The mi-eye 3 needlescope™ with cannula, mi-eye tablet 3™ is indicated for use in diagnostic and operative arthroscopic and endoscopic procedures to provide illumination of an interior cavity of the body through either a natural or surgical opening.
Device Story
Portable visualization system comprising sterile, single-use needlescope probe and reusable mi-tablet 3 LCD monitor. Probe features integrated camera, LED light source, and rigid shaft (75/95mm lengths; 0/25-degree angles). System connects via cable; probe powered by tablet. Tablet includes internal battery, power supply, and external charging. Irrigation provided through probe distal end via user-supplied solution. Monitor displays real-time video; supports connection to external ultrasound transducers. Used in clinical settings for arthroscopic/endoscopic procedures. Physician operates device to visualize internal anatomy, aiding diagnostic and operative decision-making. Image stabilization feature maintains orientation relative to probe position, with handle-top indicator displayed on screen. Benefits include minimally invasive visualization of joint/body cavities.
Clinical Evidence
Bench testing only. Testing included biocompatibility (ISO 10993), electrical safety/EMC (IEC 60601 series), packaging (ASTM D4169, F2096, F88), irrigation flow/tube strength (ISO 9626), shelf life (ASTM F1980), software V&V (IEC 62304), dimensional analysis (ISO 594, 7864, 80369-7), sterilization (ISO 11135, 11737, 10993-7), and usability validation (IEC 60601-1-6, 62366-1). All tests passed.
Indicated for diagnostic and operative arthroscopic and endoscopic procedures requiring illumination and visualization of interior body cavities via natural or surgical openings.
Regulatory Classification
Identification
An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.
{0}------------------------------------------------
September 15, 2021
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Trice Medical, Inc. David Vancelette Vice President of Ouality Assurance & Regulatory Affairs 40 General Warren Blvd, Suite 100 Malvern, Pennsylvania 19355
Re: K212556
Trade/Device Name: mi-eye 3 needlescope™ with cannula, mi-tablet 3TM Regulation Number: 21 CFR 888.1100 Regulation Name: Arthroscope Regulatory Class: Class II Product Code: HRX Dated: August 12, 2021 Received: August 13, 2021
### Dear David Vancelette:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Laura C. Rose, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
510(k) Number (if known) K212556
Device Name
mi-eye 3 needlescope™ with cannula, mi-tablet 3TM
Indications for Use (Describe)
The mi-eye 3 needlescope™ with cannula, mi-eye tablet 3™ is indicated for use in diagnostic and operative arthroscopic and endoscopic procedures to provide illumination of an interior cavity of the body through either a natural or surgical opening.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
[ ] Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Trice Medical. The logo consists of a blue circular icon on the left and the words "Trice Medical" in black on the right. The icon is a stylized design with three curved shapes arranged in a circular pattern.
## 1. 510(K) SUMMARY
| DATE | September 14, 2021 |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| APPLICANT | Trice Medical, Inc.<br>40 General Warren Blvd<br>Suite 100<br>Malvern, PA 19355<br>Tel: (610) 989-8080<br>Fax: (610) 706-5063 |
| OFFICIAL<br>CORRESPONDENT | David Vancelette<br>Trice Medical, Inc.<br>40 General Warren Blvd<br>Suite 100<br>Malvern, PA 19355<br>dvancelette@tenexhealth.com<br>Tel: (949) 238-8220<br>Fax: (949) 580-1270 |
| TRADE NAME | mi-eye 3 needlescope™ with cannula, mi-tablet 3™ |
| COMMON NAME | Arthroscope |
| DEVICE<br>CLASSIFICATION | Class II, 21 CFR §888.1100 |
| PRODUCT CODE | HRX: Arthroscope |
| PREDICATE DEVICE | mi-eye 2, mi-eye 2 monitor (K162475), now rebranded<br>as mi-eye 3, mi-tablet 3 |
| SUBMISSION TYPE | Special 510(k). The subject device is a modification to<br>the previously cleared mi-eye 2, mi-eye 2 monitor<br>(K162475). |
## SUBSTANTIALLY EQUIVALENT TO:
The mi-eye 3 needlescope™ with cannula, mi-tablet 3™ is substantially equivalent to the previously cleared mi-eye 2, mi-eye 2 monitor (K162475).
# DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION:
The mi-eye 3 needlescope™ with cannula, mi-tablet 3™ is a portable visualization device that uses a probe with integrated camera and separate LCD monitor attached via a cable. The sterile, single-use needlescope probe includes the camera and image capture features with
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Trice Medical. The logo consists of a blue circular icon on the left and the words "Trice Medical" in black font on the right. The icon appears to be a stylized representation of three interconnected shapes.
LED light source. The mi-eye 3 probe connects to, and is powered by, the reusable mi-tablet 3. The mi-tablet 3 includes an internal battery and power supply, along with a cable for external charging. The mi-tablet 3 LCD Monitor displays a real-time image from the probe. The Monitor is also capable of connecting to separate ultrasound transducer, linear and convex, imaging probes and displaying their visual output. The mi-eye 3 needlescope™ with cannula has a rigid shaft that extends from the handle. The distal tip of the probe contains the camera, illumination, and imaging optics. Irrigation may be provided through the distal end of the probe from user supplied solution attached to the handle.
The mi-eye 3 needlescope™ with cannula is available in 75 and 95mm lengths with 0 and 25 degree viewing angles, relative to the axis of the probe is packaged with a scalpel, cannula, obturator and switching stick for use in creating and maintaining the channel for the probe. The monitor has a 12" (diagonal) screen. The probe weight is under 300 grams.
#### INDICATIONS FOR USE:
The mi-eye 3 needlescope™ with cannula, mi-tablet 3™ is indicated for use in diagnostic and operative arthroscopic and endoscopic procedures to provide illumination and visualization of an interior cavity of the body through either a natural or surgical opening.
| Property | Subject Device<br>mi-eye 3 needlescope™ with cannula, mi-<br>tablet 3™ | Predicate Device<br>mi-eye 2, mi-eye 2 monitor |
|----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|
| 510(k) Number | N/A | K162475 |
| Indications for Use | The mi-eye 3 with cannula is indicated for use<br>in diagnostic and operative arthroscopic and<br>endoscopic procedures to provide illumination<br>and visualization of an interior cavity of the<br>body through either a natural or surgical<br>opening. | Same |
| Operating Principles | Transmission of light to illuminate<br>and image an arthroscopic joint, then<br>relaying the image out of the surgical<br>site for processing and display. | Same |
| Property | Subject Device<br>mi-eye 3 needlescope™ with cannula, mi-<br>tablet 3™ | Predicate Device<br>mi-eye 2, mi-eye 2 monitor |
| Principle Components | • Imaging Probe with camera, rigid shaft, handle and output cable.<br>• LCD Display Monitor<br>• Irrigation channel for user supplied solution<br>• Scalpel, Cannula, Obturator and Switching Stick (for establishing arthroscope channel) | Same<br>Same<br>Same<br>N/A |
| Materials | | |
| Probe & Patient<br>contacting features | 304 Stainless Steel | Same |
| Battery | Lithium Ion | Same |
| Technological Characteristics | | |
| Probe Handle Size | 35 - 55mm diameter | Same |
| Probe Needle Outer<br>Diameter | 2.3mm | Same |
| Probe Cannula Outer<br>Diameter | 2.9mm | None |
| Working Probe<br>Lengths | 75 and 95mm | 50, 95, 120 and 160mm |
| Camera Angle | 0 and 25 degrees | 0 and 15 degrees |
| Camera + Display<br>Resolution | 400 x 400 | Same |
| Image leveling | Image stabilized for probe's position (display<br>does not rotate). Arrow indicating the top side<br>of probe handle rotates in the image. | Top of probe always<br>oriented at top of display<br>(display rotates as probe<br>rotates) |
| Tablet Operating<br>System | MS Windows 10 Pro | Same |
| Display output | HDMI and SDI | Same |
| Media capture | USB, DB9 and TRS | Same |
| Irrigation with Angled<br>View Probe | User attached solution is channeled through<br>the annulus between the two tubes to the distal<br>end of the probe | None |
| Packaging for Single<br>Use Probe | Sealed Tyvek lid and thermal formed tray | Same |
| Sterilization Method | Ethylene Oxide | Same |
| Sterility Assurance<br>Level | 10-6 | Same |
### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE:
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Trice Medical. The logo consists of a blue circular icon on the left and the words "Trice Medical" in black on the right. The icon is a stylized design with three curved shapes converging in the center, resembling a three-leaf clover or a stylized medical symbol.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Trice Medical. The logo consists of a blue circular icon with three curved shapes inside, resembling a stylized flower or a three-way intersection. To the right of the icon, the words "Trice Medical" are written in a simple, sans-serif font, with the first letter of each word capitalized.
## PERFORMANCE DATA:
The following performance data were provided in support of the substantial equivalence determination.
| Test | Method | Result |
|---------------------------------------------------------|--------------------------------------------------------------------|--------|
| Biocompatibility | ISO 10993-1<br>ISO 10993-5 | Pass |
| Electrical Safety & EMC | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-18<br>IEC 60529 | Pass |
| Packaging | ASTM D4169, ASTM F2096, ASTM<br>F88/F88M, ISO 11607-1, ISO 11607-2 | Pass |
| Performance - Bench<br>Irrigation Flow<br>Tube strength | ISO 9626 | Pass |
| Shelf Life | ASTM F1980, ASTM F88/F88M, ISO 11607-1,<br>ISO 11607-2 | Pass |
| Software V&V | IEC 62304 | Pass |
| Specification &<br>Dimensional Analysis | ISO 594-1, ISO 594-2, ISO 7864, ISO 80369-7,<br>ISO 9626 | Pass |
| Sterilization | ISO 11135, ISO 11737-1, ISO 10993-7 | Pass |
| Usability Validation | IEC 60601-1-6, IEC 62366-1 | Pass |
## CONCLUSION:
The mi-eye 3 needlescope™ with cannula, mi-tablet 3™ Arthroscope is substantially equivalent in design, intended use, principles of operation, technological characteristics and safety features to the predicate device. There are no new issues of safety and/or effectiveness introduced by the mi-eye 3 needlescope™ with cannula, mi-tablet 3™ Arthroscope when used as instructed.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.