K171794 · Smith & Nephew, Inc. · HRX · Jun 27, 2017 · Orthopedic
Device Facts
Record ID
K171794
Device Name
Smith & Nephew 25 Fluid Management System
Applicant
Smith & Nephew, Inc.
Product Code
HRX · Orthopedic
Decision Date
Jun 27, 2017
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.1100
Device Class
Class 2
Indications for Use
The Smith & Nephew DYONICS 25 Fluid Management System is intended for use during arthroscopic joint surgery to regulate flow of irrigation fluids to maintain intra-articular pressure for uniform distension and clear visualization of the surgical site.
Device Story
Centrifugal fluid management pump for arthroscopic surgery; regulates irrigation fluid flow to maintain intra-articular pressure for joint distension and surgical site visualization. System comprises control unit, various tube sets (inflow, inflow/outflow, day-tube, patient-tube), and optional wired/wireless remote control. Operates via electronic pressure control feedback loop between control unit and tube set cassettes. Used in OR by surgical staff. Provides real-time pressure regulation to ensure clear visualization; assists surgeon in maintaining stable joint environment during procedures. Modified version includes updated PCBs, touch-screen LCD, and software/OS updates.
Clinical Evidence
Bench testing only. Performance data included package/performance testing, environmental testing, fluids testing, EMC testing, safety testing, and software verification and validation. Device met all specified performance criteria and FDA recognized consensus standards (IEC 60601-1, IEC 60601-1-2, IEC 62304).
Indicated for use during arthroscopic joint surgery in the knee, shoulder, hip, and small joints to regulate irrigation fluid flow and maintain intra-articular pressure for distension and visualization.
Regulatory Classification
Identification
An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.
Predicate Devices
Smith & Nephew DYONICS 25 Fluid Management System (K051326)
Reference Devices
Smith & Nephew LEVELERT II Fluid Level Sensors (K060123)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 27, 2017
Smith & Nephew, Inc. Ms. Kathleen Solomon Senior Regulatory Affairs Specialist 150 Minuteman Road Andover, Massachusetts 01810
Re: K171794
Trade/Device Name: Smith & Nephew DYONICS 25 Fluid Management System Regulation Number: 21 CFR 888.1100 Regulation Name: Arthroscope Regulatory Class: Class II Product Code: HRX Dated: June 14, 2017 Received: June 16, 2017
Dear Ms. Solomon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Actinclude requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Jennifer R. Stevenson -S3
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known)
K171794
Device Name Smith & Nephew DYONICS 25 Fluid Management System
#### Indications for Use (Describe)
The Smith & Nephew DYONICS 25 Fluid Management System is indicated for use during arthroscopic joint surgery to regulate flow of irrigation fluids in the knee, shoulder, hip, and small joints to maintain intra-articular pressure for uniform distension and clear visualization of the surgical site.
| Type of Use ( <i>Select one or both, as applicable</i> ) |
|----------------------------------------------------------|
|----------------------------------------------------------|
| <span style="text-decoration: underline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: underline;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|-------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement below. -------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
{3}------------------------------------------------
K171794 p 1/3
Endoscopy
Smith & Nephew, Inc.
150 Minuteman Road
Andover, MA 01810
USA
nith-nenhew com
We are smith&nephew
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION
as required by the Safe Medical Devices Act of 1990 and codified in 21 CFR 807.92 upon which the substantial equivalence is based.
Smith & Nephew DYONICS 25 Fluid Management System Date Prepared: June 26, 2017
#### Submitter's Name:
Smith & Nephew, Inc., Endoscopy Division 150 Minuteman Road, Andover MA. 01810
#### Company Contact:
Kathleen Solomon Sr. Regulatory Affairs Specialist T 978-749-1605 F 978-749-1443
## Device Name
| Trade Name: | Smith & Nephew DYONICS 25 Fluid Management System |
|----------------------|---------------------------------------------------|
| Common Name: | Arthroscopic Fluid Management System |
| Classification Name: | Arthroscope |
| Regulatory class: | II |
| Product Code: | HRX |
## Predicate Devices
The Smith & Nephew DYONICS 25 Fluid Management System is substantially equivalent in Intended Use and Fundamental Scientific Technology to the following legally marketed device in commercial distribution: Smith & Nephew DYONICS 25 Fluid Management System-K051326.
{4}------------------------------------------------
## Description of Device
The Smith & Nephew DYONICS 25 Fluid Management System is a centrifugal fluid management pump used in arthroscopic surgery to regulate the flow of irrigation fluids through the joint to maintain intra-articular pressure for distension and visualization of the surgical site. This is accomplished using an electronic pressure control feedback loop between the control unit and the tube set cassettes.
The Smith & Nephew DYONICS 25 Fluid Management System includes -
- . control unit
- . inflow tube set
- . inflow /outflow tube set
- inflow/outflow tube set with forked suction lines
- . day-tube set
- . patient tube set
- . wired/wireless (IR) remote control (optional)
- . Luer to colder Connector (optional)
Additionally, the Smith & Nephew LEVELERT II Fluid Level Sensors (K060123) as well as a shaver system interface cable, U.S. and international power cords can be used with the Smith & Nephew DYONICS 25 Fluid Management System (DYONICS 25).
## Intended Use
The Smith & Nephew DYONICS 25 Fluid Management System is intended for use during arthroscopic joint surgery to regulate flow of irrigation fluids to maintain intra-articular pressure for uniform distension and clear visualization of the surgical site.
## Comparison of Technological Characteristics
The design changes to the modified Smith & Nephew DYONICS 25 Fluid Management System include updated printed circuit boards (PCBs), new touch-screen LCD display, updated operating system, drivers and software updates. The modified Smith & Nephew DYONICS 25 Fluid Management System has the following similarities as the predicate device cleared in K151326. In that:
- The same indications for use/intended use.
- Utilizes the same principle of operation and fundamental scientific technology
- Incorporates the same basic design
- Incorporates the same material
There are no major differences between the modified Smith & Nephew DYONICS 25 Fluid Management System and the currently cleared predicate.
{5}------------------------------------------------
K171794 p 3/3
Endoscopy
Smith & Nephew, Inc.
150 Minuteman Road
Andover, MA 01810
USA
T 978-749-1000
F 978-749-1443
www.smith-nephew.com
We are smith&nephew
## Performance Data
Testing demonstrated that the Smith & Nephew DYONICS 25 Fluid Management System has met the performance specifications and therefore, is substantially equivalent to the predicate device cleared in K051326.
The following performance data and validations were conducted:
- Package/Performance Testing
- . Environmental Testing
- Fluids Testing
- EMC Testing
- Safety Testing
- Software Verification & Validation ●
The Smith & Nephew DYONICS 25 Fluid Management System meets the following FDA recognized consensus standards:
- IEC 60601-1-2:2015 Part 1-2: General requirements for basic safety and essential ● performance -Collateral Standard: Electromagnetic disturbances - Requirements and tests
- IEC 60601-1 Medical Electrical Equipment Part 1: General Requirements for Safety ●
- IEC 62304 Medical Device Software - Lifecycle:2006
## Conclusion:
The Smith & Nephew DYONICS 25 Fluid Management System met all specified criteria and did not raise new safety or effectiveness questions. The substantial equivalence of the modified device is based on same indications for use, fundamental technology including design, and operational principles. Based on the similarities to the predicates and the performance data, the Smith & Nephew DYONICS 25 Fluid Management System is substantially equivalent to its predicate (K051326).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.