POWER DRIVE ART1 2304 WITH MOTOR HANDLES, TOOLS AND ACCESSORIES

K080617 · Richard Wolf Medical Instruments Corp. · HRX · Jun 27, 2008 · Orthopedic

Device Facts

Record IDK080617
Device NamePOWER DRIVE ART1 2304 WITH MOTOR HANDLES, TOOLS AND ACCESSORIES
ApplicantRichard Wolf Medical Instruments Corp.
Product CodeHRX · Orthopedic
Decision DateJun 27, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.1100
Device ClassClass 2
AttributesTherapeutic

Intended Use

In conjunction with a motorized handle (e.g. Power Stick M5/3), the POWER DRIVE ART1 2304 serves as a drive unit for Richard Wolf rotary blades/abraders for the removal of tissue during endoscopic intervention. Simultaneous evacuation (suction) allows continuous removal of the ablated tissue. In conjunction with Power Drill MI, bones can be severed or machined. Indications and field of use: The product is used in conjunction with endoscopic accessories for the following purposes: - In arthroscopy, e.g. resection of the meniscus, for removing soft tissue as well as intraarticular severing or abrasion of bone tissue (e.g. ACL or shoulder applications). - In thoracic surgery, e.g. for the removal of hematomas. - In sinus surgery, e.g. for the removal of polyps. - In spinal surgery (arthroscopic microdiskectomy (AMD) for removing degenerated tissue.

Device Story

POWER DRIVE ART1 2304 is a surgical motor control unit for endoscopic procedures. System comprises generator, motor handles, rotary blades/abraders, saws, and drills. Input: user control via touch screen or integrated keypads on handles. Operation: generator powers motor handles to drive surgical tools; RFID technology enables automatic tool recognition. Output: mechanical rotation/oscillation for tissue resection or bone machining; simultaneous suction for tissue evacuation. Used in OR by surgeons. Integration: CAN-BUS interface allows connection to R. Wolf RIWO NET SYSTEM for remote/speech control. Benefits: improved user comfort, reduced risk of incorrect settings, and efficient tissue removal/bone machining.

Clinical Evidence

No clinical data. Bench testing only.

Technological Characteristics

Generator-based motor control system. Materials: DLC (diamond-like carbon) coating on rotary blade/abrader sheaths; anodized aluminum for drill housing. Connectivity: CAN-BUS interface for RIWO NET SYSTEM integration. Features: RFID-based automatic tool recognition, touch screen interface, integrated keypads. Power: electrical motor drive. Sterilization: multiple-use components require manual sterilization; single-use components provided sterile.

Indications for Use

Indicated for patients undergoing endoscopic procedures requiring tissue removal or bone machining, including arthroscopy (meniscus resection, soft tissue removal, bone abrasion for ACL/shoulder), thoracic surgery (hematoma removal), sinus surgery (polyp removal), and spinal surgery (arthroscopic microdiskectomy for degenerated tissue).

Regulatory Classification

Identification

An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Ko 80617 Page 1 of 3 # 510(k) Su | Submitter: | Date of Preparation: | | | |----------------------------------------|------------------------------------------------|----------|--------------------| | | 25. February 2008 | | | | | Revised: 05 June 2008 | | | | Company / Institution name: | FDA establishment registration | | | | RICHARD WOLF MEDICAL INSTRUMENTS CORP. | number: | | | | | 14 184 79 | | | | Division name (if applicable): | Phone number (include area code): | | | | N.A. | (847) 913 1113 | | | | Street address: | FAX number (include area code): | | | | 353 Corporate Woods Parkway | (847) 913-0924 | | | | City: | State/Province: | Country: | ZIP / Postal Code: | | Vernon Hills | Illinois | USA | IL 60061 | | Contact name: | Mr. Ron Haselhorst | | | | Contact title: | Quality Assurance / Regulatory Affairs Manager | | | Product Information: | Trade name: | Model number: 2304.xxx, | |--------------------------------------------------------------------------|----------------------------------------------------------------| | POWER DRIVE ART1 2304 with motor handles, tools and Accessories | 8995X.xxx, 8564.xxx, 8996X.xxx,8997X.xxx, 8998X.xxx, 4997X.xxx | | Common name: | Classification name: | | Surgical instrument Motor Unit and accessories/attachments for endoscopy | Surgical instrument motors and accessories/attachments. | Information on devices to which substantial equivalence is claimed: | 510(k) Number | Trade or proprietary or model name | Manufacturer | |---------------|-------------------------------------------------------------------------------------------------|-------------------| | 1 K030082 | 1 Power Control 2303 with Power Stick M4 | 1 Richard Wolf | | 2 K871250 | 2 Intra-articulated arthro power system 2161 | 2 Richard Wolf | | 3 K970088 | 3 RIWO DRIVE Generator with Footswitch, Motor handles and Single Use Rotary Blades and Abraders | 3 Richard Wolf | | 4 K984304 | 4 RIWO DRIVE Generator, single use or reusable Rotary Cutters and Burrs | 4 Richard Wolf | | 5 K984521 | 5 RIWO DRIVE small handle | 5 Richard Wolf | | 6. K030009 | 6. Powershaver SL | 6. Karl Storz | | 7. K062849 | 7. DYONICS Power II Shaver System | 7. Smith & Nephew | #### 1.0 Description The POWER DRIVE ART1 2304 is an enhanced version of our POWER CONTROL 2303, cleared in K030082. The new device has new electronics, new software, additional motor handles, tools and tool holders such as rotary blades/ abraders, saws, sagittal and reciprocal saw attachments, drill chucks, drills and surgical wires. {1}------------------------------------------------ Kos0617 2.0 Intended Use Page 283 In conjunction with a motorized handle (e.g. Power Stick M5/3), the POWER Intended use: DRIVE ART1 2304 serves as a drive unit for Richard Wolf rotary blades/abraders for the removal of tissue during endoscopic intervention. Simultaneous evacuation (suction) allows continuous removal of the ablated tissue. In conjunction with Power Drill MI, bones can be severed or machined. Indications and field of use: The product is used in conjunction with endoscopic accessories for the following purposes: - . In arthroscopy, e.g. resection of the meniscus, for removing soft tissue as well as intraarticular severing or abrasion of bone tissue (e.g. ACL or shoulder applications). - In thoracic surgery, e.g. for the removal of hematomas. . - In sinus surgery, e.g. for the removal of polyps. ● - In spinal surgery (arthroscopic microdiskectomy (AMD) for removing . degenerated tissue. #### 3.0 Technological Characteristics The POWER DRIVE ART1 2304, consisting of: generator, motor handles, tools (such as rotary blades/ abraders, sagittal and reciprocal saws, drills and surgical wires) and tool holders (such as sagittal and reciprocal saw attachments and drill chucks). Additional features are new motor handles, some with integrated keypad as well as automatic tool recognition with the new rotary blades and abraders (via RFID technology) and a touch screen for user control. Multiple use devices are not provided sterile; they must be startlized per their instruction before each use. Single use devices are provided sterife and designed for only one use on or in-a single patient. Via the integrated CAN-BUS interface, the POWER CONTROL 2304 can be integrated into the R. Wolf RIWO NET SYSTEM with remote control, speech control, and touch-screen monitor. #### 4.0 Substantial Equivalence The submitted POWER Drive 2304 generator have same indications, equivalent design, functions and materials as the previous devices sold by Richard Wolf. The new device has new electronics, new software, additional motor handles, tools and tool holders. Additional features are new motor handles, some with integrated keypad as well as automatic tool recognition for the new rotary blades and abraders. Together with the touch screen control, this increases the user comfort and avoids damage caused by incorrect settings. The new inner sheaths of the rotary blades and abraders have a DLC (diamond-like carbon) coating to reduce friction. The housing of the new drill is made of anodized Aluminum. . Submitted devices pose the same type of questions about safety or effectiveness as the compared devices and the new technological characteristics have not diminished safety or effectiveness. The submitted devices are substantially equivalent to existing 510(k)-cleared devices sold by Richard Wolf, Karl Storz, and Smith & Nephew {2}------------------------------------------------ Ko 80617 Page 383 ### 5.0 Performance Data No performance standards are known. - 6.0 Clinical Tests No clinical tests performed. ### 7.0 Conclusions Drawn These devices are designed and tested to guarantee the safety and effectiveness, when used according to the instructions manual. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle with its wings spread, rendered in thick, black lines. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle symbol. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUN 2 7 2008 Richard Wolf Medical Instruments Corporation % Mr. Ron Haselhorst Quality Assurance/Regulatory Affairs Manager 353 Corporate Woods Parkway Vernon Hills, Illinois 60061 Re: K080617 Trade/Device Name: POWER DRIVE ART1 2304 with motor handles, tools and Accessories Regulation Number: 21 CFR 888.1100 Regulation Name: Arthroscope Regulatory Class: II Product Code: HRX, HWE, ERL, GEY Dated: June 5, 2008 Received: June 13, 2008 Dear Mr. Haselhorst: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 {4}------------------------------------------------ Page 2 - Mr. Ron Haselhorst CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket, Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely vours. Mark N Mulhearn Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ Kofobi7 ## Indications for Use Statement 510(k) Number (if known): Device Name: POWER DRIVE ARTI 2304 with motor handles, tools and Accessories - Intended use: In conjunction with a motorized handle (c.g. Power Stick M5/3), the POWER DRIVE ART1 2304 serves as a drive unit for Richard Wolf rotary blades/abraders for the removal of tissue during endoscopic intervention. Simultaneous evacuation (suction) allows continuous removal of the ablated tissue. In conjunction with Power Drill M1, bones can be severed or machined. Indications and field of use: The product is used in conjunction with endoscopic accessories for the following purposes: - . In arthroscopy, e.g. resection of the meniscus, for removing soft tissue as well as intraarticular severing or abrasion of bone tissue (e.g. ACL or shoulder applications). - In theracte surgery, e.g. for the removal of hematomas. . - � In sinus surgery, e.g. for the removal of polyps. - . In spinal surgery (arthroscopic microdiskectomy (AMD) for removing degenerated tissue. Prescription Use > (Part 21CFR 801 Subpart D) AND/OR Over-The Counter Use (Part 21CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH Office of Device Evaluation (ODE) **Division Sign-Off** Division of General, Restorative, and Neurological Devi **510(k) Number** K080617
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...