<br>through natural or surgically<br>created passages<br>to grasp, manipulate, and<br>cut, as well as for the<br>dissection and biopsy of<br>tissue, organs, or foreign<br>bodies.<br>punching and removal of<br>tissue through natural or<br>surgically created passages. | The Joimax TESSYS Spinal<br>Stenosis System is a set of<br>endoscopic instruments that<br>facilitate removal of<br>osseous/bony material<br>under direct endoscopic<br>imaging. The EndoReamer<br>and EndoChisel instruments<br>are intended to remove<br>osteophytes and calcifying<br>tissues, including coarse<br>stenosis fragments. | enCise Bone Cutter is<br>intended to facilitate<br>removal of remove<br>osteophytes and calcifying<br>tissues under direct<br>endoscopic visualization. | | Product Code | HRX, Class II | LXH, Class I - 510(k)<br>Exempt | HRX, Class II | | Design | The punch-featured<br>instruments are designed<br>with a sliding jaw, which is<br>forced into the material to be<br>cut by depressing the<br>mechanical actuator in the<br>handle | The chisel-featured<br>instruments are designed<br>with a characteristically<br>shaped cutting edge, which<br>are forced into the material<br>to be cut by use of a<br>surgical mallet or manual<br>manipulation | Same | | Dimensions | 2.5 - 4.0 mm OD, WL 290 -<br>360 mm (=2.5-4.0 mm<br>cutting element tip width) | EndoReamer: Single<br>Working Channel for<br>endoscope and endoscopic<br>instruments<br>3.0-7.5 mm ID; 4.0-8.5 mm<br>OD; WL 210 mm (4.0 -<br>8.5mm circular serrated tip<br>diameter)<br>EndoChisel: 2.6 or 3.0 mm<br>OD, WL 370 mm (=2.6 or<br>3.0 mm chisel tip width) | Single Working Channel for<br>endoscope and endoscopic<br>instruments<br>6.12 mm ID, 7.16 mm OD,<br>196 &198 mm WL (3.76 mm<br>chisel tip width) | | Patient Contact Materials | Stainless Steel | Same | Same | | Target Anatomy | Intervertebral procedures | Same | Same | | Supplied Sterile? | No | No | Yes | | Single Use? | No | No | Yes | | | Predicate | Subject | | | Company, K# | Spine View, MIS Decompression System<br>K113362 | Spine View, MIS Decompression System<br>K121548 | | | Device Name | enVue Cannula | enVue Cannula | | | Intended Use | 1. Access cannula with working channel for<br>endoscope, endoscopic instruments, and<br>irrigation and drainage<br>2. Distal feature for retracting tissue away<br>from end of cannula | 1. Same<br>2. Same | | | Product Code | HRX, Class II | Same | | | Design | Articulating Top Jaw can be mechanically<br>actuated to retract tissue away from end of<br>cannula Distal tip: Articulating jaw (Nylon) for<br>tissue retraction (Standard and Long jaw<br>configurations) | Same | | | Dimensions | 1. OD: 0.282"<br>2. Working Length: 7.5"<br>3. Length, Tissue Retraction Feature 0.394"<br>4. OD, Tissue Retraction Feature: 0.354 "<br>5. Single Working Channel 0.240" for enVue<br>Sheath (for endoscopic instruments and<br>endoscope [2 mm OD] | 1. Same<br>2. Same<br>3. Same<br>4. Same<br>5. Single Working Channel 0.240" for<br>endoscopic instruments and endoscope [5.9<br>mm OD] | | | Patient Contact<br>Materials | Stainless Steel, Nylon-12 | Same | | | Target Anatomy | Intervertebral procedures | Same | | | Supplied<br>Sterile? | Yes | Same | | | Single Use? | Yes | Same | | . : {3}------------------------------------------------ K1215468 g. 4 c of 5 Spine View, Inc. # Special 510(k) Notification SpineVu Endoscopic Spine System (SESS) ### enVue Cannula The technological characteristics and principals of operation of the modified Spine View Minimally Invasive Surgical (MIS) Decompression System are substantially equivalent to the named predicate devices. . ### J. Non-Clinical Performance Data The following non-clinical testing was conducted to support a determination of substantial equivalence to the predicate device. | • Device to Device Compatibility Testing | • Jaw Cycle Integrity Testing | |------------------------------------------|-------------------------------| | • Joint Integrity Testing | • Jaw Activation Testing | | • Tip Rotation | • Biocompatibility Testing | | • Tip Robustness | • Design Validation Testing | | • Handle/Joint Integrity | • Packaging Testing | | • Jaw & Trigger Force Testing | • Sterility Testing | {4}------------------------------------------------ 1215486- 5 of J Spine View, Inc. The above testing confirmed that the Spine View Minimally Invasive Surgical (MIS). Decompression System performs according to the stated intended use. All data fell well within product specifications and external standard requirements. Results of non-clinical testing demonstrated that the Spine View Minimally Invasive Surgical (MIS) Decompression System is substantially equivalent to the predicate devices for its intended use. ### K. Conclusions The Spine View Minimally Invasive Surgical (MIS) Decompression System has been carefully compared to legally marketed devices with respect to intended use and technological characteristics. In addition, non-clinical testing was conducted to validate the performance of the device and ensure the modified Spine View Minimally Invasive Surgical (MIS) Decompression System functions as intended and meets design specifications. The comparison and non-clinical results demonstrate that the device is substantially equivalent to the predicate device for its intended use. {5}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Spine View, Incorporated % Ms. Diana DeGregorio Regulatory Affairs Consultant 48810 Kato Road, Suite 100E Fremont, California 94538 Re: K121548 Trade/Device Name: Minimally Invasive Surgical (MIS) Decompression System Regulation Number: 21 CFR 888.1100 Regulation Name: Arthroscope Regulatory Class: Class II Product Code: HRX Dated: July 15, 2012 Received: July 18, 2012 Dear Ms. DeGregorio: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set AUG 16 2012 {6}------------------------------------------------ # Page 2 - Ms. Diana DeGregorio forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours For Duth. Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ Spine View, Inc. K121548 p. 1 o f 1 # Indications for Use Statement 510(k) Number (if known): Device Name: Minimally Invasive Surgical (MIS) Decompression System K ### Indications for Use: The Spine View Minimally Invasive Surgical (MIS) Decompression System is indicated for use to facilitate access and visualization in the surgical area of the cervical, thoracic, or lumbar spine and is accessorized with surgical tools for interventional spinal procedures such as discectomy, nucleotomy and foraminotomy. Prescription Use × Over-The-Counter Use (per 21 CFR 801.109) Or PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED Concurrence of CDRH, Office of Device Evaluation (ODE)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...