The SOPRO 670 irrigation pump is an arthroscopic pump system intended to provide fluid distension and irrigation of knee, shoulder, hip, elbow, ankle and wrist joint cavities and fluid suction during diagnostic and operative arthroscopic procedures that may only be used by qualified physicians.
Device Story
SOPRO 670 is a microprocessor-controlled peristaltic pump system for arthroscopic procedures. It provides fluid distension, irrigation, and aspiration of joint cavities. The system includes a housing, power supply, roller wheel, pump head, servo-motor, setting keys, and display elements. It operates with specific irrigation/suction tubing sets and a foot-remote control. A redundant pressure measurement system continuously monitors joint cavity pressure, comparing it against a pre-set nominal pressure to ensure conformity. Used in clinical settings by qualified physicians, the device maintains joint distension to facilitate visualization and surgical access during diagnostic and operative arthroscopy. The automated pressure control helps maintain stable joint conditions, potentially improving surgical outcomes and patient safety by preventing over-pressurization.
Clinical Evidence
Bench testing only. The device complies with IEC 601-1 (Electrical Safety) and UL 60601-1 standards. No clinical data provided.
Technological Characteristics
Peristaltic pump system; microprocessor-controlled; servo-motor driven. Components include roller wheel, pump head, display, and foot-remote. Connectivity includes specialized irrigation/suction tubing. Complies with IEC 601-1 and UL 60601-1. Sterilization method not specified.
Indications for Use
Indicated for fluid distension, irrigation, and suction of knee, shoulder, hip, elbow, ankle, and wrist joint cavities during diagnostic and operative arthroscopic procedures in patients requiring such intervention. For use by qualified physicians only.
Regulatory Classification
Identification
An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.
{0}------------------------------------------------
# 10. Summary of Safety and Effectiveness - "510 (k) Summary"
A. Submitter Information SOPRO JUL 1 1 2008 ZAC Athélia IV Avenue des Genévriers 13705 La Ciotat Cedex FRANCE Telephone: 33 (0) 442 98 01 01 Fax: 33 (0) 442 71 76 90 Contact Person: Steve Salesky SOPRO c/o ACTEON, Inc. 124 Gaither Drive, Suite 140 Mt. Laurel. NJ 08054 Tel: 609 828-2428 Fax: 856 222-4726 E-mail: steve.salesky@us.acteongroup.com Date Prepared: January 11, 2008 B. Device Identification Classification Name: Arthroscope and Accessories Common Usual Name: Arthroscopic Pump
Common Sample Name: Form Sample Strip
Proprietary Name: SOPRO 670 Arthroscopic Pump
C. Identification of Predicate Device
| Device | Applicant | 510(k) No. | Date Cleared |
|-----------------------------------|---------------------------|------------|--------------|
| Future Medical<br>Systems FMS DUO | Future Medical<br>Systems | K954465 | Nov. 9, 1995 |
The SOPRO 670 Arthroscopic Pump is substantially equivalent to the arthroscopic pump functions of the predicate device by Future Medical Systems, arthroscopic pump and shaver model FMS DUO (K954465) previously cleared by the FDA and currently marketed.
## D. Device Description
The SOPR0 670 irrigation pump is a peristaltic pump system designed to automatically provide and control distension and irrigation of the operative site during arthroscopic procedures and fluid irrigation of the operative site during laparoscopic procedures using sterile fluids. The pump is used in conjunction with specific tubing sets designed for arthroscopic procedures.
18
{1}------------------------------------------------
- 1080122
The Arthroscopy Pump SOPRO 670 is a microprocessor controlled pump system designed to provide liquid distension and irrigation of joint cavities and aspiration of liquids out of the joint cavities during diagnostic and operative arthroscopy. The Arthroscopy SOPRO 670 consists of the following main components: housing, power supply, roller wheel, pump head, a servo-motor, various setting keys and display elements. The device is to be used with special designed irrigation and suction tubings and a foot-remote control. A constant performed redundant pressure measurement controls the conformity of the actual pressure in the joint cavity with the pre-set nominal pressure.
### E. Intended Use
The SOPRO 670 irrigation pump is an arthroscopic pump system intended to provide fluid distension and irrigation of knee, shoulder, hip, elbow, ankle and wrist joint cavities and fluid suction during diagnostic and operative arthroscopic procedures that may only be used by qualified physicians.
#### F. Substantial Equivalence
The SOPRO 670 is substantially equivalent to the Arthroscopic pump functions of the predicate device by Future Medical Systems, Arthroscopy Pump and Shaver Model FMS DUO (K954465) previously cleared by the FDA and currently marketed. Both the Arthroscopy Pump SOPRO 670 and the predicate device are intended to provide fluid distension and irrigation of knee, shoulder, elbow, hip, ankle and wrist joint cavities during diagnostic and operative arthroscopic procedures.
Furthermore, the Arthroscopy Pump SOPRO 670 and the predicate device FMS DUO (K954465) are both intended to provide fluid suction during arthroscopic procedures. In addition, the device described in this notification is similar in design and technical characteristics to the Arthroscopic pump of the predicate device.
Differences that exist between the devices relating to technical specifications, performances and intended use are minor and do not affect the safety and effectiveness of the SOPRO 670 Arthroscopic Pump.
#### G. Performance Data
The device complies with the International Standard IEC 601-1 (Electrical Safety) .In addition, the device meets the requirements of the Underwriter Laboratories Standard UL60601-1 and bears the CE mark in accordance with the Medical Device Directive 93/42/EEC.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SOPRO % ACTEON, Inc. Mr. Steve Salesky 124 Gaither Drive, Suite 140 Mt. Laurel, New Jersey 08054
JUL 11 2008
Re: K080122
Trade/Device Name: SOPRP 670 Arthroscopic Pump Regulation Number: 21 CFR 888.1100 Regulation Name: Arthroscope Regulatory Class: II Product Code: HRX Dated: June 6, 2008 Received: June 9, 2008
Dear Mr. Salesky:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Mr. Steve Salesky
This letter will allow you to begin marketing your device as described in your Section 510/k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Mark M. Wilkinson
Mark N. Melkerson Director Division of General. Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
1080122
p. 1 of 1
# Indications for Use
510(k) Number:
SOPRO 670 Arthroscopic Pump Device Name:
Indications for Use:
The SOPRO 670 irrigation pump is a arthroscopic pump system intended to provide fluid distension and irrigation of knee, shoulder, hip, elbow, ankle and wrist joint cavities and fluid suction during diagnostic and operative arthroscopic procedures that may only be used by qualified physicians
Prescription Use _ X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
510(k) Number korap
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.